  • 📎Raptor OIML R 85 Report No. SP-P905645.pdf170 KB
  • 📎Tüübikinnituse taotlemine Raptor süsteemile.msg3248 KB

Sisu (failidest)

Report number SP-P905645 OIML R 85:2008 (E) Page 1(51) TYPE APPROVAL ROSEMOUNT RAPTOR TANK GAUGING SYSTEM A. AUTHORITY RESPONSIBLE FOR THIS REPORT .......................................3 B. SYNOPSIS OF THE RESULTS OF THE EXAMINATIONS AND TESTS....4 C. SUMMARY OF THE RESULTS OF THE EXAMINATIONS AND TESTS ..5 C.1 Examinations .......................................................................................................5 C.2 Performance tests ................................................................................................6 D. GENERAL INFORMATION .............................................................................7 D.1 Manufacturer .......................................................................................................7 D.2 Applicant .............................................................................................................7 D.3 Testing laboratories involved in the tests ............................................................8 D.4 General information concerning the type and the sample(s) submitted for the tests (as stated on the instrument / provided by the manufacturer) .............................................9 D.5 Accessories supplied by the applicant ................................................................10 D.6 Selection of sample(s) tested...............................................................................10 D.7 Adjustments and modifications ...........................................................................11 D.8 Choice of the manufacturer concerning the severity levels ................................11 D.9 Full details about the electrical power requirements...........................................11 D.10 Additional information concerning the type .......................................................11 D.11 Documentation supplied by the applicant ...........................................................12 D.12 Information concerning the test equipment used for the type evaluation ...........12 E. EXAMINATION ................................................................................................13 E.1 Constituents (4) ...................................................................................................13 E.2 Units of measurements (5) ..................................................................................14 E.3 Rated operating conditions (6.1) .........................................................................14 E.4 Special conditions (6.1).......................................................................................15 E.5 Indicating device (7.1) ........................................................................................15 E.6 Printer(s) (7.1.9) ..................................................................................................16 E.7 Movable sensor (7.2)...........................................................................................16 E.8 Installation (7.3) ..................................................................................................17 E.9 Ancillary devices (7.4) ........................................................................................17 E.10 Markings (7.5) .....................................................................................................18 E.11 Verification marks (7.6) ......................................................................................18 E.12 Mechanical sealing (7.7) .....................................................................................19 E.13 Electronic sealing (7.7) .......................................................................................19 E.14 Safeguarding the integrity of the measurement (7.8) ..........................................20 F. PERFORMANCE TESTS ..................................................................................21 F.1 Accuracy (8.1.5.2) ...............................................................................................21 F.1.1 Parabolic antenna ................................................................................................21 Report number SP-P905645 OIML R 85:2008 (E) Page 2(51) F.1.2 Horn antenna .......................................................................................................23 F.1.3 Array antenna 5” .................................................................................................25 F.1.4 Array antenna 6” .................................................................................................27 F.1.5 Array antenna 8” .................................................................................................29 F.1.6 Array antenna 10” ...............................................................................................31 F.1.7 Array antenna 12” ...............................................................................................33 F.1.8 LPG/LNG antenna ..............................................................................................35 F.2 Discrimination (8.1.5.3) ......................................................................................37 F.3 Hysteris (8.1.5.4) .................................................................................................38 F.4 Influence factor tests ...........................................................................................39 F.4.1 Static environmental temperatures (8.1.6.2) .......................................................39 F.4.2 Mains voltage variations (8.1.6.3 and 8.1.6.4) ....................................................40 F.5 Disturbance tests .................................................................................................41 F.5.1 Damp heat cyclic (condensing) (8.1.7.1) ............................................................41 F.5.2 Radiated, radio frequency, electromagnetic fields (8.1.7.2.1) ............................42 F.5.3 Conducted, radio frequency, electromagnetic fields (8.1.7.2.2) .........................43 F.5.4 Electrostatic discharge (8.1.7.2.3) .......................................................................44 F.5.5 Bursts on signal, data and control lines (8.1.7.2.4) .............................................45 F.5.6 Surges on signal, data and control lines (8.1.7.2.5) ............................................46 F.5.7 AC mains voltage dips, short interruptions and voltage variations (8.1.7.2.6) ...47 F.5.8 Bursts (transients) on AC and DC mains (8.1.7.2.7) ..........................................48 F.5.9 Voltage dips, short interruptions and voltage variations on DC mains (8.1.7.2.8) 49 F.5.10 Ripple on DC mains power (8.1.7.2.9) ...............................................................50 F.5.11 Surges on AC and DC mains power lines (8.1.7.2.10) .......................................51 Report number SP-P905645 OIML R 85:2008 (E) Page 3(51) A. AUTHORITY RESPONSIBLE FOR THIS REPORT Name SP Address Brinellgatan 4. Box 857. SE-501 15 Borås. Sweden Report No. SP-P905645 Application No. 2010-06-10RosemountTankRadarAB Date/period of tests 2009-08-01 – 2010-12-31 Date of issuing this report 2011-04-12 Name and signature of the Måns Ackerholm, SP Measurement Technology responsible person Stamp(s) (if applicable) Report number SP-P905645 OIML R 85:2008 (E) Page 4(51) B. SYNOPSIS OF THE RESULTS OF THE EXAMINATIONS AND TESTS The tested specimen fulfills ALL the applicable requirements in OIML R 85 YES NO (2008) X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 5(51) C. SUMMARY OF THE RESULTS OF THE EXAMINATIONS AND TESTS C.1 Examinations For details, refer to the examination: Clause E of this report as indicated in the first column Sub- Test (with reference to requirements in + - Remarks Page clause Part 1) E.1 Constituents (4) X 13 E.2 Units of measurement (5) X 14 E.3 Rated operating conditions (6.1) X 14 E.4 Special conditions (6.1) X 15 E.5 Indicating device (7.1) X 15 E.6 Printer(s) (7.1.9) X 16 E.7 Moveable sensor (7.2) X 16 E.8 Installation (7.3) X 17 E.9 Ancillary devices (7.4) X 17 E.10 Markings (7.5) X 18 E.11 Verification marks (7.6) X 18 E.12 Mechanical sealing (7.7) X 19 E.13 Electronic sealing (7.8) X 19 E.14 Safeguarding the integrity of the X 20 measurement (7.9) Report number SP-P905645 OIML R 85:2008 (E) Page 6(51) C.2 Performance tests For details, refer to the tests: clause F in this report as indicated in first column Sub- Test (with reference to requirements in + - Remarks Page clause Part 1) F.1 Accuracy (8.1.5.2) X 21-36 F.2 Discrimination (8.1.5.3) Not applicable (37) F.3 Hysteresis (8.1.5.4) Not applicable (38) F.4.1 Temperature tests (8.1.6.2) X 39 F.4.2 Main voltage variations (8.1.6.3 + 8.1.6.4) X 40 F.5.1 Damp heat cyclic (condensing) (8.1.7.1) X 41 F.5.2 Radiated radio-frequency electromagnetic X 42 fields (8.1.7.2.1) F.5.3 Conducted radio-frequency electromagnetic X 43 fields (8.1.7.2.2) F.5.4 Electrostatic discharge (8.1.7.2.3) X 44 F.5.5 Bursts on signal. data and control lines X 45 (8.1.7.2.4) F.5.6 Surges on signal. data and control lines X 46 (8.1.7.2.5) F.5.7 AC mains voltage dips. short interruptions X 47 and voltage variations (8.1.7.2.6) F.5.8 Bursts (transients) on AC and DC mains X 48 (8.1.7.2.7) F.5.9 Voltage dips. short interruptions and voltage X 49 variations on DC mains power (8.1.7.2.8) F.5.10 Ripple on DC mains power (8.1.7.2.9) X 50 F.5.11 Surges on AC and DC mains power lines X 51 (8.1.7.2.10) Report number SP-P905645 OIML R 85:2008 (E) Page 7(51) D. GENERAL INFORMATION D.1 Manufacturer Company Emerson Process Management, Rosemount Tank Radar AB Address Box 130 45. SE-402 51 Göteborg. Sweden D.2 Applicant Company Emerson Process Management, Rosemount Tank Radar AB Representative N/A Address Box 130 45. SE-402 51 Göteborg. Sweden Reference Christian Skaug, Technical product manager Date of Application 2010-06-10 Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 8(51) D.3 Testing laboratories involved in the tests Name SP Swedish technical research institute Address Brinellgatan 4. SE-504 62 Borås. Sweden Application No. 2010-06-10RosemountTankRadarAB Tests by this laboratory All tests Date/period of tests 2009-08-01-2010-12-31 Name(s) of test engineer(s) Måns Ackerholm Jörgen Eriksson Magnus Holmgren Magnus Herbertsson Mattias Engström Bengt Andersson Joakim Lindeblom Reine Johansson Accredited by SWEDAC Accreditation includes R85 Yes Details of relevant peer assessment or N/A assessment by other means In case tests were performed at another Test F1 and F4 were performed at Rosemount location than the premises of this laboratory. Tank Radar AB, in their measuring range (see give details here D.12). Name of the responsible person Måns Ackerholm Date of signature 2011-04-08 Stamp (if applicable) and signature of the responsible person Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 9(51) D.4 General information concerning the type and the sample(s) submitted for the tests (as stated on the instrument / provided by the manufacturer) Manufacturer’s Rosemount Tank Radar AB trademark/corporate name Year of manufacture 2009 Type designation Raptor Tank Gauging System Model No. - 5900S Radar level gauge with: Parabolic antenna 1P Horn antenna 1H Still-pipe array antenna 1A5 Still-pipe array antenna 1A6 Still-pipe array antenna 1A8 Still-pipe array antenna 1AA Still-pipe array antenna 1AB LPG/LNG antenna G1/G2 - 2410 Tank Hub - 2230 Graphical Field Display - 2160/61 Field Communication Unit - 2180 Field Bus Modem - General purpose PC with TankMaster for display of measured level - Printer. HP LaserJet 2100M Serial number(s) 5900S: SP marked SPL 62/09 of the samples 2410: SP marked SPL 64/09 2230: SP marked 65/09 2160/61: Serial No. 6109 2180: SPL 80/09 PC: SAAB-SME-1609 and SAAB-SME-1327 Measuring For all antenna types: From 0.8 to 30 meter. Distances range(s) measured from installation flange. Electrical power Power supply to 2410 Tank Hub: 24- 48 VDC or 48- 240 VAC 50/60 Hz 5900S and 2230 are powered from 2410 with intrinsically safe power. Identification of software Version number(s) 5900S: 1.A8 (0xCFD5) 2410: 1.A4 2230: 1.A1 Checksum/identification TankMaster: 6.xx Report number SP-P905645 OIML R 85:2008 (E) Page 10(51) Remarks: D.5 Accessories supplied by the applicant Operating instructions 5900S Radar level gauge Reference Manual 2410 TankHub Reference Manual 2230 Graphical Field Display Reference Manual Rosemount Raptor Tank Gauging System Data printer Network printer Cables Cables for connection of system units Ancillary facilities Generic FF device that fulfills ITK-tests (such as 2240S Multi-input Temperature Transmitter) Power supply unit DC power supply for 2410 Other accessories/remarks: 1. 2240S has no legal metrology function related to R85. Only connected to show no influence on level measuring results. 2. Coaxial cable SP number SPL 215/99 used as target under disturbance tests. D.6 Selection of sample(s) tested The 5900S Radar Level Gauge is available with different antennas for different tank types. See D4. Accuracy tests were performed with a separate test protocol for each antenna type and each antenna type has to be within MPE. Two different electrical interfaces between 2410 and FCU2160 were be tested. One consists of a standard RS485 interface and the other consists of an FSK modulated signal (the so called “TRL/2 interface”). Both results were registered on two separate PC’s connected to FCU2160. Justification for the selection of the samples: The sample of the 5900S Radar Level Gauge was selected randomly from 3 produced units by SP. The 5900S unit was engraved with identifier for safe identification during the whole test. Other units as 2410, 2230 etc has no measuring function, but instead a special function such as: communication of data. power supply or digital display of measured value. These were selected from available units for the test and marked with identifier as above. Report number SP-P905645 OIML R 85:2008 (E) Page 11(51) D.7 Adjustments and modifications FCU 2160 was modified for test F.5.11 D.8 Choice of the manufacturer concerning the severity levels Influence Severity level Dry Heat (1, 2, 3, 4): 4 Cold (1, 2, 3, 4): 4 DC mains voltage High Level: -15%, +10% D.9 Full details about the electrical power requirements 24-48 VDC or 48-240 VAC, 50/60 Hz Max 20W depending on configuration D.10 Additional information concerning the type Additional remarks and/or information (connection equipment. interfaces etc) Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 12(51) D.11 Documentation supplied by the applicant Technical manuals and descriptions Name Ref. no Revision Technical Description 704 010 EN Rev AA, May 2010 Section 1. Raptor system Section 2. 5900S series Radar Level Gauge Section 3. 2240S Multi-input Temperature Transmitter Section 6. 2230 Graphical Field Display Section 7. 2410 Tank Hub Section 8. 2160 Field Communication Unit Section 10. 2180 Field Bus Modem Raptor System Revision History DE-0056 Issue 8d Raptor System Configuration Manual 300 510 EN Rev AA, June 2010 5900 Reference Manual 300 520 EN Rev AA, June 2010 2410 Reference Manual 300 530 EN Rev AA, June 2010 2230 Reference Manual 300 560 EN Rev AA, June 2010 2240 Reference Manual 300 550 EN Rev AA, June 2010 D.12 Information concerning the test equipment used for the type evaluation 1) Leica Lasertracker SP 602084, calibrated at SP. 2) Measuring Range M2. Located at Rosemount Tank Radar AB. Calibrated against Laser interferometer (SP inventory no. 601046). Evaluated at SP with an expanded measurement uncertainty of ±0.06mm. Report MTmP904224-K01. 3) Keithley 2700, s/n 09224830, calibrated at SP 4) Measurement equipment for EMC tests, routinely used for accredited methods 5) Measurement equipment and chambers for environmental tests routinely, used at accredited methods 6) Various enviromental indicators (temperature, pressure, humidity), accredited calibrated at SP Report number SP-P905645 OIML R 85:2008 (E) Page 13(51) E. EXAMINATION E.1 Constituents (4) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Liquid level sensor Microwave antennas, type varies with application. Non- movable type Transducer 5900S, electronic type, emits microwave sweep, type FMCW. Correction sensor 1. 2240S with temperature sensor for correction of gauge reference height. 2. 2051 pressure sensor for calculation of vapor density in LPG tanks. Calculator 1. In 5900S: Calculates level from microwave signal, can correct with data from correction sensor(s) 2. Tankmaster software in PC: Calculates volumes and mass from measured and operator entered data. Indicating device(s) 1. 2230 Grafical Field Display, typically installed at tank 2. 2410 TankHub Display, optional field display in Tank Hub 3. Tankmaster software in PC, typically installed in control room Printer Standard printer Ancillary devices Various FF devices that fulfills ITK-tests can be connected to the Tank Bus Checking facilities Automatic checking facility of both Permanent and Intermittent type Others FCU 2160/61, FBM 2180 Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 14(51) E.2 Units of measurements (5) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Display Print Dip m m Ullage m m Other Indication(s) #-sign = legal value #-sign (2230 and Tankmaster) *-sign = legal value (2410) Remarks: Displays and printers are programmable to show meter, mm, ft-inch-1/16, feet decimal. Normal mode for legal tank gauging system is meter. E.3 Rated operating conditions (6.1) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Ambient temperatures Low -40 °C High +70 °C Relative humidity up to 100% DC mains voltage 24-48 VDC (-15%...+10%) AC mains voltage 48-240 VAC (-15%...+10%) Temperature extreme values Liquid N/A antenna not in contact with liquid Medium above -40 – 230°C the liquid (LPG/LNG antenna -170°C to 90°C, 55°C – 90°C close to ball-valve) Pressure extreme values Varies with antenna type -0.2 – 2 bar (horn- and still-pipe antenna), - 0.2 -10 bar (parabolic antenna), up to 25 bar (LPG antenna) Liquid characteristics N/A, liquid does not affect measurement performance Liquid density. extreme N/A, liquid density does not affect measurement performance values Medium characteristics Typical petroleum products Medium density. extreme values Remarks: Normal petroleum products have no measurable influence on performance. Certain special products (e.g. ketons) may influence velocity of the radar wave and may result in performance outside m.p.e. See manufacturer’s recommendation on which products to avoid. Tank design with internal floating roof may reduce influence for critical products. Report number SP-P905645 OIML R 85:2008 (E) Page 15(51) E.4 Special conditions (6.1) Date: Observer: Serial No. Remarks: E.5 Indicating device (7.1) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Sub-clause Requirement Remarks + - 7.1.1 Distance between analog scale marks N/A, digital indicating device 7.1.2 Number of indicating devices 4 types (incl printer), see D.4 X 7.1.3 Common indicating devices Tankmaster screen and Printers X 7.1.4 Remote indication duly identified Yes X 7.1.2-3-4 Other indicating devices See D.4 X 7.1.2 Alarm when outside limits Yes, flagged as not approved X value 7.1.5 Default display of dip / ullage Dip X 7.1.6 Display of measurement Yes X 7.1.7 Symbol or name of unit present Yes X Scale interval N/A, digital indicating device 7.1.8 Digital display Yes X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 16(51) E.6 Printer(s) (7.1.9) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Sub-clause Requirement Remarks + - 7.1.2 Number of printers None, one or more X 7.1.4 Printer duly identified to Tank name to which a certain ALG X ALG is configured is printed on the same row as a specific level value for indication 7.1.5 Default printer of dip/ullage Dip X 7.1.6 Print of measurement Yes X 7.1.7 Symbol or name of unit Yes X present Scale interval N/A, digital indicating device Remarks: E.7 Movable sensor (7.2) Date: Observer: Serial No. Sub-clause Requirement Remarks + - 7.2.1 Suspension mechanism 7.2.2 Static position Remarks: Not applicable Report number SP-P905645 OIML R 85:2008 (E) Page 17(51) E.8 Installation (7.3) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Sub-clause Requirement Remarks + - 7.3.1.1 Accessible and legible Yes X 7.3.1.2 Verification possible Yes X 7.3.1.3 Gauge hatch 7.3.1.3 No obstacles 7.3.1.4 Affect ALG measurement Installation to be in accordance 7.3.1.5 Influence eddies etc with installation drawings. 7.3.1.6 Compensation movement tank 7.3.1.7 Location correction sensor 7.3.1.8 Thermal expansion Supported in the software X Remarks: E.9 Ancillary devices (7.4) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Description Remarks + - 2240S Multi-input Temperature Transmitter X Generic FF device that fulfills ITK-test X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 18(51) E.10 Markings (7.5) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Requirements Remarks + - Locations of the markings Top or side on each unit X Name of the manufacturer / trademark Rosemount Tank Radar AB X Type designation 5900S, 2410 X Serial number Yes X Year of manufacture Yes X Type approval mark (provisions) Yes (space available on X complementary plate) Ranges defining the field of operation Yes X Visibility Yes X Additional information (if required) Remarks: National marking requirements will be taken into consideration for each country. E.11 Verification marks (7.6) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Requirements Remarks + - Location Top or side on each unit X Fit for easy application Yes X Impossible to remove without damage Yes X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 19(51) E.12 Mechanical sealing (7.7) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Requirements Remarks + - Data plate Yes X Other components Housings: Yes X Impossible to remove without damage Yes X Remarks: Various solutions possible. Filling or wire solutions E.13 Electronic sealing (7.7) Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Requirements Remarks + - Access Using Tankmaster WinSetup, access level X as administrator Password changeable Yes X Configuration mode Indicated by write protection X Event counter Yes X Date of change parameter Yes X Value of parameter Yes X Identification authorized person Yes X Traceability last intervention Yes X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 20(51) E.14 Safeguarding the integrity of the measurement (7.8) Date: 2011-03-15 Observer: Måns Ackerholm Serial No. (See D.4) Sub-clause Requirement Remarks + - 7.8.1 Possibility of testing checking Can’t be disabled. #-sign indicate X facility “no error” Checking facility type I or P Background Diagnostics are type I X where they are asynchronous with the actual measurement. The #sign checking against limits, write protect etc are of type P, where the check is done prior the data for #-sign is delivered. 7.8.2.1 Checking facility disturbances OK X 7.8.2.3 Durability checking facility OK X 7.8.2.5 Action by checking facility Removal of #-sign X 7.8.3.1 Protection of data OK X 7.8.3.2 Integrity of permanent stored CRC-16 or checksum X instructions 7.8.3.3 Transfer/storage of data CRC-16 X 7.8.3.4 Checking facility calculator Both software and hardware types of X wathdogs 7.8.3.5 Checking facility indicating OK X device 7.8.3.6 Checking facility ancillary device Checked by 2410 Tank Hub X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 21(51) F. PERFORMANCE TESTS F.1 Accuracy (8.1.5.2) F.1.1 Parabolic antenna Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4) Test conditions Begin End Time 09:30 10:45 Environmental temperature 19.5 °C 19.6 °C Relative humidity 56.9 % RH 56,8 %RH Atmospheric pressure 995 hPa 992 hPa Observations in m: upwards Level Indication Error (mm) MPE (mm) + - 0.7998 0.800 0.2 ±1 X 1.0001 1.000 -0.1 ±1 X 3.9999 4.000 0.1 ±1 X 6.9996 6.999 -0.6 ±1 X 10.0003 10.001 0.7 ±1 X 12.9994 13.000 0.6 ±1 X 16.0004 16.001 0.6 ±1 X 18.9996 19.000 0.4 ±1 X 22.0001 22.000 -0.1 ±1 X 24.9998 25.000 0.2 ±1 X 27.9999 28.000 0.1 ±1 X 29.5003 29.500 -0.3 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 22(51) Observations in m: downwards Level Indication Error (mm) MPE (mm) + - 29.5003 29.500 -0.3 ±1 X 27.9997 28.000 0.3 ±1 X 25.0001 25.000 -0.1 ±1 X 22.0001 22.000 -0.1 ±1 X 18.9998 19.000 0.2 ±1 X 16.0004 16.001 0.6 ±1 X 12.9995 13.000 0.5 ±1 X 10.0005 10.001 0.5 ±1 X 6.9996 6.999 0.3 ±1 X 4.0003 4.000 -0.3 ±1 X 1.0000 1.000 0.0 ±1 X 0.7998 0.800 0.2 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 23(51) F.1.2 Horn antenna Date of test: 2010-09-27 Observer: Magnus Holmsten Serial No. (See D.4) Test conditions Begin End Time 09:25 09:55 Environmental temperature 20.8 °C 20.5 °C Relative humidity 41.8 % RH 41.5 %RH Atmospheric pressure 1012 hPa 1012 hPa Observations in m: upwards Level Indication Error (mm) MPE (mm) + - 0.8002 0.800 -0.2 ±1 X 1.0001 1.000 -0.1 ±1 X 4.0000 4.000 0.0 ±1 X 6.9997 7.000 0.3 ±1 X 10.0003 10.001 0.7 ±1 X 12.9994 12.999 -0.4 ±1 X 16.0004 16.000 -0.4 ±1 X 18.9996 18.999 -0.6 ±1 X 22.0002 22.000 -0.2 ±1 X 24.9998 24.999 -0.8 ±1 X 27.9999 27.999 -0.9 ±1 X 29.5003 29.500 -0.3 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 24(51) Observations in m: downwards Level Indication Error (mm) MPE (mm) + - 29.5003 29.500 -0.3 ±1 X 27.9998 27.999 -0.8 ±1 X 25.0001 25.000 -0.1 ±1 X 22.0001 22.000 -0.1 ±1 X 18.9998 19.000 -0.8 ±1 X 16.0004 16.000 -0.4 ±1 X 12.9995 12.999 -0.5 ±1 X 10.0006 10.001 0.4 ±1 X 6.9996 7.000 0.4 ±1 X 4.0002 4.000 -0.2 ±1 X 1.0000 1.000 0.0 ±1 X 0.7998 0.800 0.2 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 25(51) F.1.3 Array antenna 5” Date of test: 2010-10-12 Observer: Måns Ackerholm Serial No. (See D.4) Test conditions Begin End Time 11:02 11:54 Environmental temperature 22.2 °C 22.2 °C Relative humidity 31.5 % RH 31.5%RH Atmospheric pressure 1018 hPa 1018 hPa Observations in m: upwards Level Indication Error (mm) MPE (mm) + - 0.7933 0.793 -0.3 ±1 X 1.0068 1.007 0.2 ±1 X 3.9991 4.000 0.9 ±1 X 6.9986 6.999 0.4 ±1 X 9.9996 10.000 0.4 ±1 X 12.9979 12.998 0.1 ±1 X 15.9978 15.998 0.2 ±1 X 18.9965 18.996 -0.5 ±1 X 21.9978 21.998 0.2 ±1 X 24.9975 24.998 0.5 ±1 X 27.9973 27.998 0.7 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 26(51) Observations in m: downwards Level Indication Error (mm) MPE (mm) + - 27.9951 27.995 -0.1 ±1 X 25.0054 25.006 0.6 ±1 X 22.0034 22.004 0.6 ±1 X 19.0040 19.004 0.0 ±1 X 16.0030 16.003 0.0 ±1 X 13.0032 13.003 -0.2 ±1 X 10.0045 10.005 0.5 ±1 X 7.0032 7.004 0.8 ±1 X 4.0041 4.005 0.9 ±1 X 1.0025 1.003 0.5 ±1 X 0.7930 0.793 0.0 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 27(51) F.1.4 Array antenna 6” Date of test: 2010-10-11 Observer: Måns Ackerholm Serial No. (See D.4) Test conditions Begin End Time 12:03 12:40 Environmental temperature 21.8 °C 21.8 °C Relative humidity 35 % RH 35 %RH Atmospheric pressure 1021 hPa 1021 hPa Observations in m: upwards Level Indication Error (mm) MPE (mm) + - 0.7992 0.800 0.8 ±1 X 1.0022 1.002 -0.2 ±1 X 4.0026 4.002 -0.6 ±1 X 7.0041 7.004 -0.1 ±1 X 10.0039 10.004 0.1 ±1 X 13.0030 13.003 0.0 ±1 X 16.0004 16.000 -0.4 ±1 X 19.0028 19.003 0.2 ±1 X 22.0001 22.000 -0.1 ±1 X 25.0007 25.000 -0.7 ±1 X 28.0007 28.000 -0.7 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 28(51) Observations in m: downwards Level Indication Error (mm) MPE (mm) + - 28.0007 28.000 -0.7 ±1 X 24.9996 24.999 -0.6 ±1 X 21.9984 21.998 -0.4 ±1 X 18.9997 18.999 -0.7 ±1 X 15.9995 15.999 -0.5 ±1 X 12.9995 12.999 -0.5 ±1 X 10.0000 10.000 0.0 ±1 X 6.9989 6.999 0.1 ±1 X 3.9995 3.999 -0.5 ±1 X 1.0005 1.001 0.5 ±1 X 0.7958 0.796 0.2 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 29(51) F.1.5 Array antenna 8” Date of test: 2010-10-11 Observer: Måns Ackerholm Serial No. (See D.4) Test conditions Begin End Time 13:42 14:22 Environmental temperature 21.8 °C 21.8 °C Relative humidity 34.1 % RH 34.1 %RH Atmospheric pressure 1021 hPa 1021 hPa Observations in m: upwards Level Indication Error (mm) MPE (mm) + - 0.8003 0.800 -0.3 ±1 X 1.0075 1.008 0.5 ±1 X 3.9981 3.998 -0.1 ±1 X 6.9975 6.998 0.5 ±1 X 9.9985 9.999 0.5 ±1 X 12.9999 13.000 0.1 ±1 X 16.0008 16.001 0.2 ±1 X 19.0016 19.002 0.4 ±1 X 22.0011 22.001 -0.1 ±1 X 25.0014 25.001 -0.4 ±1 X 26.5024 26.503 0.6 ±1 X 27.7700 27.770 0.0 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 30(51) Observations in m: downwards Level Indication Error (mm) MPE (mm) + - 27.5088 27.509 0.2 ±1 X 26.4844 26.484 -0.4 ±1 X 25.0000 25.000 0.0 ±1 X 22.0000 22.000 0.0 ±1 X 19.0018 19.002 0.2 ±1 X 16.0022 16.002 -0.2 ±1 X 13.0038 13.004 0.2 ±1 X 10.0035 10.004 0.5 ±1 X 7.0032 7.003 -0.2 ±1 X 4.0033 4.004 0.7 ±1 X 1.0000 1.000 0.0 ±1 X 0.7972 0.797 -0.2 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 31(51) F.1.6 Array antenna 10” Date of test: 2010-10-12 Observer: Måns Ackerholm Serial No. (See D.4) Test conditions Begin End Time 09:07 09:31 Environmental temperature 21.9 °C 21.9 °C Relative humidity 31.2 % RH 31.2 %RH Atmospheric pressure 1018 hPa 1018 hPa Observations in m: upwards Level Indication Error (mm) MPE (mm) + - 0.7907 0.791 0.3 ±1 X 1.0095 1.009 -0.5 ±1 X 3.9962 3.996 -0.2 ±1 X 6.9989 6.999 -0.1 ±1 X 9.9993 9.999 -0.3 ±1 X 12.9984 12.998 -0.4 ±1 X 15.9991 15.999 -0.1 ±1 X 18.9970 18.997 0.0 ±1 X 22.0004 22.000 -0.4 ±1 X 24.9992 24.999 -0.2 ±1 X 27.9970 27.997 0.0 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 32(51) Observations in m: downwards Level Indication Error (mm) MPE (mm) + - 27.9970 27.997 0.0 ±1 X 25.0053 25.005 -0.3 ±1 X 22.0052 22.005 -0.2 ±1 X 19.0035 19.003 -0.5 ±1 X 16.0060 16.006 0.0 ±1 X 13.0034 13.003 -0.4 ±1 X 10.0045 10.004 -0.5 ±1 X 7.0045 7.004 -0.5 ±1 X 4.0054 4.005 -0.4 ±1 X 1.0057 1.006 0.3 ±1 X 0.7926 0.793 0.4 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 33(51) F.1.7 Array antenna 12” Date of test: 2010-10-12 Observer: Måns Ackerholm Serial No. (See D.4) Test conditions Begin End Time 15:02 15:25 Environmental temperature 21.8 °C 21.8 °C Relative humidity 34.5 % RH 34.5 %RH Atmospheric pressure 1020 hPa 1020 hPa Observations in m: upwards Level Indication Error (mm) MPE (mm) + - 0.7951 0.796 0.9 ±1 X 1.0052 1.005 -0.2 ±1 X 4.0005 4.001 0.5 ±1 X 7.0003 7.000 -0.3 ±1 X 9.9980 9.998 0.0 ±1 X 13.0001 13.000 -0.1 ±1 X 16.0005 16.001 0.5 ±1 X 19.0024 19.002 -0.4 ±1 X 22.0006 22.001 0.4 ±1 X 25.0004 25.000 -0.4 ±1 X 27.9997 28.000 0.3 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 34(51) Observations in m: downwards Level Indication Error (mm) MPE (mm) + - 27.9997 28.000 0.3 ±1 X 25.0043 25.004 -0.3 ±1 X 22.0052 22.005 -0.2 ±1 X 19.0009 19.001 0.1 ±1 X 16.0034 16.003 -0.4 ±1 X 13.0027 13.003 0.3 ±1 X 10.0039 10.004 -0.1 ±1 X 7.0040 7.004 0.0 ±1 X 4.0049 4.005 0.1 ±1 X 1.0027 1.003 0.3 ±1 X 0.7939 0.794 0.1 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 35(51) F.1.8 LPG/LNG antenna Date of test: 2010-12-20 Observer: Magnus Herbertsson Serial No. (See D.4) Test conditions Begin End Time 11:36 12:43 Environmental temperature 21.3 °C 21.3 °C Relative humidity 16 % RH 16 %RH Atmospheric pressure 998 hPa 998 hPa Observations in m: upwards Level Indication Error (mm) MPE (mm) + - 0.8003 0.800 -0.3 ±1 X 0.9968 0.997 0.2 ±1 X 4.0074 4.007 -0.4 ±1 X 7.0050 7.005 0.0 ±1 X 10.0054 10.005 -0.4 ±1 X 13.0055 13.005 -0.5 ±1 X 16.0051 16.005 -0.1 ±1 X 19.0054 19.005 -0.4 ±1 X 22.0053 22.005 -0.3 ±1 X 25.0056 25.005 -0.6 ±1 X 28.0055 28.005 -0.5 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 36(51) Observations in m: downwards Level Indication Error (mm) MPE (mm) + - 28.0081 28.008 -0.1 ±1 X 24.9951 24.995 -0.1 ±1 X 21.9950 21.995 0.0 ±1 X 18.9943 18.994 -0.3 ±1 X 15.9943 15.994 -0.4 ±1 X 12.9944 12.995 0.6 ±1 X 9.9948 9.994 -0.8 ±1 X 6.9946 6.995 0.4 ±1 X 3.9949 3.995 0.1 ±1 X 0.9945 0.995 0.5 ±1 X 0.8017 0.802 0.3 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 37(51) F.2 Discrimination (8.1.5.3) Date of test: Observer: Serial No. Test conditions Begin End Time Environmental temperature °C °C Relative humidity % RH %RH Atmospheric pressure hPa hPa Discrimination Level Indication Level change Indication change + - Upwards Downwards Remarks: This test is not applicable since the ALG has no moving liquid detection element. Report number SP-P905645 OIML R 85:2008 (E) Page 38(51) F.3 Hysteris (8.1.5.4) Date of test: Observer: Serial No. Test conditions Begin End Time Environmental temperature °C °C Relative humidity % RH %RH Atmospheric pressure hPa hPa Upwards Level 1 Level 2 Level 3 Maximum hysteresis MPE + - Level up Indication Level down Indication Hysteresis Downwards Level 1 Level 2 Level 3 Maximum hysteresis MPE + - Level down Indication Level up Indication Hysteresis Remarks: This test is not applicable since the ALG has no moving liquid detection element. Report number SP-P905645 OIML R 85:2008 (E) Page 39(51) F.4 Influence factor tests F.4.1 Static environmental temperatures (8.1.6.2) Date of test: Observer: Magnus Holmsten Serial No. (See D.4) 2010-09-15 to 2010-09-17 Test conditions Begin End Time 2010-09-15 at 15:03 2010-09-17 at 13:57 Environmental temperature 21.3 °C 21.2 °C Relative humidity 44.9 % RH 44.6 %RH Atmospheric pressure 989 hPa 989 hPa F.4.1.1 Dry Heat (8.1.6.2.1) Temperature Level (m) Indication (m) Error (mm) MPE (mm) + - 6.9997 6.999 -0.7 ±1 X Reference +20 °C 14.9998 15.000 0.2 ±1 X temperature 23.0005 23.000 -0.5 ±1 X 6.9998 6.999 -0.8 ±1 X High +70 °C 14.9998 15.000 0.2 ±1 X temperature 23.0005 23.000 -0.5 ±1 X 6.9996 6.999 -0.6 ±1 X Reference +20 °C 14.9997 15.000 0.3 ±1 X temperature 23.0005 23.000 -0.5 ±1 X F.4.1.1 Cold (8.1.6.2.2) Temperature Level (m) Indication (m) Error (mm) MPE (mm) + - 6.9997 6.999 -0.7 ±1 X Reference +20 °C 14.9998 15.000 0.2 ±1 X temperature 23.0005 23.000 -0.5 ±1 X 7.0002 7.000 -0.2 ±1 X Low -40 °C 15.0394 15.040 0.6 ±1 X temperature 23.0297 23.030 0.3 ±1 X 6.9998 7.000 0.2 ±1 X Reference +20 °C 14.9998 15.000 0.2 ±1 X temperature 23.0005 23.001 0.5 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 40(51) F.4.2 Mains voltage variations (8.1.6.3 and 8.1.6.4) Date of test: 2010-09-27 Observer: Magnus Holmsten Serial No. (See D.4) Test conditions Begin End Time 10:15 11:03 Environmental temperature 20.9 °C 20.9 °C Relative humidity 40.8 % RH 40.6 %RH Atmospheric pressure 1013 hPa 1012 hPa X DC mains voltage variations (8.1.6.3) X AC mains voltage variations (8.1.6.4) Voltage Level (m) Indication (m) Error (mm) MPE (mm) + - 6.9997 7.000 0.3 ±1 X Reference 240 VAC 16.0004 16.000 -0.4 ±1 X voltage 22.0002 22.000 -0.2 ±1 X 6.9998 7.000 0.2 ±1 X High 265 VAC 16.0002 16.000 -0.2 ±1 X voltage 21.9998 21.999 -0.8 ±1 X 6.9998 7.000 0.2 ±1 X Low 40 VAC 16.0002 16.000 -0.2 ±1 X voltage 21.9998 22.000 0.2 ±1 X 6.9998 7.000 0.2 ±1 X Reference 240 VAC 16.0002 16.000 -0.2 ±1 X voltage 21.9998 22.000 0.2 ±1 X 6.9998 7.000 0.2 ±1 X Reference 36 VDC 16.0002 16.000 -0.2 ±1 X voltage 21.9998 22.000 0.2 ±1 X 6.9998 7.000 0.2 ±1 X High 65 VDC 16.0001 16.000 -0.1 ±1 X voltage 21.9998 21.999 -0.8 ±1 X 6.9998 7.000 0.2 ±1 X Low 20 VDC 16.0002 16.000 -0.2 ±1 X voltage 21.9998 21.999 -0.8 ±1 X 6.9998 7.000 0.2 ±1 X Reference 36 VDC 16.0002 16.000 -0.2 ±1 X voltage 21.9998 21.999 -0.8 ±1 X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 41(51) F.5 Disturbance tests F.5.1 Damp heat cyclic (condensing) (8.1.7.1) Observer: Jörgen Eriksson Serial No. (See D.4) Test conditions Begin End Date 2010-10-18 2010-10-20 Time 13:35 15:30 Atmospheric pressure 1013 hPa 1013 hPa Cycle no. Time Low temp Humidity Time High temp Humidity °C %RH °C %RH 1 9h +25 98 9h +55 93 2 9h +25 98 9h +55 93 Level Indication Error MPE Action checking + - (m) (m) (mm) (mm) facility At start of test 16.992 16.992 0 ±1 None X After cycle no. 1 16.992 16.992 0 ±1 None X After cycle no. 2 16.992 16.992 0 ±1 None X After finish of test 16.992 16.992 0 ±1 None X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 42(51) F.5.2 Radiated, radio frequency, electromagnetic fields (8.1.7.2.1) Date of test: 2009-10-28 Observer: Reine Johansson Serial No. (See D.4) Test conditions Begin End Time 09:02 09:21 Environmental temperature 22.3 °C 22.4 °C Relative humidity 28 % RH 31 %RH Atmospheric pressure 999 hPa 999 hPa EM field Level Indication Error MPE Action + - Frequency Field (m) (m) (mm) (mm) checking strength facility 80-1000 MHz 10 V/m 16.992 16.992 0 ±1 None X Vertical 0° 80-1000 MHz 10 V/m 16.992 16.992 0 ±1 None X Vertical 90° 80-1000 MHz 10 V/m 16.992 16.992 0 ±1 None X Vertical 180° 80-1000 MHz 10 V/m 16.992 16.992 0 ±1 None X Vertical 270° 80-1000 MHz 10 V/m 16.992 16.992 0 ±1 None X Horizontal 270° 80-1000 MHz 10 V/m 16.992 16.992 0 ±1 None X Horizontal 180° 80-1000 MHz 10 V/m 16.992 16.992 0 ±1 None X Horizontal 90° 80-1000 MHz 10 V/m 16.992 16.992 0 ±1 None X Horizontal 0° 1000-2000 MHz 10 V/m 16.992 16.992 0 ±1 None X Vertical 0° 1000-2000 MHz 10 V/m 16.992 16.992 0 ±1 None X Vertical 90° 1000-2000 MHz 10 V/m 16.992 16.992 0 ±1 None X Vertical 180° 1000-2000 MHz 10 V/m 16.992 16.992 0 ±1 None X Vertical 270° 1000-2000 MHz 10 V/m 16.992 16.992 0 ±1 None X Horizontal 270° 1000-2000 MHz 10 V/m 16.992 16.992 0 ±1 None X Horizontal 180° 1000-2000 MHz 10 V/m 16.992 16.992 0 ±1 None X Horizontal 90° Report number SP-P905645 OIML R 85:2008 (E) Page 43(51) EM field Level Indication Error MPE Action + - Frequency Field (m) (m) (mm) (mm) checking strength facility 1000-2000 MHz 10 V/m 16.992 16.992 0 ±1 None X Horizontal 0° Remarks: F.5.3 Conducted, radio frequency, electromagnetic fields (8.1.7.2.2) Date of test: 2009-10-29 Observer: Reine Johansson Serial No. (See D.4) Test conditions Begin End Time 11:00 16:34 Environmental temperature 22.4 °C 21.3 °C Relative humidity 24.3 % RH 23.8 %RH Atmospheric pressure 1007 hPa 1009 hPa Field injected on port(s): AC mains. DC mains. TRL/2 bus. RS-485. Relay. RS-232 EM field Level Indication Error MPE Action checking + - Frequency Amplitude (m) (m) (mm) (mm) facility 150kHz - 10 V 16.992 16.992 0 ±1 None X 80 MHz Remarks: Performed at EUT 2230, 2180 and 2160 specified in section D.4. Report number SP-P905645 OIML R 85:2008 (E) Page 44(51) F.5.4 Electrostatic discharge (8.1.7.2.3) Date of test: 2009-12-01 Observer: Bengt Andersson Serial No. (See D.4) Test conditions Begin End Time 12:50 14:18 Environmental temperature 22.3 °C 21.4 °C Relative humidity 31.3 % RH 32.5 %RH Atmospheric pressure 1011 hPa 1002 hPa X Contact discharge X Direct application Test voltage C/A 6/8 kV X Air discharge X Indirect application Number of discharges 10 Discharge Level Indication Error MPE Action checking + - applied on (m) (m) (mm) (mm) facility 2240S 16.992 16.992 0 ±1 None X 5900S 16.992 16.992 0 ±1 None X 2230 16.992 16.992 0 ±1 None X 2410 16.992 16.992 0 ±1 None X 2160+2180 16.992 16.992 0 ±1 None X Remarks: The time between the discharges was 1 second. 2240S has no legal metrology function related to R85. Only connected to show no influence on level measuring results. Report number SP-P905645 OIML R 85:2008 (E) Page 45(51) F.5.5 Bursts on signal, data and control lines (8.1.7.2.4) Date of test: 2009-11-04 Observer: Bengt Andersson Serial No. (See D.4) Test conditions Begin End Time 10:40 13:55 Environmental temperature 22.3 °C 22.4 °C Relative humidity 25.2 % RH 24.1 %RH Atmospheric pressure 1002 hPa 1010 hPa Test voltage 2 kV Repetition rate 5 kHz Discharge Level Indication Error MPE Action checking + - applied on (m) (m) (mm) (mm) facility 2240S 16.992 16.992 0 ±1 None X 5900S 16.992 16.992 0 ±1 None X 2230 16.992 16.992 0 ±1 None X 2410 16.992 16.992 0 ±1 None X 2160+2180 16.992 16.992 0 ±1 None X Remarks: 2240S has no legal metrology function related to R85. Only connected to show no influence on level measuring results. Report number SP-P905645 OIML R 85:2008 (E) Page 46(51) F.5.6 Surges on signal, data and control lines (8.1.7.2.5) Date of test: 2009-11-06 Observer: Bengt Andersson Serial No. (See D.4) Test conditions Begin End Time 13:32 15:51 Environmental temperature 21.7 °C 22.4 °C Relative humidity 28.4 % RH 24.9 %RH Atmospheric pressure 997 hPa 1002 hPa Surge Voltage Number Level Indication Error MPE Action + - applied on and of (m) (m) (mm) (mm) checking polarity surges facility 2240S ± 0.5 kV 5 16.992 16.992 0 ±1 None X ± 1.0 kV 5 16.992 16.992 0 ±1 None X 5900S ± 0.5 kV 5 16.992 16.992 0 ±1 None X ± 1.0 kV 5 16.992 16.992 0 ±1 Note 1 X 2230 ± 0.5 kV 5 16.992 16.992 0 ±1 None X ± 1.0 kV 5 16.992 16.992 0 ±1 Note 1 X 2410 ± 0.5 kV 5 16.992 16.992 0 ±1 None X ± 1.0 kV 5 16.992 16.992 0 ±1 None X 2160+2180 ± 0.5 kV 5 16.992 16.992 0 ±1 None X ± 1.0 kV 5 16.992 16.992 0 ±1 None X Remarks: Note 1: The test object restarted automatically, a message on the AE has to be granted 2240S has no legal metrology function related to R85. Only connected to show no influence on level measuring results. Report number SP-P905645 OIML R 85:2008 (E) Page 47(51) F.5.7 AC mains voltage dips, short interruptions and voltage variations (8.1.7.2.6) Date of test: 2009-11-04 Observer: Bengt Andersson Serial No. (See D.4) Test conditions Begin End Time 15:00 16:43 Environmental temperature 22.4 °C 22.2 °C Relative humidity 25.4 % RH 22.1 %RH Atmospheric pressure 1005 hPa 977 hPa Mains voltage and frequency 48, 65, 120, 230 V 50Hz Test Reduction Duration Level Indication Error MPE Action + - to (cycles) (m) (m) (mm) (mm) checking facility Dips a 0% 0.5 16.992 16.992 0 ±1 None X Dips b 0% 1 16.992 16.992 0 ±1 None X Dips c 40 % 10 16.992 16.992 0 ±1 None X Dips d 70 % 25 16.992 16.992 0 ±1 None X Dips e 80 % 250 16.992 16.992 0 ±1 None X Short 0% 5s 16.992 16.992 0 ±1 None X interruption Remarks: Tests performed on 2410. 2160 and 2180. Report number SP-P905645 OIML R 85:2008 (E) Page 48(51) F.5.8 Bursts (transients) on AC and DC mains (8.1.7.2.7) Date of test: 2009-11-04 Observer: Bengt Andersson Serial No. (See D.4) Test conditions Begin End Time 10:40 13:55 Environmental temperature 22.3 °C 22.4 °C Relative humidity 25.2 % RH 24.1 %RH Atmospheric pressure 1002 hPa 1010 hPa Test voltage 2 kV Repetition rate 5 kHz Polarity ± Exposure Level Indication Error MPE Action checking + - time (m) (m) (mm) (mm) facility +0.5 1 min 16.992 16.992 0 ±1 None X -0.5 1 min 16.992 16.992 0 ±1 None X +1.0 1 min 16.992 16.992 0 ±1 None X -1.0 1 min 16.992 16.992 0 ±1 None X +2.0 1 min 16.992 16.992 0 ±1 None X -2.0 1 min 16.992 16.992 0 ±1 None X Remarks: Tests performed on units 2410. 2160 and 2180. Report number SP-P905645 OIML R 85:2008 (E) Page 49(51) F.5.9 Voltage dips, short interruptions and voltage variations on DC mains (8.1.7.2.8) Date of test: 2009-12-02 Observer: Bengt Andersson Serial No. (See D.4) Test conditions Begin End Time 10:45 15:30 Environmental temperature 22.1 °C 22.4 °C Relative humidity 22.9 % RH 23.0 %RH Atmospheric pressure 994 hPa 998 hPa Test at 21 V: Test Reduction Duration Level Indication Error MPE Action + - to (s) (m) (m) (mm) (mm) checking facility Voltage dips 40 % 0.1 16.992 16.992 0 ±1 Note 1 X 70 % 0.1 16.992 16.992 0 ±1 None X Short 0% 0.01 16.992 16.992 0 ±1 Note 1 X interruptions “High imp” 0% 0.01 16.992 16.992 0 ±1 Note 1 X “Low imp” Voltage 85% 10 16.992 16.992 0 ±1 None X variations 120% Tested only at 60 V Remarks: Note 1: Power off. start-up after ~30 s, no influence on measured value (display: 16.992 m) DC mains, rated input power 21-60 VDC Test at 60 V: Test Reduction Duration Level Indication Error MPE Action + - to (s) (m) (m) (mm) (mm) checking facility Voltage dips 40 % 0.1 16.992 16.992 0 ±1 None X 70 % 0.1 16.992 16.992 0 ±1 None X Short 0% Tested only at 21 V interruptions Voltage 85% 10 16.992 16.992 0 ±1 None X variations 120% 10 16.992 16.992 0 ±1 None X Remarks: DC mains, rated input power 21-60 VDC Report number SP-P905645 OIML R 85:2008 (E) Page 50(51) F.5.10 Ripple on DC mains power (8.1.7.2.9) Date of test: 2009-12-02 Observer: Bengt Andersson Serial No. (See D.4) Test conditions Begin End Time Environmental temperature 22.1 °C 22. 0 °C Relative humidity 22.9 % RH 24.0 %RH Atmospheric pressure 994 hPa 995 hPa Frequency 100 Hz Test time 10 min Voltage Level Indication Error MPE Action checking + - (m) (m) (mm) (mm) facility 24 VDC 16.992 16.992 0 ±1 None X 48 VDC 16.992 16.992 0 ±1 None X Remarks: Report number SP-P905645 OIML R 85:2008 (E) Page 51(51) F.5.11 Surges on AC and DC mains power lines (8.1.7.2.10) Date of test: 2009-11-05 Observer: Bengt Andersson Serial No. (See D.4) Test conditions Begin End Time 13:50 16:13 Environmental temperature 23.1 °C 21.4 °C Relative humidity 27.8 % RH 24.9 %RH Atmospheric pressure 1001 hPa 977 hPa Angle Voltage Number Level Indicatio Error MPE Action + - and of (m) n (mm) (mm) checking polarity surges (m) facility 0° 1 kV 3+3 16.992 16.992 0 ±1 None X Line 90° 1 kV 3+3 16.992 16.992 0 ±1 None X to 180° 1 kV 3+3 16.992 16.992 0 ±1 None X line 270° 1 kV 3+3 16.992 16.992 0 ±1 None X 0° 2 kV 3+3 16.992 16.992 0 ±1 None X Line 90° 2 kV 3+3 16.992 16.992 0 ±1 None X to 180° 2 kV 3+3 16.992 16.992 0 ±1 None X earth 270° 2 kV 3+3 16.992 16.992 0 ±1 None X Remarks Tests were performed on 2410, 2160 and 2180. Unit 2160 was modified to pass the test. Saatja: Ando Vendel / ProREX <[email protected]> Saaja: Triinu Sillamaa Teema: Tüübikinnituse taotlemine Raptor süsteemile Lugupeetud Triinu Sillamaa, Soovime taodelda tüübikinnitust mahutite radarnivoomõõtesüsteemile Raptor (firmalt Rosemount Tank Radar). Manusena lisatud esialgne dokumentatsioon. Palun andke tagasisisdet, kas see on piisav tüübikinnituse saamiseks või on vaja veel midagi juurde. Kui palju dokumentatsiooni on vaja füüsiliselt paberile välja printida? Kas piisab, kui mahukama tehnilise dokumentatsiooni (kataloogid, juhendid, joonised jms) esitame CD-l? Midagi veel? PS! Varasemast olemas meil tüübikinnitused (TJI 5-3/15 31.02; TJI 5-3/15 32.02; TJI 5.3-1/15.11.07, mis paraku küll juba aegunud) sama tootja eelmise põlvkonna samalaadsetele toodetele, mis osaliselt kattuvad ka uue süsteemiga. Ette tänades! Lugupidamisega, Ando Vendel Projektijuht ProREXoil OÜ Peterburi tee 47 11415 Tallinn Tel.: +372 60 624 60 Fax: +372 60 624 61 GSM: +372 56 900 810 e-mail: [email protected] <mailto:[email protected]> www.prorexoil.ee <http://www.prorexoil.ee/> Olge valmis uueks ajastuks mahutite mõõtmises! Tutvuge lähemalt Raptor süsteemiga » <http://www2.emersonprocess.com/en-US/brands/rosemounttankgauging/Pages/ index.aspx>
Allikas: Tarbijakaitse ja Tehnilise Järelevalve Amet dokumendiregister →
dokumendiregister.eeAsutusedEesti avalike dokumendiregistrite otsing · nimistu.ee andmetel