2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
EMC TEST REPORT
Test report No. : TRECEE19-0126
Applicant : COIZ Co.,Ltd.
Address : (07650) 37-4, Hwagok-ro 59-gil, Gangseo-gu, Seoul,
Korea
Manufacturer : COIZ Co.,Ltd.
Address : (07650) 37-4, Hwagok-ro 59-gil, Gangseo-gu, Seoul,
Korea
Type of equipment : AstrodomeFacial
Model name : AD19A
Variant model name : Not applicable
Date of incoming : June 17, 2019
Date of test : July 31, 2019 ~ August 01, 2019
Date of issue : September 04, 2019
Test standards : EN 55014-1:2017
EN 55014-2:2015
EN 61000-3-2:2014
EN 61000-3-3:2013
Test result : Comply Not comply
Summary
The Test results presented in this test report are limited only to the sample supplied by
applicant and the use of this test report is inhibited other than its purpose. This test
report shall not be reproduced except in full, without the written approval of Lab-T, Inc.
Prepared by Approved by
Jae-Hee Song / EMC test engineer Sung-Min Yang / Technical manager
If this test report is required to confirmation of authenticity, please contact to
[email protected]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 1 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
- CONTENTS -
1. REVISION HISTORY .............................................................................................................................. 3
2. INFORMATION OF TEST LABORATORY ............................................................................................. 4
3. APPLICANT INFORMATION ................................................................................................................. 5
4. DESCRIPTION OF EUT (EQUIPMENT UNDER TEST) .......................................................................... 6
4.1 Product description.....................................................................................................................................6
4.2 Product specification ..................................................................................................................................6
4.3 EUT internal operating frequency ..............................................................................................................6
4.4 Details information of multi-listing model....................................................................................................6
4.5 Peripheral equipment .................................................................................................................................7
4.6 Connection cable........................................................................................................................................7
4.7 Test set-up configuration ............................................................................................................................7
4.8 EUT operating test mode(s) .......................................................................................................................7
4.9 EUT modifications ......................................................................................................................................7
5. TEST STANDARDS AND PERFORMANCE CRITERIA ......................................................................... 9
5.1 Standards ...................................................................................................................................................9
5.2 Performance Criteria ................................................................................................................................11
5.3 Classification of apparatus .......................................................................................................................11
5.4 Application of tests for the different categories of apparatus ...................................................................13
6. EMISSION TEST RESULTS ................................................................................................................. 14
6.1 Continuous disturbance voltage ...............................................................................................................14
6.2 Continuous disturbance power.................................................................................................................19
6.3 Discontinuous disturbance voltage ..........................................................................................................22
6.4 Radiated disturbance (30 MHz ~ 1 000 MHz) ........................................................................................29
6.5 Harmonics current emission.....................................................................................................................33
6.6 Voltage fluctuations and flicker emission .................................................................................................42
7. IMMUNITY TEST RESULTS ................................................................................................................. 48
7.1 Electrostatic discharge .............................................................................................................................48
7.2 Radio-frequency electromagnetic field (80 MHz ~ 1 000 MHz) ...............................................................53
7.3 Electrical fast transients ...........................................................................................................................58
7.4 Surges ......................................................................................................................................................62
7.5 Injected currents (0.15 MHz to 230 MHz) ................................................................................................65
7.6 Voltage dips and interruptions ..................................................................................................................68
8. PHOTOGRAPHS OF THE PRODUCT .................................................................................................. 71
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 2 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
1. Revision history
Issued report No. Version Issued date Revisions
TRECEE19-0126 Rev. 00 September 04, 2019 Original
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 3 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
2. Information of test laboratory
Corporate name Lab-T, Inc.
Representative Jong-Young Kim
Address 2182-42, 2182-40 Baegok-daero, Mohyeon-eup, Cheoin-gu, Yongin-si
Gyeonggi-do 17036, Korea
Telephone +82-31-322-6767
Fax +82-31-322-6768
E-mail
[email protected]
* Lab-T, Inc. has been accredited / filed / authorized by the agencies listed in the following table.
Certificate Nation Agency Code Mark
Accreditation Korea KOLAS KT703
USA FCC 133186
R-14282, C-14764,
Site filing Japan VCCI
T-12276, G-10886, G-10887
Canada IC 22000
Korea KC KR0159
Certification
CARAT 15
EU TUV
09 93449 001
- Quality control in the testing laboratory is implemented as per ISO/IEC 17025 which is the
“General requirements for the competent of calibration and testing laboratory”.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 4 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
3. Applicant information
Applicant COIZ Co.,Ltd.
Address (07650) 37-4, Hwagok-ro 59-gil, Gangseo-gu, Seoul, Korea
Manufacturer COIZ Co.,Ltd.
Address (07650) 37-4, Hwagok-ro 59-gil, Gangseo-gu, Seoul, Korea
Country of origin Korea
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 5 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
4. Description of EUT (Equipment under test)
4.1 Product description
Name of EUT AstrodomeFacial
Model name AD19A
Rating AC 100 ~ 240 V, 50 / 60 Hz
Input rating Not applicable
(Use for adaptor)
4.2 Product specification
4.3 EUT internal operating frequency
Frequency Description Frequency Description
8 MHz - - -
4.4 Details information of multi-listing model
Division Model name
- -
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 6 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
4.5 Peripheral equipment
Product Model name Serial No. Manufacturer
AstrodomeFacial
AD19A - COIZ Co.,Ltd. / Korea
(EUT)
Remote control
- - COIZ Co.,Ltd. / Korea
(EUT)
4.6 Connection cable
Start-up device Connected end device Cable specifications
Name I/O port Name I/O port Length(m) Spec.
AstrodomeFacial
AC IN AC MAIN - 1.5 Unshield
(EUT)
4.7 Test set-up configuration
AC MAIN
AstrodomeFacial
(EUT)
Remote
Controller
(EUT)
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 7 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
4.8 EUT operating test mode(s)
- The test was conducted by checking the purple light and wind coming out of the EUT.
4.9 EUT modifications
- Not modifications
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 8 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
5. Test standards and performance criteria
5.1 Standards
- The following standards have been applied:
Phenomena Applied standard Result
Emission _ EN 55014-1:2017
Continuous disturbance voltage EN 55014-1:2017 C
N/A
Continuous disturbance power EN 55014-1:2017
(Note 1)
Discontinuous disturbance voltage EN 55014-1:2017 C
Radiated disturbance (30 MHz ~ 1 000 MHz) EN 55014-1:2017 C
Harmonics current EN 61000-3-2:2014 C
Voltage fluctuations and flickers EN 61000-3-3:2013 C
Immunity _ EN 55014-2:2015 (Single mode: Category III, Charge mode: Category IV)
Electrostatic discharge EN 61000-4-2:2009 C
Radio-frequency electromagnetic field N/A
EN 61000-4-3:2006+A2:2010
(80 MHz to 1 000 MHz) (Note 2)
Fast transients EN 61000-4-4:2012 C
Surges EN 61000-4-5:2014+A1:2017 C
Injected currents (0.15 MHz to 230 MHz) EN 61000-4-6:2014 C
Voltage dips and interruptions EN 61000-4-11:2004+A1:2017 C
* C=Comply, N/A=Not applicable
* Note 1: Replaced by the radiated disturbance test.
* Note 2: Excluded from the test as it was classified as Category II.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 9 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
* Measurement uncertainty
All measurements involve certain levels of uncertainties, especially in field of EMC.
The factors contributing to uncertainties are test receiver, cable loss, antenna factor calibration,
Antenna directivity, antenna factor variation with height, antenna phase center variation, antenna
frequency interpolation, measurement distance variation, site imperfection, mismatch, and system
repeatability. Based on CISPR 16-4-2, the measurement uncertainty level with a 95 % confidence
level was applied.
Confidence level of
Test item Uncertainty
approximately
Continuous disturbance voltage 9 ㎑ ~ 30 ㎒ 2.36 ㏈ Least about 95 %, k = 2
Discontinuous disturbance voltage 9 ㎑ ~ 30 ㎒ 3.28 ㏈ Least about 95 %, k = 2
Radiated disturbance
30 ㎒ ~ 1 000 ㎒ 4.67 ㏈ Least about 95 %, k = 2
(30 MHz ~ 1 000 MHz)
Injected currents (0.15 MHz to 230 MHz) 2.08 ㏈ Least about 95 %, k = 2
Harmonics current Least about 95 %, k = 2
Voltage fluctuations and flickers Least about 95 %, k = 2
Electrostatic discharge Least about 95 %, k = 2
Fast transients Least about 95 %, k = 2
Surges Least about 95 %, k = 2
Voltage dips and interruptions Least about 95 %, k = 2
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 10 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
5.2 Performance Criteria
A functional description and a definition of performance criteria, during or as a consequence of the
EMC testing, shall be provided by the manufacturer and noted in the test report, based on the
following criteria.
Performance criterion A:
The apparatus shall continue to operate as intended during the test. No degradation of performance
or loss of function is allowed below a performance level (or permissible loss of performance) specified
by the manufacturer, when the apparatus is used as intended. If the minimum performance level or
the permissible performance loss is not specified by the manufacturer, then either of these may be
derived from the product description and documentation, and from what the user may reasonably
expect from the apparatus if used as intended.
Performance criterion B:
The apparatus shall continue to operate as intended after the test. No degradation of performance or
loss of function is allowed below a performance level (or permissible loss of performance) specified
by the manufacturer, when the apparatus is used as intended. During the test, degradation of
performance is allowed, however no change of actual operating state or stored data is allowed to
persist after the test. If the minimum performance level or the permissible performance loss is not
specified by the manufacturer, then either of these may be derived from the product description and
documentation, and from what the user may reasonably expect from the apparatus if used as
intended.
Performance criterion C:
Temporary loss of function is allowed, provided the function is self- recoverable or can be restored by
the operation of the controls, or by any operation specified in the instructions for use.
The selection, the specification of functions, and the permissible degradation is left to the
responsibility of the manufacturer.
Annex A serves as a guide to formulate the permissible degradation of the equipment under test
(EUT) caused by electromagnetic phenomena.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 11 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
5.3 Classification of apparatus
The apparatus covered by this standard is subdivided into categories. For each category, specific
requirements are formulated.
Category I: apparatus containing no electronic control circuitry.
All appliances having no electronic control circuitry are considered to be category I.
Electric circuits consisting of passive components (such as radio interference suppression capacitors
or inductors, mains transformers, mains frequency rectifiers and heating elements) are not considered
to be electronic control circuitry.
EXAMPLES Appliances operated with a motor and mechanical switch only; lighting toys with a battery and a LED or
incandescent lamp without additional electronic control circuitry; track sets without electronic control circuitry; heating
or cooling appliances without electronic control circuitry; tools without electronic controls and all other apparatus
containing only electromechanical components (e. g. switches or thermostats).
Category II: transformer toys, dual supply toys, mains powered motor operated appliances, tools,
heating appliances and similar electric apparatus (for example – UV radiators, IR radiators and
microwave ovens) containing electronic control circuitry with no clock frequency higher than 15 MHz.
NOTE For toys, examples include educational computers, organs, track sets with electronic control units.
Category III: equipment which in normal use, is not connected to a power network and has no
cables attached.
This category includes apparatus provided with rechargeable batteries, solar or other similar d.c.
power sources which can be charged or operated by connecting the apparatus to the mains power.
However, this apparatus shall also be tested as an apparatus in category II while it is connected to the
mains network.
NOTE For toys, examples include musical soft toys, cord-controlled toys and motor-operated electronic toys.
Category IV: all other apparatus covered by the scope of this standard.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 12 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
5.4 Application of tests for the different categories of apparatus
Category I
Category I apparatus is deemed to fulfill the relevant immunity requirements without testing
Category II
Category II apparatus shall fulfill the following requirements;
- Electrostatic discharge with performance criterion B
- Fast transients with performance criterion B
- Injected currents up to 230 ㎒ with performance criterion A
- Surge with performance criterion B
- Voltage dips and interruptions with performance criterion C
Category III
Category III apparatus shall fulfill the following requirements;
- Electrostatic discharge with performance criterion B
(A performance criterion C could be applied to toys not using score or data entered by the user.
Examples are musical soft toys, sounding toys, etc.)
- Radio frequency electromagnetic fields, with performance criterion A.
This test is only applicable to the ride on toys operating with electronic devices.
Category IV
Category IV apparatus shall fulfill the following requirements;
- Electrostatic discharge with performance criterion B
- Fast transients with performance criterion B
- Injected currents up to 80 ㎒ with performance criterion A
- Radiofrequency EM fields with performance criterion A
- Surge with performance criterion B
- Voltage dips and interruptions with performance criterion C
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 13 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6. Emission test results
6.1 Continuous disturbance voltage
Test standard EN 55014-1:2017
Test date 2019.07.31
Test facility Shielded room (#1)
Test voltage AC 230 V, 50 Hz
Temperature (20.8 ~ 22.0) C
Relative humidity (48.7 ~ 50.4) % R.H.
Test result Comply
6.1.1 Measurement procedure
If the EUT is table top equipment, it was placed on a wooden table with a height of 0.8 m above the
reference ground plane and 0.4 m from the conducting wall of the shielded room.
Also if the EUT is floor-standing equipment, it was placed on a non-conducted support with a height
up to 0.15 m ± 25 % above the reference ground plane. Connect the EUT’s power source lines to the
appropriate power mains / peripherals through the LISN. All the other peripherals are connected to
the 2nd LISN & ISN, if any. Unused measuring port of the LISN & ISN was resistively terminated by
50 ohm terminator. The measuring port of the LISN for EUT was connected to spectrum analyzer.
Using conducted emission test software, the emissions were scanned with peak detector mode.
After scanning over the frequency range, suspected emissions were selected to perform final
measurement. When performing final measurement, the receiver was used which has quasi-peak
detector and CISPR average detector. By varying the configuration of the test sample and the cable
routing it was attempted to maximize the emission.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 14 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.1.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
EMI Test Receiver ESR7 R&S 101440 2019.12.14
PULSE LIMITER VTSD 9561-F Schwarzbeck 9561-F189 2020.04.23
LISN ENV216 R&S 101883 2020.04.24
LISN Control Unit LISN Controller TSJ 04494-1 -
LISN Control Unit LISN Controller TSJ 04494-2 -
* All test equipment used is calibrated on a regular basis.
6.1.3 Frequency range 148.5 kHz to 30 MHz (Terminal voltages)
* Household appliances and equipment causing similar disturbances and regulating controls
incorporating semiconductor devices
Main ports Associated ports
Frequency Disturbance voltage Disturbance voltage Disturbance current
(㎒) (㏈(㎶)) (㏈(㎶)) (㏈(㎶)) (㏈(㎶)) (㏈(㎶)) (㏈(㎶))
Quasi-peak Average * Quasi-peak Average * Quasi-peak Average*
Decreasing linearly with the Decreasing linearly with
logarithm of the frequency the logarithm of the
0.15 to 0.50 from: 80 70 frequency from:
66 to 56 59 to 46 40 to 30 30 to 20
0.50 to 5 56 46 74 64
30 20
5 to 30 60 50 74 64
* The lower limit applies at the transition frequencies.
The test report shall state which test method was used and which limits were applied.
* Main terminals of tools
P ≤ 700 W 700 W < P ≤ 1 000 W P > 1 000 W
Frequency
(㎒) (㏈(㎶)) (㏈(㎶)) (㏈(㎶)) (㏈(㎶)) (㏈(㎶)) (㏈(㎶))
Quasi-peak Average * Quasi-peak Average * Quasi-peak Average*
Decreasing linearly with the logarithm of the frequency from:
0.15 to 0.35
66 to 59 59 to 49 70 to 63 63 to 53 76 to 69 69 to 59
0.35 to 5 59 49 63 53 69 59
5 to 30 64 54 68 58 74 64
* The lower limit applies at the transition frequencies.
P = rated power of the motor only
* Note: The limits for the measurement with the average detector are tentative and may be modified after a
period of experience.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 15 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.1.4 Continuous disturbance voltage test data
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 16 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
* Remark: “L1”: (Line), “N”: (Neutral)
* Results [㏈(㎶)] = Reading [㏈(㎶)] + C.FACTOR [㏈]
* C.FACTOR [㏈] = LISN insertion Loss [㏈] + Cable loss [㏈] + Pulse limiter factor [㏈]
* Margin [㏈] =Limit [㏈(㎶)] - Result [㏈(㎶)]
* QP: Quasi-peak , CAV: CISPR average
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 17 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.1.5 Test setup photos
[Front]
[Rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 18 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.2 Continuous disturbance power
Test standard EN 55014-1:2017
Test date -
Test facility -
Test voltage -
Temperature -
Relative humidity -
Test result Not applicable
6.2.1 Measurement procedure
The distance between the clamp test set-up (the appliance, the lead to be measured and the absorbing
clamp) and any other conductive objects (including persons, walls and ceiling, but excluding the floor)
shall be at least 0.8 m.
The appliance to be tested shall be placed on a non-metallic support table parallel to the floor.
The height of the table shall be 0.1 m ± 0.025 m for appliances primarily intended to be positioned on
the floor in normal use, and 0.8 m ± 0.05 m for other appliances. The lead to be measured is placed in
a straight line for a distance sufficient to accommodate the absorbing clamp, and to permit the necessar
measuring adjustment of position for tuning. The clamp is placed around the lead.
The absorbing clamp is positioned for maximum indication at each test frequency: the clamp shall be
moved along the lead until the maximum value is found between a position adjacent to the appliance
and a distance of about a half-wavelength from it. The straight portion of the lead to be measured on
should therefore be about 6 m long, this being equal to λ max/2 + 0.6 m in order to allow at any time
the positioning of the absorbing clamp and a possible second clamp for additional isolation.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 19 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.2.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
- - - - -
* All test equipment used is calibrated on a regular basis.
6.2.3 Disturbance power measurement in the frequency range 30 MHz to 300 MHz
* Disturbance power limits for the frequency range 30 MHz to 300 MHz
Tools
General
Frequency P ≤ 700 W 700 W < P ≤1 000 W P > 1 000 W
range ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺)
㏈(㎺)
(㎒) Quasi- Average Quasi- Average Quasi- Average Quasi-
Average
peak peak peak peak
30 to 300 45 to 55 35 to 45 45 to 55 35 to 45 49 to 59 39 to 49 55 to 65 45 to 55
P = rated power of the motor only
* Margin when performing disturbance power measurement in the frequency range
30 MHz to 300 MHz
Tools
General
Frequency P ≤ 700 W 700 W < P ≤1 000 W P > 1 000 W
range ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺) ㏈(㎺)
(㎒) Quasi- Average Quasi- Average Quasi- Average Quasi- Average
peak peak peak peak
Increasing linearly with the frequency from:
200 to 300 0 to 10 - 0 to 10 - 0 to 10 - 0 to 10 -
P = rated power of the motor only
Note This table only applies if method a) specified in 4.3.4.2 is followed.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 20 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.2.4 Continuous disturbance power test data
N/A
(replaced by a radiation test.)
6.2.6 Test setup photos
N/A
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 21 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.3 Discontinuous disturbance voltage
Test standard EN 55014-1:2017
Test date 2019.07.31
Test facility Shielded room (#1)
Test voltage AC 230 V, 50 Hz
Temperature (20.5 ~ 23.0) C
Relative humidity (48.5 ~ 50.8) % R.H.
Test result Comply
6.3.1 Measurement procedure
The initial test was made to count discontinuous disturbance classified as clicks (n) over the Q-peak
limits of mains terminals disturbance voltage at each frequencies. The test was observed for 120 min
(T) for each measurement. The click rate “N” was determined from the formula N = n * f/T.
The relevant click limit “Lq” for discontinuous disturbance was determined in according the formula
given in EN 55014-1, Clause 4.2.2.2. The second test run was made to determine how many clicks
exceed the click limit “Lq” the time for this second run was the same as the time taken for initial test
6.3.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
LISN LS16C/10 AFJ 16011808404 2020.01.31
CLICK ANALYZER DDA55 AFJ 14041750127 2020.01.30
* All test equipment used is calibrated on a regular basis.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 22 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.3.3 Discontinuous disturbance voltage
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 23 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.3.4 Discontinuous disturbance voltage test data
- Click L1 -
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 24 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 25 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
- Click N -
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 26 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 27 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.3.5 Test setup photos
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 28 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.4 Radiated disturbance (30 MHz ~ 1 000 MHz)
Test standard EN 55014-1:2017
Test date 2019.07.31
Test facility 10 m chamber
Test voltage AC 230 V, 50 Hz
Temperature (20.0 ~ 21.7) C
Relative humidity (49.5 ~ 51.8) % R.H.
Test result Comply
6.4.1 Measurement procedure
If the EUT is tabletop equipment, it was placed on a wooden table with a height of 0.8 m above the
reference ground plane and 10 m away from the interference receiving antenna in the 10 m semi-
anechoic chamber.
Also if the EUT is floor-standing equipment, it was placed on a non-conducted support with a height
up to 0.15 m ± 25 % above the reference ground plane. Rotate the EUT from (0 - 360)° and position
the receiving antenna at heights from (1 - 4) m above the reference ground plane continuously to
determine associated with higher emission levels and record them. The measurement was made in
both the vertical and horizontal polarization, and the maximum value is presented in the report.
For 30 MHz ~ 1 000 MHz frequency range, quasi-peak detector with 120 KHz RBW was used.
.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 29 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.4.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
EMI Test Receiver ESW44 R&S 101839 2020.06.17
Low Noise
MLA-10K01-B01-14 TSJ 2060297 2020.06.17
Preamplifier
Bi-Log Antenna VULB9160 Schwarzbeck 9160-3380 2021.04.09
Attenuator 50FPE-006 N 6dB FIXED - 2020.04.23
Controller C3000 Innco 45450119 -
Antenna Mast MA4000-EP Innco - -
Turn Table - - - -
* All test equipment used is calibrated on a regular basis.
6.4.3 Radiated disturbance limits
Limit a
Frequency range
Testing method Basic standard (㏈(㎶/m)) Remark
(㎒)
Quasi-peak
30 - 230 30 Measurement
OATS or SAC b CISPR 16-2-3
230 - 1 000 37 distance 10 m
30 - 230 42 to 35 d Measurement
FAR c CISPR 16-2-3
230 - 1 000 42 distance 3 m
30 - 230 42 to 35 d Measurement
FAR c IEC 61000-4-22
230 - 1 000 42 distance 3 m
TEM- 30 - 230 30
IEC 61000-4-20 -
Waveguide e
230 - 1 000 37
NOTE: The test report shall state which test method was used and which limits were applied.
a
The lower limit is applies at the transition frequency.
b Measurements may be made at closer distance, down to 3 m. An inverse proportionality factor of 20 dB per
decade shall be used to normalize the measured data to the specified distance for determining the limit.
In this case the recommendations of the CISPR basic standards shall be considered when testing large EUT
at frequency approaching 30 MHz, due to near field effects.
C All equipment shall be measured within the test volume as described in 5.3.4.3 and shown in Figures 12 to 19.
d Decreasing linearly with the logarithm of the frequency.
e The TEM waveguide method shall be limited to battery operated EUT without cables attached and with a
maximum size according to 6.2 of IEC 61000-4-20:2010 (the largest dimension of the enclosure is equal to the
wavelength at the maximum measurement frequency, 300 mm at 1 GHz).
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 30 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.4.4 Radiated emission test data
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 31 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
* Results [㏈(㎶/m)] = Reading [㏈(㎶)] + Antenna factor [㏈/m] - Loss [㏈]
* Loss = Cable loss [㏈] - Amp gain [㏈]
* Margin [㏈] = Limit [㏈(㎶/m)] - Results [㏈(㎶/m)]
* QP: Quasi-peak
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 32 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.4.5 Test setup photos
[Single mode Front]
[Single mode Rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 33 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.5 Harmonics current emission
Test standard EN 61000-3-2:2014
Test date 2019.07.31
Test facility Shielded room (#1)
Test voltage AC 230 V, 50 Hz
Temperature (20.5 ~ 22.0) C
Relative humidity (52.0 ~ 54.7) % R.H.
Test result Comply
6.5.1 Measurement procedure
Harmonics of the fundamental current were measured up to 2 ㎑ using a universal power analyzer.
The measurements were carried out under steady conditions and using average. Before making
measurements, it is necessary for the EUT to decide which class the EUT fulfills into; A, B, C or D.
The test set-up photo is included In appendix 3. (Remark: The input power of the EUT is lower than
75 W. According to the cause 7 in EN 61000-3-2:2014, no limit applies for the EUT. See next page
for harmonics on AC mains test data.)
Observation time 150 s
Windows width 10 periods
6.5.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
Digital power analyzer
DPA 500N EMTEST P1437139796 2020.04.24
(Harmonics & Flicker)
AC power supply PCR4000M KIKUSUI UK001881 2020.05.10
* All test equipment used is calibrated on a regular basis.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 34 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.5.3 Measurement result
Report title: T1906-0077
Company Name: LAB-T
Date of test: 2019.07.31
Measurement file name: Harmonics_3_2_Ed4.rsd
Tester: SJH
Standard used: EN/IEC 61000-3-2 Ed.4 Quasi-stationary
Equipment class A <= 150% of the limit (Limit factor: 1.00)
Observation time: 150s
Windows width: 10 periods - (EN/IEC 61000-4-7 Edition 2002 + A1:2008)
Customer: COIZ
E. U. T.: AD19A
Temperature : (20.5 ~ 22.0) ℃
Humidity : (52.0 ~ 54.7) % R.H.
Test Result
E. U. T.: PASS
Power Source: PASS
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 35 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
E. U. T. Result
Check harmonics 2..40 [exception odd 21..39]:
Harmonic(s) > 150%:
Order (n): None
Harmonic(s) with average > 100%:
Order (n): None
Check odd harmonics 21..39:
All Partial Odd Harmonics below partial limits.
Harmonic(s) > 150%:
Order (n): None
Harmonic(s) with average > 150%:
Order (n): None
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 36 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
Average harmonic current results
Hn Ieff [A] % of Limit Limit [A] Result
1 57.128E-3
2 10.194E-3 1.049 972.00E-3 PASS
3 52.471E-3 2.535 2.07 PASS
4 8.632E-3 2.230 387.00E-3 PASS
5 50.933E-3 4.964 1.03 PASS
6 9.490E-3 3.515 270.00E-3 PASS
7 49.517E-3 7.145 693.00E-3 PASS
8 9.436E-3 4.559 207.00E-3 PASS
9 47.036E-3 13.066 360.00E-3 PASS
10 8.862E-3 5.351 165.60E-3 PASS
11 44.192E-3 14.879 297.00E-3 PASS
12 8.519E-3 6.173 138.00E-3 PASS
13 40.909E-3 21.645 189.00E-3 PASS
14 8.107E-3 6.853 118.29E-3 PASS
15 37.106E-3 27.486 135.00E-3 PASS
16 7.617E-3 7.359 103.50E-3 PASS
17 33.142E-3 27.823 119.11E-3 PASS
18 7.097E-3 7.714 92.00E-3 PASS
19 29.160E-3 27.361 106.58E-3 PASS
20 6.607E-3 7.980 82.80E-3 PASS
21 25.227E-3 26.162 96.43E-3 PASS
22 6.139E-3 8.156 75.28E-3 PASS
23 21.357E-3 24.256 88.05E-3 PASS
24 5.685E-3 8.239 68.99E-3 PASS
25 17.636E-3 21.773 81.00E-3 PASS
26 5.310E-3 8.338 63.69E-3 PASS
27 14.152E-3 18.870 75.00E-3 PASS
28 4.950E-3 PASS
29 10.974E-3 15.715 69.83E-3 PASS
30 4.638E-3 PASS
31 8.216E-3 12.578 65.32E-3 PASS
32 4.360E-3 PASS
33 6.043E-3 9.848 61.36E-3 PASS
34 4.082E-3 PASS
35 4.488E-3 PASS
36 3.827E-3 PASS
37 3.660E-3 PASS
38 3.628E-3 PASS
39 3.551E-3 PASS
40 3.360E-3 PASS
Harmonic currents less than 0.6% of the input current measured under the test conditions, or less than 5 mA, whichever is greater, are
disregarded.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 37 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
Maximum harmonic current results
Hn Ieff [A] % of Limit Limit [A] Result
1 58.802E-3
2 10.335E-3 0.478 2.16 PASS
3 54.051E-3 1.175 4.60 PASS
4 8.787E-3 1.022 860.00E-3 PASS
5 52.345E-3 2.296 2.28 PASS
6 9.676E-3 1.613 600.00E-3 PASS
7 50.977E-3 3.310 1.54 PASS
8 9.571E-3 2.081 460.00E-3 PASS
9 48.524E-3 6.066 800.00E-3 PASS
10 8.982E-3 2.441 368.00E-3 PASS
11 45.446E-3 6.886 660.00E-3 PASS
12 8.656E-3 2.823 306.66E-3 PASS
13 41.883E-3 9.972 420.00E-3 PASS
14 8.235E-3 3.133 262.86E-3 PASS
15 37.910E-3 12.637 300.00E-3 PASS
16 7.719E-3 3.356 230.00E-3 PASS
17 33.825E-3 12.779 264.70E-3 PASS
18 7.197E-3 3.520 204.44E-3 PASS
19 29.762E-3 12.566 236.84E-3 PASS
20 6.704E-3 3.643 184.00E-3 PASS
21 25.706E-3 11.997 214.28E-3 PASS
22 6.247E-3 3.734 167.28E-3 PASS
23 21.633E-3 11.056 195.66E-3 PASS
24 5.784E-3 3.773 153.32E-3 PASS
25 17.806E-3 9.892 180.00E-3 PASS
26 5.406E-3 3.820 141.54E-3 PASS
27 14.289E-3 8.574 166.66E-3 PASS
28 5.037E-3 3.833 131.42E-3 PASS
29 11.119E-3 7.165 155.18E-3 PASS
30 4.748E-3 PASS
31 8.386E-3 5.777 145.16E-3 PASS
32 4.452E-3 PASS
33 6.213E-3 4.556 136.36E-3 PASS
34 4.185E-3 PASS
35 4.611E-3 PASS
36 3.946E-3 PASS
37 3.853E-3 PASS
38 3.717E-3 PASS
39 3.805E-3 PASS
40 3.453E-3 PASS
Harmonic currents less than 0.6% of the input current measured under the test conditions, or less than 5 mA, whichever is greater, are
disregarded.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 38 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
Maximum harmonic voltage results
Hn Ueff [V] Ueff [%] Limit [%] Result
1 230.38 100.165
2 167.13E-3 0.073 0.2 PASS
3 91.97E-3 0.040 0.9 PASS
4 20.01E-3 0.009 0.2 PASS
5 11.30E-3 0.005 0.4 PASS
6 8.37E-3 0.004 0.2 PASS
7 9.87E-3 0.004 0.3 PASS
8 5.05E-3 0.002 0.2 PASS
9 9.31E-3 0.004 0.2 PASS
10 7.59E-3 0.003 0.2 PASS
11 8.48E-3 0.004 0.1 PASS
12 7.02E-3 0.003 0.1 PASS
13 8.96E-3 0.004 0.1 PASS
14 10.30E-3 0.004 0.1 PASS
15 8.48E-3 0.004 0.1 PASS
16 9.54E-3 0.004 0.1 PASS
17 9.02E-3 0.004 0.1 PASS
18 7.10E-3 0.003 0.1 PASS
19 10.60E-3 0.005 0.1 PASS
20 5.05E-3 0.002 0.1 PASS
21 9.80E-3 0.004 0.1 PASS
22 7.90E-3 0.003 0.1 PASS
23 10.71E-3 0.005 0.1 PASS
24 6.72E-3 0.003 0.1 PASS
25 8.12E-3 0.004 0.1 PASS
26 6.72E-3 0.003 0.1 PASS
27 8.82E-3 0.004 0.1 PASS
28 7.45E-3 0.003 0.1 PASS
29 8.41E-3 0.004 0.1 PASS
30 5.52E-3 0.002 0.1 PASS
31 5.16E-3 0.002 0.1 PASS
32 6.23E-3 0.003 0.1 PASS
33 7.70E-3 0.003 0.1 PASS
34 4.25E-3 0.002 0.1 PASS
35 7.93E-3 0.003 0.1 PASS
36 4.94E-3 0.002 0.1 PASS
37 8.65E-3 0.004 0.1 PASS
38 4.75E-3 0.002 0.1 PASS
39 5.83E-3 0.003 0.1 PASS
40 6.01E-3 0.003 0.1 PASS
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 39 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 40 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.5.4 Setup photos
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 41 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.6 Voltage fluctuations and flicker emission
Test standard EN 61000-3-3:2013
Test date 2019.07.31
Test facility Shielded room (#1)
Test voltage AC 230 V, 50 Hz
Temperature (20.8 ~ 21.7) C
Relative humidity (49.5 ~ 51.9) % R.H.
Test result Comply
6.6.1 Measurement procedure
The voltage changes at the supply terminals were measured across the complex reference
impedance Z = (0.4 + j0.25) Ω. The short-term flicker values are measured during a time interval of
10 min. Dc = relative voltage change between two steady states and dmax = maximum single voltage
change, are measured over the reference impedance. The test set-up photo Is included in appendix 3.
Short time (Pst) 10 min
Observation time 120 min (12 Flicker measurement)
Flicker impedance Zref (IEC 60725)
6.6.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. Date
Digital power analyzer
DPA 500N EMTEST P1437139796 2020.04.24
(Harmonics & Flicker)
AC Power Supply PCR4000M KIKUSUI UK001881 2020.05.10
* All test equipment used is calibrated on a regular basis.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 42 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.6.3 Measurement result
Report title: T1906-0077
Company Name: LAB-T
Date of test: 2019.07.31
Tester: SJH
Standard used: EN/IEC 61000-3-3 Ed.3 Flicker
Short time (Pst): 10 min
Observation time: 120 min (12 Flicker measurements)
Flickermeter: 230V / 50Hz according IEC 61000-4-15 Ed.2
Flicker Impedance: Zref (IEC 60725)
Customer: COIZ
E. U. T.: AD19A
Temperature : (20.8 ~ 21.7) ℃
Humidity : (49.5 ~ 51.9) % R.H.
Test Result PASS
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 43 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
Maximum Flicker results
EUT values Limit Result
Pst 0.028 1.00 PASS
Plt 0.028 0.65 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.159 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Detail Flicker data
Flicker EUT values Limit Result
measurement 1
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.159 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 2
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.065 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 3
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.063 4.00 PASS
Tmax [s] 0.000 0.50 PASS
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 44 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
Flicker EUT values Limit Result
measurement 4
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.062 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 5
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.063 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 6
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.064 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 7
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.063 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 8
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.064 4.00 PASS
Tmax [s] 0.000 0.50 PASS
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 45 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
Flicker EUT values Limit Result
measurement 9
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.062 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 10
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.063 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 11
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.063 4.00 PASS
Tmax [s] 0.000 0.50 PASS
Flicker EUT values Limit Result
measurement 12
Pst 0.028 1.00 PASS
dc [%] 0.000 3.30 PASS
dmax [%] 0.062 4.00 PASS
Tmax [s] 0.000 0.50 PASS
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 46 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
6.6.4 Setup photos
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 47 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7. Immunity Test results
7.1 Electrostatic discharge
EN 55014-2:2015
Test standard
EN 61000-4-2:2009
Test date 2019.08.01
Test facility Shielded room (#2)
Test voltage AC 230 V, 50 Hz
Temperature (20.8 ~ 21.7) C
Relative humidity (48.1 ~ 52.4) % R.H
Air pressure 101.1 kPa
Performance criterion B
Test result Comply
7.1.1 Measurement procedure
A ground reference plane was located on the floor, and connected to earth via a low impedance
connection. The return cable of the ESD generator was connected to the reference plane. In case of
table top equipment, EUT was placed on the reference plane on 80 cm of insulating support.
Also if the EUT is floor-standing equipment, it was placed on a non-conducted support with a height
up to between 0.05 m and 0.15 m the reference ground plane. And a vertical coupling plane (VCP) of
(0.5 x 0.5) m was located 10 cm from the EUT’s sides. The VCP was connected to the reference
plane via a cable with a 470 kΩ (2ea) resistor. The test was made by applying contact and air
discharges to the EUT and contact discharges to the VCP/HCP. When applying the discharges to the
VCP the tip of the generator was located at the middle edge of the VCP. The VCP was located 10 cm
from each side of the EUT.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 48 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
* Conditions during testing
Discharge Test levels Number of discharge
HCP/VCP ±4 ㎸
Contact ±4 ㎸ 10
Air ±8 ㎸
7.1.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
ESD SIMULATOR ESD NX30 EMTEST 23028 2020.06.21
* All test equipment used is calibrated on a regular basis.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 49 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.1.3 Test point and result
[Front]
2
5
3
4
[Rear]
1
3
4
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 50 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Remote control - Front]
5
[Remote control - Rear]
5
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 51 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
* Indirect applied
ESD point Test levels Discharge Results
HCP ±4 ㎸ - -
VCP ±4 ㎸ Contact A
* Directly applied
ESD point Test levels Discharge Results
(1) external case(Metal) ±4 ㎸ Contact A
(2) control part ±8 ㎸ Air A
(3) a connecting line ±4 ㎸ Contact A
(4) prop ±4 ㎸ Contact A
(5) external case(Nonmetal) ±8 ㎸ Air A
* Comment / Performance
- Passed as EUT was in normal operation during the test.
- HCP test is excluded as the floor-Installation type products.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 52 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.1.4 Test setup photos
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 53 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.2 Radio-frequency electromagnetic field
EN 55014-2:2015
Test standard
EN 61000-4-3:2006+A2:2010
Test date -
Test facility -
Test voltage -
Temperature -
Relative humidity -
Air pressure -
Performance criterion -
Test result Not applicable
7.2.1 Measurement procedure
The EUT has been placed in center of a wooden turntable. The height of this table was 0.8 m.
Also if the EUT is floor-standing equipment, it was placed on a non-conducted support with a height
up to between 0.05 m and 0.15 m the reference ground plane. The field strength was monitored by an
isotropic sensor during the complete test. The isotropic sensor was located beside the equipment. The
antenna has been orientated for both horizontal and vertical polarization. The distance between
antennas the equipment under testing was at least 3 m. The tests have been performed with the
antenna facing each of the four side of the EUT.
* Conditions during testing
Frequency range 80 ㎒ ~ 1 000 ㎒
Test level 3 V/m (unmodulated, r.m.s)
Modulation AM 80 % (1 kHz Sine wave)
Step 1 % step
Dwell time 1s
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 54 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.2.2 Test configuration
N/A
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 55 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.2.3 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
- - - - -
* All test equipment used is calibrated on a regular basis.
7.2.4 Measurement result
Performance result
Test level Exposed Side
Horizontal Vertical
Front side - -
Back side - -
3 V/m
Left side - -
Right side - -
* Comment / Performance
- Not applicable
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 56 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.2.5 Test setup photos
N/A
(It was excluded from the test because it was classified as Category II)
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 57 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.3 Electrical fast transients
EN 55014-2:2015
Test standard
EN 61000-4-4:2012
Test date 2019.08.01
Test facility Shielded room (#2)
Test voltage AC 230 V, 50 Hz
Temperature (20.7 ~ 22.2) C
Relative humidity (46.8 ~ 51.0) % R.H.
Air pressure 101.1 kPa
Performance criterion B
Test result Comply
7.3.1 Measurement procedure
The EUT has been placed on a wooden table 10 cm above the reference ground plane.
The reference ground plane exceeded the projected geometry of the EUT and the capacitive clamp
by more than 20 cm. The clamp has placed directly on the reference ground plane. The distance
between the EUT and all other conductive structures except the ground plane beneath the EUT was
more than 50 cm. The distance between noise generator and EUT was about 50 cm.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 58 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
* Conditions during testing
Input and output a.c. power ports: ± 1 ㎸
Test level Input and output d.c power ports: ± 0.5 ㎸
Signal ports and control lines ports: ± 0.5 ㎸ (Peak)
Repetition rate 5 ㎑
Calibration of the coupling/decoupling network (5.5 ± 1.5) ns
Rise time
Calibration of the capacitive coupling clamp (5 ± 1.5) ns
Calibration of the coupling/decoupling network (45 ± 15) ns
Impulse duration
Calibration of the capacitive coupling clamp (50 ± 15) ns
Test duration 15 ms
Test period 300 ms
Coupling time 120 s
7.3.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
Ultra Compact
UCS 500N7 EMTEST P1823221015 2020.06.17
Simulator
Motorized Variac NX1-260-16 EMTEST P1909227468 -
* All test equipment used is calibrated on a regular basis.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 59 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.3.3 Measurement result
* Input and output a.c. power ports
Results
Coupling point Test level
(Positive / Negative)
L ±1 ㎸ A
N ±1 ㎸ A
L-N ±1 ㎸ A
PE ±1 ㎸ A
L-PE ±1 ㎸ A
N-PE ±1 ㎸ A
L-N-PE ±1 ㎸ A
* Signal ports and control lines ports
Results
Coupling point Test level
(Positive / Negative)
Not applicable - -
* Comment / Performance
- Passed as EUT was in normal operation during the test.
- Telecommunication test is excluded as the signal cable is less than 3m.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 60 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.3.4 Setup photo
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 61 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.4 Surges
EN 55014-2:2015
Test standard
EN 61000-4-5:2014+A1:2017
Test date 2019.08.01
Test facility Shielded room (#2)
Test voltage AC 230 V, 50 Hz
Temperature (21.1 ~ 22.6) C
Relative humidity (49.8 ~ 51.9) % R.H.
Air pressure 101.1 kPa
Performance criterion B
Test result Comply
7.4.1 Measurement procedure
The test consists of the injection of slow high energy transients in the AC/DC mains supply lines in
both line-to-line and line-to-ground coupling mode, and into the signal and extra low voltage supply
lines in line-to-ground coupling mode. The impedance of the transient generator is characterized by
the shape of the open-circuit voltage and the circuit current pulses.
To simulate typical installation impedances, 40 are inserted when the generator when extra low
voltage and signal lines are tested, and 10 are inserted when the line-to-ground test is conducted on
the AC/DC mains lines. The test pulses are coupled into the leads to be tested by means of
appropriate coupling networks, which maintain the test pulses within their specification. The reference
ground plane exceeded the projected geometry of the EUT and the back filler by more than 20 cm.
The back filler has been placed directly on a separated reference ground plane. Both ground planes
were connected together. The ground terminal of the back filler has been connected directly with its
reference ground plane.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 62 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
* Conditions during testing
± 1 ㎸ (Line - Line)
Test level
± 2 ㎸ (Line - PE)
Open-circuit voltage 1.2 / 50 ㎲
Short-circuit current 8 / 20 ㎲
Number of surge / time 1 time / 60 s
Angle 90° / 270°
Number of surge 5 surge / polarity
7.4.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
Ultra Compact Simulator UCS 500N7 EMTEST P1823221015 2020.06.17
Motorized Variac NX1-260-16 EMTEST P1909227468 -
* All test equipment used is calibrated on a regular basis.
7.4.3 Measurement result
* Input and output a.c. power ports
Results
Coupling point Test level
(Positive / Negative)
L-N ±1 ㎸ A
L-PE ±2 ㎸ A
N-PE ±2 ㎸ A
* Comment / Performance
- Passed as EUT was in normal operation during the test.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 63 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.4.4 Setup photos
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 64 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.5 Injected currents (0.15 MHz to 230 MHz)
EN 55014-2:2015
Test standard
EN 61000-4-6:2014
Test date 2019.08.01
Test facility Shielded room (#2)
Test voltage AC 230 V, 50 Hz
Temperature (21.4 ~ 22.6) C
Relative humidity (49.8 ~ 53.2) % R.H.
Air pressure 101.1 kPa
Performance criterion A
Test result Comply
7.5.1 Measurement procedure
The EUT has been placed on a wooden table 10 cm above the reference ground plane.
The reference ground plane exceeded the projected geometry of the EUT and the Coupling
/Decoupling Network (CDN) by more than 30 cm. The CDN has been placed directly on the
reference ground plane. The distance between CDN and EUT has a length of 30 cm.
* Conditions during testing
Test rang 0.15 ㎒ to 80 ㎒ 0.15 ㎒ to 230 ㎒ (Category II)
Input and output a.c. power ports: 3 V (r.m.s.) (Unmodulated)
Test level Input and output d.c. power ports: 1 V (r.m.s.) (Unmodulated)
Signal lines and control lines ports:1 V (r.m.s.) (Unmodulated)
Test modulation 80 % AM (1 kHz sine wave)
Source impedance 150 Ω
Step 1 % step
Dwell time 1s
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 65 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.5.2 Test equipment used
Equipment Model Serial number Manufacturer Next cal. Date
Compact immunity test
NGC 4070C TESEQ 540128 2020.06.17
system
Attenuator ATN 6150 TESEQ 18120512 2020.06.17
CDN CDN M332 TESEQ 53239 2020.06.17
* All test equipment used is calibrated on a regular basis.
7.5.3 Measurement result
* Input and output a.c. power ports
Coupling point Coupling device Results
L-N-PE M3 A
* Signal lines and control lines ports
Coupling point Coupling device Results
Not applicable - -
* Comment / Performance
- Passed as EUT was in normal operation during the test.
- Signal test is excluded as the signal cable is less than 3 m.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 66 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.5.4 Setup photos
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 67 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.6 Voltage dips and interruptions
EN 55014-2:2015 (Category II)
Test standard
EN 61000-4-11:2004+A1:2017
Test date 2019.08.01
Test facility Shielded room (#2)
AC 100 V, 50/60 Hz
Test voltage
AC 240 V, 50/60 Hz
Temperature (20.6 ~ 22.1) C
Relative humidity (49.7 ~ 51.7) % R.H.
Air pressure 101.1 kPa
Performance criterion C
Test result Comply
7.6.1 Measurement procedure
The EUT is tested to each of the specified voltage variations, three times at 10 s interval for the most
representative modes of operations.
The EUT shall be tested for each selected combination of test level and duration with a sequence of
three dips/interruptions with intervals of 10 s minimum (between each test event). Each representative
mode of operation shall be tested.
* Conditions during testing
Environmental Test level Durations for voltage dips
Test set-up
phenomena in % UT 50 Hz 60 Hz
100 0 0.5 cycle 0.5 cycle
Voltage IEC 61000-4-11
dips in % 60 40 10 cycle 12 cycle Voltage change shall
UT occur at zero crossing
30 70 25 cycle 30 cycle
UT is the rated voltage of the equipment under test.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 68 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.6.2 Test equipment used
Equipment Model Manufacturer Serial number Next cal. date
Ultra Compact Simulator UCS 500N7 EMTEST P1823221015 2020.06.17
Motorized Variac NX1-260-16 EMTEST P1909227468 -
* All test equipment used is calibrated on a regular basis.
7.6.3 Measurement result
* AC 100 V, 50/60 Hz
Voltage dips in (%UT) Duration(Period) Angle(°) Count Number Results
100 0.5 A
60 10 / 12 0 / 180 3T A
30 25 / 30 A
* AC 240 V, 50/60 Hz
Voltage dips in (%UT) Duration(Period) Angle(°) Count Number Results
100 0.5 A
60 10 / 12 0 / 180 3T A
30 25 / 30 A
* Comment / Performance
- Passed as EUT was in normal operation during the test.
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 69 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
7.6.4 Setup photos
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 70 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
8. Photographs of the product
[All set]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 71 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Front]
[Rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 72 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Left]
[Right]
[Inside 1]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 73 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Inside 2]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 74 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Inside 3]
[Remote control Inside]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 75 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Remote control Front]
[Remote control Rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 76 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[SMPS Front]
[SMPS Rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 77 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[SMPS Inside]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 78 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[SMPS board front]
[SMPS board rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 79 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Inside fan front]
[Inside fan rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 80 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Inside motor front]
[Inside motor rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 81 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Inside motor left]
[Inside motor right]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 82 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Inside motor top]
[Inside motor bottom]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 83 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Noise filter 1 front]
[Noise filter 1 rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 84 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Noise filter 2 front]
[Noise filter 2 rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 85 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Remote control board front]
[Remote control board rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 86 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Controller board 1 front]
[Controller board 1 rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 87 / 88 ]
http://www.lab-t.net
2182-42, 2182-40 Baegok-daero, Mohyeon-eup Test report No.:
Cheoin-gu, Yongin-si, Gyeonggi-do 17036, Korea TRECEE19-0126
Main: +82-31-322-6767 / Fax: +82-31-322-6768
[Controller board 2 front]
[Controller board 2 rear]
TRF-E-016(02) It shall not be reproduced except in full, without the written approval of Lab-T, Inc. [Page 88 / 88 ]
Barcelona
Pol. Ind. La Valldan
Camí de Garreta 92,
08600 Berga
Barcelona – Spain
T. +34 93 824 90 20
E.
[email protected]
W. www.weelko.com
B-EQUIPMENT
Cold and hot hammer instrument. The heat relaxes the Cold & Hot Hammer
Input voltage: 220 ~240V
muscles and opens the pores. The cold reduces Ref. F318A Input frequency: 50Hz/60Hz
wrinkles and smoothes skin.
Weight: 2Kg Working temperature: min. 5ºC/max. 40ºC
Equipo con martillo de frío y calor. El calor se usa para Size: 20,5x22x7 cm
relajar los músculos y abrir los poros. El frío se utiliza
para reducir las arrugas y suavizar la piel.
PACKING SIZE cm
Appareil à marteau froid et chaud. La chaleur
37x27,5x14
décontracte les muscles et ouvre les pores. Le froid
N. W. Kg
réduit les rides et adoucit la peau. 2
G. W. Kg
Kälte- und Wärmehammer. Die Wärme entspannt die 2,5
Muskeln und öffnet die Poren. Die Kälte mindert Falten
und macht die Haut weicher.
Apparecchio con martello freddo/caldo. Il calore si usa
per rilassare i muscoli e aprire i pori. Il freddo si utilizza
per ridurre le rughe e rendere soave la pelle.
Technical Zoom
Beauty February 2015
INDEX
Description multifunctional tower B-System 2
Components B-Equipment and functions 4
How to make the most of your B-Equipment? 6
Assembly and installation 7
Technical speci cations 12
Webside information 15
The multifunctional tower B-System
from Weelko
Do you aim to offer a complete service to your customers but you count on a small
space? Make use of the versatile B-System and build up your workspace to measure.
Enlarge your treatment range at your
own pace!
Our technical ZOOM reveals all the details and
functions and affords you the possibility to know our
equipment in great depth. The compact multifunctio-
nal trolley B-system is put in the Weelko spotlight.
Discover everything here inside!
Beauty| Multifunctional tower B-System from Weelko Page 2
DESCRIPTION AND COMPONENTS OF THE B-SYSTEM EQUIPMENT (REF. 3021)
FROM WEELKO
All about the multifunctional tower B-System
To keep yourself healthy and young to the maximum, you should take care of your health at each
level and carry out regular treatments that will doubtlessly help maintain a much healthier skin.
The multifunctional tower B-System from Weelko (Ref. 3021) offers you the possibility to combine
different functions so as to create an ideal work tool that exactly adapts to the speci c work needs
of each professional.
Components and B-System functions
The indispensable tools are standardly adapted to the set which consists of a steamer, a mag-
nifying lamp, a multifunctional trolley and a towel warmer. The tower is likewise composed by
6 compartments to make your B-System to measure, according to the certain needs at any time.
In our catalogue, you will nd the series B-EQUIPMENT, consisting of 9 instruments of different
functions so that you can complete your multifunctional tower.
At your disposal there is the skin scrubber, the ultrasound, the galvanic, the high frequency, the
rotary brushes, the vacuum and spray, the high-conductivity glove and bio-lifting, the hot and
cold hammer and the microdermabrasion by diamond heads. Find out here all the possibilities
and create your own B-System!
Beauty| Multifunctional tower B-System from Weelko Page 3
Standard components
B-SYSTEM STEAMER (Ref. F-300DT)
In any cutaneous treatment, the ozone
steamer is an essential tool that provides
multiple bene ts to the skin. For example,
it softens the skin and the dead cells befo-
re the facial cleansing treatment, it helps
open the pores so as to obtain a deeper
exfoliation, it contributes to eliminate grea-
se, blackheads, traces of make-up and dirt,
it helps release natural toxins of the skin, it
smoothes wrinkles before the treatments,
it increases the blood circulation, it hydra-
tes the skin, it improves its metabolism and helps renew it. The ozone applied through the stea-
mer has a germicide and anti-in ammatory effect. To provide extra comfort to the patient, the
steamer F-300DT includes the aromatherapy option. The essential fragrances can be soaked up
in a small pad placed next to the steamer vent in a practical accessible compartment.
Steamer treatment table according to the skin type
(Guiding instructions to be considered by
the professional at each treatment)
Skin type Distance Time
Greasy skin or with circulation problems 35-40cm 15 Minutes aprox.
Normal or healthy skin 35-40cm 10 Minutes aprox.
Dry or sensitive skin 40-45cm 5 Minutes aprox.
Dilated capillary or in amed blood vessels 40cm 5 Minutes aprox.
MAGNIFYING LAMP AMPLI (Ref. 1005T)
The magnifying lamp AMPLI (Ref. 1005T) is designed with a
LED system that ensures a longer lifespan, apart from saving
more energy (compared to other cold-light lamps) and being
less liable to be defective. This lamp is conceived for any kind
of treatment (skin diagnose, skin examination before and after
treatments, semipermanent make-up and eyelash tinting).
TOWEL WARMER WARMEX (Ref. T-01)
An essential tool to offer a relaxing and
comfortable moment during the treatments.
WARMEX is equipped with ultraviolet light
that acts as a germicide so that you can have
wet and hot towels in perfect condition.
Beauty| Multifunctional tower B-System from Weelko Page 4
Components B-EQUIPMENT to complete the unit
Weelko has made 9 basic functions in facial treatments available for you in order you to custo-
mise the equipment. The multifunctional trolley can lodge 6 devices (to be chosen amongst the 9
functions shown here below). Another advantage of B-System is that you do not need to make up
your mind at once, but you can complete your work station step by step!
SKIN SCRUBBER (REF. 311A)
The exfoliating spatula cleans at a very great
depth and stimulates the cell regeneration. It
emits very low frequency waves and vibrations
that break the dirt adhered to the skin into tiny
particles and soften without the steamer help.
This makes very easy eliminating dead cells,
blackheads, the dirt and multiple traces of the
skin without damaging it and minimising the
pain. The result: hydrated and radiant skin with
a younger aspect. The skin must be slightly wet
during the treatment, a facial tonic must be used.
ULTRASOUND (REF. 312A)
The application of ultrasounds favors the re-
duction of rides and scars, contributes to skin
hydration and helps cosmetic actives to be
well absorbed into the skin. The treatment
with ultrasounds improves blood circulation
and increases cell regeneration. The Weelko
device allows choosing the type of wave: con-
tinuous for normal type skin and thicker tissue
areas; combined wave for smoother skin areas
and intermittent for sensitive skins. Always use
a conductive lotion on the skin. Never apply on
dry skin.
GALVANIC (REF. 313A)
The Weelko system for treatments with gal-
vanic current means a progress regarding the
traditional equipment, in which the patient ne-
eds to hold the pole on his/her hand during the
treatment. The B-Equipment Weelko works when
both handles are in touch with the area to treat.
Multiple bene ts are highlighted, such as rinsing
treatments that soften and blend the grease de-
posits accumulated as fat in the follicles and also
ionisation treatments that promote the products
to be introduced into the skin. Apply a polarised
lotion and select the pole in the corresponding
handle, since the instrument needs to be set up
with the same polarity as the product.
Beauty| Multifunctional tower B-System from Weelko Page 5
HIGH FREQUENCY (REF. 314A)
This function is used to clean the skin, kill skin
bacteria, regulate the glandular activity and
prepare the skin before the galvanic treatment
by improving like that the penetration of the
product into the skin. The conductive element
that runs inside the electrodes can be argon,
with its particular purple colour, used to disin-
fect or the neon, which gives it s orange colour,
applied to give tone to the skin. No product is
needed during the treatment.
ROTARY BRUSHES (REF. 315A)
The mechanical exfoliating process by means
of brushes eliminates outer dead cells and the
skin dirt, as well as the sweat excess or make-
up traces. Through a pleasant massage it also
stimulates blood circulation. The rotation
speed is adjustable. An exfoliating product is
needed to dampen the brush before starting
the treatment.
VACUUM AND SPRAY (REF. 316A)
The vacuum function is the great ally of blackhead
and whitehead removal, as well as acne or pore
obstruction, daily dirt and persistent grease, such
as make-up. By spraying a tonic, you can reduce
the visibly reddened or in amed areas. It also helps
calm the skin.
HIGH CONDUCTIVITY GLOVE AND BIO-LIFTING
(F-317A)
A massage using high conductivity gloves or electrodes
helps improve the skin texture due to the activation of
collagen and elastin. Its function is to stimulate the
production of ATP, and provide the cells with more energy,
so they can increase the production of fibroblasts, and by
doing so, they give more energy to the muscles and tissue,
resulting in a reaffirmation and strengthening effect.
This is a painless treatment and the client won't feel any
muscular contraction, because it's a current very similar to
the one produced by the body. The intensity is very low and
its charge is not enough to stimulate the peripheral nerve
fibres, therefore its application is at a sub sensorial level.
The client usually can't feel the passage of current, but will
see the effects of the tissue repair soon.
In both cases it's necessary to use conducting gel and
dampen the gloves.
Beauty| Multifunctional tower B-System from Weelko Page 6
COLD AND HOT HAMMER (REF. 318A)
The heat that gives the hammer improves the
blood circulation, relaxes the muscles and opens
the pores for both improve the product penetra-
tion and prepare the skin for a deep cleansing.
The cold side is used to calm the skin after the
microdermabrasion treatments, to close the po-
res and to reduce rides. No product is needed in
both cases.
MICRODERMABRASION BY DIAMOND
HEADS (REF.319A)
Microdermabrasion is a precise and progressi-
ve controlled exfoliation on the horny layer (the
most external layer of the epidermis), its result is
a smooth, fresh and clean skin. It is specially de-
signed to treat ageing, scars as a consequence of
acne, stretch marks, blackheads and to prepare
the skin for applying nutrients. It is a non-invasive
and painless treatment for both young and adult
skins and it does not require any product. Results
can be noticed at the end of each session.
Remember that all the instruments of the series B-Equipment are also designed
as table-top devices, so you can also place them on another trolley if you wish
so.
Make the most of your B-System!
Here below we offer you, as an example, some advices to combine functions and to take advantage of all
the possibilities that your multifunctional tower offers you:
•Warm up a towel and place it on the patient’s skin to prepare and relax it before the treatment.
•Dampen the small pads of the B-System steamer with some fragrance, but remember to always use
these pads. Never pour fragrances directly into the water reservoir.
•Prepare the skin and open the pores with the B-System steamer or use the heat of the hammer.
•You can now carry out the spot removal with the help of the magnifying lamp Ampli.
•Eliminate the spots on the skin and the daily accumulated dirt with a cleansing lotion with the help of
the skin scrubber.
•Start the deep exfoliating session with the microdermabrasion by diamond heads.
•In order to close the pores, calm and soften the skin, use the cold side of the hammer.
•Apply a refreshing and hydrating lotion and sun lter cream (depending on the time of the year and
weather).
Beauty| Multifunctional tower B-System from Weelko Page 7
ASSEMBLY AND INSTALLATION
Here below the elements that compose the multifunctional trolley are listed
MAIN PART 3
2
1
1. Support for accessories (left)
2. Support for accessories (right)
3. Top shelf
4
5 6
4. Trolley base
5. Wheel trim
6. Support for steamer
7 8 9
7. Cable
8. Wheels
9. Support for lamp
Assembly of the wheels at the base:
Screw the wheels at the base. You can use for that the tools included in the packaging
(Image 1 and 2)
Image 2
Image 1
Beauty| Multifunctional tower B-System from Weelko Page 8
Place the corner pieces in the correct position and screw (Image 3 and 4)
Image 3
Image 4
Turn the base to assemble the tower on it. Place the screws to attach the support for
(Image 5) the lamp. (Image 6)
Place the tower laterally and screw the
base and the support for the lamp (Image
7).
Screw the glass shelf (Image 8)
Beauty| Multifunctional tower B-System from Weelko Page 9
Once the tower is assembled, screw the lateral supports for the accessories. Screw the support
for the steamer with the Allen key too (Image 9). Now you can place the steamer on its support
(Image 10).
Image 10
Image 9
Unpack the rest of instruments B-Equipment that form your B-System choice. Introduce them
into the lodgements carefully.
Make sure that you properly place them up to the rear to be plugged at the column (Image 11).
Once you have all the equipment introduced, screw them (Image 12).
Image 12
Image 11
Beauty| Multifunctional tower B-System from Weelko Page 10
Now you can place the towel warmer. Check that all packing has been removed from the base.
Tighten the equipment against the column to make sure they are well plugged (Image 13).
Image 13
Plug all the accessories to their corresponding device (Image14). For that, keep the trolley
doors opened and make the cables go through the lower opening of the door. Check that
none of the cables has been caught and close the tower doors (Image 15).
Image 14
Image 15
Beauty| Multifunctional tower B-System from Weelko Page 11
Introduce the lamp into its support and plug it at the rear of the column (Image 16). Plug the
power cable at the rear of the column. Your B-System is ready for use! Read attentively the ins-
tructions of use of each device and sterilise the accessories before using them.
Image 16
Beauty| Multifunctional tower B-System from Weelko Page 12
TECHNICAL SPECIFICATIONS
STEAMER B-SYSTEM (REF. F-300DT):
Adjustments Head height (arm at 45º) 140 cm
Arm rotation 150º
Head rotation 3D
Time for warming-up 6 minutes (aprox.)
Features Capacity of water reservoir 1,3 litres
Ozone to be switched on independently
Aromatic oil receptacle
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 800 W
MAGNIFYING LAMP AMPLI (REF. 1005T)
Features 3 dioptres with a focal spot of 5 dioptres
Intensity adjustment in 2 xed positions
Características eléctricas Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 6W
TOWEL WARMER WARMEX (REF. T-01)
Features Capacity 7 litres
Electrical speci cations Rated voltage 220V—240V
Rated frequency 50 Hz/60 Hz
Rated power 164 W
PSKIN SCRUBBER B-EQUIPMENT (REF. F-311A)
Working features Working power 6W
Working frequency 25-27 KHz
Established working time 15 minutes
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 20 W
ULTRASOUND B-EQUIPMENT (REF. F-312A)
Working features Working power 0-3 W
Working frequency 1 MHz
Established working time 15 minutes
Características eléctricas Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 20 W
Beauty| Multifunctional tower B-System from Weelko Page 13
GALVANIC B-EQUIPMENT (REF. F-313A)
Working features Working intensity 30-500 µA
Working power 0-3 W
Working voltage 30 V
Established working time 15 minutes
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 20 W
HIGH FREQUENCY B-EQUIPMENT (REF. F-314A)
Working features Working power 12 W
Working frequency 650.000 GHz
Established working time 15 minutes
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 20 W
ROTARY BRUSHES B-EQUIPMENT (REF. F-315A)
Working features Rotation speed 0 - 550 RPM
Established working time 3 minutes
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 20 W
VACUUM AND SPRAY B-EQUIPMENT (REF. F-316A)
Working features Aspiration pressure 0 ~ -0,07 MPa
Established working time 15 minutes
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 120 W
HIGH CONDUCTIVITY GLOVE AND BIO-LIFTING B-EQUIPMENT (REF. F-317A)
Working features Working intensity 50 to 600 µA
Working power 2W
Working voltage 30V
Established working time 3 minutes
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 20 W
Beauty| Multifunctional tower B-System from Weelko Page 14
COLD AND HOT HAMMER B-EQUIPMENT (REF. F-318A)
Working features Temperature 5ºC to 40ºC
Established working time 15 minutes
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 20 W
MICRODERMABRASION WITH DIAMOND HEAD B-EQUIPMENT (REF. F-319A)
Working features Working power 65 W
Working pressure 0-0,8 bar
Established working time 15 minutes
Electrical speci cations Rated voltage 220 V—240 V
Rated frequency 50 Hz/60 Hz
Rated power 20 W
Visit our TECHNICAL INFO and download all the technical
information that you may need. Pictures, dossiers,
TECHNICAL INFO technical sheets, etc. Do not miss them!
Descargas
Dossiers | Technical Zoom
Weelko Products Beauty Spa Physiotherapy Podiatry News Downloads
Turn on the WEELKO TV at weelko.com. The best way to
learn how to use our equipment. Because a picture is
worth a hundred words!
weelko.com
Barcelona
Pol. Ind. La Valldan
Camí de Garreta,92
08600 Berga
Barcelona – Spain
T. +34 93 824 90 20
E.
[email protected]
Kasutusjuhend
Sisukord
MesopZoratsiooni
Mesoporatsiooni kirjeldus kirjeldus
Aparaadi komplektatsioon
Tehnilised näitajad ja sertifikaat
Seadme installeerimine
Aparaadi ettevalmistamine protseduuriks
Seadme intensiivsuse määramine
Protseduuri ülesehitus
Intensiivne mesoporatsiooni hooldus
Seadme osade puhastamine
Vastunäidustused
Garantii
Võimalikud vigade allikad
Mesoporatsiooni kirjeldus:
Mesoporatsioon- innovaatiline lähenemine noorusliku väljanägemise jaoks
Dermatseutilises ja esteetilises kosmeetikas on mesoporatsioon
kõige innovaatilisem ja efektiivsem protseduur, transportimaks suure
molekuliga aktiivaineid sügavale dermisesse ilma nahakihte vigastamata.
Töö põhimõte on lihtne ja geniaalne. Peamiste uuringute põhjal võib julgelt
väita, et mesoporatsioon on uusimaid leiutisi elektroporatsiooni valdkonnas.
Piirkonnale tekitatakse kiired madalasageduslikud elektriimpulsid , mis
avavad niiskuse kanalid murdosa sekundi jooksul.
Selle meetodiga saame transportida sügavamale naha sisse suure molekuliga
aktiivaineid. Senini oli see osaliselt võimalik ainult ultraheliga ja
ionoforeesiga, kuigi mitte nii sügavale, kui nüüd mesoporatsiooniga. Selle
meetodiga saate märkimisväärselt efektiivsema ja pikaajalisema tulemuse.
Neid erakordseid tulemusi saate tunda ja näha juba peale esimest
protseduuri. Mitmeid aastaid on ilumaailmas teada olnud mesoteraapia,
selle meetodiga süstitakse naha sisse aktiivaineid. Erinevalt sellest on
mesoporatsioon äärmiselt õrn meetod ilma kõrvalmõjudeta, nahk pole
kahjustatud, puuduvad soovimatud verevalumid. Taastumisaeg ei ole vajalik.
Mesoporatsiooni kasutatakse :
• Naha noorendamiseks
• Naha struktuuri ühtlustamiseks
• Naha vananemise ennetamiseks
• Naha toonuse parandamiseks
• Näokontuuride taastamiseks
• Naha ebapuhtus, rasu, akne
• Kuperosa
• Hüperpigmentatsioon
• Tseluliit
Aparaadi komplektatsioon:
Veenduge, et aparaat oleks õigesti komplekteeritud
1. 1x Mesoporatsiooni aparaat
2. 1x rullik Roll-on
3. 1x Meso roller
4. 1x Elektroodi randmepael
5. 1x Toitekaabel
6. 2x Ühenduskaabel
7. 1x Elektrooniline klamber
Tehnilised näitajad:
LED ekraan koos taustavalgustusega ja
mikropriotsessori juhtimine
Esialgne tase : 37V
Poratsiooni sagedus : 2kHz
Toiteallikas : peamine: 100-240 V ( 50-60 Hz )
sekundaarne 12V DC/ 1,25 A 500 mA
Energiatarve : 7,2W
Mõõtmed: 310 x 221 x 65 mm ( laius x sügavus x kõrgus )
Kaal: 25 kg ( sisaldab ka lisatarvikuid )
Sertifitseeritud
Seadme installeerimine:
1. Ühendage kaabel ( punase otsaga ) Roll-on´i külge ja Meso roller
mesoporatsiooni seadme külge nii , et panete juhtme punase otsa
punasesse ühenduspessa ja teise otsa Roll-on või Meso rolleri
külge.
2. Ühendage kaabel ( musta otsaga ) mis on mõeldud sinise
elektroonilise randmepaela jaoks, mesoporatsiooni seadmega nii,
et panete juhtme musta otsa ühenduspessa ja teise otsa
randmepaela külge.
3. Ühendage toitekaabel seadmega vooluvõrku.
Aparaadi ettevamistamine protseduuriks:
1.Keera spetsiaalne Roll-on pea aktiivaine ampullile
2. Ühenda kaabel ( jämedam punane ) elektoonilise
klambri külge ja klambri kinnitad Roll-on´i külge.
3. Niisutaga sinine randmepael põhjalikult ( ainult pael, mitte
valget klemmi )
4. Asetage sinine randmepael ümber kliendi randme ja
ühendage selle juhe ( jämedam must ) valge klemmiga.
Valge klemm ühendage sinise randmepaelaga. Veenduge, et
randmepael oleks korrektselt tihedalt ümber käe ( kuid ei
tohi avaldada kliendi randmele survet )
5. Puhastage korrektselt naha pind, kus hakkate protseduuri teostama.
6. Keerake jaoturi pead ( roll-on ) ülemist osa natuke vasakule, et avada
Jaotur toimeaine väljutamisks.
7. Lülitage sisse Mesoporatsiooni süsteem
Seadme intensiivsuse määramine:
Sobilik seadme intensiivsus määratakse kooskõlas kliendiga.
Erilist rõhku tuleks pöörata kliendi otsmikul ( „ kolmanda silma „ )
piirkonnale, mis on väga tundlik.
1. Alustuseks niisutage kliendi nahka, kasutades selleks roll-on´i , mis
on kinnitatud Age Defense Liposom aktiivaine ampulli külge. Samal ajal
Intensiivsust ei tohi muuda. Alustaga liposoomi pealekandmist ülejäänud
näopiirkonnale, siis muutke aparaadi intensiivsus madalamaks ja teostage
pealekandmine otsaesise piirkonnas.
2. Seejärel rullige aeglaste ringjate liigutustega kliendi otsaesist ja
suurendage aeglaselt aparaadi intensiivsust.
3. Protseduuri jooksul veendu, et intensiivsus oleks kliendile vastuvõetav.
Kui vajadus, siis saate intensiivsust vastavalt tundlikkusele muuta.
Protseduuri ülesehitus:
Alustustuseks hoolda terve nägu aeglaste ringjate
liigutustega ( u 5 min ) kasutades selleks roll-on seadet,
samal ajal kohanda dosaatorit nii, et nahk oleks stabiilselt
niisutatud kogu protseduuri vältel
n
Järgnevalt keera aparaadi intensiivsus „0“ , enne
kui eemaldad voolujuhtme ( valge otsik ) ja
ühendad selle meso-rollerile.
Osutades õrna survet meso-rollerile töötle
aktiivainet seni, kuni kõik on naha sisse imendunud.
Järgnevalt teosta samad sammud 1-3 kasutades
Selleks roll-on, Meso-rollerit ja spetsiaalset Age
Defense aktiivaine ampulli, vastavalt näonaha tüübile ja
hoolduse eesmärgile.
Nüüd kanna näole Hydrogel Face and Neck Mask,
Hydro marin Alginat Mask või Mask Powder. Peale
maski eemaldamist, eemalda niiske vatipadjaga jäägid.
Eemaldage aktiivaine jäägid niiskete vatipatjadega.
Selleks sovbivad hästi AW Facial Tonic või
SLM Cleansing Tonic
Kui olete teostanud enne mesoporatsiooni teemantlihvi
protseduuri, siis kandke näole aesthetic world Laminar
Protection Shield
Kandke silmadele vastavalt nahatüübile silmakreem ja
näokreem
Protseduuri lõpetuseks kandke näole päikesekaitse
White Secret Anti-Agening Sun Fluid LPF 50
Intensiivne mesoporatsiooni protseduur:
1. Kauakestva tulemuse saavutate ainult Intensiivse Meso-
poratsiooni hooldusega.
2. Ideaalne moodus nahka ettevalmistada on CNC Fruit Acid Treatmentiga,
CNC MicroSilver BG Peelinguga või CNC Teemantlihviga. Seda teha umbes
nädal enne intensiivset mesoporatsiooni hooldust.
3. Umbes 4-6 hooldust nädalase intervalliga on vajalikud lähtuvalt
individuaalselt naha seisukorrast ( parima efekti saavutate
2 hooldust nädalas ).
4. Et säilitada intensiivse Mesoporatsiooni protseduuri tulemusi
soovitame teostada 1-2 protseduuri 8-10 nädalase intervalliga
Silmameso hooldus
Kanna lehvikpintsliga Age Defense Eye Concentrate silma piirkonnale.
( Eelnevalt Age Defence Liposomi pole vaja näole kanda ). Ühenda
Silmameso seade aparaadiga ja reguleeri seadme intensiivsust
vastavalt kliendi tundlikkusele. Õrnalt masseerida Meso Eye-Ball seadmega
silma alumist ja ülemist piirkonda ( mitte liikuvat osa ) ringjate liigutustega.
Liikumise suund seest poolt väljapoole ( 5 min mõlemat silma ).
Aparaadi intensiivsus max 25/26 kulmude alumises piirkonnas.
Et tagada Meso Eye-Ball libisemine ja toimimine lisage protseduuri jooksul
juurde Eye Concentratet.
Ühekordne hooldus
1. Ühekordse hoolduse tulemused ilmnevad 3-7 päeva jooksul,
olenevalt nahatüübist.
2. Ühekordset hooldust on tihtipeale soovitud just enne mõnda
tähtsat sündmust ( pulmad, aastapäevad jne. )
Puhastus:
NB! Kindlasti mitte panna lisatarvikuid sterilisaatorisse või Ultraheliaparaati.
Piisab täielikult kui puhastate neid kuuma veega ja seejärel desinfitseerite.
1. Hoolduse tarvikuid ( spetsiaalne roll-on, meso- roller, randmepael )
puhastada kuuma kraani veega.
2. Peale igat hooldust kindlasti puhastada roller. Asetage see kuuma vette,
Kuhu on lisatud paar tilka puhastusvahendit ( gigansepti ). Loputada
põhjalikult külma veega. Kui rollerit pole puhastatud koheselt peale
hooldust, siis aktiivaine võib kleepuda rolleri sisse kinni ja peale seda
rolleri pea ei pruugi enam vabalt liikuda ja muutub kasutuks. Sellisele juhtumile
garantii tingimused ei laine.
3. Soovitame kasutada meso rolleri puhastuseks spetsiaalset väikest harja
Millega saate rolleri edukalt aktiivaine jääkidest puhastada.
4. Roll-on ja meso-rollerit saate lisaks hiljem õrnalt hooldada nahasõbraliku
desinfitseeriva spreiga. ( soovitatavalt haava desinfiteerimise vahendiga )
5. Mitte puhastada randme paela desinfitseerimise vahendiga, see
võib mõjutada elektriimpulsi juhtivust. Õrn puhastusvahend või seep
on selle puhastamiseks ideaalne vahend.
Vastunäidustused:
Mitte teostada protseduuri klientidele, kellel on alljärgnevad naha
probleemid või vastunäidustused.
( alati eelnevalt konsulteerida arstiga, kui esineb kahtlusi )
*Kahjustatud nahk ( ka.lahtised haavad, deramtiit, ekseem, herpes, skleroderma jt )
*Bakteriaalne infektsioon nahal ( ka. tubergoloos )
*Akuutne põletik piirkonnas, kus soovite teostada hooldust
*Näokirurgia ja süstimine
*Südamestimulaator
*Metallist osad organites ( va. Hambaimplantaadid )
*Maovähendus, klambrid sooltes ( kehaprotseduuri korral )
*Rasedus
*Epilepsia
*Pahaloomulised kasvajad ( vähk ) keemiaravi
*Värsked armid
*Operatsiooni haavad, mis pole täielikult paranenud
NB: Mesoporatsiooni hooldust ei tohi teostada kõri ja silmalau liikuval osal.
Garantii
Garantii kehtib 2 a
Kui aparaat ei tööta korrektselt, siis ärge proovige seda ise
parandada. Juhul, kui olete aparaadi ise avanud, siis garantii kaotab kehtivuse.
Kaabel, randmepael ja elektrooniline klamber on kuluvad osad ja nendele
garantii ei laiene.
Garantii kehtib, kui aparaati kasutatakse vastavalt kasutusjuhendile.
Aparaadi kahjustuse korral tuleb seade saata tehasesse tagasi ja
niipea, kui seade on parandatud saadetakse see kliendile tagasi. Tehas annab
endast alati parima, et seda võimalikult kiiresti korda teha ja tagastada kliendile.
Võimalikud vigade allikad
1. Aparaadi intensiivsus on liiga madal või ei ole märgatav
- Randmepael pole korrektselt ühendatud pistikusse
- Roll-on pole korrektselt ühendatud pistikusse
- Aparaadi tagant on juhtmed pistikust väljas
- Randmepael pole ühendatud pistikusse
- Randmepael on liiga kuiv
- Klient on just kreemitanud enda käsi ja selle tõttu on ühendus
paela ja naha vahel häiritud ( sellisel puhul puhastage randmepaela
piirkond koheselt.
- Väga haruldasel juhul klient ei pruugi tunda impulsse , kuid see ei
mõjuta protseduuri efektiivsust .
2. Aparaadi intensiivsust ei saa muuta, kuigi seade on korrektselt sisse lülitatud.
- Kliendi kaitseks on aparaat üles ehitatud nii, et kui intensiivsuse
reguleerimise nupp on jäänud üle 10%, siis aparaat ei lähe käima. Keerake
Intensiivsuse reguleerimise nupp nii palju vasakule, kuni olete saavutanud
madalaima taseme ja peale seda alustage intensiivsuse suurendamist.
Kontaktid:
Eurowest CNC OÜ
Tartu Lai 23
Tallinn Põhja pst 21c
Tel:+3725118332; +3725104323
Tootja:
CNC Cosmetic GmbH
Bruchstücker 9
D-76661 Philippsburg
Germany
Tel:+49725693200
-844550 -899795 0 0 3580765 2169160 0 0 -899795 -897255 0 0 Näo & peanaha hooldus T opelt LED-paneelid näo ja peanaha s amaaegseks hoolduseks 3587750 19685 0 0 Hapniku & aniooniteraapia H apniku ja aniooni udu infusioon hapnikuvahu lahusega 3640455 51435 0 0 NIR Sügavam tungimine dermise kihini A ktiveerige mitokondrid ATP tekitamiseks LPE teraapia ( valgusimpulssenergia ) 34290 400050 0 0 O ptimaalse LED-lainepikkuse kasutamine hoolduse efektiivsuse tagamiseks PUNANE 630nm SININE 465nm VIOLETNE 465 * 630nm helendav ja pinguldav leevendab nahaprobleeme pärast protseduure rahustav (akne, liigne rasu) 3875405 2056765 0 0 -860425 -932180 0 0 U du pihustamise funktsioon H apnikku ja aniooni sisaldav udulahus, et vähendada kuivust, mida põhjustab LED-kiirgus nahale PGM režiim -1905 354965 0 0 Mugav eelseadistatud programm, mis lülitab automaatselt sisse udu- ja LED-funktsioonid 2494280 91440 0 0 Kasutajaliides K asutajaliidese haldamiseks peamise juhtpaneeli või kaugjuhtimispuldi abil -114935 475615 0 0 -899795 -899795 0 0 E emaldada meik ja mustus Kasuta: 1 pump Pore Mask + Cryogenic Activator 15-20 ml -113030 10795 0 0 Puhasta ja valmista ette Kasuta: Cryogenic Activator 15-20 ml -113030 9525 0 0 I nfundeerida toitaineid nahatüübi alusel Kasuta: Oxygen Ceuticals Formula 5-10 ml -184785 307975 0 0 Lisa HA Gel & Toning Gel -208280 152400 0 0 0 0 0 0 S üstida 6 kuni 8 cc Gryogenic Activator -208280 203200 0 0 V alige sobiv programm -208032 113361 0 0 CryoGel Velvet mask või Oxygen Mask (RT/CT/VC) -265430 298450 0 0 Kasuta: Cryogenic Activator Vali sobivad viimistlustooted -899795 -909955 0 0 Akne Sinine: leevendab nahaprobleeme (akne, liigne rasu) Lõpetuseks: Toning Gel -> Herbal Fluid -> Purifying Wintergreen Serum -> Purifying Wintergreen Cream või Cell Couperose Cream Helendav Pun ane: helendav ja pinguldav Lõpetuseks: Toning Gel -> Fermented Fluid -> Moisture Aqua Serum -> PP cream või Hydrating cream Järelhooldus Violetne : pärast protseduure rahustav Lõpetuseks: Toning Gel -> Couperose Fluid -> Moisture Aqua Serum -> Cell Couperose Cream -> PP Cream -886460 -913765 0 0 Puhastamine -558800 88900 0 0 Pore Mask (pooride mask) P oorimask on puhastusvaht, mis oma rikkaliku vahuga puhastab peh melt. S teariinhape, mis on saadud looduslikest rasvhapetest, adsorbeerib erinevaid väliskeskkonna saastunud aineid ja eemaldab tõhusalt jäägid. -407035 310515 0 0 Cryogenic Activator (k rüogeenne aktivaator ) Krüogeenne aktivaator sisaldab suures koguses lahustunud hapnikku ja süvamerevee mineraale, et rahustada ja taastada nahka pärast plastilist kirurgiat või näohooldust, nagu laser, koorimine, microneedling. Ampull -407035 255905 0 0 Couperouse Formula (Kuperoosa valem) Tomati tüvirakkude ja oligopeptiidi eelistega ampull aitab taastada kahjustatu d nahka ja säilitada niiskust, samal ajal leevendades tundlikku nahka. -1673225 575310 0 0 Oxygen Formula (Hapniku valem) L ooduslike aminohapetega rikastatud antioksüdatiivsed ampullid, mis niisutavad ja pehmendavad nahka, siidiselt pehme ja särava välimuse esiletoomiseks -486410 -346710 0 0 Exfoliant Formula ( Kooriv valem ) Panthenooliga täidetud ampull leevendab kuiva naha pinguloleku tunnet ja niatsiinamiid valgendab ja ühtlustab nahatooni . -438785 309245 0 0 Hydrating Formula ( Niisutav valem ) K ortse parandav ampull ühendab endas loodusliku niisutava faktori B etaiini, mis suurendab niisutust ja adenosiini, mis parandab naha elastsust . Pakid -438785 224790 0 0 HA gel (HA geel) Kolm ühes n iisutav geel intensiivse niiskuse suurendamiseks, kortsude ja peente j oonte parandamiseks ning nahatooni ühtlustamiseks , et muuta teie välimus nooruslikuks . -393700 60325 0 0 Toning Gel (toniseeriv geel) Centella asiatica, Glycyrrhiza glabra juure ekstraktid toonivas geelis leevendavad päikesepõletust, atoopiat, ärritust ja valu nahal pärast näohooldusi . -454660 31750 0 0 Clarifying Mask (Puhastav mask) Nahka puhastav mask, mis eemaldab liigse rasu ja surnud naharakud, samal ajal nahka niisutades -375920 274955 0 0 RT/CT/VC MASK E smaklassiline modelleeriv mask rohkem kui 15% merevetika ekstraktiga. A lginaat koor ib surnud naharakud ja niisutab ning toidab nahka rahustava toimega
Saatja: Rigrado OÜ <
[email protected]>
Saadetud: 28.01.2022 13:00
Adressaat: TA Info <
[email protected]>
Teema: Re: Kiri
Manused: Näidustused_F318A_Technical_file.pdf; näidustused_WEB ZT_3021_ENG_.pdf;
AstrodomeFacial_Protocol.pdf; astrodome_kasutus_täiendatud_est_.docx; DMD SEERIA
NUMBER.jpg; Meso saksa keelne kasutusjuhend.pdf; mesoporatsioon EK 2018 eesti
keeles.pdf; AD19A_COIZ Co.,Ltd._EMC_CE.pdf
Tähelepanu! Tegemist on väljastpoolt asutust saabunud kirjaga. Tundmatu saatja korral
palume linke ja faile mitte avada.
Lugupeetud Terviseamet
Käesolevaga saadame vastuse Teie kirjale.
1. Manusena on lisatud Cryoteraapia F-318A seadme täiendav dokumentatsioon kus on
nähtavad seadme kasutamise näidustused.
Lisa (näidustused_WEB ZT_3021_ENG) ja Lisa (Näidustused_F318A_Technical_file).
2. Manusena on lisatud AstrodomeFacial seadme kohta täiendavad dokumentatsioon, kus on
nähtavad seadme kasutamise näidustused inglise keeles (AstrodomeFacial_Protocol) ja eesti
keeles (astrodome_kasutus_täiendatud_est_)
Vastavusdeklaratsiooni täiendamiseks on manusena lisatud EMC TEST RAPORT Lisa
( AD19A_COIZ Co.,Ltd._EMC_CE).
3. Manusena on lisatud CNC Cosmetic teemantlihvi aparaadi tagaosa pilt aparaadi numbriga
Lisa (DMD SEERIA NUMBER.jgp).
Suhtlesime CNC aparaatide maaletoojaga, kellel ei ole esitada teisi dokumente.
Meie ettevõttes on Terviseamet ka varem CNC Cosmetic aparaate kontrollinud ning siis kõik
meie poolt esitatud dokumendid olid Terviseameti jaoks korrektsed.
Teame, et Terviseamet on CNC Cosmetic aparaate ka teistes ettevõtetes kontrollinud,
arvestades, et maaletooja on neid aparaate müünud vähemalt 5 aastat, varustab kõiki kliente
identsete dokumentidega ja Rigrado OÜ-le teadaolevalt ei ole teistel ettevõttetel nende
aparaatide dokumentatsiooni esitamisega Terviseametile probleeme olnud, siis kas esitatavad
nõudmised Rigrado OÜ- le on muutunud ja erinevad teistele ettevõtetele esitatavate
nõudmistega?
4. Manusena on lisatud CNC Cosmetic mesoporatsiooni
aparaadi korrektsed kasutusjuhendid saksa keeles Lisa (Meso saksa keelne kasutusjuhend) ja
eesti keeles Lisa (mesoporatsioon EK 2018 eesti keeles).
Lugupidamisega
Ruslan Ivanov
Rigrado OÜ
juhatuse liige
Kontakt <
[email protected]> kirjutas kuupäeval N, 20. jaanuar 2022 kell 15:04:
Pealkiri: Vastus esitatud teabe kohta Rigrado OÜ
Registreerimise kuupäev: 18.01.2022
Registreerimise number: 12.4-2/21/1301-10
Juurdepääsupiirang: AK AvTS § 35 lg 1 p 2
Lugupidamisega
Terviseamet
Tel 794 3500
E-post
[email protected]
Käesolev kiri on konfidentsiaalne ning mõeldud kasutamiseks
kirja päises nimetatud isikule/asutusele. Käesoleva kirja mistahes
viisil kasutamine või kopeerimine isiku poolt, kes ei ole märgitud
selle adressaadiks, on keelatud. Kui te olete saanud käesoleva
kirja ekslikult, palume sellest koheselt teavitada kirja saatjat ning
kustutada saadud kiri koos kõikide lisadega.
This e-mail is confidential and meant for use by the person
named in the letterhead. Any use in any way or copying of it by a
person not marked as the addressee thereof is prohibited. If you
have got this e-mail by mistake, please notify of it the sender
without delay and delete the received e-mail together with all its
attachments.
Данное письмо является конфиденциальным и
предназначено для использования адресатом. Любое
использование данного письма каким-либо образом или
копирование лицом, которое не обозначено адресатом,
запрещено. Если Вы получили данное письмо по ошибке,
пожалуйста, немедленно сообщите об этом отправителю
письма и удалите полученное письмо вместе с
прикреплёнными к нему файлами.
www.terviseamet.ee
Saatja: Neti Eva Grepp </o=Sotsiaalministeerium/ou=Exchange Administrative Group (FYDIBOHF23SPDLT)/cn=Recipients/cn=490e8060a39a464a810a17be3bc5c851-Net>
Saaja: Meditsiiniseadmed (Medical Devices)
Teema: Vs: CNC seadmete dokumendid
Tere
Kas Saksamaalt on CNC seadmete kohta uut infot tulnud?
Lugupidamisega
Neti Eva Grepp
inspektor
Põhja regionaalosakond
+372 794 3729 / +372 55943258
[email protected] <mailto:
[email protected]>
Terviseamet
+372 794 3500
[email protected] <mailto:
[email protected]>
www.terviseamet.ee <http://www.terviseamet.ee/>
Paldiski mnt 81, 10614 Tallinn
Saatja: Neti Eva Grepp
Saatmisaeg: reede, 29. juuli 2022 11:04
Adressaat: Meditsiiniseadmed (Medical Devices) <
[email protected]>
Teema: Vs: CNC seadmete dokumendid
Tere
Suur aitäh kõikide vastuste eest!
Jään Saksamaa pädeva asutuse vastust ootama.
Lugupidamisega
Neti Eva Grepp
inspektor
Põhja regionaalosakond
+372 794 3729 / +372 55943258
[email protected] <mailto:
[email protected]>
Terviseamet
+372 794 3500
[email protected] <mailto:
[email protected]>
www.terviseamet.ee <http://www.terviseamet.ee/>
Paldiski mnt 81, 10614 Tallinn
Saatja: Meditsiiniseadmed (Medical Devices) <
[email protected] <mailto:
[email protected]> >
Saatmisaeg: reede, 29. juuli 2022 11:02
Adressaat: Neti Eva Grepp <
[email protected] <mailto:
[email protected]> >
Teema: Vs: CNC seadmete dokumendid
Tere
Viimane info laekus 26.04.2022. Saksamaa pädeva asutuse vastus:
Intermediate information: The former clerk from 2018/2019 is temporarily not on duty. This requires the complete revision of the operation that was run at that time. Inquiries and obtaining opinions are still in progress. This time it takes a little more time; follow-up information can be expected in about six weeks (≈ end of June).
28.07.2022 kirjutasin Saksamaa pädevale asutusele uuesti ning väga ootame nende vastust.
Lugupidamisega
Kristina Kübar
peaspetsialist
meditsiiniseadmete osakond
+372 5330 5236
[email protected] <mailto:
[email protected]> |
[email protected] <mailto:
[email protected]>
Terviseamet
+372 794 3500
[email protected] <mailto:
[email protected]>
www.terviseamet.ee <http://www.terviseamet.ee/>
Paldiski mnt 81, 10614 Tallinn
Käesolev kiri on konfidentsiaalne ning mõeldud kasutamiseks kirja päises nimetatud isikule/asutusele. Käesoleva kirja mistahes viisil kasutamine või kopeerimine isiku poolt, kes ei ole märgitud selle adressaadiks, on keelatud. Kui te olete saanud käesoleva kirja ekslikult, palume sellest koheselt teavitada kirja saatjat ning kustutada saadud kiri koos kõikide lisadega.
This e-mail is confidential and meant for use by the person named in the letterhead. Any use in any way or copying of it by a person not marked as the addressee thereof is prohibited. If you have got this e-mail by mistake, please notify of it the sender without delay and delete the received e-mail together with all its attachments.
Данное письмо является конфиденциальным и предназначено для использования адресатом. Любое использование данного письма каким-либо образом или копирование лицом, которое не обозначено адресатом, запрещено. Если Вы получили данное письмо по ошибке, пожалуйста, немедленно сообщите об этом отправителю письма и удалите полученное письмо вместе с прикреплёнными к нему файлами.
Saatja: Neti Eva Grepp <
[email protected] <mailto:
[email protected]> >
Saatmisaeg: kolmapäev, 20. juuli 2022 11:44
Adressaat: Meditsiiniseadmed (Medical Devices) <
[email protected] <mailto:
[email protected]> >
Teema: Vs: CNC seadmete dokumendid
Tere
Arusaadav, ootame Egle ja Kristina puhkuste lõpuni.
Lugupidamisega
Neti Eva Grepp
inspektor
Põhja regionaalosakond
+372 794 3729 / +372 55943258
[email protected] <mailto:
[email protected]>
Terviseamet
+372 794 3500
[email protected] <mailto:
[email protected]>
www.terviseamet.ee <http://www.terviseamet.ee/>
Paldiski mnt 81, 10614 Tallinn
Saatja: Meditsiiniseadmed (Medical Devices) <
[email protected] <mailto:
[email protected]> >
Saatmisaeg: kolmapäev, 20. juuli 2022 11:33
Adressaat: Neti Eva Grepp <
[email protected] <mailto:
[email protected]> >
Teema: Vs: CNC seadmete dokumendid
Tere
Vabandan, et vastused kirjadele on viibinud.
Nendest teemadest teavad kõige rohkem Egle ja Kristina, kuid nad mõlemad puhkavad veel see nädal. Egle on tagasi teisipäeval ja Kristina veel veidi hiljem.
Kas võiks sobida, et me vastame neile (ja kahes eelmises kirjas olnud) küsimustele, kui nad on puhkuselt tagasi? Või on nendega kiire?
Tervitades
Piret
juhataja
meditsiiniseadmete osakond
+372 5853 3610
[email protected] <mailto:
[email protected]> |
[email protected] <mailto:
[email protected]>
Terviseamet
+372 794 3500
[email protected] <mailto:
[email protected]>
www.terviseamet.ee <http://www.terviseamet.ee/>
Paldiski mnt 81, 10614 Tallinn
Selles e-kirjas sisalduv teave (kaasa arvatud manused) on mõeldud ametialaseks kasutamiseks ning seda võivad kasutada vaid e-kirja adressaadid. E-kirjas sisalduvat teavet ei tohi ilma saatja selgelt väljendatud loata edasi saata ega mistahes viisil kõrvalistele isikutele avaldada. Juhul, kui Te olete saanud käesoleva e-kirja eksituse tõttu, teavitage sellest koheselt saatjat ning kustutage e-kiri oma arvutist.
Saatja: Neti Eva Grepp <
[email protected] <mailto:
[email protected]> >
Saatmisaeg: esmaspäev, 18. juuli 2022 14:31
Adressaat: Meditsiiniseadmed (Medical Devices) <
[email protected] <mailto:
[email protected]> >
Teema: Vs: CNC seadmete dokumendid
Tere
Kas Saksamaalt on CNC seadmete kohta vastused tulnud?
Lugupidamisega
Neti Eva Grepp
inspektor
Põhja regionaalosakond
+372 794 3729 / +372 55943258
[email protected] <mailto:
[email protected]>
Terviseamet
+372 794 3500
[email protected] <mailto:
[email protected]>
www.terviseamet.ee <http://www.terviseamet.ee/>
Paldiski mnt 81, 10614 Tallinn
Saatja: Meditsiiniseadmed (Medical Devices) <
[email protected] <mailto:
[email protected]> >
Saatmisaeg: teisipäev, 8. märts 2022 10:13
Adressaat: Neti Eva Grepp <
[email protected] <mailto:
[email protected]> >
Teema: Vs: CNC seadmete dokumendid
Tere
Esitatud dokumendi põhjal ei saa veenduda seadmete ELi nõuetele vastavuses. Manusena lisatud dokument „certificate of conformity.pdf“ ei ole nõuetekohane ning võltsimistunnustega.
Hinnangu andmiseks on vajalikud järgmised dokumendid:
a) ELi vastavusdeklaratsioon (inglise keeles Declaration of Confomity);
b) võimalusel pilt seadme märgistusest (et näha oleksid seadme andmed ja valmistamise kuupäev).
MSO esmane hinnang:
1)CNC Cosmetic DiamondPeel (teemantlihvimise aparaat)
Eestikeelse kasutusjuhendi alusel tegemist on meditsiiniseadmega.
ELi vastavusdeklaratsiooni ei ole. Seega ei saa hinnata, kas meditsiiniseade on nõuetekohaselt EL turule viidud.
2)CNC Cosmetic Mesoporation (mesoporatsiooni aparaat)
Kasutusjuhendit ei ole. ELi vastavusdeklaratsiooni ei ole. Seega ei saa hinnata. Määruse 2017/745 mesoporatsiooni seadmed peavad olema meditsiiniseadmed.
3)CNC Cosmetic ColdPLASMA
Eestikeelse- ja saksakeelse kasutusjuhendite alusel tegemist on meditsiiniseadmega. ELi vastavusdeklaratsiooni ei ole ning seega ei saa hinnata, kas seade on korrektselt turule viidud.
Palun küsige ilusalongide käest ELi vastavusdeklaratsioon (inglise keeles Declaration of Confomity) iga seadme kohta.
Meie uurime Saksamaa kolleegide käest CNC seadmete kohta, kuid tavaliselt läheb 1-2 kuud vastuse saamiseks.
Lugupidamisega
Kristina Kübar
peaspetsialist
meditsiiniseadmete osakond
+372 5330 5236
[email protected] <mailto:
[email protected]> |
[email protected] <mailto:
[email protected]>
Terviseamet
+372 794 3500
[email protected] <mailto:
[email protected]>
www.terviseamet.ee <http://www.terviseamet.ee/>
Paldiski mnt 81, 10614 Tallinn
Saatja: Neti Eva Grepp <
[email protected] <mailto:
[email protected]> >
Saatmisaeg: esmaspäev, 7. märts 2022 15:55
Adressaat: Meditsiiniseadmed (Medical Devices) <
[email protected] <mailto:
[email protected]> >
Teema: CNC seadmete dokumendid
Tere
Seoses kahe poolelioleva menetlusega ilusalongide suhtes saadeti Saksamaalt CNC seadmete dokumendid.
Ühe salongi (Rigrado OÜ, RoyalGrace) puhul olen Teile erinevaid dokumente juba saatnud ning viimased puuduolevad dokumendid on nüüd loodetavasti manuses.
Teise salongi esindaja saatis originaalsed saksakeelsed kasutusjuhendid fotodena, suure mahu tõttu panin need järgmisesse kausta:
V:\1. Kõigile_töötlemiseks\Keskkonnatervis\Teenuste valdkonna materjalid\CNC seadmed 2022.
Palun Teie seisukohta, kas seadmed on nõuetele vastavad, järgmiste seadmete suhtes:
1. CNC Cosmetic DiamondPeel (teemantlihvimise aparaat)
2. CNC Cosmetic Mesoporation (mesoporatsiooni aparaat)
3. CNC Cosmetic ColdPLASMA
Lugupidamisega
Neti Eva Grepp
inspektor
Põhja regionaalosakond
+372 794 3729 / +372 55943258
[email protected] <mailto:
[email protected]>
Terviseamet
+372 794 3500
[email protected] <mailto:
[email protected]>
www.terviseamet.ee <http://www.terviseamet.ee/>
Paldiski mnt 81, 10614 Tallinn