Certificate of
measurement 4005
Hard Drinking Water – Anions
Certified Reference Material LGC6012
Certified Values
Number of Units Certified Coverage
Constituent1 Uncertainty3
laboratories value2 factor, k3
Ammonium (NH4) 17 mg/L 0.412 0.024 2.05
Chloride (Cl) 19 mg/L 246.5 5.8 2.04
Fluoride (F) 17 mg/L 1.36 0.11 2.07
Nitrate (NO3) 18 mg/L 49.4 1.1 2.10
Sulfate (SO4) 19 mg/L 246.9 8.8 2.05
Notes:
1. The constituents were certified using the results of an inter-laboratory comparison using different
methods (see page 3). Robust estimates of location and dispersion were used as the values of
the concentration determined by participant laboratories. Each data set was obtained in a different
laboratory and/or using a different method of measurement.
2. The results are traceable to the SI through the physical and chemical standards used by the inter-
laboratory study participant laboratories. The certified values are reported to the same number of
decimal places as the uncertainties (which are reported to 2 significant figures).
3. The quoted uncertainty is the half-width of the expanded uncertainty interval, calculated using a
coverage factor (k), which gives a level of confidence of approximately 95 %.
Date of Issue: March 2024
Signed: _____________________
Gill Holcombe (Mrs)
for the Government Chemist
Material number: LGC6012
Batch number: 004
Page 1 of 5
Additional Information
The results of the inter-laboratory study provided a median pH of 8.0. The range of results reported
was 7.4 to 8.2.
The results of the inter-laboratory study provided a median conductivity of 1580 µS/cm. The range
of results reported was 1400 µS/cm to 1800 µS/cm.
The levels of calcium and magnesium were measured to calculate the hardness level of the
material. The total hardness was calculated to be 265 mg/L expressed as CaCO3, which equates to
a hardness classification of ‘Hard’ [1].
The density of the material was measured in duplicate on three units, using an Anton Paar DMA
5000 five place density meter at (20.00 ± 0.01) °C. A mean value of 0.99814 g/cm3 was obtained
for the density.
The values quoted for pH, conductivity and density are not with LGC’s scope of accreditation to
ISO 17034.
Material Preparation
Hard drinking water was sourced from the Teddington (Greater London, UK) potable mains supply.
The water was filtered through a set of 8 μm, 1.2 μm and 0.45 μm in-line filters and preserved with
the addition of copper sulfate to a final concentration of 1 mg/L as copper.
High purity salt standards were used to spike the base material to reach the target levels, which
reflect the regulatory limits in the European Drinking Water Directive [2] and the UK Water Supply
(Water Quality) Regulations 2016 [3].
The material was bottled in 250 mL portions in tamper-evident screw cap amber glass bottles and
stored at (5 ± 4) ºC.
Homogeneity Assessment
The homogeneity of the material was assessed for each of the parameters of interest. Analysis was
carried out using methods accredited to ISO/IEC 17025.
The results showed the material was fit-for-purpose for a sample size of 5 mL for ammonium,
chloride, nitrate and sulfate and for a sample size of 50 mL for fluoride.
Stability
The nature of the material is such that deterioration is not anticipated over its lifetime when stored
under the recommended conditions. The material will be monitored at LGC, and customers will be
notified of any changes in the certified values.
Certification
The material was certified using results from an inter-laboratory study organised by LGC.
Units of the candidate material were distributed to laboratories that had previously agreed to
participate in an inter-laboratory study. Participant laboratories chose suitable methods with which
they were familiar. The number of laboratories that used a particular method and submitted
acceptable data is shown in Table 1.
The data from the inter-laboratory study were processed using a robust statistical approach after
screening laboratories based on their performance in analysing a separate QC sample. The
certified value for each constituent was assigned as the robust estimate of location of the accepted
laboratory data. Uncertainties were based on the robust estimate of dispersion (taking into account
the number of laboratories and corrected for the efficiency of the estimator) and combined with the
uncertainties related to homogeneity and stability.
Material number: LGC6012
Batch number: 004
Page 2 of 5
Table 1: Number of laboratories using each method.
Constituent
Ammonium Nitrate Sulfate
Chloride (Cl) Fluoride (F)
(NH4) (NO3) (SO4)
Colorimetry 16 12 2 12 9
Ion
1 6 9 6 6
Chromatography
Ion Selective
0 0 5 0 0
Electrode
Titrimetry 0 1 1 0 0
Turbidimetry 0 0 0 0 1
ICP-MS 0 0 0 0 2
ICP-OES 0 0 0 0 1
Total 17 19 17 18 19
Participants
The following laboratories participated in the inter-laboratory study for this material:
Anglian Water Central Laboratory UK
British Geological Survey UK
Dwr Cymru Welsh Water UK
Edinburgh Scientific Services (City of Edinburgh Council) UK
Eurofins Chemtest UK
Glasgow Scientific Services UK
I2 Analytical Ltd Poland
Instituto Nacional de Saúde Dr. Ricardo Jorge (INSA) - Lisbon Portugal
Instituto Nacional de Saúde Dr. Ricardo Jorge (INSA) - Porto Portugal
Intertek Caleb Brett UK
Laboratorio Tecnológico del Uruguay Uruguay
Northern Ireland Water UK
Portsmouth Water UK
Public Analysts Laboratory Ireland
Severn Trent Water UK
South East Water UK
South West Water Ltd UK
States Analysts’ Laboratory - Guernsey UK
Water Services Corporation Laboratory Malta
Accreditation
Property values on this certificate are within LGC’s scope of accreditation to ISO 17034 unless
otherwise marked.
Material number: LGC6012
Batch number: 004
Page 3 of 5
Intended Use
This reference material is primarily intended for use in the development, validation, or quality
control of analytical methods for the determination of anions in hard drinking water. The material
may also be applicable to other similar matrices where more closely matched reference materials
are not available.
Instructions for Use
Before use, the material should be allowed to equilibrate to room temperature and thoroughly
mixed by inversion. The recommended minimum amount of sample to be used is 5 mL for
ammonium, chloride, nitrate and sulfate, and 50 mL for fluoride, based on the volume used for the
homogeneity assessment.
After use, the bottle should be securely closed and returned to storage at (5 ± 4) °C. Any remaining
sample should be used within one month from the date of opening, or discarded.
Storage
The material should be stored at (5 ± 4) °C in its original packaging.
Shelf Life
If stored unopened under the recommended conditions, the certified values will remain valid for 12
months from the date of shipment (see page 5 for shipment date).
Metrological Traceability
The results are considered traceable to the SI through the physical and chemical standards used
by the participant laboratories.
References
[1] UK Drinking Water Inspectorate website: https://www.dwi.gov.uk/consumers/learn-more-about-
your-water/water-hardness-hard-water/
[2] Drinking Water Directive. Council Directive 98/83/EC of 3 November 1998 on the quality of
water intended for human consumption.
[3] Statutory Instrument 2016 No. 614, Water, England and Wales, The Water Supply (Water
Quality) Regulations 2016.
Material number: LGC6012
Batch number: 004
Page 4 of 5
Unit Number Date of Shipment
Legal Notice
The values quoted in this certificate are the best estimate of the true values within the stated
uncertainties and based on the techniques described herein. No warranty or representation, express
or implied, is made that the use of the product or any information, material, apparatus, method or
process which is the subject of or referred to in this certificate does not infringe any third party rights.
Further, save to the extent: (a) prohibited by law; or (b) caused by a party's negligence; no party shall
be liable for the use made of the product, any information, material, apparatus, method or process
which is the subject of or referred to in this certificate. In no event shall the liability of any party exceed
whichever is the lower of: (i) the value of the product; or (ii) £500,000; and any liability for loss of
profit, loss of business or revenue, loss of anticipated savings, depletion of goodwill, any third-party
claims or any indirect or consequential loss or damage in connection herewith is expressly excluded.
_________________________________________________________________________
This material was produced by LGC in its role as the UK’s National Measurement Laboratory and
Designated Institute for Chemical and Bio-measurement.
This certificate may not be published except in full, unless permission for the publication of
an approved extract has been obtained in writing from LGC Limited. It does not of itself impute
to the subject of measurement any attributes beyond those shown by the data contained
herein.
Material number: LGC6012
Batch number: 004
Page 5 of 5
Certificate of
measurement 4005
Soft drinking water – Anions
Certified Reference Material LGC6013
Certified Values
Number of Units Certified Coverage
Constituent1 Uncertainty3
laboratories value2 factor, k3
Ammonium (NH4) 17 mg/L 0.441 0.015 2.07
Chloride (Cl) 19 mg/L 243.4 7.0 2.08
Fluoride (F) 17 mg/L 1.449 0.099 2.07
Nitrate (NO3) 18 mg/L 48.89 0.91 2.07
Sulfate (SO4) 19 mg/L 248.1 8.4 2.04
Notes:
1. The constituents were certified using the results of an inter-laboratory comparison using
different methods (see page 3). Robust estimates of location and dispersion were used as the
values of the concentration determined by participant laboratories. Each data set was obtained
in a different laboratory and/or using a different method of measurement.
2. The results are traceable to the SI through the physical and chemical standards used by LGC
and the inter-laboratory study participant laboratories. The certified values are reported to the
same number of decimal places as the uncertainties (reported to 2 significant figures).
3. The quoted uncertainty is the half-width of the expanded uncertainty interval, calculated using
a coverage factor (k), which gives a level of confidence of approximately 95 %.
Date of Issue: March 2024
Signed: _____________________
Gill Holcombe (Mrs)
for the Government Chemist
Material number: LGC6013
Batch number: 004
Page 1 of 5
Additional Information
The results of the inter-laboratory study provided a median pH of 7.2, the range of results reported
was 6.9 to 7.8.
The results of the inter-laboratory study provided a median conductivity of 1380 µS/cm, the range
of results reported was 1270 µS/cm to 1540 µS/cm.
The levels of calcium and magnesium were measured to calculate the hardness level of the
material. The total hardness was calculated to be 37 mg/L expressed as CaCO3, which equates to
a hardness classification of ‘Soft’ [1].
The density of the material was measured in duplicate on three units, using an Anton Paar DMA
5000 five-place density meter at (20.00 ± 0.01) °C. A mean value of 0.99789 g/cm3 was obtained
for the density.
The values quoted for pH, conductivity and density are not with LGC’s scope of accreditation to
ISO 17034.
Material Preparation
Soft drinking water was sourced from the Bury (Greater Manchester, UK) potable mains supply.
The water was filtered through a set of 8 μm, 1.2 μm and 0.45 μm in-line filters and preserved with
the addition of copper sulfate to a final concentration of 1 mg/L as copper.
High purity salt standards were used to spike the base material to reach the target levels, reflecting
the regulatory limits in the European Drinking Water Directive [2] and the UK Water Supply (Water
Quality) Regulations 2016 [3].
The material was bottled in 250 mL portions in tamper-evident screw cap amber glass bottles and
stored at (5 ± 4) ºC.
Homogeneity Assessment
The homogeneity of the material was assessed for each of the parameters of interest and analysis
was carried out, using methods accredited to ISO/IEC 17025.
The results showed the material was fit-for-purpose for a sample size of 5 mL for ammonium,
chloride, nitrate and sulfate, and for a sample size of 50 mL for fluoride.
Stability
The nature of the material is such that deterioration is not anticipated over its lifetime when stored
under the recommended conditions. The material will be monitored at LGC, and customers will be
notified of any changes in the certified values.
Certification
The material was certified using results from an inter-laboratory study organised by LGC.
Units of the candidate material were distributed to laboratories that had previously agreed to
participate in an inter-laboratory study. Participant laboratories chose suitable methods with which
they were familiar. The number of laboratories that used a particular method and submitted
acceptable data is shown in Table 1.
The data from the inter-laboratory study were processed using a robust statistical approach after
screening laboratories based on their performance in analysing a separate QC sample. The
certified value for each constituent was assigned as the robust estimate of location of the accepted
laboratory data. Uncertainties were based on the robust estimate of dispersion (taking into account
the number of laboratories and corrected for the efficiency of the estimator) and combined with the
uncertainties related to homogeneity and stability.
Material number: LGC6013
Batch number: 004
Page 2 of 5
Table 1: Number of laboratories using each method.
Constituent
Ammonium Nitrate Sulfate
Chloride (Cl) Fluoride (F)
(NH4) (NO3) (SO4)
Colorimetry 16 12 2 12 9
Ion
1 6 9 6 6
Chromatography
Ion Selective
0 0 5 0 0
Electrode
Titrimetry 0 1 1 0 0
Turbidimetry 0 0 0 0 1
ICP-MS 0 0 0 0 2
ICP-OES 0 0 0 0 1
Total 17 19 17 18 19
Participants
The following laboratories participated in the inter-laboratory study for this material:
Anglian Water Central Laboratory UK
British Geological Survey UK
Dwr Cymru Welsh Water UK
Edinburgh Scientific Services (City of Edinburgh Council) UK
Eurofins Chemtest UK
Glasgow Scientific Services UK
I2 Analytical Ltd Poland
Instituto Nacional de Saúde Dr. Ricardo Jorge (INSA) - Lisbon Portugal
Instituto Nacional de Saúde Dr. Ricardo Jorge (INSA) - Porto Portugal
Intertek Caleb Brett UK
Laboratorio Tecnológico del Uruguay Uruguay
Northern Ireland Water UK
Portsmouth Water UK
Public Analysts Laboratory Ireland
Severn Trent Water UK
South East Water UK
South West Water Ltd UK
States Analysts’ Laboratory - Guernsey UK
Water Services Corporation Laboratory Malta
Accreditation
Property values on this certificate are within LGC’s scope of accreditation to ISO 17034 unless
otherwise marked.
Material number: LGC6013
Batch number: 004
Page 3 of 5
Intended Use
This reference material is primarily intended for use in the development, validation, or quality
control of analytical methods for the determination of anions in soft drinking water. The material
may also be applicable to other similar matrices where more closely matched reference materials
are not available.
Instructions for Use
Before use, the material should be allowed to equilibrate to room temperature and thoroughly
mixed by inversion. The recommended minimum amount of sample to be used is 5 mL for
ammonium, chloride, nitrate and sulfate, and 50 mL for fluoride, based on the volume used for
homogeneity assessment.
After use, the bottle should be securely closed and returned to storage at (5 ± 4) °C. Any remaining
sample should be used within one month from the date of opening or discarded.
Storage
The material should be stored at (5 ± 4) °C in its original packaging.
Shelf Life
If stored unopened under the recommended conditions the certified values will remain valid for 12
months from the date of shipment (see page 5 for shipment date).
Metrological Traceability
The results are considered traceable to the SI through the physical and chemical standards used
by the participant laboratories.
References
[1] UK Drinking Water Inspectorate Website: https://www.dwi.gov.uk/consumers/learn-more-about-
your-water/water-hardness-hard-water/
[2] Drinking Water Directive. Council Directive 98/83/EC of 3 November 1998 on the quality of
water intended for human consumption.
[3] Statutory Instrument 2016 No. 614, Water, England and Wales, The Water Supply (Water
Quality) Regulations 2016.
Material number: LGC6013
Batch number: 004
Page 4 of 5
Unit Number Date of Shipment
Legal Notice
The values quoted in this certificate are the best estimate of the true values within the stated
uncertainties and based on the techniques described herein. No warranty or representation, express
or implied, is made that the use of the product or any information, material, apparatus, method or
process which is the subject of or referred to in this certificate does not infringe any third party rights.
Further, save to the extent: (a) prohibited by law; or (b) caused by a party's negligence; no party shall
be liable for the use made of the product, any information, material, apparatus, method or process
which is the subject of or referred to in this certificate. In no event shall the liability of any party exceed
whichever is the lower of: (i) the value of the product; or (ii) £500,000; and any liability for loss of
profit, loss of business or revenue, loss of anticipated savings, depletion of goodwill, any third-party
claims or any indirect or consequential loss or damage in connection herewith is expressly excluded.
_________________________________________________________________________
This material was produced by LGC in its role as the UK’s National Measurement Laboratory and
Designated Institute for Chemical and Bio-measurement.
This certificate may not be published except in full, unless permission for the publication of
an approved extract has been obtained in writing from LGC Limited. It does not of itself impute
to the subject of measurement any attributes beyond those shown by the data contained
herein.
Material number: LGC6013
Batch number: 004
Page 5 of 5
Saatja: "Adam Kochanowski" <
[email protected]>
Saaja: "TA Info" <
[email protected]>
Teema: Fw: LGC Standards: Introduction and new alternatives for Drinking Waters
Kuupäev: 2024-06-07 11:22
Tähelepanu! Tegemist on väljastpoolt asutust saabunud kirjaga. Tundmatu
saatja korral palume linke ja faile mitte avada.
Dear Customer,
I sent the below e-mail to Marleen but she's away.
Please take a look at this! :)
Have a good day,
Adam
<https://www2.lgcgroup.com/l/31922/2019-03-08/gxvzcy/31922/192630/Lgc_logo.png>
Adam Kochanowski | Sales Specialist | LGC Standards |
LGC | ul. Ogrodowa 27/29 | Kiełpin | 05-092 Lomianki | Poland |
Tel: +48 22 751 31 40 | Mob: +48 721 739 797 | Fax: +48 22 751 58 45 |
LinkedIn <https://www.linkedin.com/in/adam-kochanowski/>
Book time with me
<https://outlook.office.com/bookwithme/user/
[email protected]?anonymous&ep=pcard>
Zachęcamy do odwiedzenia naszej strony/Visit our website
www.lgcstandards.com <http://www.lgcstandards.com/>
--------------- Forwarded Message ---------------
From: Adam Kochanowski <
[email protected]>
Date: 7 June 2024 at 13:18
Subject: LGC Standards: Introduction and new alternatives for Drinking
Waters
To: Marleen Alttoa <
[email protected]>
Dear Marleen,
Hope you are doing well!
Firstly, I would like to introduce myself to you. My name is Adam and I have
just taken responsibility over your customer account. Do not hesitate to
contact me when you need to talk about current or future cooperation 😊
This time, I am contacting you because you have been interested in the
following products in the past:
ERM-CA015 - Hard drinking water – Anions
ERM-CA016 - Soft drinking water – Anions
These materials have been sold out for a while.
Now there are the new replacements available:
<https://url.uk.m.mimecastprotect.com/s/GLAIC2GNDHKwqqXt1qJin?domain=lgcstandards.com>
LGC6012 - Hard drinking water - Anions
<https://url.uk.m.mimecastprotect.com/s/VuLZC3YXVt7Q11lC2W0nH?domain=lgcstandards.com>
LGC6013 - Soft drinking water - Anions
Attached you can find the Certificates with the values for Ammonium,
Chloride, Fluoride, Nitrate and Sulfate.
If you are interested in these materials, please let us know and we can send
a quote.
Attached is a list of further matrix materials for water analysis, that
might be of interest.
By clicking on the links you will be forwarded to our webshop with further
detailed information.
In case of any questions, please let me know.
Kind regards,
Adam
<https://www2.lgcgroup.com/l/31922/2019-03-08/gxvzcy/31922/192630/Lgc_logo.png>
Adam Kochanowski | Sales Specialist | LGC Standards |
LGC | ul. Ogrodowa 27/29 | Kiełpin | 05-092 Lomianki | Poland |
Tel: +48 22 751 31 40 | Mob: +48 721 739 797 | Fax: +48 22 751 58 45 |
LinkedIn <https://www.linkedin.com/in/adam-kochanowski/>
Book time with me
<https://outlook.office.com/bookwithme/user/
[email protected]?anonymous&ep=pcard>
Zachęcamy do odwiedzenia naszej strony/Visit our website
www.lgcstandards.com <http://www.lgcstandards.com/>
<https://lgcgroup.my.salesforce.com/servlet/servlet.ImageServer?oid=00DD0000000pza9&esid=018P300000A6t2Y&from=ext>