Measles and Rubella Laboratory
Check-list for Annual WHO Accreditation
Date of Review: 01.01 – 31.12.2016 By correspondence Accreditation for calendar year: 2018
Laboratory: Laboratory of Communicable Diseases
Institute: Health Board
Address: Kotka 2, 11315, Tallinn
Country Estonia
Phone: +3726943635 Email:
[email protected] Website: www.terviseamet.ee
+3726943671
[email protected]
Head of Institute: Dr Tiiu Aro Head of Department: Külli Rae
Head of Laboratory: Külli Rae
Head Mea/Rub Lab: Svetlana Rudõka Chief Technician: Svetlana Rudõka
Name of Laboratory used for Virus Isolation/Detection: HB Laboratory of Communicable Diseases
Name of Laboratory used for Virus Sequencing: HB Laboratory of Communicable Diseases in
collaboration with Estonian Biocentre
(Laboratories that act as a reference laboratory for sub-national laboratories should complete annex I to the checklist)
Reviewers: Dr. Myriam Ben Mamou & Dovile Videbaek, WHO EURO
Name of National Accrediting Authority (if appropriate) and current accreditation status from this
authority (e.g. ISO15189 or other, provide documentation): Estonian Accreditation Centre certified on 30.06.2014 that
Communicable Diseases Laboratory for HB has competence according to ISO 15189:2007 as a medical
laboratory to conduct researches in the field of clinical and microbiological research. The accreditation
shall remain in effect until 29.06.2019.
Part I: Summary of Review (assessor to complete)
Recommendations:
Measles Rubella
Serology RT-PCR Sequencing Serology RT-PCR Sequencing
Accredited1
Provisionally accredited2
Do not accredit3
1
Laboratory meets all criteria
2
Laboratory passed the most recent serology PT, but needs to strengthen one or more of the remaining criteria
3
Laboratory did not pass the most recent serology PT or molecular EQA
Findings:
Measles and rubella IgM test results are reported by the laboratory to the Measles: 100%
1.
surveillance system within 4 days for ≥ 80% of specimens: Rubella: 95%
Completeness 100%
2. The laboratory reports to MRLDMS weekly and/or to CISID monthly
Timeliness 92%
Measles: 176
3. IgM tests are performed on at least 50 specimens annually:
Rubella: 176
Measles: 100%
4. The accuracy of IgM detection is ≥ 90% (38 samples)
Rubella: 100%
5. Internal quality control (QC) procedures are implemented: (Yes/No/Partially) Partially
The score on the most recent WHO proficiency test is ≥ 90%: Measles: 97.73%
6.
PT panel number: 01602 Date reported: 03/02/17 Rubella: 97.73%
Performance on the most recent WHO Molecular proficiency test (if
7. NA
appropriate): Round: Date reported: dd/mm/yy
Virus genotyping (if performed) is completed AND data reported to WHO
MeaNS:
8. through MeaNS or RubeNS within 2 months of receipt of specimens, for ≥80% NA
RubeNS :
of the specimens appropriate for genetic analysis:
The score from the on-site review of laboratory operating procedures and practices is ≥
9. NA
80%:
Key accreditation criteria
5- Routine internal quality control procedures should be strengthened.
6- Measles and rubella in-house should be used. (Collaborate with RRL if needed).
Conclusion
• Excellent performance.