Report number SP-P905645 OIML R 85:2008 (E) Page 1(51)
TYPE APPROVAL
ROSEMOUNT RAPTOR TANK GAUGING SYSTEM
A. AUTHORITY RESPONSIBLE FOR THIS REPORT .......................................3
B. SYNOPSIS OF THE RESULTS OF THE EXAMINATIONS AND TESTS....4
C. SUMMARY OF THE RESULTS OF THE EXAMINATIONS AND TESTS ..5
C.1 Examinations .......................................................................................................5
C.2 Performance tests ................................................................................................6
D. GENERAL INFORMATION .............................................................................7
D.1 Manufacturer .......................................................................................................7
D.2 Applicant .............................................................................................................7
D.3 Testing laboratories involved in the tests ............................................................8
D.4 General information concerning the type and the sample(s) submitted for the tests
(as stated on the instrument / provided by the manufacturer) .............................................9
D.5 Accessories supplied by the applicant ................................................................10
D.6 Selection of sample(s) tested...............................................................................10
D.7 Adjustments and modifications ...........................................................................11
D.8 Choice of the manufacturer concerning the severity levels ................................11
D.9 Full details about the electrical power requirements...........................................11
D.10 Additional information concerning the type .......................................................11
D.11 Documentation supplied by the applicant ...........................................................12
D.12 Information concerning the test equipment used for the type evaluation ...........12
E. EXAMINATION ................................................................................................13
E.1 Constituents (4) ...................................................................................................13
E.2 Units of measurements (5) ..................................................................................14
E.3 Rated operating conditions (6.1) .........................................................................14
E.4 Special conditions (6.1).......................................................................................15
E.5 Indicating device (7.1) ........................................................................................15
E.6 Printer(s) (7.1.9) ..................................................................................................16
E.7 Movable sensor (7.2)...........................................................................................16
E.8 Installation (7.3) ..................................................................................................17
E.9 Ancillary devices (7.4) ........................................................................................17
E.10 Markings (7.5) .....................................................................................................18
E.11 Verification marks (7.6) ......................................................................................18
E.12 Mechanical sealing (7.7) .....................................................................................19
E.13 Electronic sealing (7.7) .......................................................................................19
E.14 Safeguarding the integrity of the measurement (7.8) ..........................................20
F. PERFORMANCE TESTS ..................................................................................21
F.1 Accuracy (8.1.5.2) ...............................................................................................21
F.1.1 Parabolic antenna ................................................................................................21
Report number SP-P905645 OIML R 85:2008 (E) Page 2(51)
F.1.2 Horn antenna .......................................................................................................23
F.1.3 Array antenna 5” .................................................................................................25
F.1.4 Array antenna 6” .................................................................................................27
F.1.5 Array antenna 8” .................................................................................................29
F.1.6 Array antenna 10” ...............................................................................................31
F.1.7 Array antenna 12” ...............................................................................................33
F.1.8 LPG/LNG antenna ..............................................................................................35
F.2 Discrimination (8.1.5.3) ......................................................................................37
F.3 Hysteris (8.1.5.4) .................................................................................................38
F.4 Influence factor tests ...........................................................................................39
F.4.1 Static environmental temperatures (8.1.6.2) .......................................................39
F.4.2 Mains voltage variations (8.1.6.3 and 8.1.6.4) ....................................................40
F.5 Disturbance tests .................................................................................................41
F.5.1 Damp heat cyclic (condensing) (8.1.7.1) ............................................................41
F.5.2 Radiated, radio frequency, electromagnetic fields (8.1.7.2.1) ............................42
F.5.3 Conducted, radio frequency, electromagnetic fields (8.1.7.2.2) .........................43
F.5.4 Electrostatic discharge (8.1.7.2.3) .......................................................................44
F.5.5 Bursts on signal, data and control lines (8.1.7.2.4) .............................................45
F.5.6 Surges on signal, data and control lines (8.1.7.2.5) ............................................46
F.5.7 AC mains voltage dips, short interruptions and voltage variations (8.1.7.2.6) ...47
F.5.8 Bursts (transients) on AC and DC mains (8.1.7.2.7) ..........................................48
F.5.9 Voltage dips, short interruptions and voltage variations on DC mains (8.1.7.2.8) 49
F.5.10 Ripple on DC mains power (8.1.7.2.9) ...............................................................50
F.5.11 Surges on AC and DC mains power lines (8.1.7.2.10) .......................................51
Report number SP-P905645 OIML R 85:2008 (E) Page 3(51)
A. AUTHORITY RESPONSIBLE FOR THIS REPORT
Name SP
Address Brinellgatan 4. Box 857. SE-501 15 Borås. Sweden
Report No. SP-P905645
Application No. 2010-06-10RosemountTankRadarAB
Date/period of tests 2009-08-01 – 2010-12-31
Date of issuing this report 2011-04-12
Name and signature of the Måns Ackerholm, SP Measurement Technology
responsible person
Stamp(s)
(if applicable)
Report number SP-P905645 OIML R 85:2008 (E) Page 4(51)
B. SYNOPSIS OF THE RESULTS OF THE EXAMINATIONS AND TESTS
The tested specimen fulfills ALL the applicable requirements in OIML R 85 YES NO
(2008) X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 5(51)
C. SUMMARY OF THE RESULTS OF THE EXAMINATIONS AND TESTS
C.1 Examinations
For details, refer to the examination: Clause E of this report as indicated in the first column
Sub- Test (with reference to requirements in + - Remarks Page
clause Part 1)
E.1 Constituents (4) X 13
E.2 Units of measurement (5) X 14
E.3 Rated operating conditions (6.1) X 14
E.4 Special conditions (6.1) X 15
E.5 Indicating device (7.1) X 15
E.6 Printer(s) (7.1.9) X 16
E.7 Moveable sensor (7.2) X 16
E.8 Installation (7.3) X 17
E.9 Ancillary devices (7.4) X 17
E.10 Markings (7.5) X 18
E.11 Verification marks (7.6) X 18
E.12 Mechanical sealing (7.7) X 19
E.13 Electronic sealing (7.8) X 19
E.14 Safeguarding the integrity of the X 20
measurement (7.9)
Report number SP-P905645 OIML R 85:2008 (E) Page 6(51)
C.2 Performance tests
For details, refer to the tests: clause F in this report as indicated in first column
Sub- Test (with reference to requirements in + - Remarks Page
clause Part 1)
F.1 Accuracy (8.1.5.2) X 21-36
F.2 Discrimination (8.1.5.3) Not applicable (37)
F.3 Hysteresis (8.1.5.4) Not applicable (38)
F.4.1 Temperature tests (8.1.6.2) X 39
F.4.2 Main voltage variations (8.1.6.3 + 8.1.6.4) X 40
F.5.1 Damp heat cyclic (condensing) (8.1.7.1) X 41
F.5.2 Radiated radio-frequency electromagnetic X 42
fields (8.1.7.2.1)
F.5.3 Conducted radio-frequency electromagnetic X 43
fields (8.1.7.2.2)
F.5.4 Electrostatic discharge (8.1.7.2.3) X 44
F.5.5 Bursts on signal. data and control lines X 45
(8.1.7.2.4)
F.5.6 Surges on signal. data and control lines X 46
(8.1.7.2.5)
F.5.7 AC mains voltage dips. short interruptions X 47
and voltage variations (8.1.7.2.6)
F.5.8 Bursts (transients) on AC and DC mains X 48
(8.1.7.2.7)
F.5.9 Voltage dips. short interruptions and voltage X 49
variations on DC mains power (8.1.7.2.8)
F.5.10 Ripple on DC mains power (8.1.7.2.9) X 50
F.5.11 Surges on AC and DC mains power lines X 51
(8.1.7.2.10)
Report number SP-P905645 OIML R 85:2008 (E) Page 7(51)
D. GENERAL INFORMATION
D.1 Manufacturer
Company Emerson Process Management, Rosemount Tank Radar AB
Address Box 130 45. SE-402 51 Göteborg. Sweden
D.2 Applicant
Company Emerson Process Management, Rosemount Tank Radar AB
Representative N/A
Address Box 130 45. SE-402 51 Göteborg. Sweden
Reference Christian Skaug, Technical product manager
Date of Application 2010-06-10
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 8(51)
D.3 Testing laboratories involved in the tests
Name SP Swedish technical research institute
Address Brinellgatan 4. SE-504 62 Borås. Sweden
Application No. 2010-06-10RosemountTankRadarAB
Tests by this laboratory All tests
Date/period of tests 2009-08-01-2010-12-31
Name(s) of test engineer(s) Måns Ackerholm
Jörgen Eriksson
Magnus Holmgren
Magnus Herbertsson
Mattias Engström
Bengt Andersson
Joakim Lindeblom
Reine Johansson
Accredited by SWEDAC
Accreditation includes R85 Yes
Details of relevant peer assessment or N/A
assessment by other means
In case tests were performed at another Test F1 and F4 were performed at Rosemount
location than the premises of this laboratory. Tank Radar AB, in their measuring range (see
give details here D.12).
Name of the responsible person Måns Ackerholm
Date of signature 2011-04-08
Stamp (if applicable) and signature of the
responsible person
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 9(51)
D.4 General information concerning the type and the sample(s)
submitted for the tests (as stated on the instrument / provided by the
manufacturer)
Manufacturer’s Rosemount Tank Radar AB
trademark/corporate name
Year of manufacture 2009
Type designation Raptor Tank Gauging System
Model No. - 5900S Radar level gauge with:
Parabolic antenna 1P
Horn antenna 1H
Still-pipe array antenna 1A5
Still-pipe array antenna 1A6
Still-pipe array antenna 1A8
Still-pipe array antenna 1AA
Still-pipe array antenna 1AB
LPG/LNG antenna G1/G2
- 2410 Tank Hub
- 2230 Graphical Field Display
- 2160/61 Field Communication Unit
- 2180 Field Bus Modem
- General purpose PC with TankMaster for display of measured
level
- Printer. HP LaserJet 2100M
Serial number(s) 5900S: SP marked SPL 62/09
of the samples 2410: SP marked SPL 64/09
2230: SP marked 65/09
2160/61: Serial No. 6109
2180: SPL 80/09
PC: SAAB-SME-1609 and SAAB-SME-1327
Measuring For all antenna types: From 0.8 to 30 meter. Distances
range(s) measured from installation flange.
Electrical power Power supply to 2410 Tank Hub:
24- 48 VDC or 48- 240 VAC 50/60 Hz
5900S and 2230 are powered from 2410 with intrinsically safe
power.
Identification of software Version number(s) 5900S: 1.A8 (0xCFD5)
2410: 1.A4
2230: 1.A1
Checksum/identification TankMaster: 6.xx
Report number SP-P905645 OIML R 85:2008 (E) Page 10(51)
Remarks:
D.5 Accessories supplied by the applicant
Operating instructions 5900S Radar level gauge Reference Manual
2410 TankHub Reference Manual
2230 Graphical Field Display Reference Manual
Rosemount Raptor Tank Gauging System
Data printer Network printer
Cables Cables for connection of system units
Ancillary facilities Generic FF device that fulfills ITK-tests
(such as 2240S Multi-input Temperature
Transmitter)
Power supply unit DC power supply for 2410
Other accessories/remarks:
1. 2240S has no legal metrology function related to R85. Only connected to show no
influence on level measuring results.
2. Coaxial cable SP number SPL 215/99 used as target under disturbance tests.
D.6 Selection of sample(s) tested
The 5900S Radar Level Gauge is available with different antennas for different tank types. See
D4. Accuracy tests were performed with a separate test protocol for each antenna type and each
antenna type has to be within MPE.
Two different electrical interfaces between 2410 and FCU2160 were be tested.
One consists of a standard RS485 interface and the other consists of an FSK modulated signal
(the so called “TRL/2 interface”). Both results were registered on two separate PC’s connected
to FCU2160.
Justification for the selection of the samples:
The sample of the 5900S Radar Level Gauge was selected randomly from 3 produced units by
SP.
The 5900S unit was engraved with identifier for safe identification during the whole test.
Other units as 2410, 2230 etc has no measuring function, but instead a special function such as:
communication of data. power supply or digital display of measured value. These were selected
from available units for the test and marked with identifier as above.
Report number SP-P905645 OIML R 85:2008 (E) Page 11(51)
D.7 Adjustments and modifications
FCU 2160 was modified for test F.5.11
D.8 Choice of the manufacturer concerning the severity levels
Influence Severity level
Dry Heat (1, 2, 3, 4): 4
Cold (1, 2, 3, 4): 4
DC mains voltage High Level: -15%, +10%
D.9 Full details about the electrical power requirements
24-48 VDC or 48-240 VAC, 50/60 Hz
Max 20W depending on configuration
D.10 Additional information concerning the type
Additional remarks and/or information (connection equipment. interfaces etc)
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 12(51)
D.11 Documentation supplied by the applicant
Technical manuals and descriptions
Name Ref. no Revision
Technical Description 704 010 EN Rev AA, May 2010
Section 1. Raptor system
Section 2. 5900S series Radar Level Gauge
Section 3. 2240S Multi-input Temperature
Transmitter
Section 6. 2230 Graphical Field Display
Section 7. 2410 Tank Hub
Section 8. 2160 Field Communication Unit
Section 10. 2180 Field Bus Modem
Raptor System Revision History DE-0056 Issue 8d
Raptor System Configuration Manual 300 510 EN Rev AA, June 2010
5900 Reference Manual 300 520 EN Rev AA, June 2010
2410 Reference Manual 300 530 EN Rev AA, June 2010
2230 Reference Manual 300 560 EN Rev AA, June 2010
2240 Reference Manual 300 550 EN Rev AA, June 2010
D.12 Information concerning the test equipment used for the type
evaluation
1) Leica Lasertracker SP 602084, calibrated at SP.
2) Measuring Range M2. Located at Rosemount Tank Radar AB. Calibrated against Laser
interferometer (SP inventory no. 601046). Evaluated at SP with an expanded
measurement uncertainty of ±0.06mm. Report MTmP904224-K01.
3) Keithley 2700, s/n 09224830, calibrated at SP
4) Measurement equipment for EMC tests, routinely used for accredited methods
5) Measurement equipment and chambers for environmental tests routinely, used at
accredited methods
6) Various enviromental indicators (temperature, pressure, humidity), accredited calibrated
at SP
Report number SP-P905645 OIML R 85:2008 (E) Page 13(51)
E. EXAMINATION
E.1 Constituents (4)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Liquid level sensor Microwave antennas, type varies with application. Non-
movable type
Transducer 5900S, electronic type, emits microwave sweep, type
FMCW.
Correction sensor 1. 2240S with temperature sensor for correction of
gauge reference height.
2. 2051 pressure sensor for calculation of vapor density
in LPG tanks.
Calculator 1. In 5900S: Calculates level from microwave signal,
can correct with data from correction sensor(s)
2. Tankmaster software in PC: Calculates volumes and
mass from measured and operator entered data.
Indicating device(s) 1. 2230 Grafical Field Display, typically installed at tank
2. 2410 TankHub Display, optional field display in Tank
Hub
3. Tankmaster software in PC, typically installed in
control room
Printer Standard printer
Ancillary devices Various FF devices that fulfills ITK-tests can be connected to
the Tank Bus
Checking facilities Automatic checking facility of both Permanent and
Intermittent type
Others FCU 2160/61, FBM 2180
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 14(51)
E.2 Units of measurements (5)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Display Print
Dip m m
Ullage m m
Other Indication(s) #-sign = legal value #-sign
(2230 and Tankmaster)
*-sign = legal value (2410)
Remarks: Displays and printers are programmable to show meter, mm, ft-inch-1/16, feet
decimal. Normal mode for legal tank gauging system is meter.
E.3 Rated operating conditions (6.1)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Ambient temperatures Low -40 °C
High +70 °C
Relative humidity up to 100%
DC mains voltage 24-48 VDC (-15%...+10%)
AC mains voltage 48-240 VAC (-15%...+10%)
Temperature extreme values Liquid N/A antenna not in contact with liquid
Medium above -40 – 230°C
the liquid (LPG/LNG antenna -170°C to 90°C, 55°C –
90°C close to ball-valve)
Pressure extreme values Varies with antenna type -0.2 – 2 bar (horn- and still-pipe antenna), -
0.2 -10 bar (parabolic antenna), up to 25 bar (LPG antenna)
Liquid characteristics N/A, liquid does not affect measurement performance
Liquid density. extreme N/A, liquid density does not affect measurement performance
values
Medium characteristics Typical petroleum products
Medium density. extreme
values
Remarks:
Normal petroleum products have no measurable influence on performance. Certain special products
(e.g. ketons) may influence velocity of the radar wave and may result in performance outside m.p.e.
See manufacturer’s recommendation on which products to avoid. Tank design with internal floating
roof may reduce influence for critical products.
Report number SP-P905645 OIML R 85:2008 (E) Page 15(51)
E.4 Special conditions (6.1)
Date: Observer: Serial No.
Remarks:
E.5 Indicating device (7.1)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Sub-clause Requirement Remarks + -
7.1.1 Distance between analog scale marks N/A, digital indicating device
7.1.2 Number of indicating devices 4 types (incl printer), see D.4 X
7.1.3 Common indicating devices Tankmaster screen and Printers X
7.1.4 Remote indication duly identified Yes X
7.1.2-3-4 Other indicating devices See D.4 X
7.1.2 Alarm when outside limits Yes, flagged as not approved X
value
7.1.5 Default display of dip / ullage Dip X
7.1.6 Display of measurement Yes X
7.1.7 Symbol or name of unit present Yes X
Scale interval N/A, digital indicating device
7.1.8 Digital display Yes X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 16(51)
E.6 Printer(s) (7.1.9)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Sub-clause Requirement Remarks + -
7.1.2 Number of printers None, one or more X
7.1.4 Printer duly identified to Tank name to which a certain ALG X
ALG is configured is printed on the same
row as a specific level value for
indication
7.1.5 Default printer of dip/ullage Dip X
7.1.6 Print of measurement Yes X
7.1.7 Symbol or name of unit Yes X
present
Scale interval N/A, digital indicating device
Remarks:
E.7 Movable sensor (7.2)
Date: Observer: Serial No.
Sub-clause Requirement Remarks + -
7.2.1 Suspension mechanism
7.2.2 Static position
Remarks:
Not applicable
Report number SP-P905645 OIML R 85:2008 (E) Page 17(51)
E.8 Installation (7.3)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Sub-clause Requirement Remarks + -
7.3.1.1 Accessible and legible Yes X
7.3.1.2 Verification possible Yes X
7.3.1.3 Gauge hatch
7.3.1.3 No obstacles
7.3.1.4 Affect ALG measurement
Installation to be in accordance
7.3.1.5 Influence eddies etc
with installation drawings.
7.3.1.6 Compensation movement
tank
7.3.1.7 Location correction sensor
7.3.1.8 Thermal expansion Supported in the software X
Remarks:
E.9 Ancillary devices (7.4)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Description Remarks + -
2240S Multi-input Temperature Transmitter X
Generic FF device that fulfills ITK-test X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 18(51)
E.10 Markings (7.5)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Requirements Remarks + -
Locations of the markings Top or side on each unit X
Name of the manufacturer / trademark Rosemount Tank Radar AB X
Type designation 5900S, 2410 X
Serial number Yes X
Year of manufacture Yes X
Type approval mark (provisions) Yes (space available on X
complementary plate)
Ranges defining the field of operation Yes X
Visibility Yes X
Additional information (if required)
Remarks:
National marking requirements will be taken into consideration for each country.
E.11 Verification marks (7.6)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Requirements Remarks + -
Location Top or side on each unit X
Fit for easy application Yes X
Impossible to remove without damage Yes X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 19(51)
E.12 Mechanical sealing (7.7)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Requirements Remarks + -
Data plate Yes X
Other components Housings: Yes X
Impossible to remove without damage Yes X
Remarks:
Various solutions possible. Filling or wire solutions
E.13 Electronic sealing (7.7)
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Requirements Remarks + -
Access Using Tankmaster WinSetup, access level X
as administrator
Password changeable Yes X
Configuration mode Indicated by write protection X
Event counter Yes X
Date of change parameter Yes X
Value of parameter Yes X
Identification authorized person Yes X
Traceability last intervention Yes X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 20(51)
E.14 Safeguarding the integrity of the measurement (7.8)
Date: 2011-03-15 Observer: Måns Ackerholm Serial No. (See D.4)
Sub-clause Requirement Remarks + -
7.8.1 Possibility of testing checking Can’t be disabled. #-sign indicate X
facility “no error”
Checking facility type I or P Background Diagnostics are type I X
where they are asynchronous with
the actual measurement. The #sign
checking against limits, write protect
etc are of type P, where the check is
done prior the data for #-sign is
delivered.
7.8.2.1 Checking facility disturbances OK X
7.8.2.3 Durability checking facility OK X
7.8.2.5 Action by checking facility Removal of #-sign X
7.8.3.1 Protection of data OK X
7.8.3.2 Integrity of permanent stored CRC-16 or checksum X
instructions
7.8.3.3 Transfer/storage of data CRC-16 X
7.8.3.4 Checking facility calculator Both software and hardware types of X
wathdogs
7.8.3.5 Checking facility indicating OK X
device
7.8.3.6 Checking facility ancillary device Checked by 2410 Tank Hub X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 21(51)
F. PERFORMANCE TESTS
F.1 Accuracy (8.1.5.2)
F.1.1 Parabolic antenna
Date: 2010-10-15 Observer: Måns Ackerholm Serial No. (See D.4)
Test conditions Begin End
Time 09:30 10:45
Environmental temperature 19.5 °C 19.6 °C
Relative humidity 56.9 % RH 56,8 %RH
Atmospheric pressure 995 hPa 992 hPa
Observations in m: upwards
Level Indication Error (mm) MPE (mm) + -
0.7998 0.800 0.2 ±1 X
1.0001 1.000 -0.1 ±1 X
3.9999 4.000 0.1 ±1 X
6.9996 6.999 -0.6 ±1 X
10.0003 10.001 0.7 ±1 X
12.9994 13.000 0.6 ±1 X
16.0004 16.001 0.6 ±1 X
18.9996 19.000 0.4 ±1 X
22.0001 22.000 -0.1 ±1 X
24.9998 25.000 0.2 ±1 X
27.9999 28.000 0.1 ±1 X
29.5003 29.500 -0.3 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 22(51)
Observations in m: downwards
Level Indication Error (mm) MPE (mm) + -
29.5003 29.500 -0.3 ±1 X
27.9997 28.000 0.3 ±1 X
25.0001 25.000 -0.1 ±1 X
22.0001 22.000 -0.1 ±1 X
18.9998 19.000 0.2 ±1 X
16.0004 16.001 0.6 ±1 X
12.9995 13.000 0.5 ±1 X
10.0005 10.001 0.5 ±1 X
6.9996 6.999 0.3 ±1 X
4.0003 4.000 -0.3 ±1 X
1.0000 1.000 0.0 ±1 X
0.7998 0.800 0.2 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 23(51)
F.1.2 Horn antenna
Date of test: 2010-09-27 Observer: Magnus Holmsten Serial No. (See D.4)
Test conditions Begin End
Time 09:25 09:55
Environmental temperature 20.8 °C 20.5 °C
Relative humidity 41.8 % RH 41.5 %RH
Atmospheric pressure 1012 hPa 1012 hPa
Observations in m: upwards
Level Indication Error (mm) MPE (mm) + -
0.8002 0.800 -0.2 ±1 X
1.0001 1.000 -0.1 ±1 X
4.0000 4.000 0.0 ±1 X
6.9997 7.000 0.3 ±1 X
10.0003 10.001 0.7 ±1 X
12.9994 12.999 -0.4 ±1 X
16.0004 16.000 -0.4 ±1 X
18.9996 18.999 -0.6 ±1 X
22.0002 22.000 -0.2 ±1 X
24.9998 24.999 -0.8 ±1 X
27.9999 27.999 -0.9 ±1 X
29.5003 29.500 -0.3 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 24(51)
Observations in m: downwards
Level Indication Error (mm) MPE (mm) + -
29.5003 29.500 -0.3 ±1 X
27.9998 27.999 -0.8 ±1 X
25.0001 25.000 -0.1 ±1 X
22.0001 22.000 -0.1 ±1 X
18.9998 19.000 -0.8 ±1 X
16.0004 16.000 -0.4 ±1 X
12.9995 12.999 -0.5 ±1 X
10.0006 10.001 0.4 ±1 X
6.9996 7.000 0.4 ±1 X
4.0002 4.000 -0.2 ±1 X
1.0000 1.000 0.0 ±1 X
0.7998 0.800 0.2 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 25(51)
F.1.3 Array antenna 5”
Date of test: 2010-10-12 Observer: Måns Ackerholm Serial No. (See D.4)
Test conditions Begin End
Time 11:02 11:54
Environmental temperature 22.2 °C 22.2 °C
Relative humidity 31.5 % RH 31.5%RH
Atmospheric pressure 1018 hPa 1018 hPa
Observations in m: upwards
Level Indication Error (mm) MPE (mm) + -
0.7933 0.793 -0.3 ±1 X
1.0068 1.007 0.2 ±1 X
3.9991 4.000 0.9 ±1 X
6.9986 6.999 0.4 ±1 X
9.9996 10.000 0.4 ±1 X
12.9979 12.998 0.1 ±1 X
15.9978 15.998 0.2 ±1 X
18.9965 18.996 -0.5 ±1 X
21.9978 21.998 0.2 ±1 X
24.9975 24.998 0.5 ±1 X
27.9973 27.998 0.7 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 26(51)
Observations in m: downwards
Level Indication Error (mm) MPE (mm) + -
27.9951 27.995 -0.1 ±1 X
25.0054 25.006 0.6 ±1 X
22.0034 22.004 0.6 ±1 X
19.0040 19.004 0.0 ±1 X
16.0030 16.003 0.0 ±1 X
13.0032 13.003 -0.2 ±1 X
10.0045 10.005 0.5 ±1 X
7.0032 7.004 0.8 ±1 X
4.0041 4.005 0.9 ±1 X
1.0025 1.003 0.5 ±1 X
0.7930 0.793 0.0 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 27(51)
F.1.4 Array antenna 6”
Date of test: 2010-10-11 Observer: Måns Ackerholm Serial No. (See D.4)
Test conditions Begin End
Time 12:03 12:40
Environmental temperature 21.8 °C 21.8 °C
Relative humidity 35 % RH 35 %RH
Atmospheric pressure 1021 hPa 1021 hPa
Observations in m: upwards
Level Indication Error (mm) MPE (mm) + -
0.7992 0.800 0.8 ±1 X
1.0022 1.002 -0.2 ±1 X
4.0026 4.002 -0.6 ±1 X
7.0041 7.004 -0.1 ±1 X
10.0039 10.004 0.1 ±1 X
13.0030 13.003 0.0 ±1 X
16.0004 16.000 -0.4 ±1 X
19.0028 19.003 0.2 ±1 X
22.0001 22.000 -0.1 ±1 X
25.0007 25.000 -0.7 ±1 X
28.0007 28.000 -0.7 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 28(51)
Observations in m: downwards
Level Indication Error (mm) MPE (mm) + -
28.0007 28.000 -0.7 ±1 X
24.9996 24.999 -0.6 ±1 X
21.9984 21.998 -0.4 ±1 X
18.9997 18.999 -0.7 ±1 X
15.9995 15.999 -0.5 ±1 X
12.9995 12.999 -0.5 ±1 X
10.0000 10.000 0.0 ±1 X
6.9989 6.999 0.1 ±1 X
3.9995 3.999 -0.5 ±1 X
1.0005 1.001 0.5 ±1 X
0.7958 0.796 0.2 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 29(51)
F.1.5 Array antenna 8”
Date of test: 2010-10-11 Observer: Måns Ackerholm Serial No. (See D.4)
Test conditions Begin End
Time 13:42 14:22
Environmental temperature 21.8 °C 21.8 °C
Relative humidity 34.1 % RH 34.1 %RH
Atmospheric pressure 1021 hPa 1021 hPa
Observations in m: upwards
Level Indication Error (mm) MPE (mm) + -
0.8003 0.800 -0.3 ±1 X
1.0075 1.008 0.5 ±1 X
3.9981 3.998 -0.1 ±1 X
6.9975 6.998 0.5 ±1 X
9.9985 9.999 0.5 ±1 X
12.9999 13.000 0.1 ±1 X
16.0008 16.001 0.2 ±1 X
19.0016 19.002 0.4 ±1 X
22.0011 22.001 -0.1 ±1 X
25.0014 25.001 -0.4 ±1 X
26.5024 26.503 0.6 ±1 X
27.7700 27.770 0.0 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 30(51)
Observations in m: downwards
Level Indication Error (mm) MPE (mm) + -
27.5088 27.509 0.2 ±1 X
26.4844 26.484 -0.4 ±1 X
25.0000 25.000 0.0 ±1 X
22.0000 22.000 0.0 ±1 X
19.0018 19.002 0.2 ±1 X
16.0022 16.002 -0.2 ±1 X
13.0038 13.004 0.2 ±1 X
10.0035 10.004 0.5 ±1 X
7.0032 7.003 -0.2 ±1 X
4.0033 4.004 0.7 ±1 X
1.0000 1.000 0.0 ±1 X
0.7972 0.797 -0.2 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 31(51)
F.1.6 Array antenna 10”
Date of test: 2010-10-12 Observer: Måns Ackerholm Serial No. (See D.4)
Test conditions Begin End
Time 09:07 09:31
Environmental temperature 21.9 °C 21.9 °C
Relative humidity 31.2 % RH 31.2 %RH
Atmospheric pressure 1018 hPa 1018 hPa
Observations in m: upwards
Level Indication Error (mm) MPE (mm) + -
0.7907 0.791 0.3 ±1 X
1.0095 1.009 -0.5 ±1 X
3.9962 3.996 -0.2 ±1 X
6.9989 6.999 -0.1 ±1 X
9.9993 9.999 -0.3 ±1 X
12.9984 12.998 -0.4 ±1 X
15.9991 15.999 -0.1 ±1 X
18.9970 18.997 0.0 ±1 X
22.0004 22.000 -0.4 ±1 X
24.9992 24.999 -0.2 ±1 X
27.9970 27.997 0.0 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 32(51)
Observations in m: downwards
Level Indication Error (mm) MPE (mm) + -
27.9970 27.997 0.0 ±1 X
25.0053 25.005 -0.3 ±1 X
22.0052 22.005 -0.2 ±1 X
19.0035 19.003 -0.5 ±1 X
16.0060 16.006 0.0 ±1 X
13.0034 13.003 -0.4 ±1 X
10.0045 10.004 -0.5 ±1 X
7.0045 7.004 -0.5 ±1 X
4.0054 4.005 -0.4 ±1 X
1.0057 1.006 0.3 ±1 X
0.7926 0.793 0.4 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 33(51)
F.1.7 Array antenna 12”
Date of test: 2010-10-12 Observer: Måns Ackerholm Serial No. (See D.4)
Test conditions Begin End
Time 15:02 15:25
Environmental temperature 21.8 °C 21.8 °C
Relative humidity 34.5 % RH 34.5 %RH
Atmospheric pressure 1020 hPa 1020 hPa
Observations in m: upwards
Level Indication Error (mm) MPE (mm) + -
0.7951 0.796 0.9 ±1 X
1.0052 1.005 -0.2 ±1 X
4.0005 4.001 0.5 ±1 X
7.0003 7.000 -0.3 ±1 X
9.9980 9.998 0.0 ±1 X
13.0001 13.000 -0.1 ±1 X
16.0005 16.001 0.5 ±1 X
19.0024 19.002 -0.4 ±1 X
22.0006 22.001 0.4 ±1 X
25.0004 25.000 -0.4 ±1 X
27.9997 28.000 0.3 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 34(51)
Observations in m: downwards
Level Indication Error (mm) MPE (mm) + -
27.9997 28.000 0.3 ±1 X
25.0043 25.004 -0.3 ±1 X
22.0052 22.005 -0.2 ±1 X
19.0009 19.001 0.1 ±1 X
16.0034 16.003 -0.4 ±1 X
13.0027 13.003 0.3 ±1 X
10.0039 10.004 -0.1 ±1 X
7.0040 7.004 0.0 ±1 X
4.0049 4.005 0.1 ±1 X
1.0027 1.003 0.3 ±1 X
0.7939 0.794 0.1 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 35(51)
F.1.8 LPG/LNG antenna
Date of test: 2010-12-20 Observer: Magnus Herbertsson Serial No. (See D.4)
Test conditions Begin End
Time 11:36 12:43
Environmental temperature 21.3 °C 21.3 °C
Relative humidity 16 % RH 16 %RH
Atmospheric pressure 998 hPa 998 hPa
Observations in m: upwards
Level Indication Error (mm) MPE (mm) + -
0.8003 0.800 -0.3 ±1 X
0.9968 0.997 0.2 ±1 X
4.0074 4.007 -0.4 ±1 X
7.0050 7.005 0.0 ±1 X
10.0054 10.005 -0.4 ±1 X
13.0055 13.005 -0.5 ±1 X
16.0051 16.005 -0.1 ±1 X
19.0054 19.005 -0.4 ±1 X
22.0053 22.005 -0.3 ±1 X
25.0056 25.005 -0.6 ±1 X
28.0055 28.005 -0.5 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 36(51)
Observations in m: downwards
Level Indication Error (mm) MPE (mm) + -
28.0081 28.008 -0.1 ±1 X
24.9951 24.995 -0.1 ±1 X
21.9950 21.995 0.0 ±1 X
18.9943 18.994 -0.3 ±1 X
15.9943 15.994 -0.4 ±1 X
12.9944 12.995 0.6 ±1 X
9.9948 9.994 -0.8 ±1 X
6.9946 6.995 0.4 ±1 X
3.9949 3.995 0.1 ±1 X
0.9945 0.995 0.5 ±1 X
0.8017 0.802 0.3 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 37(51)
F.2 Discrimination (8.1.5.3)
Date of test: Observer: Serial No.
Test conditions Begin End
Time
Environmental temperature °C °C
Relative humidity % RH %RH
Atmospheric pressure hPa hPa
Discrimination Level Indication Level change Indication change + -
Upwards
Downwards
Remarks:
This test is not applicable since the ALG has no moving liquid detection element.
Report number SP-P905645 OIML R 85:2008 (E) Page 38(51)
F.3 Hysteris (8.1.5.4)
Date of test: Observer: Serial No.
Test conditions Begin End
Time
Environmental temperature °C °C
Relative humidity % RH %RH
Atmospheric pressure hPa hPa
Upwards Level 1 Level 2 Level 3 Maximum hysteresis MPE + -
Level up
Indication
Level down
Indication
Hysteresis
Downwards Level 1 Level 2 Level 3 Maximum hysteresis MPE + -
Level down
Indication
Level up
Indication
Hysteresis
Remarks:
This test is not applicable since the ALG has no moving liquid detection element.
Report number SP-P905645 OIML R 85:2008 (E) Page 39(51)
F.4 Influence factor tests
F.4.1 Static environmental temperatures (8.1.6.2)
Date of test: Observer: Magnus Holmsten Serial No. (See D.4)
2010-09-15 to 2010-09-17
Test conditions Begin End
Time 2010-09-15 at 15:03 2010-09-17 at 13:57
Environmental temperature 21.3 °C 21.2 °C
Relative humidity 44.9 % RH 44.6 %RH
Atmospheric pressure 989 hPa 989 hPa
F.4.1.1 Dry Heat (8.1.6.2.1)
Temperature Level (m) Indication (m) Error (mm) MPE (mm) + -
6.9997 6.999 -0.7 ±1 X
Reference
+20 °C 14.9998 15.000 0.2 ±1 X
temperature
23.0005 23.000 -0.5 ±1 X
6.9998 6.999 -0.8 ±1 X
High
+70 °C 14.9998 15.000 0.2 ±1 X
temperature
23.0005 23.000 -0.5 ±1 X
6.9996 6.999 -0.6 ±1 X
Reference
+20 °C 14.9997 15.000 0.3 ±1 X
temperature
23.0005 23.000 -0.5 ±1 X
F.4.1.1 Cold (8.1.6.2.2)
Temperature Level (m) Indication (m) Error (mm) MPE (mm) + -
6.9997 6.999 -0.7 ±1 X
Reference
+20 °C 14.9998 15.000 0.2 ±1 X
temperature
23.0005 23.000 -0.5 ±1 X
7.0002 7.000 -0.2 ±1 X
Low
-40 °C 15.0394 15.040 0.6 ±1 X
temperature
23.0297 23.030 0.3 ±1 X
6.9998 7.000 0.2 ±1 X
Reference
+20 °C 14.9998 15.000 0.2 ±1 X
temperature
23.0005 23.001 0.5 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 40(51)
F.4.2 Mains voltage variations (8.1.6.3 and 8.1.6.4)
Date of test: 2010-09-27 Observer: Magnus Holmsten Serial No. (See D.4)
Test conditions Begin End
Time 10:15 11:03
Environmental temperature 20.9 °C 20.9 °C
Relative humidity 40.8 % RH 40.6 %RH
Atmospheric pressure 1013 hPa 1012 hPa
X DC mains voltage variations (8.1.6.3) X AC mains voltage variations (8.1.6.4)
Voltage Level (m) Indication (m) Error (mm) MPE (mm) + -
6.9997 7.000 0.3 ±1 X
Reference
240 VAC 16.0004 16.000 -0.4 ±1 X
voltage
22.0002 22.000 -0.2 ±1 X
6.9998 7.000 0.2 ±1 X
High
265 VAC 16.0002 16.000 -0.2 ±1 X
voltage
21.9998 21.999 -0.8 ±1 X
6.9998 7.000 0.2 ±1 X
Low
40 VAC 16.0002 16.000 -0.2 ±1 X
voltage
21.9998 22.000 0.2 ±1 X
6.9998 7.000 0.2 ±1 X
Reference
240 VAC 16.0002 16.000 -0.2 ±1 X
voltage
21.9998 22.000 0.2 ±1 X
6.9998 7.000 0.2 ±1 X
Reference
36 VDC 16.0002 16.000 -0.2 ±1 X
voltage
21.9998 22.000 0.2 ±1 X
6.9998 7.000 0.2 ±1 X
High
65 VDC 16.0001 16.000 -0.1 ±1 X
voltage
21.9998 21.999 -0.8 ±1 X
6.9998 7.000 0.2 ±1 X
Low
20 VDC 16.0002 16.000 -0.2 ±1 X
voltage
21.9998 21.999 -0.8 ±1 X
6.9998 7.000 0.2 ±1 X
Reference
36 VDC 16.0002 16.000 -0.2 ±1 X
voltage
21.9998 21.999 -0.8 ±1 X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 41(51)
F.5 Disturbance tests
F.5.1 Damp heat cyclic (condensing) (8.1.7.1)
Observer: Jörgen Eriksson Serial No. (See D.4)
Test conditions Begin End
Date 2010-10-18 2010-10-20
Time 13:35 15:30
Atmospheric pressure 1013 hPa 1013 hPa
Cycle no. Time Low temp Humidity Time High temp Humidity
°C %RH °C %RH
1 9h +25 98 9h +55 93
2 9h +25 98 9h +55 93
Level Indication Error MPE Action checking + -
(m) (m) (mm) (mm) facility
At start of test 16.992 16.992 0 ±1 None X
After cycle no. 1 16.992 16.992 0 ±1 None X
After cycle no. 2 16.992 16.992 0 ±1 None X
After finish of test 16.992 16.992 0 ±1 None X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 42(51)
F.5.2 Radiated, radio frequency, electromagnetic fields (8.1.7.2.1)
Date of test: 2009-10-28 Observer: Reine Johansson Serial No. (See D.4)
Test conditions Begin End
Time 09:02 09:21
Environmental temperature 22.3 °C 22.4 °C
Relative humidity 28 % RH 31 %RH
Atmospheric pressure 999 hPa 999 hPa
EM field Level Indication Error MPE Action + -
Frequency Field (m) (m) (mm) (mm) checking
strength facility
80-1000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Vertical 0°
80-1000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Vertical 90°
80-1000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Vertical 180°
80-1000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Vertical 270°
80-1000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Horizontal 270°
80-1000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Horizontal 180°
80-1000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Horizontal 90°
80-1000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Horizontal 0°
1000-2000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Vertical 0°
1000-2000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Vertical 90°
1000-2000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Vertical 180°
1000-2000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Vertical 270°
1000-2000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Horizontal 270°
1000-2000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Horizontal 180°
1000-2000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Horizontal 90°
Report number SP-P905645 OIML R 85:2008 (E) Page 43(51)
EM field Level Indication Error MPE Action + -
Frequency Field (m) (m) (mm) (mm) checking
strength facility
1000-2000 MHz
10 V/m 16.992 16.992 0 ±1 None X
Horizontal 0°
Remarks:
F.5.3 Conducted, radio frequency, electromagnetic fields (8.1.7.2.2)
Date of test: 2009-10-29 Observer: Reine Johansson Serial No. (See D.4)
Test conditions Begin End
Time 11:00 16:34
Environmental temperature 22.4 °C 21.3 °C
Relative humidity 24.3 % RH 23.8 %RH
Atmospheric pressure 1007 hPa 1009 hPa
Field injected on port(s): AC mains. DC mains. TRL/2 bus. RS-485. Relay. RS-232
EM field Level Indication Error MPE Action checking + -
Frequency Amplitude (m) (m) (mm) (mm) facility
150kHz -
10 V 16.992 16.992 0 ±1 None X
80 MHz
Remarks:
Performed at EUT 2230, 2180 and 2160 specified in section D.4.
Report number SP-P905645 OIML R 85:2008 (E) Page 44(51)
F.5.4 Electrostatic discharge (8.1.7.2.3)
Date of test: 2009-12-01 Observer: Bengt Andersson Serial No. (See D.4)
Test conditions Begin End
Time 12:50 14:18
Environmental temperature 22.3 °C 21.4 °C
Relative humidity 31.3 % RH 32.5 %RH
Atmospheric pressure 1011 hPa 1002 hPa
X Contact discharge X Direct application Test voltage C/A 6/8 kV
X Air discharge X Indirect application Number of discharges 10
Discharge Level Indication Error MPE Action checking + -
applied on (m) (m) (mm) (mm) facility
2240S 16.992 16.992 0 ±1 None X
5900S 16.992 16.992 0 ±1 None X
2230 16.992 16.992 0 ±1 None X
2410 16.992 16.992 0 ±1 None X
2160+2180 16.992 16.992 0 ±1 None X
Remarks:
The time between the discharges was 1 second.
2240S has no legal metrology function related to R85. Only connected to show no influence on
level measuring results.
Report number SP-P905645 OIML R 85:2008 (E) Page 45(51)
F.5.5 Bursts on signal, data and control lines (8.1.7.2.4)
Date of test: 2009-11-04 Observer: Bengt Andersson Serial No. (See D.4)
Test conditions Begin End
Time 10:40 13:55
Environmental temperature 22.3 °C 22.4 °C
Relative humidity 25.2 % RH 24.1 %RH
Atmospheric pressure 1002 hPa 1010 hPa
Test voltage 2 kV
Repetition rate 5 kHz
Discharge Level Indication Error MPE Action checking + -
applied on (m) (m) (mm) (mm) facility
2240S 16.992 16.992 0 ±1 None X
5900S 16.992 16.992 0 ±1 None X
2230 16.992 16.992 0 ±1 None X
2410 16.992 16.992 0 ±1 None X
2160+2180 16.992 16.992 0 ±1 None X
Remarks:
2240S has no legal metrology function related to R85. Only connected to show no influence on
level measuring results.
Report number SP-P905645 OIML R 85:2008 (E) Page 46(51)
F.5.6 Surges on signal, data and control lines (8.1.7.2.5)
Date of test: 2009-11-06 Observer: Bengt Andersson Serial No. (See D.4)
Test conditions Begin End
Time 13:32 15:51
Environmental temperature 21.7 °C 22.4 °C
Relative humidity 28.4 % RH 24.9 %RH
Atmospheric pressure 997 hPa 1002 hPa
Surge Voltage Number Level Indication Error MPE Action + -
applied on and of (m) (m) (mm) (mm) checking
polarity surges facility
2240S ± 0.5 kV 5 16.992 16.992 0 ±1 None X
± 1.0 kV 5 16.992 16.992 0 ±1 None X
5900S ± 0.5 kV 5 16.992 16.992 0 ±1 None X
± 1.0 kV 5 16.992 16.992 0 ±1 Note 1 X
2230 ± 0.5 kV 5 16.992 16.992 0 ±1 None X
± 1.0 kV 5 16.992 16.992 0 ±1 Note 1 X
2410 ± 0.5 kV 5 16.992 16.992 0 ±1 None X
± 1.0 kV 5 16.992 16.992 0 ±1 None X
2160+2180 ± 0.5 kV 5 16.992 16.992 0 ±1 None X
± 1.0 kV 5 16.992 16.992 0 ±1 None X
Remarks:
Note 1: The test object restarted automatically, a message on the AE has to be granted
2240S has no legal metrology function related to R85. Only connected to show no influence on
level measuring results.
Report number SP-P905645 OIML R 85:2008 (E) Page 47(51)
F.5.7 AC mains voltage dips, short interruptions and voltage variations (8.1.7.2.6)
Date of test: 2009-11-04 Observer: Bengt Andersson Serial No. (See D.4)
Test conditions Begin End
Time 15:00 16:43
Environmental temperature 22.4 °C 22.2 °C
Relative humidity 25.4 % RH 22.1 %RH
Atmospheric pressure 1005 hPa 977 hPa
Mains voltage and frequency 48, 65, 120, 230 V 50Hz
Test Reduction Duration Level Indication Error MPE Action + -
to (cycles) (m) (m) (mm) (mm) checking
facility
Dips a 0% 0.5 16.992 16.992 0 ±1 None X
Dips b 0% 1 16.992 16.992 0 ±1 None X
Dips c 40 % 10 16.992 16.992 0 ±1 None X
Dips d 70 % 25 16.992 16.992 0 ±1 None X
Dips e 80 % 250 16.992 16.992 0 ±1 None X
Short
0% 5s 16.992 16.992 0 ±1 None X
interruption
Remarks:
Tests performed on 2410. 2160 and 2180.
Report number SP-P905645 OIML R 85:2008 (E) Page 48(51)
F.5.8 Bursts (transients) on AC and DC mains (8.1.7.2.7)
Date of test: 2009-11-04 Observer: Bengt Andersson Serial No. (See D.4)
Test conditions Begin End
Time 10:40 13:55
Environmental temperature 22.3 °C 22.4 °C
Relative humidity 25.2 % RH 24.1 %RH
Atmospheric pressure 1002 hPa 1010 hPa
Test voltage 2 kV
Repetition rate 5 kHz
Polarity ± Exposure Level Indication Error MPE Action checking + -
time (m) (m) (mm) (mm) facility
+0.5 1 min 16.992 16.992 0 ±1 None X
-0.5 1 min 16.992 16.992 0 ±1 None X
+1.0 1 min 16.992 16.992 0 ±1 None X
-1.0 1 min 16.992 16.992 0 ±1 None X
+2.0 1 min 16.992 16.992 0 ±1 None X
-2.0 1 min 16.992 16.992 0 ±1 None X
Remarks:
Tests performed on units 2410. 2160 and 2180.
Report number SP-P905645 OIML R 85:2008 (E) Page 49(51)
F.5.9 Voltage dips, short interruptions and voltage variations on DC mains
(8.1.7.2.8)
Date of test: 2009-12-02 Observer: Bengt Andersson Serial No. (See D.4)
Test conditions Begin End
Time 10:45 15:30
Environmental temperature 22.1 °C 22.4 °C
Relative humidity 22.9 % RH 23.0 %RH
Atmospheric pressure 994 hPa 998 hPa
Test at 21 V:
Test Reduction Duration Level Indication Error MPE Action + -
to (s) (m) (m) (mm) (mm) checking
facility
Voltage dips 40 % 0.1 16.992 16.992 0 ±1 Note 1 X
70 % 0.1 16.992 16.992 0 ±1 None X
Short 0% 0.01 16.992 16.992 0 ±1 Note 1 X
interruptions “High imp”
0% 0.01 16.992 16.992 0 ±1 Note 1 X
“Low imp”
Voltage 85% 10 16.992 16.992 0 ±1 None X
variations 120% Tested only at 60 V
Remarks:
Note 1: Power off. start-up after ~30 s, no influence on measured value (display: 16.992 m)
DC mains, rated input power 21-60 VDC
Test at 60 V:
Test Reduction Duration Level Indication Error MPE Action + -
to (s) (m) (m) (mm) (mm) checking
facility
Voltage dips 40 % 0.1 16.992 16.992 0 ±1 None X
70 % 0.1 16.992 16.992 0 ±1 None X
Short
0% Tested only at 21 V
interruptions
Voltage 85% 10 16.992 16.992 0 ±1 None X
variations 120% 10 16.992 16.992 0 ±1 None X
Remarks:
DC mains, rated input power 21-60 VDC
Report number SP-P905645 OIML R 85:2008 (E) Page 50(51)
F.5.10 Ripple on DC mains power (8.1.7.2.9)
Date of test: 2009-12-02 Observer: Bengt Andersson Serial No. (See D.4)
Test conditions Begin End
Time
Environmental temperature 22.1 °C 22. 0 °C
Relative humidity 22.9 % RH 24.0 %RH
Atmospheric pressure 994 hPa 995 hPa
Frequency 100 Hz
Test time 10 min
Voltage Level Indication Error MPE Action checking + -
(m) (m) (mm) (mm) facility
24 VDC 16.992 16.992 0 ±1 None X
48 VDC 16.992 16.992 0 ±1 None X
Remarks:
Report number SP-P905645 OIML R 85:2008 (E) Page 51(51)
F.5.11 Surges on AC and DC mains power lines (8.1.7.2.10)
Date of test: 2009-11-05 Observer: Bengt Andersson Serial No. (See D.4)
Test conditions Begin End
Time 13:50 16:13
Environmental temperature 23.1 °C 21.4 °C
Relative humidity 27.8 % RH 24.9 %RH
Atmospheric pressure 1001 hPa 977 hPa
Angle Voltage Number Level Indicatio Error MPE Action + -
and of (m) n (mm) (mm) checking
polarity surges (m) facility
0° 1 kV 3+3 16.992 16.992 0 ±1 None X
Line
90° 1 kV 3+3 16.992 16.992 0 ±1 None X
to
180° 1 kV 3+3 16.992 16.992 0 ±1 None X
line
270° 1 kV 3+3 16.992 16.992 0 ±1 None X
0° 2 kV 3+3 16.992 16.992 0 ±1 None X
Line
90° 2 kV 3+3 16.992 16.992 0 ±1 None X
to
180° 2 kV 3+3 16.992 16.992 0 ±1 None X
earth
270° 2 kV 3+3 16.992 16.992 0 ±1 None X
Remarks
Tests were performed on 2410, 2160 and 2180. Unit 2160 was modified to pass the test.
Saatja: Ando Vendel / ProREX <
[email protected]>
Saaja: Triinu Sillamaa
Teema: Tüübikinnituse taotlemine Raptor süsteemile
Lugupeetud Triinu Sillamaa,
Soovime taodelda tüübikinnitust mahutite radarnivoomõõtesüsteemile
Raptor (firmalt Rosemount Tank Radar).
Manusena lisatud esialgne dokumentatsioon.
Palun andke tagasisisdet, kas see on piisav tüübikinnituse saamiseks või
on vaja veel midagi juurde.
Kui palju dokumentatsiooni on vaja füüsiliselt paberile välja printida?
Kas piisab, kui mahukama tehnilise dokumentatsiooni (kataloogid,
juhendid, joonised jms) esitame CD-l?
Midagi veel?
PS! Varasemast olemas meil tüübikinnitused (TJI 5-3/15 31.02; TJI 5-3/15
32.02; TJI 5.3-1/15.11.07, mis paraku küll juba aegunud) sama tootja
eelmise põlvkonna samalaadsetele toodetele, mis osaliselt kattuvad ka
uue süsteemiga.
Ette tänades!
Lugupidamisega,
Ando Vendel
Projektijuht
ProREXoil OÜ
Peterburi tee 47
11415 Tallinn
Tel.: +372 60 624 60
Fax: +372 60 624 61
GSM: +372 56 900 810
e-mail:
[email protected] <mailto:
[email protected]>
www.prorexoil.ee <http://www.prorexoil.ee/>
Olge valmis uueks ajastuks mahutite mõõtmises!
Tutvuge lähemalt Raptor süsteemiga »
<http://www2.emersonprocess.com/en-US/brands/rosemounttankgauging/Pages/
index.aspx>