Dear Katažyna,
Thank you for your inquiry concerning medical device registration in Estonia. I hope I understood you correctly that your question concerns distribution of medical devices in Estonia.
Indeed, there is a distribution notification obligation in Estonia and it is based on Medical Device Act § 26 (4):<https://www.riigiteataja.ee/en/eli/ee/515032023005/consolide/current>
Every undertaking who distributes on the market of Estonia a system or procedure pack of medical devices according to Article 22 of Regulation (EU) 2017/745 of the European Parliament and of the Council or a class IIa, IIb or III medical device classified according to Regulation (EU) 2017/745 of the European Parliament and of the Council or Council Directive 90/385/EEC on the approximation of the laws of the Member States relating to active implantable medical devices (OJ L 189, 20.07.1990, p 17–36) or Council Directive 93/42/EEC concerning medical devices (OJ L 169, 12.07.1993, p 1–43) or class B, C or D in vitro diagnostic medical devices classified according to Annex VIII of Regulation (EU) 2017/746 of the European Parliament and of the Council, shall notify the Health Board thereof within 10 days after distribution of the relevant medical device for the first time.
Notification about distribution can be completed through our national Medical Devices and Appliances Database, https://msa.sm.ee/eng and it is free of charge. For actors outside of Estonia we can make an account manually, so it is possible to log into the database with username and password. For creating the account we need a contact person’s full name, e-mail and the position in the company. If there is a need for an account, please let me know.
Documents that need to be supplied depend on which class of devices you plan to distribute. For class I medical devices they are the following: Declaration of Conformity (as described in annex IV of MDR), instructions for use and copy of the labelling or packaging of the device. For class IIa and IIb devices you also need to present the EC Certificate of Quality Management System. For class III devices we need two EC Certificates.
A few explanatory words about instructions for use as well. If the devices that you plan to distribute are meant to be used by lay users, you have to provide the IFU in both English and Estonian on the application taking note of that these are of the same version (Medical Device Act §16 (3)). If, however, the devices are for professional use only (which has to be clearly stated in the documentation you provide), IFU can be provided only in English if the end user has agreed to it (Estonian Language Act<https://www.riigiteataja.ee/en/eli/528072023001/consolide> §17).
Also, please use a short guide for using the Estonian Medical Device Database that can be found on database homepage under „Forms<https://msa.sm.ee/eng/form/?>“. It is a step-by-step guide on how to carry out the notification process successfully in the database.
Please do not hesitate to contact us for further information or with any questions.
Best regards
Janne Aule
Chief Specialist
Department of Medical Devices
+37258668664
[email protected]<mailto:
[email protected]>
[cid:
[email protected]]
Republic of Estonia
Terviseamet/Health Board
+372 794 3500
[email protected]
www.terviseamet.ee
Paldiski mnt 81, 10614 Tallinn
Estonia
This e-mail is confidential and meant for use by the person named in the letterhead. Any use in any way or copying of it by a person not marked as the addressee thereof is prohibited. If you have got this e-mail by mistake, please notify of it the sender without delay and delete the received e-mail together with all its attachments.
From: Katažyna Stankevič <
[email protected]>
Sent: teisipäev, 25. juuni 2024 16:00
To: Meditsiiniseadmed (Medical Devices) <
[email protected]>
Subject: Medical devices
Tähelepanu! Tegemist on väljastpoolt asutust saabunud kirjaga. Tundmatu saatja korral palume linke ja faile mitte avada.
Good day,
We would like to know about the registration of medical devices. What is the registration process if the manufacturer is from Lithuania?
What documents need to be submitted? What forms do I need to fill out?
Waiting for your reply.
Pagarbiai,
_________________
Katažyna Stankevič
Administratorė
+37065377167