11.06.2025
VAIDLUSTUS
Riigihange: „Traumatoloogia metallosteosünteesi vahendid ja liigeste
endoproteesid“
viitenumber: 295377
Vaidlustaja: OneMed OÜ
registrikood: 10891247
aadress: Veerenni 38, 10138 Tallinn
Esindaja: Erki Fels, vandeadvokaat
Gregor Saluveer, vandeadvokaadi abi
Advokaadibüroo TGS Baltic
e-post:
[email protected]
Hankija: sihtasutus Tartu Ülikooli Kliinikum
registrikood: 90001478
aadress: L. Puusepa 1a, Tartu 50406
e-post:
[email protected]
TAOTLUSED
1. Peatada riigihanke „Traumatoloogia metallosteosünteesi vahendid ja liigeste
endoproteesid“ (vr 295377) menetlus kõikides riigihanke osades.
2. Kohustada sihtasutust Tartu Ülikooli Kliinikum viima õigusaktidega kooskõlla järgmised
riigihanke alusdokumentide tingimused:
2.1. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ punkt 3;
2.2. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ punkt 4.1;
2.3. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ punkt 5.1;
2.4. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ punkt 6.1;
2.5 riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 6.1;
2.6. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 8.1;
2.7. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 8.2;
2.8. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 9.2;
2.9. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 9.3;
2.10. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 9.6;
Advokaadibüroo TGS Baltic AS Ahtri 6a, 10151 Tallinn Kaluri 2, 51004 Tartu
[email protected] Telefon +372 626 4300 Telefon +372 730 1610
2.11. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 1.5;
2.12. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 1.6;
2.13. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 2.5;
2.14. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 2.8;
2.15. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 3.1;
2.16. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 1.1;
2.17. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 1.2;
2.18. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 1.6;
2.19. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 1.7;
2.20. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 2.4;
2.21. riigihanke osa 1 alusdokumendis „Lisa 1 osa 1 - tsemeneteeritav standardne
puusaprotees (põhi)“ sätestatud nõuded, sealhulgas peatükk „Muud täiendavad
komponendid“;
2.22. riigihanke osa 2 alusdokumendis „Lisa 2 osa 2 - tsemeneteeritav standardne
puusaprotees (alternatiiv)“ sätestatud nõuded, sealhulgas peatükk „Muud
täiendavad komponendid“;
2.23. riigihanke osa 3 alusdokumendis „Lisa 3 osa 3 - puusaliigese eriprotees“
sätestatud nõuded, sealhulgas peatükid „Muud täiendavad komponendid“ ja
„Lisavahendid“;
2.24. riigihanke osa 5 alusdokumendis „Lisa 5 osa 5 - Tsemeneteeritav põlveprotees
tagumise ristatisideme stabiliseerimisega“ sätestatud nõuded, sealhulgas
peatükid „Lisavõimalused“ ja „Teised lisavõimalused“;
2.25. riigihanke osa 7 alusdokumendis „Lisa 7 osa 7 - Põlveliigese eriprotees“ sätestatud
nõuded, sealhulgas punkt 8;
2.26. riigihanke osa 13 alusdokumendis „Lisa 13 osa 13 - Megaprotees“ sätestatud
nõuded;
2.27. riigihanke alusdokumendi „Lisa_24_pakkumuse_maksumus“ punkt 2.3;
2.28. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ peatükid „Muud täiendavad komponendid“,
„Lisavahendid“ ja „Teised lisavõimalused“;
2.29. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ peatükid „Lisavõimalused“, „Teised
lisavõimalused“ ja „Muud täiendavad komponendid ja võimalused“;
2.30. riigihanke osa 7 alusdokumendi „Lisa 7 osa 7 - Põlveliigese eriprotees“ peatükid
„Lisavõimalused“, „Teised lisavõimalused“ ja „Muud täiendavad komponendid ja
võimalused“;
2.31. riigihanke osa 8 alusdokumendi „Lisa 8 osa 8 - Õlaliigese protees“ peatükk „Muud
täiendavad komponendid ja võimalused“;
2/13
2.32. riigihanke osa 9 alusdokumendi „Lisa 9 osa 9 - Küünarliigese protees“ peatükk
„Muud täiendavad komponendid ja võimalused“;
2.33. riigihanke osade 1—14 tehniliste kirjelduste tingimused, mis algavad sõnadega
„Endoproteesi komplekti baashinna arvutamiseks /…/“;
2.34 riigihanke osade 1—23 tehniliste kirjelduste tingimused, mis algavad sõnadega
„Pakkuja peab hankijale tagama spetsiifiliste instrumentide komplektid /…/“;
2.35. riigihanke osade 1—23 tehniliste kirjelduste tingimused, mis algavad sõnadega
„Garanteeritud instrumentide komplektide arv arvestatakse /…/“ ja „Lisaks
garanteeritud instrumentide komplektidele /…/“;
2.36. riigihanke alusdokumendi „Juhised pakkujale“ punktid 6.18—6.18.2 ja 7.2.
3. Jätta OneMed OÜ menetluskulud sihtasutuse Tartu Ülikooli Kliinikum kanda.
1. ASJAOLUD
1.1. 19.05.2025 avaldas sihtasutus Tartu Ülikooli Kliinikum (hankija) avatud hankemenetlusena
läbiviidava riigihanke „Traumatoloogia metallosteosünteesi vahendid ja liigeste endoproteesid“ (vr
295377) (riigihange) hanketeate ja tegi kättesaadavaks riigihanke alusdokumendid (RHAD).
Tehnilise kirjelduse (TK) järgi on hankelepingu esemeks traumatoloogia metallosteosünteesi
vahendite ja liigeste endoproteeside ost.
1.2. Pakkumuste esitamise tähtpäev on 19.06.2025.
1.3. TK on vastuolus riigihangete seaduse (RHS) § 88 lg-ga 7 ja RHS § 3 p-dega 1—3 ja 5. Riigihanke
osades 1, 2, 3, 4, 5, 6, 7, 10, 11 ja 13 on hankija koostanud TK nii, et neile vastab kindel toode.
OneMed OÜ (vaidlustaja) osalemine riigihankes on põhjendamatult takistatud. Ülejäänud
riigihanke osades on tingimused ebaselged, mistõttu pole vaidlustajal võimalik aru saada, kas
tema pakutavad tooted tingimustele vastavad või mitte, või ebaproportsionaalsed.
1.4. Vaidlustaja taotleb riigihangete vaidlustuskomisjonilt (VAKO) kohustada hankijat viima TK
õigusaktidega kooskõlla (RHS § 185 lg 2 p 1).
2. RIIGIHANKE PEATAMISE TAOTLUS
2.1. VAKO võib vaidlustaja põhjendatud taotluse alusel teha igas vaidlustusmenetluse staadiumis
otsuse riigihanke peatamise kohta (RHS § 193 lg 1). Vaidlustaja taotleb riigihanke peatamist
kõikides riigihanke osades, kuna kõikides osades on õigusvastaseid tingimusi, mis vaidlustaja
õiguseid riivavad. Riigihanke peatamine on põhjendatud, kuna pakkumuste esitamise tähtpäev
saabub enne VAKO otsuse tegemist (RHS § 200 lg 1).1
2.2. Vaidlustuse eesmärk on saavutada olukord, kus hankija kehtestaks õigusvastaste
hanketingimuste asemel õiguspärased tingimused. Sellise eesmärgi saavutamine ei ole enam
võimalik, kui pakkumused on esitatud. Kui pakkumuste esitamise tähtaeg möödub, on RHADis
sätestatud tingimused kohustuslikud nii hankijale kui pakkujatele, kellel tuleb edasises
menetluses nendest tingimustest lähtuda. VAKO ega kohus ei saa kohustada hankijat pärast
pakkumuste esitamise tähtpäeva möödumist RHADi muutma.2 Vaidlustaja on valmis taotlusest
loobuma, kui hankija soostub ise pakkumuste esitamise tähtpäeva edasi lükkama nii, et tagatud
oleks VAKO otsuse tegemine ja ka kaebetähtaeg.
1
VAKO 93-18/194960, p 4.
2
TlnRKm 3-15-1817, p 10; VAKO 166-17/190812, p-d 8, 11—14; VAKO 187-17/192426, p-d 8, 11—14.
3/13
3. PÕHJENDUSED
3.1. Vaidlustaja vaidlustab tingimusi kahel põhjusel. Esiteks, TK on koostatud kindlate toodete järgi
(ptk A). Teiseks, TK on ebaselge (ptk 3.6).
A. TEHNILINE KIRJELDUS ON KOOSTATUD KINDLATE TOODETE JÄRGI
3.1. Hankija on koostanud TK-d vaidlustatud osades kindlate toodete spetsifikatsioonide järgi
(täpsemalt toodud allolevas tabelis). TK-d tervikuna lugedes on selge, et hankija on lihtsalt
kopeerinud TK-sse nende toodete omadused.
3.2. Riigihangete korraldamise üheks üldpõhimõtteks on konkurentsi efektiivne ärakasutamine (RHS
§ 3 p 3). Hankija on vaidlusaluse riigihanke kirjutanud n-ö lukku nendele pakkujatele, kes pakuvad
neid tooteid, mille omadusi on TK-s kirjeldatud. TK ei tohi aga tekitada objektiivselt
põhjendamatuid takistusi riigihangete avamisel konkurentsile (RHS § 88 lg 7). TK ei tohi olla
koostatud nii, et tingimustele vastaks üksnes üks toode või pakkuja.3
3.3. Hanketingimused peavad olema põhjendatud. Praegune TK, millega suunatakse hankeleping
kindlate toodete pakkujatele, kindlasti põhjendatud ei ole. Hankija ei tohi seada hankelepingu eset
silmas pidades põhjendamatuid piiranguid, mis välistavad ausa ja läbipaistva konkurentsi või on
mõistliku põhjenduseta piiravad. 4 Hanketingimus ei tohi minna kaugemale, kui riigihankes
taotletava eesmärgi saavutamiseks on mõistlikult vajalik.5
3.4. Vaidlustaja juhib tähelepanu, et 2021. aastal korraldas hankija endoproteeside riigihanke
(238201) sarnase ülesehitusega ja tulemuseks oli olukord, kus hankija maksis endoproteeside
eest 4 aasta jooksul ebamõistlikult kõrget hinda võrreldes samade või paremate tulemustega
Eestis kasutatavate endoproteesidega. Hankija tegevus ei vasta seega ka üldpõhimõttele
kasutada rahalisi vahendeid säästlikult ja otstarbekalt (RHS § 3 p 5).
3.5. Allolevas tabelis on välja toodud tingimused, mis takistavad vaidlustajal pakkumust esitamast.
Paremalt teises lahtris on viide dokumendile, millest nähtub, et hankija valitud tootel on olemas
täpselt kehtestatud nõudele vastav omadus. Parempoolses lahtris selgitab vaidlustaja, miks ei
ole nõuded sellised kujul põhjendatud.
TK Viide hankija valitud Miks ei ole tingimus
TK punkti sisu
punkt tootele põhjendatud
Osa 4: Põlveliigese kondülaarne protees tagumise ristatisideme säilitamisega
Hankija nõuded kirjeldavad OÜ Wimberg poolt müüdava Waldemar Link tooteid
„SymphoKnee“ ja „Gemini SL“ kombinatsioonis veel mõne Waldemar Linki tootega,
mille edasimüügiõigus on vaid OÜ-l Wimberg.
3 „Reieluu ja sääreluu Dokument „SymphoKnee Nõue on objektiivselt
komponentide väljavõte“ (lisa 1) põhjendamatu ja valitud kindla
suuruse ühilduvus toote järgi. Kõigil tootjatel on
vähemalt +/- 2 omad kombineeritavuse tabelid,
suurust.“ mis täidavad samaväärselt
hankija eesmärki.
3
TlnRnKo 3-10-3225, p 18.
4
TlnHKo 3-20-582, p 12.
5
VAKO 34-19/206183, p 15.
4/13
4.1 “mobiilse ja Dokument „Gemini SL Nõue on objektiivselt
toestatud kande surgical technique“ (lisa põhjendamatu ja valitud kindla
komponendid;” 2), lk 2 toote järgi. Mobiilne kanne võib
käia roteeruva tibia platoo
mõistena, mitte reieosa
komponendi nõudena.
5.1 “mobiiline kanne ja Dokument „Gemini SL Nõue on objektiivselt
fikseeritud kanne;” surgical technique“ (lisa põhjendamatu ja valitud kindla
2), lk 2 toote järgi. Mobiilse kande nõude
eesmärk on füsioloogiline liigese
rotatsioon, eri tootjad lahendavad
seda erinevalt süsteemipõhiselt,
näiteks mediaalselt roteeruv
protees. See täidab samaväärselt
hankija eesmärki.
6.1 „Ülikõrge Dokument „Link Nõue on objektiivselt
molekulmassiga Symphoknee Surgical põhjendamatu ja valitud kindla
polüetüleenist technique“ (lisa 3), lk 65 toote järgi. Laialtlevinud standard
(UHMWPE) on HXLPE või tootjaspetsiifiline
liugpinnad.“ variant, mis täidab samaväärselt
või paremini hankija eesmärki.
UHMWPE materjal antud nõude
kombineerimisel ülejäänud
nõuetega on konkurentsi välistav.
Osa 6: Erikonstruktsiooniga tsemeneteeritav põlveprotees
Hankija nõuded kirjeldavad OÜ Ortolink Invest poolt müüdavat Smith & Nephew toodet
„Primary Knee System“.
6.1 „reisosa materjal Dokument „OXINIUM- “Keramiseeritud metall” on üks
keramiseeritud Knee-Brochure“ (lisa 4), võimalustest lahendada
metall (ZrNb), et lk 4 probleemi, mis esineb 1-l
sobiks ka patsiendil 100st (metalliallergia).
metalliallergiaga Tegemist on Smith & Nephew
patsientidele“ patenteeritud tehnoloogiaga
(OXINIUM). On olemas ka
alternatiive, nt Ti-Nidium, mis
täidavad samaväärselt hankija
eesmärki.
8.1 “plastikosa igal Dokument „Legion TKS „High flex insert“ on nimetus, mida
mõõdul vähemalt 9 surgical technique“ (lisa kasutab ainult Smith & Nephew
erinevat kõrgust, 5), lk 46 oma plastikosade kirjeldamiseks.
high flex ja Teised tootjad võivad selle
constrained tingimuse eesmärki täita näiteks
variandid;” erilise disainiga reie
komponendiga.
5/13
8.2 “vähemalt 6 Dokument „Legion TKS Nõue on objektiivselt
erinevat surgical technique“ (lisa põhjendamatu ja valitud kindla
polüetüleenist 5), lk-d 59 ja 61 toote järgi. Enamik süsteeme
liigestuva pinna algab suurusest 10mm, mis
suurust, CCK täidab samaväärselt hankija
versioonil eesmärki.
paksusega alates 9
mm;”
9.2 „Monteeritavad Dokument „Legion TKS Nõue objektiivselt põhjendamatu
intramedullaarsed surgical technique“ (lisa ja valitud kindla toote järgi. Igal
pikendused: varre 5), lk 59 tootjal on oma suuruste ja
pikendused alates pikkustega varred, mis vastavad
diameetrist 10 mm, süsteemile ja täidavad
pikkusega vähemalt samaväärselt hankija eesmärki.
vahemikus 80 mm –
275 mm, sirged.
Tsemendivabad ja
tsemendiga
versioonid.“
9.3 “Revisjonvarred: Dokument „Legion TKS Nõue on objektiivselt
pikendused koos surgical technique“ (lisa põhjendamatu ja valitud kindla
mõõdetava Offset- 5), lk 59 toote järgi. Igal tootjal on oma
ga , alates suuruste ja pikkustega varred, mis
diameetrist 9 mm ja vastavad süsteemile ja täidavad
pikkusega vähemalt samaväärselt hankija eesmärki.
vahemikus 120 –
270 mm. Nii
tsementeeritavad
kui tsemendivabad.”
9.6 „Põlvekedra Dokument „Legion TKS Nõue on objektiivselt
protees: vähemalt surgical technique“ (lisa põhjendamatu ja valitud kindla
kolme erineva 5), lk 48 toote järgi. Igal tootjal on oma
kujuga ja vähemalt suuruste ja variatsioonidega
6 erineva põlvekedra proteesid, mis
paksusega.“ vastavad süsteemile ja täidavad
samaväärselt hankija eesmärki.
Osa 10: Mittetsemeneteeritav puusaprotees (põhiproteesina)
Hankija nõuded kirjeldavad OÜ Ortolink Invest poolt müüdavaid Smith & Nephew
tooteid „SL Plus MIA“ ja „R3“ kombinatsioonis veel mõne Smith & Nephew tootega,
mille edasimüügiõigus on vaid OÜ-l Ortolink Invest.
1.5 “reieosa 2 erineva Dokument „SL PLUS Nõue on objektiivselt
kaela nurgaga- MIA Surgical technique“ põhjendamatu ja valitud kindla
standard (131°) ja (lisa 6), lk 16 toote järgi (täpselt 131°).
Vaidlustaja toote puhul on reieosa
6/13
lateraliseeritud 2 erineva kaela nurgaga: standard
(123°);” (133°) ja lateraliseeritud (123°).
See täidab hankija eesmärki
samaväärselt.
1.6 ”reieosa kaela Dokument „SL PLUS Nõue on objektiivselt
pikkus standard MIA Surgical technique“ põhjendamatu ja valitud kindla
varrel 17-49mm ja (lisa 6), lk 17 toote järgi (täpselt 17-49mm ja
lateraliseeritud 26-58mm). Vaidlustaja toote puhul
variandil 26-58mm” on reieosa kaela pikkus standard
varrel 28,3-49,4mm ja
lateraliseeritud variandil 33,6-
54,6mm. See täidab hankija
eesmärki samaväärselt.
2.5 ”napa sisemine - Nõue on objektiivselt
vooder saadaval nii põhjendamatu. PE vooder on
PE kui XLPE vananenud tehnoloogia. Kuna PE
variandis;” voodriga implantaatide kliinilised
tulemused on kehvemad kui
näiteks XLPE versioonil, siis ei
paku enamik vastutustundlikuid
tootjad enam PE versiooni. Kuna
hankija nõuab nii üht kui teist,
teeb hankija võimatuks XLPE
variandis tooteid pakkuda (mis ei
täida hankija eesmärki isegi
samaväärselt, vaid paremini).
2.8 „Pressfit napaosa Dokument „R3 Surgical Nõue on objektiivselt
suurused 40mm – technique“ (lisa 7) lk-d 13 põhjendamatu ja valitud kindla
80 mm millele ja 17. toote järgi. 0° ja 10°
sobivad XLPE kõrgendusega on saavutatav
plastikosad 0° ja samaväärne tulemus, mis 0° ja
20° kõrgendusega;“ 20° kõrgendusega.
3.1 „reiepea materjal Dokument „OXINIUM- „Keramiseeritud metall” on Smith
keramiseeritud Knee-Brochure“ (lisa 4), & Nephew patenteeritud
metall (ZrNb);“ lk 4 tehnoloogia (OXINIUM). On
olemas ka alternatiive, nt Ti-
Nidium, mis täidavad
samaväärselt hankija eesmärki.
Osa 11: Mittetsemeneteeritav puusaprotees (alternatiiv)
Hankija nõuded kirjeldavad OÜ Wimberg poolt müüdavat Waldemar Link toodet „SP-
CL“ kombinatsioonis veel mõne Waldemar Link tootega, mille edasimüügiõigus on
vaid OÜ-l Wimberg.
7/13
1.1 ”bioloogiliselt/anato Dokument „SP-CL Nõue on objektiivselt
omiliselt sobilik, Product Rationale“ (lisa põhjendamatu. S-kõverus on
mille S-kõverus 8), lk 3 ainult Waldemar Linki toodetele
järgib reieluu omane omadus. Samaväärseid
loomulikku või paremaid kliinilisi tulemusi on
anatoomilist kuju;” võimalik saavutada ka teiste
kujudega proteesidega.
1.2 ”Vasaku ja parema Dokument „SP-CL Vasak ja parem versioon on
poole versioonid” Anatomically Adapted vajalikud anatoomilise disaini
Cementless Hip System“ puhul. Samaväärsete või
(lisa 9), lk 3 paremate tulemustega teise
disainiga vartel ei ole vaja vasakut
ja paremat versiooni.
1.6 ”Kahe CCD nurga Dokument „SP-CL Nõue on objektiivselt
võimalus – 126 ja Anatomically Adapted põhjendamatu ja valitud kindla
135 kraadi” Cementless Hip System“ toote järgi (täpselt 126 ja 135
(lisa 9), lk 3 kraadi, nagu Waldemar Link
tootel).
1.7 ”Reieosal vähemalt Dokument „SP-CL Vasak ja parem versioon on
18 erinevat suurust Anatomically Adapted vajalikud anatoomilise disaini
vasakule ja Cementless Hip System“ puhul. Samaväärsete või
paremale” (lisa 9), lk 3 paremate tulemustega teise
disainiga vartel ei ole vaja vasakut
ja paremat versiooni.
2.4 ”napaosa võimalik - Nõue on objektiivselt
modulaarsete põhjendamatu ja valitud kindla
komponentidega toote järgi. 0° ja 10°
10° ja 20° serva kõrgendusega on saavutatav
kõrgendada” samaväärne tulemus, mis 0° ja
20° kõrgendusega.
3.6. Lisaks eelnevale vaidlustab vaidlustaja TK-sid tervikuna osades 1, 2, 3, 5, 7 ja 13. Riigikohus on
selgitanud, et tingimuste kogum on õigusvastane, kui selle eesmärk on ebakohastel eesmärkidel
konkreetse ettevõtja soosimine.6 Viidatud osade TK-d on kogumis koostatud nii, et neile vastavad
vaid ühe tootja tooted:
a. Osa 1 („tsementeeritav standardne puusaprotees (põhiproteesina)“) nõuded kirjeldavad
OÜ Wimberg poolt müüdavat Waldemar Link toodet „Lubinus SPII“ kombinatsioonis veel
mõne Waldemar Link tootega, mille edasimüügiõigus on vaid OÜ-l Wimberg.
b. Osa 2 („tsementeeritav standardne puusaprotees (alternatiivse proteesina)“) nõuded
kirjeldavad OÜ Ortolink Invest poolt müüdavaid Smith & Nephew tooteid „Polar Stem“ ja
„Polar Cup“ kombinatsioonis veel mõne Smith & Nephew tootega, mille edasimüügiõigus
on vaid OÜ-l Ortolink Invest.
c. Osa 3 („puusaliigese eriprotees“) nõuded kirjeldavad OÜ Wimberg poolt müüdavat
Waldemar Link toodet „Link MP Reconstruction“ kombinatsioonis veel mõne Waldemar
6
RKHKo 3-20-718, p 18.
8/13
Link tootega, mille edasimüügiõigus on vaid OÜ-l Wimberg.
d. Osa 5 („tsemeneteeritav põlveprotees tagumise ristatisideme stabiliseerimisega“) nõuded
kirjeldavad Johnson&Johnson AB Eesti filiaali poolt müüdava DePuy Synthese toodet
„Attune“.
e. Osa 7 („põlveliigese eriprotees“) nõuded kirjeldavad OÜ Wimberg poolt müüdava
Waldemar Link põlveproteesi kombinatsioonis veel mõne Waldemar Link tootega, mille
edasimüügiõigus on vaid OÜ-l Wimberg.
f. Osa 13 („megaprotees“) nõuded kirjeldavad OÜ Wimberg poolt müüdava Waldemar Link
tooteid. Eriti ilmselt avaldub see tõsiasi TK punktis 5, kus nõuab hankija, et pakutav toode
peab olema kombineeritav põlveliigese Endo-model põlve implantaadi ning puusaliigese
MP rekonstruktsiooni proteesiga. Tegemist on Waldemar Link toodetega. Lisaks, Eestis on
erinevatel tootjatel oluliselt kaasaegsemaid lahendusi, mille kliinilised tulemused on
paremad kui mainitud Endo-model proteesil.
B. TEHNILINE KIRJELDUS ON EBASELGE
3.7. Teatud TK tingimusi vaidlustab vaidlustaja ka põhjusel, et nende sisu on ebaselge ning seega
vastuolus läbipaistvuse põhimõttega (RHS § 3 p 1). Vaidlustajal ei ole nende tingimuste sisu
ebaselguse tõttu võimalik aru saada, kas tema pakutavad tooted tingimustele vastavad või mitte.
3.8. Vaidlustaja hinnangul on hankija teadlikult koostanud segaste ning asjatundmatute mõistetega,
tihti omavahel vastuolevad tingimused, et varjata soovi osta kindlaid tooteid kindlatelt
edasimüüjatelt.
TK punkt TK punkti sisu Miks on tingimus ebaselge
Osa 7 TK punkt „Peab ühilduma Pakkujal pole võimalik aru saada, millega
8 megaproteesi (osa 13) peab osas 7 hangitav põlveliigese
komponentidega.“ eriprotees ühilduma. See, millist toodet
hankija osa 13 („megaprotees“) raames
hangib, sõltub sellest, milline pakkumus
osutub edukaks. Pakkumuse esitamise
hetkel ei ole olemas „megaproteesi (osa
13) komponente“, millega oleks üldse
ühilduda võimalik. Jääb mulje, nagu
hakkab hankija osas 7 kontrollima
pakkumuste vastavust selle alusel,
milline pakkumus osutus edukaks osas
13. Selline teguviis on ilmselgelt
õigusvastane.
Lisa 24 Esitada tuleb „reiepea Osas 2, millega pakkumuse maksumuse
(pakkumuse komplekti“ hind. vormi (lisa 24) punkt 2.3 korreleerub, ei
maksumuse ole sellist asja nagu „reiepea komplekt“.
tabel), punkt 2.3 Tsemendilise puusaproteesi varre külge
ei paigaldata komplekte, vaid reiepäid.
Terminist “reiepea komplekt” võib aru
saada kui mitmest osast koosnevast
lahendusest, mis on oluliselt kallim
tavapärasest reiepeast. Hankija on
teabevahetuses kinnitanud, et tegemist
9/13
on „inimliku eksimusega“ ja ta peab
silmas reiepead, kuid ei ole
hankedokumente parandanud. Hankijal
ei ole võimalik teabevahetuse teel
hankedokumente muuta. Maksumuse
vormi ebaselgus võib põhjustada
pakkumuste tagasi lükkamise või
maksumuste võrreldamatuse.
Osade 1—4 ja TK-des on osad nõuded Ebaselgeks jääb, mida need nõuded
10 TK-de liigitatud kui „Muud endast kujutavad – kas tegemist on
peatükid „Muud täiendavad komponendid“, kohustusliku nõuetega või mitte.
täiendavad „Lisavahendid“, „Teised Dokumendi „Juhised pakkujale“ p-s 1.4
komponendid“ lisavõimalused“, on kirjutatud, et mistahes dokumendis
„Lisavõimalused“, „Muud sätestatud kohustus, tingimus või nõue
Osade 3–4 TK-
täiendavad komponendid ja on pakkujale kohustuslik. Sama
de peatükid
võimalused“. dokumendi p-s 6.9 on kirjas, et
„Lisavahendid“
pakutavad tooted peavad vastama
Osade 4—7 TK- kõikidele TK-s esitatud nõuetele. See on
de peatükid ka loogiline, kuna tehniline kirjeldus on
„Teised „asjade omaduste ja neile esitatavate
lisavõimalused“ nõuete loetelu“ (RHS § 87 lg 1 p 1).
Osade 5—7 TK- Teabevahetuses (ID 972062) on hankija
de peatükid aga andnud teada, et n-ö lisanõuete
„Lisavõimalused“ täitmine pole siiski kohustuslik: „Need ei
ole kohustuslikud, kuid kui pakkujal on
Osade 6—9 TK- neid pakkuda, peavad need olema
de peatükid sobivad ja ühilduvad nõutud
„Muud implantaatidega.“ Hankija ei saa
täiendavad teabevahetuse kaudu muuta
komponendid ja kohustuslikku tingimust vabatahtlikuks
võimalused“ (RHS § 46 lg 3). Hankija ei ole lubanud
alternatiivsete pakkumuste esitamist
(„Juhised pakkujale“ p 3.4). Ka praktilises
mõttes jääb täiesti arusaamatuks, kuidas
saavad olla teatud TK nõuded
iseloomuga „kui pakkujal on pakkuda, siis
paku“. Hankija peab hanketingimusi
muutma, et sellest tuleneks üheselt
selgelt, kuidas pakkujad peavad n-ö
lisanõuete osas käituma. Ebaselgus võib
põhjustada pakkumuste tagasi lükkamise
või maksumuste võrreldamatuse.
Seejuures on ilmselgelt asjakohatu
hankija põhjendus, justkui võimaldavad
lisavõimalused „saada pakkumusi
laiemalt“, arvestades et hankija on TK-d
kirjutanud lukku kindlatele toodetele.
Osade 1—14 „Endoproteesi komplekti Jääb arusaamatuks, miks on vaja
TK-de baashinna arvutamiseks endoproteesi komplekti baashinda. Mitte
tingimused, mis liidetakse kõige enam kusagil mujal hanketingimustes ei ole
10/13
algavad kasutatavate mõõtudega reie, viidet endoproteesi komplekti
sõnadega sääre ja liigespinna baashinnale. Millise metoodika järgi
„Endoproteesi komponentide hinnad.“ liidetake kõige enam kasutatavate
komplekti mõõtudega reie, napa ja reiepea
baashinna komponentide hinnad ja miks seda vaja
arvutamiseks on? Kust tuleb sisend „enim kasutatavate
/…/“ mõõtude jaoks“? Pole aru saada, millele
pakkuja selles tingimuses kinnituse
annab.
Osade 1—23 Jääb arusaamatuks, mida tähendab
TK-de instrumentide komplekti “tagamine”.
„Pakkuja peab hankijale
tingimused, mis Endoproteeside valdkonnas on
tagama spetsiifiliste
algavad tavapärane, et hanke võitja annab
instrumentide komplektid,
sõnadega instrumentide komplekti(d) tasuta
mille maksumus on arvutatud
„Pakkuja peab kasutamiseks. Kui tegemist on väga vähe
implantaatide osana (nn
hankijale kasutatavate implantaatidega, siis on ka
garanteeritud instrumentide
tagama võimalus, et hankija ostab instrumentide
komplektid).“
spetsiifiliste komplekti välja. Sõnakasutus „tagama“
instrumentide põhjustab ebaselgust, millele pakkuja
komplektid /…/“. oma kinnituse andma peab. „Tagama“
viitab sellele, et pakkujal peab olema
teoreetiline võimekus komplekte
hankijale tarnida, mitte ei pea ta neid
kindlasti tarnima.
Osade 1—23 „Garanteeritud instrumentide Jääb arusaamatuks, mida tähendab
TK-de komplektide arv arvestatakse “garanteeritud instrumentide komplekt”.
tingimused, mis 1 komplekt/100 endoproteesi Mujal hanketingimused eristata
algavad täiskomplekti kohta.“ „garanteeritud“ ja „garanteerimata“
sõnadega komplekte. Valdkonnas väljakujunenud
„Lisaks garanteeritud
„Garanteeritud tava on, et instrumendid kas antakse
instrumentide komplektidele
instrumentide tasuta kasutamiseks või hankija ostab
on hankijal õigus instrumente
komplektide arv instrumendid välja või kasutatakse
või komplekte ka täiendavalt
arvestatakse laenuinstrumente (instrumendid tuuakse
osta pakkumuses märgitud
/…/“ ja „Lisaks kohale, kui hankija selleks soovi
hindadega.“
garanteeritud avaldab). Pole aru saada, kes, kellele ja
instrumentide mida garanteerib.
komplektidele
Samuti on arusaamatu hankija nõue
/…/“.
saada endale 1 komplekt garanteeritud
instrumente 100 endoproteesi
täiskomplekti kohta. Kui pakkuja kinnitab
selle tingimusega nõustumust, peab ta
olema valmis tarnima hankeperioodi
jooksul kuni 20 komplekti instrumente
ühe osa kohta. Tavapäraselt saab hankija
hakkama 1—3 instrumentide komplektiga
ühe hanke osa kohta kogu hankeperioodi
jooksul (3—4 aastat). Kuna tegemist on
sedavõrd tavatu (ja ebaselge) nõudega,
11/13
ei ole aru saada, millele pakkuja peab
kinnituse andma.
Täiendavat segadust tekitab punkt,
millega hankija annab endale õiguse
lisaks garanteeritud instrumentide
komplektidele instrumente või komplekte
ka täiendavalt osta pakkumuses märgitud
hindadega. Mis mahus, mis tingimustel ja
mis alusel hankija lisatellimusi esitama
hakkab? Kuna juba mõiste „garanteeritud
instrumentide komplektid“ on ebaselge,
pole aru saada, millele pakkuja peab
kinnituse andma. Instrumentide vajadus
ja nende tarnimise tingimused peavad
olema selged.
C. TEHNILINE KIRJELDUS ON EBAPROPORTSIONAALNE
3.9. Viimaks vaidlustab vaidlustaja dokumendi „Juhised pakkujale“ punkte 6.18—6.18.2 ja 7.2:
Punkt 6.18: „Hankijal on õigus nõuda pakkuja poolt pakutud toodete näidiseid koos
paigaldamiseks vajalike instrumentidega nende toodete vastavuse tuvastamiseks (sh
hankijale eesmärgipäraseks kasutamiseks sobivuse määramiseks) läbi funktsionaalsete
omaduste praktilise testimise. Näidis peab olema originaalpakendis, varustatud toote
koodi/ref.nr-ga, toote nimetusega ja vastava pakkumuse osa numbriga.“
Punkt 6.18.1: „Näidis tuleb viia L. Puusepa 8, Tartu, ortopeediakliiniku ortopeedia osakonda
(ruum H 156) või saata postiga samale aadressile. Tellimuse täitmise tähtaeg on 7
(implantaadi näidis) või 14 (implantaadi näidis koos paigalduseks vajalike instrumentidega)
kalendripäeva alates tellimuse esitamisest. Tellimuse mittetähtaegsel täitmisel tunnistatakse
pakkumus mittevastavaks.“
Punkt 6.18.2. „Tootenäidiste maksumust ei hüvitata pakkujale. Näidise paigalduseks vajalikud
instrumendid tagastatakse pakkujale pärast hankemenetluse lõppemist, testimise tarbeks
kasutamise eest ei tasuta.“
Punkt 7.2: „Toodete vastavuse tuvastamiseks võib hankija vajadusel moodustada
ekspertkomisjoni, kes hindab näidiste vastavust praktilise testimise teel. Toodete praktiline
testimine protokollitakse. Kui toote praktilise testimise käigus ilmneb, et toode ei vasta
kirjeldatud omadustele või kasutustehnoloogiale või esineb tehnilisi tõrkeid toote kasutamisel
või toode ei täida ootuspäraselt eesmärki või omab varjatud puudusi, on hankijal õigus
tunnistada pakkumus mittevastavaks.“
3.10. Esiteks, tingimused on ebaproportsionaalsed (RHS § 3 p 1). Need nõuded annavad hankijale nt
õiguse nõuda kõikide komponentide kõikide suuruste kohaletoomist. Primaarproteeside puhul
tähendab selline nõue pakkujale orienteeruvalt 30—100 tuhande euro suurust kulutust.
Revisioon- või megaproteesi puhul võivad summad ulatuda sadade tuhandeteni. Sealjuures
peavad pakkujad suutma hankija nõude täita 7 või 14 kalendripäevaga. Olukorras, kus hankijal
on võimalik toodete vastavust kontrollida tehnilise dokumentatsiooni alusel – vajaduse korral
nõudes välja ka täiendavaid tõendeid –, on pakkumuste vastavuse kontrollimine sellisel viisil
pakkujad ilmselgelt ebamõistlikult koormav.
3.11. Teiseks, tingimused on läbipaistmatud (RHS § 3 p 1). Ebaselgeks jääb, kelle käest asub pakkuja
näidiste toomist nõudma. Vastavust tuleb kontrollida kõikide pakkumuste puhul (RHS § 114 lg 1).
Tingimuste sõnastusest tulenevalt on hankijal aga õigus nõuda näidiste toomist vaid ühe pakkuja
puhul, kusjuures nõude täitmata jätmine on eraldiseisev mittevastavuse alus ja sellega kaasnevat
kulu pakkujale ei hüvitata. On ilmselgelt õigusvastane olukord, milles hankija kontrollib
12/13
pakkumuste vastavust erineval moel. Lisaks, kui pakkuja ei pea näidiste toomisega kaasnevat
kulu äriliselt mõistlikuks (ja hankija teab seda), on hankijal sisuliselt võimalik tõrjuda
hankemenetlusest pakkuja, kes talle ei sobi. Pingereas teise pakkuja puhul pole aga hankija
kohustatud näidiste toomist nõudma, kuna tegemist on tema õigusega.
4. MENETLUSLIKUD KÜSIMUSED
4.1. Riigilõiv. Vaidlustaja on tasunud vaidlustuse eest riigilõivu 1280 eurot (RLS § 258 lg 1 p 2) (lisa
10).
4.2. Esindus. Vandeadvokaat Erki Fels ja vandeadvokaadi abi Gregor Saluveer kinnitavad, et on
vaidlustaja lepingulised esindajad (RHS § 190 lg 11).
4.3. Tähtaegsus. Pakkumuste esitamise tähtpäev on 19.06.2025. Vaidlustus on tähtaegne (RHS §
189 lg 2 p 2).
4.4. Läbivaatamine. Vaidlustaja on nõus kirjaliku menetlusega (RHS § 190 lg 1 p 6).
4.5. Esitamine. Vaidlustus on esitatud VAKO-le e-posti teel (
[email protected]).
Lugupidamisega
(digitaalselt allkirjastatud)
Erki Fels, Gregor Saluveer,
vandeadvokaat vandeadvokaadi abi
LISAD
Lisa 1 SymphoKnee väljavõte
Lisa 2 Gemini SL Surgical technique
Lisa 3 Link SymphoKnee Surgical technique
Lisa 4 OXINIUM-Knee-Brochure
Lisa 5 Legion TKS Surgical technique
Lisa 6 SL PLUS MIA surgical technique
Lisa 7 R3 Surgical technique
Lisa 8 SP-CL Product Rationale
Lisa 9 SP-CL Anatomically Adapted Cementless Hip System
Lisa 10 Riigilõivu tasumist tõendav maksekorraldus
13/13
GEMINI SL Total Knee System
Surgical Technique
Explanation of Pictograms
Manufacturer Article number
Product meets the applicable requirements, which are regulated in
Material (number)
the EU harmonization legislation for the affixing of the CE marking.
Contents
GEMINI SL Total Knee System
02 System Description GEMINI SL
03 Indications / Contraindications
Surgical Technique Implantate
04 Overview – Workflow of Surgical Steps Femoral Components (cemented/cementless)
06 Preoperative Planning 54 Fixed Bearing CR/Mobile Bearing
07 Approaches to be used with PE Articulating Surfaces
Fixed Bearing CR and Mobile Bearing
Femoral Preparation – Distal Resection
55 PS (Posterior Stabilzed) – to be used with Fixed
08 Predetermining the Femoral Size Bearing PS PE Articulating Surfaces
09 Femoral Intramedullary Alignment
56 Fixed Bearing – to be used with Fixed Bearing CR
10 Distal Femoral Resection and PE Articulating Surfaces PS
12 SIT-CUT Instruments - Tibia Plateau Resection 57 Mobile Bearing – to be used with Mobile Bearing
16 Set Tibial Resection Level PE Articulating Surfaces
17 Tibial Sizing Tibial Components (cemented/cementless)
18 Checking Extension Gap and Axes
58 Articular Surface Fixed Bearing CR
Final Femoral Preparation to be used with Femoral Components CR (Cruciate
19 Femoral Sizing Retaining) and Fixed Bearing Tibial Components
19 Rotation Setting 60 Articular Surface Fixed Bearing PS
to be used with Femoral Components PS and Fixed
21 A/P Femoral Resection and Chamfer Cuts
Bearing Tibial Components
22 Fixed Bearing PS (Posterior Stabilized)
62 Articular Surface Mobile Bearing
Tibial Preparation to be used with Femoral Components CR (Cruciate
23 Tibial Sizing Retaining) and Mobile Bearing Tibial Components
24 Tibial Stem Preparation 64 Taper Caps for Tibial Components and
Patella Resurfacing Components
Patella Preparation (Patella Resurfacing)
65 Tibial Stems: cemented and cementless
25 Patella Sizing
26 Resection Instruments
27 Reaming 66 GEMINI SL Instrument Set – Overview
30 Trial Reduction and Functional Test 67 Case 1: SIT-CUT Easy Tibia Resection Alignment
68 Case 2: Femoral Preparation
Implantation
69 Case 3: Spacer
32 Compatibility Table
70 Case 4: Tibia Preparation
33 Tibia Implant Overview
71 Case 5: Femoral Trial Prostheses
34 Implantation: Mobile Bearing
72 Case 6: Femoral PS Preparation
36 Implantation: Fixed Bearing and Fixed Bearing PS
73 Case 7: General Instruments
37 Patella Implantation
74 Case 8: Final-Spacer
38 Functional Test
75 Patella Instrument Set
39 APPENDIX 1: Precision Tibial Resection
76 Precision Tibia Resection Alignment
Alignment - Extramedullary Alignment
77 EXTRABONE Instrument Set
44 APPENDIX 2: Precision Tibial Resection
Alignment – Intramedullary Alignment 78 Additional Instruments: Adapter, Sawblades
49 APPENDIX 3: GEMINI SL EXTRABONE 79 Accessories: X-ray Templates
– Extramedullary Femoral Referencing 80 Further Information
Important Information
01
System Description
The GEMINI SL Knee Surface Replacement allows native joint reconstruction with extensive range of
motion and good kinematics.1, 2, 3
• The extended Femoral Shield and the deep Patellar Groove can lead to a physiological patellar motion
and patellar self-guidance.1
• The antomical adapted Tibial Component allows for a better cortical coverage and thereby to an
improved homogeneous force loading.4
• The Tibial Component is securely anchored by means of blades, pegs and stem, imporoving resistance
against shear, rotation and tilting force.5
• The systems consists of Fixed Bearing Cruciate Retaining (CR), Fixed Bearing Posterior Stabilized
(PS) and Mobile Bearing versions and there are optional tibial stems available which expand the range
of indications and enable a comprehensive treatment option.
• The options shown are additionally available with the LINK PorEx Technology*.
PorEx is a TiNbN = Titanium-Niobium-Nitride surface modification that significantly reduces the release of
chromium and nickel ions. It is a ceramic-like surface that is very hard improving wetting angle and results
in a low coefficient of friction.6
• All versions can be selected cemented and cementless.
Cruciate Retaining Posterior Stabilized
Mobile Bearing (MB)
Fixed Bearing (CR FB) Fixed Bearing (PS FB)
Cruciate Retaining configuration for Posterior Stabilized configuration for Rotating highly congruent artricular
use with intact ligaments and cap- use with absence of posterior cruciate surface for use with or without poste-
sule and adequate joint stability. ligament. Reduces the risk of disloca- rior cruciate ligament.7
Identical tibial metal gtray for Fixed tion and decreases contact pressure
Bearing CR and PS. in deep flexion.8 Featuring bone pre-
serving design with size specific inter-
condylar femoral box dimensions.
*
LINK PorEx: TiNbN = Titanium Niobium Nitride; hypoallergenic coating (gold color).
1
H. Thabe, „Auswirkungen verschiedener konstruktiver Prothesenmerkmale auf Langzeitergebnisse“, Akt Rheumatol 2013;38
2
Interne Daten – H. Thabe, „Aspekte zum Konzept der beweglichen Tibiaplateaukonstruktion, April 2000
3
J. Goodfellow, „The Mechanics of the Knee and Prosthesis Design”. J Bone Joint Surg Br 1978; 60:358-36
4
Figgie, H.E. et al.: Load-bearing Capacity of the Tibial Component of the Total Condylar Knee Prosthesis; Clin Orth & Rel Res, No. 183, 1984
5
Walker P.S., Hu-Ping Hsu, Zimmermann R.A.: „A comparative study of uncemented tibial components“ J of Arthrop, Vol. 5, No. 3, 1990, pp. 246-253
6
Internal Data oin file: Study of the influence of TiNbN-coating in the ion release of CoCRMO-alloys in SBF buffer simulator testing
7
Internal data on file: Innocenti B. White Paper GEMINI SL Mobile BEaring CR: Biomechnical Analysis in healthy and deficient PCL patient, 2017
8
Internal data on file: Innocenti B., White Paper GEMINI SL Fixed BEaring PS: Biomechnical Analysis of the Post-Cam System, 2017
02
Indications/Contraindications
Specified indications and contraindications: GEMINI SL Total Knee Replacement
GEMINI SL Cruciate Retaining Fixed Bearing (CR FB)
Indications:
Any form of uni-, bi- or tricompartimental arthritis of the knee joint (e.g. primary degenerative arthritis, secon-
dary arthritis resulting from rheumatoid arthritis, fracture, post-infection, gout, chondrocalcinosis and others)
Contraindications (absolute):
Acute and chronic infections, local and systemic, insofar as they may compromise the successful implantation
Moderate or severe instability or complete loss of the medial or collateral ligament
Instability or loss of the posterior cruciate ligament
Any bone defect that will result in insufficient implant fixation (based on the fact, that using stems, bone grafts
and metal bone substitutes like cones, a minimum bone stock for implant fixation cannot be defined)
Severe insufficiency or loss of extensor mechanism
Contraindications (relative):
Extension deficit >30°
Varus or valgus deformity >30°
Allergy to one of the implant materials
GEMINI SL Posterior Stabilized Fixed Bearing (PS FB)
Indications:
Any form of uni-, bi- or tricompartimental arthritis of the knee joint (e.g. primary degenerative arthritis, secon-
dary arthritis resulting from rheumatoid arthritis, fracture, post-infection, gout, chondrocalcinosis and others)
Contraindications (absolute):
Acute and chronic infections, local and systemic, insofar as they may compromise the successful implantation
Moderate or severe instability or complete loss of the medial or collateral ligament
Any bone defect that will result in insufficient implant fixation (based on the fact, that using stems, bone grafts
and metal bone substitutes like cones, a minimum bone stock for implant fixation cannot be defined)
Severe insufficiency or loss of extensor mechanism
Contraindications (relative):
Allergy to one of the implant materials
GEMINI SL Mobile Bearing (MB)
Indications:
Any form of uni-, bi- or tricompartimental arthritis of the knee joint (e.g. primary degenerative arthritis, secon-
dary arthritis resulting from rheumatoid arthritis, fracture, post-infection, gout, chondrocalcinosis and others)
Contraindications (absolute):
Acute and chronic infections, local and systemic, insofar as they may compromise the successful implantation
Moderate or severe instability or complete loss of the medial or collateral ligament
Any bone defect that will result in insufficient implant fixation (based on the fact, that using stems, bone grafts
and metal bone substitutes like cones, a minimum bone stock for implant fixation cannot be defined)
Severe insufficiency or loss of extensor mechanism
Contraindications (relative):
Allergy to one of the implant materials
Implants with TiNbN-Coating
In contrast to all other implants an allergy against one of the implant materials is not a contraindication. Otherwise
the above mentioned indications and contraindications depending on the design of the TiNbN-coated implants
remain the same.
03
Overview – Workflow of Surgical Steps
Extension Gap First
Tibial Resection
SIT-CUT
Preoperative Distal Femoral Resection
Planning
(Approach)
Tibial Resection
“Precision Tibia“
Tibia First
Tibial Resection
SIT-CUT
Preoperative Distal Femoral Resection
Planning
(Approach)
Tibial Resection Checking Extension
“Precision Tibia“ Gap and Axes
Femur First
a/p Femoral Resection Tibial Preparation
and Chamfer Cuts SIT-CUT
optional alternative
Preoperative Distal Femoral Resection
Planning
(Approach)
Preparation PS Tibial Preparation
(Posterior Stabilized) “Precision Tibia“
04
Overview – Workflow of Surgical Steps
The GEMINI SL Instruments are designed to meet the need of today’s knee surgeons.
The instrumentation assures precise and reliable bone resections and allow a variety of surgical options. 1
Preparation maybe initiated at either the femur or the tibia. Spacer blocks are provided for extension and
flexion gap evaluation. Patella instrumentation is available for compatible preparation of patella resurfacing.
Furthermore, it is available a dedicated instrumentation to perform an extramedullary femoral referencing.
a/p Femoral Resection
alternative and Chamfer Cuts
Preparation PS
(Posterior Stabilized)
Tibia Sizing Trial Reduction and Implantation
Functional Test
1
Internal data on file
05
Surgical Technique
Preoperative Planning
The anatomic landmarks in the knee joint are defined preoperatively by imaging
the whole leg on the healthy and the affected side in the standing position. The
angle between the anatomic axis (center of knee joint – intramedullary canal) and
the mechanical axis (center of femoral head – center of knee joint – center of ankle
to the second toe) determines the valgus angle (01).
These angles should be determined for both knees. The valgus angle of a healthy
knee joint is 5°–7°. In comparison with the healthy side, and for the purpose of
reconstructing the corresponding valgus angle in the affected knee joint, this
angle must be determined before carrying out the distal femoral resection.
The appropriate implant size can be selected preoperatively with X-ray templates.
The necessary resections are determined by the size of the implant.
01
01
06
Surgical Technique
Approaches
With the knee in slight flexion, a straight incision is made over the patella, as far as the tibial tuberosity (02).
A medial parapatellar incision is made through the patellar retinaculum, capsule and synovial membrane (03).
When making the parapatellar incision, the patella is pushed to one side to visualize the femoro-patellar joint.
Removal of the hypertrophic synovial membrane and parts of the fat pad allow access to the medial, lateral and
intracondylar parts of the joint. Excess synovia should be removed in order to avoid postoperative impingement
and adhesions. Some surgeons prefer total synovectomy.
02 03
Alternative Approaches:
Midvastus Subvastus
Widely used approach, performed with Special approach for a smaller incision
a simple and shorter incision
07
Surgical Technique
Femoral Preparation – Distal Femoral Resection
Predetermining the Femoral Size
The GEMINI SL Femoral Component permits size-specific distal resection:
Sizes 1-2B distal femoral cutting block "S"
Sizes 3-5 distal femoral cutting block "L"
The femoral sizer (04) is used to verify the femoral size determined in preoperative planning and to select
the distal femoral cutting block. The femoral size is determined by holding the femoral sizer against the
bone – the contour shown corresponds to size 3 (04-a). If the femur profile is smaller, the distal cutting block
"S" should be selected. If the femur profile is a perfect fit or larger, the distal cutting block "L" should be used.
The m/l dimension can be checked in addition (04-b).
Size > = L
m/l
04
m/l
04-a
04-b
08
Surgical Technique
Femoral Intramedullary Alignment
For femoral preparation, the knee is flexed to 90°. The entry point
for opening the femur can be marked with the electrocautery (05). It
is usually located approx. 3-5 mm medially above the intercondylar
fossa. The medullary canal is opened with the step drill (06).
05
05
06
Connect the T-handle to the IM guide rod and insert the
guide rod carefully. Remove the T-handle (07).
07
According to the previously selected femoral size ("S" or "L"), the distal cutting block
is connected to the alignment instrument for valgus angle. Then, the preoperatively
determined valgus angle is set, and the instrument is attached to the intramedullary
guide rod (08-a, 08-b).
08-a
08-b
09
Surgical Technique
Before fixing the distal cutting block with two drill pins, it must be ensured that at least one condyle is
in contact with the alignment instrument, and that the valgus angle of the correct side has been set.
The two drill pins are inserted through the holes labeled "0" (09).
Then the alignment instrument for valgus angle is separated from the distal cutting block and removed.
The intramedullary guide rod is then removed (10).
09 10
Distal Femoral Resection
The cutting template can be used to check
the alignment of the distal resection (11).
11
To ensure that the instrument is stable, a third drill
pin is inserted into oblique hole. There is the option of
inserting a fourth drill pin (12).
12
10
Surgical Technique
The distal saw cut (sawblade 1.27 mm) is made at
90° flexion. Then the cutting block is removed (13).
13
The parallel drill pins remain in position (14).
Note: It is possible to set the extension gap
by means of post-resection.
14
11
Surgical Technique
SIT-CUT Instruments – Tibia Plateau Resection
Extramedullary Alignment
For external alignment of the tibial resection instrument set, the extramedullary guide is assembled and
adjusted to suit the length of the tibia.
Note: For intramedullary alignment the precision tibial alignment instrument set is used. The application
is described in Appendix II.
Components
The extramedullary guide consists of E
following components:
15.5
EM Alignment Rod (15.1) F
proximal with drill pin fixation
EM Alignment Guide (15.2)
- foot clamp silicone belt fixation (15.3-a)
- foot clamp spring fixation (15.3-b) D
15.4
C 15.1
A: Locking screw foot clamp posterior slope
B: Locking screw varus/valgus
C: Locking screw EM alignment rod 15.2
Symmetrical Tibial Cutting Block (15.4) 15.4-L
(alternatively: asymmetrical left/right versions 15.4- L/R) D
D: Locking screw
Stylus (E) with guide (F) (15.5)
The stylus comes in different versions with the
following designations: 15.4-R
B D
12 for GEMINI SL Mobile Bearing
10 for GEMINI SL Fixed CR and Fixed PS
2 for GEMINI SL Defect Referencing A
Alignment is carried out in reference to the anterior
edge of the tibia. When the alignment rod is parallel
to the tibia, the resection is at 90°, which means 0° 15.3-a
posterior slope.
Note: The tibial cutting block has no integrated slope.
15.3-b
12
Surgical Technique
Assembly
The requisite components are assembled as follows:
Release locking screws (C) and (D). Push the EM alignment
rod (15.1) through the tibial cutting block (15.4-L) and into the
EM alignment guide (15.2).
Hand-tighten the locking screws (C) and (D).). Release locking
screw (A). Insert the foot clamp (15.3-a or 15.3-b) into the align-
ment guide with the flat surface facing upward, and hand-
tighten the screw (A).
Release the locking screw at the stylus guide, then insert the
foot guide (15.5) with stylus into the tibial cutting block (15.4).
15.1
Hand-tighten the locking screw.
15.5
15.4-L
D
C
15.2
15.3-a
A
15.3-b
13
Surgical Technique
Positioning
Note: Fixation of the ankle is achieved either by the silicone
belt or the optional spring clamp.
Attach the foot clamp to the ankle joint and secure with the
silicone belt (16).
16
Release the locking screw (C). Place the EM alignment rod
considering varus/valgus and rotation (17).
The axis of the EM alignment rod should be positioned with
reference to the tibial axis. Proximal fix the EM alignment rod
using a drill pin. Depending on the anatomical structure the
drill pin either can be positioned into the tibial plateau or alter-
natively into anterior tibial cortex.
Note: After inserting the drill pin there is still the possibility to
set rotational position, varus/valgus orientation and posterior
slope.
Tighten the locking screw (C).
C
17
14
Surgical Technique
Alignment
Rotation (18)
To determine the correct alignment of the rotation,
the alignment instrument is aligned centrally along
the tibial shaft axis.
Varus-valgus (18)
To set the varus/valgus, release the screw (B) and
carry out this adjustment by shifting the alignment
guide medially/laterally. Then tighten the screw (B)
again.
5°
B
7.2
18
C
Posterior Slope (19)
GEMINI SL – Tibia Slope Recommendations:
For Mobile Bearing and a Fixed Bearing Cruciate Retaining (CR) configuration
it is recommended to set the tibial posterior slope at 5 degrees. For a Fixed
Bearing Posterior Stabilized (PS) configuration, a range of 0–5 degrees
of tibial posterior slope is recommended.
After releasing the distal setting screw (A), the alignment instrument can be til-
ted by shifting the alignment rod axially, which in turn allows the requested
posterior slope to be set. Once the correct position is reached, the setting B
screw on the EM alignment guide (A) is tightened.
Finally finish by firmly tightening all screws (A,B,C). A
19
15
Surgical Technique
Set Tibial Resection Level
Use the tibial stylus (E) to set the resection height. E
Select and position the stylus:
12 for GEMINI SL Mobile Bearing
10 for GEMINI SL Fixed and Fixed PS
2 for GEMINI SL Defect Referencing
The stylus “10” respectively “12“ is positioned on the highest point of the best-
preserved area of the tibial plateau (20). Stylus “2” is choosen to align according
the defect. D
Release the locking screw (D) allowing to move the tibial cutting block until the
stylus rests on the tibial plateau.Then hand-tighten the locking screw (D). 20
Afterwards secure the tibial cutting block with two drill pins through the anterior
parallel holes with the "0" marking. Remove the tibial stylus.
The cutting template can be used to check the alignment of the tibial resection (20-a).
Any necessary corrections to the position (by ± 2 mm) can be carried out at this point.
Fix the cutting block medially with a third drill pin (20-b) and then perform the resection (20-c).
20-a 20-b 20-c
16
Surgical Technique
Tibial Sizing
Following the resection, the third drill pin and the tibial
saw guide are removed. The anterior drill pins remain
in position (21).
21
The tibial implant size can now be determined by applying the
appropriate tibial sizer. The instrument must cover the cortical
bone optimally, without projecting beyond it (22). The tibial implant
size can also serve as an indicator for the femoral implant size if
the femoral measurement is between two sizes.
22
17
Surgical Technique
Checking Extension Gap and Axes
After careful cleaning of the soft tissues, the extension spacer “S” or “L” is inserted, according to previously
used distal femur cutting block, and both the axial alignment and the stability of the joint in extension are
checked (23-a, 23-b).
23-a 23-b
If no further resections are required, the femoral and tibial drill pins can now be removed.
18
Surgical Technique
Final Femoral Preparation
Femoral Sizing
For final femoral preparation, the next step is to determine the definitive femoral
size. The femur sizer is applied to the distal femoral resection surface.
Attention: The femoral sizer is anterior refernce only.
Note: Please ensure that the femur sizer sits flush on the distal
resection. The femur sizer can be applied over the intra-medullary
guide rod to ensure stable positioning.
24
Place the femoral stylus anterior on the femur with the tip referencing
the desired exit point of the sawblade for the anterior cut. This is usually
half way up lateral, anterior prominence of the femoral trochlea. This
measurement determines the adaptation of the femoral implant size to
the A/P dimension. To check the M/L dimension of the selected size,
the appropriate femur trial can be held against it. The femoral size is
read off from the right side (24).
Two drill pins are used to fix the size and, at the same time,
to fix the frame of the measuring instrument to the distal resection
25
surface (25).
Rotation Setting
The alignment instrument allows external rotation in reference to the posterior condyle tangent,
whiteside line or epicondylar line (26).
The external rotation can be adjusted by turning the rotation setting screw until the alignment
references are matched (27).
Whiteside line: Alignment rod in the window to match the whiteside line
Epicondylar line: Collateral pegs align with the epicondylar line
Rotation setting
Posterior condyle tangent: Readable degree indicate external rotation
screw
setting aligned to posterior condyle tangent
Alignment rod for
whiteside line
Whiteside line
Scale for femoral
rotation setting
– reference to
posterior condyle
Epicondylar line tangent
26
nt
ndyle tange
Posterior co
27
Alignment of epicondylar
axis
19
Surgical Technique
Optional: Checking Flexion Gap
The flexion gap can be checked before finally defining the rotation setting and the femoral
resection.
Flex the knee joint 90°. The stepped end of the spacer for the selected size (28) is now slid
under the control block so that the higher end rests against the control block (29-a, 29-b).
The stepped design means that the condyles are exposed.
28 29-a 29-b
Once the correct rotation has been set, the instrument is fixed
with two drill pins through the medial and lateral holes (30). These
drill pins are the guide reference for the position of the femoral
cutting block.
30
The alignment instrument is removed, but the two drill pins remain
in situ (31).
31
20
Surgical Technique
A/P Femoral Resection and Chamfer Cuts
The femoral cutting block is attached over the drill pins (32-a).
Optional: The flexion gap can be checked or corrected by using the correction spacer in combination
with femoral cutting block (32-b).The spacer is with his stepped side pushed as far as under the block,
until it rests on the highest point under the femoral cutting block. The flexion gap can be adjusted by
reallocating the femoral cutting block prior to the final resection.
The femoral cutting block is fixed with drill pins with stop (Ø 3.5 mm). Optional it is possible to use the
threaded pins (Ø 3.5 mm) with stop for the fixation of the femoral cutting lock (32-c).
32-a 32-b 32-c
Note:
The cutting template (angle wing) can be used to check the alignment of the individual resections, especially
the anterior cut ensuring that notching is unlikely to occur.
The anterior cut is carried out first, followed by the posterior,
then the anterior and posterior chamfer cuts (33).
33
After that, the trochlea is prepared with the rasp (34-a, 34-b).
34-a 34-b
21
Surgical Technique
Fixed Bearing PS (Fixed Bearing Posterior Stabilized)
The box guide is selected according to the size of the femur and is
positioned on the resected end of the femur. Alignment is performed
according to the notch and the m/l dimension (35).
Note:
The box guide is symmetrically designed to permit universal use.
Marking guide lines enable orientation and alignment (36). 35
Exterior contour for Exterior contour for
aligning on the lateral, aligningon the lateral,
right femoral condyle left femoral condyle
Marking guide line Marking guide line
mL (medial left) mR (medial right)
Left implant Right implant
36
The box guide is fixed with at least two drill pins, or with three
if preferred (37).
The notch reamer is used to prepare the box:
1 Drill in Position 1 (38.1+2)
2 Drill in Position 2 (38.3)
37
3 Smooth the resection by passing the notch trephine from
Position 1 to Position 2
38.1 38.2 38.3
Femur bone after completion of resection (39).
39
22
Surgical Technique
Tibial Preparation
Tibial Sizing
The tibial implant size can now be determined by applying the appropriate
tibial preparation plate. Select the tibial size to achieve maximal tibial
coverage (40).
After applying the appropriate tibial sizer and aligning the rotation with the
aid of the extramedullary alignment rod (41), the tibial preparation plate is
fixed. Two head pins are inserted centrally into the tibial plate (42-a, 42-b). 40
Option: Additional two head pins can be inserted anterior.
42-a 42-b
41
23
Surgical Technique
Tibial Stem Preparation
The blade chisel is used to prepare the two guides and the anchoring holes for the drill pins
for the tibia base plate (43-a, 43-b).
43-a 43-b
Then the guide sleeve is inserted (44) and the taper guide is reamed (45-a, 45-b).
Note: The choice of reamer depends on which treatment option is selected: tibial
taper cap or stem extension.
44 45-a 45-b
To avoid impingement in flexion, remove any excess
bone between the posterior tibial implant and the
posterior femoral condyles in flexion. Use a curved
osteotome or gouge to remove any remaining bone
and osteophytes (46).
46
24
Surgical Technique
Patella Preparation (Patella Resurfacing)
The following operating instructions describe the use of the LINK Patella Component and assume the
use of the instrument set available for this procedure. Both the “reaming of the patella surface” technique
and the “resection of the patella surface” technique are shown.
Using the calliper allows the height of the patella to be determined
(47). The dimension is established and an amount corresponding
to the size of the selected implant subtracted. The remainder
equals the target dimension following resection. A minimal
residual dimension of 12 mm should be maintained.
Example:
For a patella size 2 (patella diameter 28 mm), the following
calculation is performed:
1. patella 25 mm thick,
2. resection 8 mm,
3. 17mm re-maining patella bone (47). 47
Patella Size Patella-Ø Patella Height
1 25 mm 7 mm
2 28 mm 8 mm
3 31 mm 9 mm
4 34 mm 10 mm
Preparation for Patella Component
The patella surface can be prepared using the onlay technique by resecting with an oscillating saw or with
the aid of the parella reamer.
Resection
The height of the bone to be resected can be adjusted b
using the height calliper (a) on the resection clamp
(48).
a
48
In doing this, it is important to ensure that the b
remaining patella is sufficiently thick. The patella
is held using the toothed jaws. The sectional plane
must lie parallel to the extended patellar tendon and
the height calliper must lie on the bone. In order to
clamp the patella firmly, the clamp is compressed
firmly and fixed using the lateral setting screw (b) (49).
49
25
Surgical Technique
The resection is carried out using an oscillating saw
with a 1.27 mm thick sawblade. The saw is guided
using the saw slots of the resection clamp (50).
50
Sizing
A patella sizing template corresponding to the implants
is available (51.1)
The template is selected that most adequately covers
the articular surface without any overhang. The handle
is positioned on the medial side of the everted patella.
If bone is deficient on the lateral side, the next smaller
size is selected, but positioned slightly to the medial
side to enhance patellar
tracking (51.2).
51.1
51.2
Option: Onlay Reaming
c
Reaming Preparation
The reaming guide (c) of the appropriate size is inserted
into the patella clamp (d) (52).
Note: Press the button on the side to insert and
remove the guide.
d
52
26
Surgical Technique
A ratchet can be activated on the clamp with the
retaining screw (e). By turning the screw to “ON” the
e
clamp is held shut. By turning the screw to “OFF”,
the ratchet is released (53).
53
Following removal of the peripheral osteophytes,
the patella is clamped with the patella holding clamp,
aiming at the most central position possible (54).
Good fixation is achieved when the spikes on the
guide grip into the bone.
54
Reaming
The patella reamer is selected (f) to match the im-
plant size which was previously determined and the f
corresponding reaming guide (55).The patella reamer
has a Hudson drill connection for direct fixation with
corresponding power tool connection. It is compatible
with other machine systems by using an adapter for
snap lock chuck.
55
The patella surface is prepared using the patella
reamer (f). Reaming depth is verified by means of f
the mechanical stop on the reamer. In addition,
there are marking grooves at 2 mm intervals. The
maximum reaming depth is reached when the
reamer collar makes contact with the reaming
guide (56).
56
27
Surgical Technique
Fixation Hole Drilling
If the saw blade resection option is selected, the g
patella is to be fixed with the patella holding clamp
as follows:
• Insert the patella reaming guide into the
patella holding clamp
• Ratchet function “ON”
• Clamp the patella firmly
Insert the drilling guide (g) for the anchoring holes
which corresponds to the chosen implant size into
the reaming guide (57). 57
In doing so, ensure that the guide pin on the drilling
h
guide lies in the opening made for it in the reaming
guide. Using the appropriately-sized patella drill (h),
6.5 mm drill through the drilling guide until stopped
by the depth stop with the appropriately sized patella
drill (58).
58
28
Surgical Technique
Option: Final Assessment of Extension and Flexion Gap and Alignment of Axis
After final preparation the final spacer according to the deter-
mined femoral size is choosen and inserted. (59-a, 59-b).
The axial alignment is evaluated in extension by means of the
extramedullary alignment rod which is referring to the mechani-
cal axis.
59-a 59-b
29
Surgical Technique
Trial Reduction and Functional Test
The femur trial is selected according to the resected femoral size and is positioned with the attachable
handle and impactor (60).
60
Procedure for GEMINI PS
The femur trial box PS is inserted into the femur trial (61) and fixed with the securing screw (62).
Then the trial is positioned with the attachable handle and impactor.
61 62
In addition, the trial can be inserted into its final position using the driver (63). Then the securing holes
are drilled with the 5.5 mm twist drill (64). A drill guide is applied for guiding the twist drill.
63 64
30
Surgical Technique
The trial plateau is selected and used according to which type of prosthesis is chosen: Fixed Bearing CR (B),
Fixed Bearing PS (C) or Mobile Bearing (A). The different heights are set by means of the shim (D) (65).
Attention!
Shims are availabe in 2 mm-steps, A
marking shows total height:
Height Marking
+ 2 mm 12 mm B
+ 4 mm 14 mm
+ 6 mm 16 mm
+ 8 mm 18 mm
C
Attention! C
Trial plateaus together with tibial preparation plate
imitate 10 mm height. For Mobile Bearing A it is
D
mandatory to assemble the shim (+2 mm) = 12 mm
65
Procedure:
Mobile Bearing
1. Insert tibial trial plateau assembled with shim into tibial preparation plate.
2. Position femoral trial prosthesis.
Fixed Bearing/Fixed Bearing PS
1. Position femoral trial prosthesis.
2. Insert tibial trail plateau (optinal with assembled shim) into tibial
preparation plate.
Note: There is a locking clip to lock the trial plateau with the tibial
preparation plate during trial reduction (66-a).
Then trial reduction is performed with the knee joint in extension 66-a
and flexion, and the ligament tension is checked (66-b).
Height correction:
Fixed Bearing/Fixed Bearing PS
For adjusting the height the trial plateau is removed. Then
the shim assembled accordingly.
Mobile Bearing
For correction of pure plateau height at lying femoral trial
prosthesis, the trial plateau must be used for Fixed Bearing.
Note: Make sure that no bony structures (e.g., osteophytes) or tissue parts
interfere with the movement. If necessary, rework the bone in the notch area,
but especially on the dorsal condyles, using a small curved chisel. 66-b
The trial implants are then removed again.
31
Surgical Technique
Implantation
Compatibility Table
The following table shows (67) possible size combinations.
Compatibility: Femoral/Tibial Components
Femoral Components
1 1B 2 2A 2B 3 4 5
PE Articula- XX = unrestricted compatibility
ting Surfaces 1 1B 2 2A/B 3 4 5 XX = recommended combination
1 XX X X – – – – – X = restricted compatibility,
depending on the patient‘s
2 XX XX XX X X X – –
Components
soft tissue situation in relation
3 XX XX XX XX XX XX X – to the articulating surface
4 XX XX XX XX XX XX XX X – = prohibited combination
Tibial
5 XX XX XX XX XX XX XX XX 67
Implantation Sequence
Mobile Bearing
1 Assembly of Mobile PE Articulating Surface (refer to the compatibility table) with Mobile Tibial Component
2 Implantation of Tibial Component complete with PE Articulating Surface (refer to the compatibility table)
3 Implantation of Femoral Component
Fixed Bearing / Fixed Bearing PS
1 Implantation of Fixed Tibial Component
2 Implantation of Femoral Component
3 Insertion of Fixed or Fixed PS PE Articulating Surface (refer to the compatibility table)
32
Surgical Technique
Tibia Implant Overview:
PE Articulating Surfaces Taper Cap +
Mobile Fixed PS version Stem Extensions
cemented/cementless
Mobile Tibial components Fixed
cemented/cementless 68
Optionally, a taper cap or stem extensions can be placed on the taper and fixed with a screw. The taper
cap and the stem extensions are available in cemented and cementless versions (68).
Note: If the tibial component with stem extension is used, ensure that the tibial medullary canal is suitably
prepared prior to placement.
Note: Installing taper cap or stem extension. The taper cap or stem extension is fixed primarily by taper
clamping, and doubly secured with securing screws.
Attention!
A first stem securing screw (grub screw with point is already in place in the taper hole. The second
securing screw is supplied, in sterile packaging, together with the tibial component.
The taper cap or stem extension is selected according to the planned type of
treatment, and is placed on the taper of the tibial component. When doing this,
align the mark on the taper cap or stem extension with the mark on the taper of
the tibial component. Then fix with a hammer blow (69).
Note: Place gauze over the surface to protect it from the hammer blow.
The first stem securing screw (grub screw with point) (70-a -1) is screwed in and
tighten it strongly by hand. Then screw in the second securing screw (70-a -2, 70-b) 69
with flat point for double securing. Use screwdriver 319-540/00 (allen wrench 2.0 mm)
for both securing screws.
-1
-2 70-a 70-b 70-c
33
Surgical Technique
Implantation: Mobile Bearing
Cemented Tibial Component
The bone cement is prepared following the specific manufacturer‘s instructions,
and applied Select the appropriate Mobile Bearing PE Articular Surface (pay attention
to compatibility chart page 33).
Onto the surgical table, place the Mobile Bearing PE Articular Surface
sideways in longitudinal direction and turn it 90° clockwise to lock it
in its final position (71).The tibial component is then inserted into the
resected tibia, and driven onto the resection surface of the tibial head
with the impactor (72, 73).
71
Attention!
Ensure that excess bone cement is completely removed and no
loose bone cement particles remain, especially in the dorsal joint
region.
72
Cementless Tibial Component
The cementless tibial component is introduced directly into the
resected tibia and driven onto the resection surface of the tibial
head with the impactor.
Note: For each type of fixation (cemented/cementless), only use
the correspondingly marked implants.
73
34
Surgical Technique
Cemented Femoral Component
The femoral component is coupled with the attachable handle and
impactor. After applying the bone cement, prepared according to the
specific manufacturer‘s instructions, the femoral component is implanted.
It is driven on with the attachable handle and impactor (74).
To finish, the femoral component can be re-driven with the
plastic impactor (75).
74
75
Important!
Ensure that excess bone cement is completely removed and no loose bone cement particles remain,
especially in the dorsal joint section.
Cementless Femoral Component
The cementless femoral component is coupled in the same way as described above for the cementable
version, using the attachable handle and impactor. Positioning of the femoral component and implantation
are performed with several measured but forceful hammer blows.
35
Surgical Technique
Implantation: Fixed Bearing and Fixed Bearing PS
Cemented Tibial Component
Attention!
The retaining screw for fixing the UHMWPE articulating surface has been screwed
into the tibial component. It must not be removed for cementing.
The bone cement is prepared following the specific manufacturer‘s instructions,
and applied. The tibial metal tray is then inserted into the resected tibia, and struck
onto the resection surface of the tibial head and driven on with the driver (76).
Attention!
Ensure that excess bone cement is completely
re-moved and no loose bone cement particles
remain, especially in the dorsal joint section (77).
Cementless Tibial Component
The cementless tibial component is introduced
directly into the resected tibia and driven on.
76
77
Cemented Femoral Component
The femoral component is coupled with the attachable handle and
impactor. After applying the bone cement, prepared according to the
specific manufacturer‘s instructions, the femoral component is implanted.
It is driven on with the attachable handle and impactor (78).
To finish, the femoral component can be re-driven with the plastic
impactor (79).
Attention!
Ensure that excess bone cement is completely removed and
no loose bone cement particles remain, especially in the dorsal
78
joint section.
Cementless Femoral Component
The cementless femoral component is coupled in the same way as
described above for the cemented version, using the attachable handle
and impactor. Positioning of the femoral component and implantation are
performed with several measured but forceful hammer blows.
79
36
Surgical Technique
PE Articulating Surface
The holding screw is removed from the tibial component. The PE articulating surface (Fixed CR or PS)
corresponding to the femoral component is selected and inserted. Ensure that the two securing pins on
the bottom surface of the articulating surface are positioned in the recesses of the component.
To finish, the PE articulating surface is fixed to the component using the retaining screw supplied
together with the component (80+81).
80 81
Note: The plateau locking screw has a „self cutting thread“. While tightened the fixation hole of
the PE articular surface will be deformed. In case the plateau locking screw will be released and
needs to be refixed again it will not be securely locked. In such a situation a new PE articular surface
has to be choosen.
Patella Implantation
The reaming guide is removed from the patella clamp.
The polyethylene insert (I) is pressed into the patella
pusher inserter (K) (82-a, 82-b) and the pusher inserter
is then inserted into the patellar clamp (82-c).
K
I
82-a
82-b 82-c
37
Surgical Technique
Following extensive rinsing and removal of all impe-
ding soft tissue, the bone cement is applied to the
back of the implant, and the implant is placed by
hand and pressed on using the patella clamp with
the pusher inserter (83).
Caution!
Prepare the bone cement according to the manu-
facturer’s instructions.
Ensure that excess bone cement is completely
removed and that no loose particles of bone cement 83
remain in the joint.
Functional Test
Implanted femoral and tibial components (84-a, 84-b).
The concluding functional test is designed to check
that the components are correctly positioned
in extension and flexion of the knee joint, and also to
check for correct ligament tension.
84-a 84-b
38
Surgical Technique
APPENDIX 1: Precision Tibial Resection Alignment
The Precision Tibial Resectipon Alignment can be used either extramedullary or intramedullary.
Extramedullary Alignment
Overview of components:
• EM Fixation Jig (85.1) 85.1
• EM Alignment Rod (85.2)
• EM Alignment Guide (85.3) with a foot clamp 85.4
- silicone belt fixation (85.3 -a)
- with spring fixation (85.3 -b)
• Stylus with Guide (85.4)
The stylus comes in the following versions with
the following designations:
- „12“ for GEMINI SL Mobile Bearing
- „10“ for GEMINI SL Fixed CR and Fixed PS 85.6
- „2“ for GEMINI SL defect referencing
85.5-a 85.5-b
• Tibial Base Guide (85.5)
The tibial base guide consists of tibial cutting block
holder (85.5-a) and tibial base guide body (85.5-b). 85.2
The tibial base guide allows fine tuning of varus-valgus,
posterior slope and resection level.
• Tibial Cutting Block (85.6)
- Symmetrical
- Asymmetrical (medial approach), left
- Asymmetrical (medial approach), right
85.3
85.3-a
85.3-b
39
Surgical Technique
Assembly
Compponents are assembled accordingly:
Step 1:
3.
First the tibial cutting jig is assembled.
85.4
According to the preferred technique the tibial cutting
block (85.6) is determined (symmetrical or asymmetrical).
The tibial cutting block is mounted with the tibial cutting block
holder (85.5 -a) and then connected with the tibial base guide
body (85.5 -b). Release the locking screw at the stylus guide
then insert the foot of the stylus guide (85.4) into the attachment 2.
of the tibial cutting block. 1.
85.6
85.5 -a 85.5 -b
Note: Preselect varus/valgus adjustment to „0“ and
posterior slope to „5°“.The device allows fine tuning of
the adjustment after the tibial cutting jig has been fixed
85.1
to the tibia.
Step 2:
The EM alignment rod (85.2) is assembled with the EM fixation jig (85.1)
and then pushed through the tibial base guide of the tibial cutting jig.
Proximal tibial cutting jig (85.7).
85.2
Step 3:
Insert the foot clamp (85.3-a) into the alignment guide (85.3)
and hand-tighten the locking screw (A).
85.3
85.7
A
85.3-a
40
Surgical Technique
Positioning
Attach the foot clamp to the ankle joint and secure
with the silicone belt (87-a).
Option: The foot clamp can be choosen either
spring clamp ot silicone belt.
The EM orientation rod with the fixation jig (86.7)
is inserted through the tibial base guide (86.3) and
into the distal guide tube, and is secured with the
connecting sleeve (87-b).
86.7
Note: The connecting sleeve secures the proximal
tibial cutting jig of anaterior release. The cutting jig
can still be moved proximal/ distal.
87-a
86.3
87-b
Proximal, the tip of the fixation jig initially rests
in the area of the point of attachment of the
anterior cruciate ligament, roughly centrally in
the region of the intercondylar eminence on the
tibial plateau (88-a). Then the tip is knocked
in with a hammer (88-b). The alignment instru-
ment can still be rotated.
88-b
88-a
41
Surgical Technique
Rotation Definition and Axial Alignment
To determine the correct alignment of the
rotation, the alignment instrument is aligned
centrally along the tibial shaft axis (89-a).
Note: For assistance, the center of the tibia, the
tibial tuberosity and the second toe can be marked
with a sterile marker pen.
When the alignment instrument, viewed from
the side, is parallel to the ventral tibia, this means
that the saw guide is set in the previously selected
posterior slope. Fine adjustment is possible following
alignment (89-b).
89-a 89-b
Then secure the tibial resection alignment proximal with two drill pins
through the dedicated holes of the tibial base guide (90-a, 90-b).
90-a 90-b
42
Surgical Technique
Resection Adjustment
The EM fixation jig and the stylus are removed and the EM Alignment
Rod ist slided distal into the EM Alignment Guide allowing now for
micro adjustment of the tibial resection.
Fine Tuning for the Tibial Resection
Correct alignment can be checked using the cutting template (91).
Posterior Slope
GEMINI SL Tibia Slope Recommendations: For a Mobile Bearing and 91
Fixed Bearing Cruciate Retaining (CR) configuration it is recommended
to set the tibial posterior slope at 5 degrees. For a Fixed Bearing Posterior
Stabilized (PS) configuration, a range of 0 – 5 degrees of tibial posterior
slope is recommended. Posterior slope can be adjusted in 1° steps with
the micro adjustment (A) (92).
Tibial Resection Height
After presetting the resection height with the selected stylus version,
the resection hight can now be micro adjusted with the screw (C) (92). A
B
C 92
Varus-valgus Adjustment
The micro adjustment (B) allows precise correction of the varus or
valgus alignment.The alignment can be checked by inserting the
EM alignment rod (93).
Then the tibial cutting block is fixed with a drill pin (94-a) and the resection
is performed (94-b). The cross-pin of the tibial cutting block is removed.
B
The fixation of the ankle clamp is taken off then the alignment device is re-
moved leaving the two pins in place which will allow a simple repositioning
if re-resection is necessary.
94-a 94-b 93
43
Surgical Technique
APPENDIX 2: Precision Tibial Resection Alignment
Intramedullary Alignment
Overview of the components:
• IM Fixation Jig (95.1)
• EM Alignment Rod (95.2)
95.6
• Stylus with Guide (95.3)
The stylus comes in the following versions
95.1
with the following designations:
- „12“ for GEMINI SL Mobile Bearing
- „10“ for GEMINI SL Fixed and Fixed PS
- „2“ for GEMINI SL defect referencing
• Tibial Cutting Block (95.4) 95.3
- Symmetrical
- Asymmetrical (medial approach), left
- Asymmetrical (medial approach), right 95.5 -b 95.7
95.5 -a
• Tibial Base Guide (95.5)
The tibial base guide consists of tibial cutting 95.4
95.2
block holder (95.5-a) and tibial base guide
body (95.5-b).The tibial base guide allows fine
tuning of varus-valgus, posterior slope and
resection level.
• T-Handle (95.6)
• IM Guide Rod (95.7)
44
Surgical Technique
Assembly:
95.3
Step 1:
3.
First the tibial cutting jig is assembled.
According to the preferred technique the tibial
cutting block (95.4) is determined (symmetrical or
asymmetrical). The tibial cutting block is mounted with
the varus/valgus adjustment (95.5 -a) and then connected
with the tibial base guide (95.5 -b). Release the locking screw
at the stylus guide then insert the foot of the stylus guide (95.3) 2.
into the attachment of the tibial cutting block. 1.
95.4 95.5-a 95.5 -b
Note: Preselect varus/valgus adjustment to „0“ and
posterior slope to „5°“. The device allows micro tuning
of the adjustment after the tibial cutting jig has been fixed
to the tibia.
96.1
Step 2: 96.2
The EM alignment rod (96.2) is assembled with the IM
fixation jig (96.1) and then pushed through the tibial base
guide of the tibial cutting jig (97-a, 97-b).
Proximal tibial cutting jig.
97-a
97-b
45
Surgical Technique
Positioning
The entrance to the meduallary canal is opened using the step
drill having a pilot tip (98).
98
Connect the T-Handle with the IM guide rod (99-a).
Insert the IM guide rod carefully into the tibial canal
until the isthmus is firmly engaged (99-b)
99-b 99-a
46
Surgical Technique
The T-Handle is removed. The previously assembled tibial cutting jig is postioned
on the IM guide rod until the tip of the stylus rests on the tibial plateau (100-a, 100-b).
Rotation Definition and Axial Alignment
To determine the correct alignment of the rotation, the alignment
instrument is aligned centrally along the tibial shaft axis (100-a).
When the alignment instrument, viewed from the side, is
parallel to the anterior tibia, this means that the sawguide
is set in the previously selected posterior slope. Fine tuning
is possible following alignment (100-b).
100-b 100-a
Determination of Resection Height
The resection height is determined using the stylus according
to preferred referencing.
Then secure the tibial cutting jig anterior with two drill pins
through the dedicated holes of the tibial base guide (101-a, 101-b).
Connect the T-Handle with the IM guide rod and remove the
IM guide rod.
101-b 101-a
47
Surgical Technique
Resection Adjustment
The IM fixation jig and the stylus with guide are removed and the
EM alignment rod is pushed distal into the tibaoil base guide.
Fine Tuning for the Tibial Resection
The tibial saw guide is now pushed against the anterior tibial
cortical bone as far as the limit stop. Correct alignment can be
checked using the cutting template (102).
Posterior Slope 102
GEMINI SL: Tibia Slope Recommendations: For Mobile Bearing
and Fixed Bearing Cruciate Retaining (CR) configuration it is recom-
mended to set the tibial posterior slope at 5 degrees. For a Fixed
Bearing Pos-terior Stabilized (PS) configuration, a range of 0 – 5
degrees of tibial posterior slope is recommended. Posterior slope
can be adjusted in 1° steps with the fine tuning (A) (103).
Tibial Resection Height
After presetting the resection height with the selected stylus version, A
the resection hight can now be fine tuned with the screw (C) (103).
B
Varus-valgus Adjustment
The fine tuning (B) allows precise correction of the varus or valgus C 103
alignment (103).The alignment can be checked by inserting the
EM alignment rod (104).
Then the tibial cutting block is fixed with a drill pin (105-a) and the B
resection is performed (105-b).
105-a 105-b 104
48
Surgical Technique
APPENDIX 3: GEMINI SL EXTRABONE – Extramedullary Femoral Referencing
Preoperative Planning
With EXTRABONE, the position and alignment of the distal femoral resection
plane are planned on the A/P femoral X-ray. A special X-ray template (106) is
available for this purpose. The long axis is aligned to the head of the femur.
The short axis is applied tangentially to the widest point of the distal femoral
condyle.Then the distance to the shorter femoral condyle is determined as the
difference. In the example (107-a, 107-b), there is a 2 mm of set between the
medial and lateral femoral condyles. With this method, the cartilage layer is not
taken into account, as it is not visualized by the imaging technique. Therefore
the cartilage layer has to be removed intraoperatively in order to create the
same preoperative condition.
Note: The basis for a correct technique is accurate preplanning based on a
X-ray of the total femur.
106
107-b
107-a
As the cartilage was not taken into account in the preceding
radiological planning, the cartilage layer – if still present – has
to be removed (108).
108
49
Surgical Technique
Femur Preparation – Distal Resection
With EXTRABONE, extramedullary referencing for the distal femoral resection is performed as follows.
The GEMINI SL femoral component permits size-specific distal resection:
Sizes 1-2B distal femoral cutting block "S"
Sizes 3-5 distal femoral cutting block "L"
The femoral sizer is used to verify the femoral size determined in preoperative planning and to select
the distal femoral cutting block (109, 109-a, 109-b). The sizer is held against the femur to determine its
size. The contour shown below corresponds to size 3. If the femur profile is smaller, the distal cutting
block "S" should be selected. If the femur profile is a perfect fit or larger, the distal cutting block "L"
should be selected. The m/l dimension can be checked in addition.
Size > = L
m/l
109
m/l
109-a
109-b
The guide for flexion/extension alignment is placed on the anterior bone (110, 110-a, 110-b).
110 110-a 110-b
50
Surgical Technique
The positioner guide is placed vertically onto the guide and fixed with two drill pins (111, 111-a,111-b,111-c).
111 111-a 111-b 111-c
The alignment instrument for holding the distal cutting block is attached with the connection piece (112) and
inserted into the positioner guide (112-a). The alignment instrument is inserted far enough that the two condyle
stop plates rest against the distal medial and lateral femoral condyles (112-b).
The condyle stop plates can be swiveled from 0° to 25° to ensure that they are precisely positioned.
112-a
112
112-b
51
Surgical Technique
The alignment instrument has a setting screw and scale medially and laterally
(113) for transferring the preoperative planning data for the offset of the distal
femoral condyles to the distal femur.
The varus/valgus orientation of the resection is achieved by adjusting the condyle
stop plates with the medial and lateral setting screws.
This alignment can be verified intraoperatively with the axial alignment guide and an
alignment rod (114).
113
114
Two drill pins are positioned for referencing the position of the distal
femoral cutting block. The distal femoral cutting block is positioned
over these two drill pins, taking into account the previously determined
size ("S" or "L") (115). 115
The cutting template can be used to check the alignment
of the distal resection (116).
116
52
Surgical Technique
To make the cutting block more stable, a third drill pin is fixed into
the lateral diagonal hole. There is the option of inserting a fourth
drill pin (117).
117
The distal saw cut is made at 90° flexion (118).
118
The parallel drill pins remain in position (119).
Note: It is possible to adjust the extension gap by repeating
the resection.
119
The rest of the procedure is performed in the same way as with the
standard surgical technique GEMINI SL Total Knee System.
53
Implants
GEMINI SL Femoral Components – Fixed Bearing CR / Mobile Bearing
to be used with Fixed Bearing CR and Mobile Bearing PE Articulating Surfaces
cemented
CoCrMo CoCrMo CoCrMo/TiNbN
SMS* SpheroGrip SMS*/LINK PorEx ** A B Identi-
Size Side
mm mm fication
318-102/01 318-402/01 318-206/01 x-small right 52 55 R1
318-102/16 318-402/16 318-206/16 x-small plus right 55 62 R1B
318-102/02 318-402/02 318-206/02 small right 59 62 R2
318-102/25 318-402/25 318-206/25 medium small right 59 69 R2A
318-102/26 318-402/26 318-206/26 medium small plus right 62 69 R2B
318-102/03 318-402/03 318-206/03 medium right 65 69 R3
318-102/04 318-402/04 318-206/04 large right 67 72 R4
318-102/05 318-402/05 318-206/05 x-large right 72 76 R5
318-103/01 318-403/01 318-207/01 x-small left 52 55 L1
318-103/16 318-403/16 318-207/16 x-small plus left 55 62 L1B
318-103/02 318-403/02 318-207/02 small left 59 62 L2
318-103/25 318-403/25 318-207/25 medium small left 59 69 L2A
318-103/26 318-403/26 318-207/26 medium small plus left 62 69 L2B
318-103/03 318-403/03 318-207/03 medium left 65 69 L3
318-103/04 318-403/04 318-207/04 large left 67 72 L4
318-103/05 318-403/05 318-207/05 x-large left 72 76 L5
cementless
CoCrMo/TiCaP*** CoCrMo/TiNbN/TiCaP***
SMS* SMS*/LINK PorEx ** A B Identi-
Size Side
mm mm fication
318-204/01 318-208/01 x-small right 52 55 R1
318-204/16 318-208/16 x-small plus right 55 62 R1B
318-204/02 318-208/02 small right 59 62 R2
318-204/25 318-208/25 medium small right 59 69 R2A
318-204/26 318-208/26 medium small plus right 62 69 R2B
318-204/03 318-208/03 medium right 65 69 R3
318-204/04 318-208/04 large right 67 72 R4
318-204/05 318-208/05 x-large right 72 76 R5
318-205/01 318-209/01 x-small left 52 55 L1
318-205/16 318-209/16 x-small plus left 55 62 L1B
318-205/02 318-209/02 small left 59 62 L2
318-205/25 318-209/25 medium small left 59 69 L2A
318-205/26 318-209/26 medium small plus left 62 69 L2B
318-205/03 318-209/03 medium left 65 69 L3
318-205/04 318-209/04 large left 67 72 L4
318-205/05 318-209/05 x-large left 72 76 L5
* SMS: Squarical Monobloc Structure.
** LINK PorEx: TiNbN = Titanium Niobium Nitride; hypoallergenic coating (gold color).
*** TiCaP double coating: Titan/Calcium Phosphate (CaP).
54
Implants
GEMINI SL Femoral Components – Fixed Bearing PS
to be used with Fixed Bearing PS PE Articulating Surfaces
cemented
CoCrMo CoCrMo CoCrMo/TiNbN
SMS* SpheroGrip SMS*/LINK PorEx ** A B Identi-
Size Side
mm mm fication
318-112/01 318-412/01 318-214/01 x-small right 52 55 R1
318-112/16 318-412/16 318-214/16 x-small plus right 55 62 R1B
318-112/02 318-412/02 318-214/02 small right 59 62 R2
318-112/25 318-412/25 318-214/25 medium small right 59 69 R2A
318-112/26 318-412/26 318-214/26 medium small plus right 62 69 R2B
318-112/03 318-412/03 318-214/03 medium right 65 69 R3
318-112/04 318-412/04 318-214/04 large right 67 72 R4
318-112/05 318-412/05 318-214/05 x-large right 72 76 R5
318-113/01 318-413/01 318-215/01 x-small left 52 55 L1
318-113/16 318-413/16 318-215/16 x-small plus left 55 62 L1B
318-113/02 318-413/02 318-215/02 small left 59 62 L2
318-113/25 318-413/25 318-215/25 medium small left 59 69 L2A
318-113/26 318-413/26 318-215/26 medium small plus left 62 69 L2B
318-113/03 318-413/03 318-215/03 medium left 65 69 L3
318-113/04 318-413/04 318-215/04 large left 67 72 L4
318-113/05 318-413/05 318-215/05 x-large left 72 76 L5
cementless
CoCrMo/TiCaP*** CoCrMo/TiNbN/TiCaP***
SMS* SMS*/LINK PorEx ** A B Identi-
Size Side
mm mm fication
318-212/01 318-216/01 x-small right 52 55 R1
318-212/16 318-216/16 x-small plus right 55 62 R1B
318-212/02 318-216/02 small right 59 62 R2
318-212/25 318-216/25 medium small right 59 69 R2A
318-212/26 318-216/26 medium small plus right 62 69 R2B
318-212/03 318-216/03 medium right 65 69 R3
318-212/04 318-216/04 large right 67 72 R4
318-212/05 318-216/05 x-large right 72 76 R5
318-213/01 318-217/01 x-small left 52 55 L1
318-213/16 318-217/16 x-small plus left 55 62 L1B
318-213/02 318-217/02 small left 59 62 L2
318-213/25 318-217/25 medium small left 59 69 L2A
318-213/26 318-217/26 medium small plus left 62 69 L2B
318-213/03 318-217/03 medium left 65 69 L3
318-213/04 318-217/04 large left 67 72 L4
318-213/05 318-217/05 x-large left 72 76 L5
* SMS: Squarical Monobloc Structure.
** LINK PorEx: TiNbN = Titanium Niobium Nitride; hypoallergenic coating (gold color).
*** TiCaP double coating: Titan/Calcium Phosphate (CaP).
55
Implants
GEMINI SL Tibial Components – Fixed Bearing
to be used with Fixed Bearing CR and PS PE Articulating Surfaces
cemented
Taper 10/12 mm
CoCrMo CoCrMo CoCrMo/TiNbN
SMS* SpheroGrip SMS*/LINK PorEx ** A B Identi-
Size Side
mm mm fication
318-372/01 318-482/01 318-328/01 x-small right 42 62 R1
318-372/02 318-482/02 318-328/02 small right 46 69 R2
318-372/03 318-482/03 318-328/03 medium right 47 74 R3
318-372/04 318-482/04 318-328/04 large right 53 78 R4
318-372/05 318-482/05 318-328/05 x-large right 56 85 R5
318-373/01 318-483/01 318-329/01 x-small left 42 62 L1
318-373/02 318-483/02 318-329/02 small left 46 69 L2
318-373/03 318-483/03 318-329/03 medium left 47 74 L3
318-373/04 318-483/04 318-329/04 large left 53 78 L4
318-373/05 318-483/05 318-329/05 x-large left 56 85 L5
incl. Locking Screw for PE Articulating Surface
cementless
Taper 10/12 mm
CoCrMo/TiCaP*** CoCrMo/TiNbN/TiCaP***
SMS* SMS*/LINK PorEx ** A B Identi-
Size Side
mm mm fication
318-376/01 318-336/01 x-small right 42 62 R1
318-376/02 318-336/02 small right 46 69 R2
318-376/03 318-336/03 medium right 47 74 R3
318-376/04 318-336/04 large right 53 78 R4
318-376/05 318-336/05 x-large right 56 85 R5
318-377/01 318-337/01 x-small left 42 62 L1
318-377/02 318-337/02 small left 46 69 L2
318-377/03 318-337/03 medium left 47 74 L3
318-377/04 318-337/04 large left 53 78 L4
318-377/05 318-337/05 x-large left 56 85 L5
incl. Locking Screw for PE Articulating Surface
* SMS: Squarical Monobloc Structure.
** LINK PorEx: TiNbN = Titanium Niobium Nitride; hypoallergenic coating (gold color).
*** TiCaP double coating: Titan/Calcium Phosphate (CaP).
56
Implants
GEMINI SL Tibial Components – Mobile Bearing
to be used with Mobile Bearing PE Articulating Surfaces
cemented
Taper 10/12 mm
CoCrMo CoCrMo CoCrMo/TiNbN
SMS* SpheroGrip SMS*/LINK PorEx ** A B Identi-
Size Side
mm mm fication
318-312/01 318-452/01 318-324/01 x-small right 42 62 R1
318-312/02 318-452/02 318-324/02 small right 46 69 R2
318-312/03 318-452/03 318-324/03 medium right 47 74 R3
318-312/04 318-452/04 318-324/04 large right 53 78 R4
318-312/05 318-452/05 318-324/05 x-large right 56 85 R5
318-313/01 318-453/01 318-325/01 x-small left 42 62 L1
318-313/02 318-453/02 318-325/02 small left 46 69 L2
318-313/03 318-453/03 318-325/03 medium left 47 74 L3
318-313/04 318-453/04 318-325/04 large left 53 78 L4
318-313/05 318-453/05 318-325/05 x-large left 56 85 L5
cementless
Taper 10/12 mm
CoCrMo/TiCaP*** CoCrMo/TiNbN/TiCaP***
SMS* SMS*/LINK PorEx ** A B Identi-
Size Side
mm mm fication
318-316/01 318-326/01 x-small right 42 62 R1
318-316/02 318-326/02 small right 46 69 R2
318-316/03 318-326/03 medium right 47 74 R3
318-316/04 318-326/04 large right 53 78 R4
318-316/05 318-326/05 x-large right 56 85 R5
318-317/01 318-327/01 x-small left 42 62 L1
318-317/02 318-327/02 small left 46 69 L2
318-317/03 318-327/03 medium left 47 74 L3
318-317/04 318-327/04 large left 53 78 L4
318-317/05 318-327/05 x-large left 56 85 L5
* SMS: Squarical Monobloc Structure.
** LINK PorEx: TiNbN = Titanium Niobium Nitride; hypoallergenic coating (gold color).
*** TiCaP double coating: Titan/Calcium Phosphate (CaP).
57
Implants
GEMINI SL Tibial Components – Fixed Bearing CR
to be used with Femoral Components CR (Cruciate Retaining) and Fixed Bearing Tibial Components
Fixed Bearing CR - PE Articulating Surfaces
UHMWPE
Size Side
H h A B Identi-
mm mm mm mm fication
318-231/10 x-small right 10 6 42 62 R1
318-238/10 x-small plus right 10 6 46 69 R1B
318-232/10 small right 10 6 46 69 R2
318-237/10 medium small / medium small plus right 10 6 47 74 R2A / R2B
318-233/10 medium right 10 6 47 74 R3
318-234/10 large right 10 6 53 78 R4
318-235/10 x-large right 10 6 56 85 R5
318-241/10 x-small left 10 6 42 62 L1
318-248/10 x-small plus left 10 6 46 69 L1B
318-242/10 small left 10 6 46 69 L2
318-247/10 medium small / medium small plus left 10 6 47 74 L2A / L2B
318-243/10 medium left 10 6 47 74 L3
318-244/10 large left 10 6 53 78 L4
318-245/10 x-large left 10 6 56 85 L5
318-231/12 x-small right 12 8 42 62 R1
318-238/12 x-small plus right 12 8 46 69 R1B
318-232/12 small right 12 8 46 69 R2
318-237/12 medium small / medium small plus right 12 8 47 74 R2A / R2B
318-233/12 medium right 12 8 47 74 R3
318-234/12 large right 12 8 53 78 R4
318-235/12 x-large right 12 8 56 85 R5
318-241/12 x-small left 12 8 42 62 L1
318-248/12 x-small plus left 12 8 46 69 L1B
318-242/12 small left 12 8 46 69 L2
318-247/12 medium small / medium small plus left 12 8 47 74 L2A / L2B
318-243/12 medium left 12 8 47 74 L3
318-244/12 large left 12 8 53 78 L4
318-245/12 x-large left 12 8 56 85 L5
318-231/14 x-small right 14 10 42 62 R1
318-238/14 x-small plus right 14 10 46 69 R1B
318-232/14 small right 14 10 46 69 R2
318-237/14 medium small / medium small plus right 14 10 47 74 R2A / R2B
318-233/14 medium right 14 10 47 74 R3
318-234/14 large right 14 10 53 78 R4
318-235/14 x-large right 14 10 56 85 R5
Further implant sizes on next page...
58
Implants
Fixed Bearing CR - PE Articulating Surfaces
UHMWPE
H h A B Identi-
Size Side
mm mm mm mm fication
318-241/14 x-small left 14 10 42 62 L1
318-248/14 x-small plus left 14 10 46 69 L1B
318-242/14 small left 14 10 46 69 L2
318-247/14 medium small / medium small plus left 14 10 47 74 L2A / L2B
318-243/14 medium left 14 10 47 74 L3
318-244/14 large left 14 10 53 78 L4
318-245/14 x-large left 14 10 56 85 L5
318-231/16 x-small right 16 12 42 62 R1
318-238/16 x-small plus right 16 12 46 69 R1B
318-232/16 small right 16 12 46 69 R2
318-237/16 medium small / medium small plus right 16 12 47 74 R2A / R2B
318-233/16 medium right 16 12 47 74 R3
318-234/16 large right 16 12 53 78 R4
318-235/16 x-large right 16 12 56 85 R5
318-241/16 x-small left 16 12 42 62 L1
318-248/16 x-small plus left 16 12 46 69 L1B
318-242/16 small left 16 12 46 69 L2
318-247/16 medium small / medium small plus left 16 12 47 74 L2A / L2B
318-243/16 medium left 16 12 47 74 L3
318-244/16 large left 16 12 53 78 L4
318-245/16 x-large left 16 12 56 85 L5
318-231/18 x-small right 18 14 42 62 R1
318-238/18 x-small plus right 18 14 46 69 R1B
318-232/18 small right 18 14 46 69 R2
318-237/18 medium small / medium small plus right 18 14 47 74 R2A / R2B
318-233/18 medium right 18 14 47 74 R3
318-234/18 large right 18 14 53 78 R4
318-235/18 x-large right 18 14 56 85 R5
318-241/18 x-small left 18 14 42 62 L1
318-248/18 x-small plus left 18 14 46 69 L1B
318-242/18 small left 18 14 46 69 L2
318-247/18 medium small / medium small plus left 18 14 47 74 L2A / L2B
318-243/18 medium left 18 14 47 74 L3
318-244/18 large left 18 14 53 78 L4
318-245/18 x-large left 18 14 56 85 L5
59
Implants
GEMINI SL Tibial Components – Fixed Bearing PS
to be used with Femoral Components PS (Posterior Stabilized) and Fixed Bearing Tibial Components
Fixed Bearing PS - PE Articulating Surfaces
UHMWPE
H h A B Identi-
Size Side
mm mm mm mm fication
318-251/10 x-small right 10 6 42 62 R1
318-258/10 x-small plus right 10 6 46 69 R1B
318-252/10 small right 10 6 46 69 R2
318-256/10 medium small / medium small plus right 10 6 47 74 R2A / R2B
318-253/10 medium right 10 6 47 74 R3
318-254/10 large right 10 6 53 78 R4
318-255/10 x-large right 10 6 56 85 R5
318-261/10 x-small left 10 6 42 62 L1
318-268/10 x-small plus left 10 6 46 69 L1B
318-262/10 small left 10 6 46 69 L2
318-266/10 medium small / medium small plus left 10 6 47 74 L2A / L2B
318-263/10 medium left 10 6 47 74 L3
318-264/10 large left 10 6 53 78 L4
318-265/10 x-large left 10 6 56 85 L5
318-251/12 x-small right 12 8 42 62 R1
318-258/12 x-small plus right 12 8 46 69 R1B
318-252/12 small right 12 8 46 69 R2
318-256/12 medium small / medium small plus right 12 8 47 74 R2A / R2B
318-253/12 medium right 12 8 47 74 R3
318-254/12 large right 12 8 53 78 R4
318-255/12 x-large right 12 8 56 85 R5
318-261/12 x-small left 12 8 42 62 L1
318-268/12 x-small plus left 12 8 46 69 L1B
318-262/12 small left 12 8 46 69 L2
318-266/12 medium small / medium small plus left 12 8 47 74 L2A / L2B
318-263/12 medium left 12 8 47 74 L3
318-264/12 large left 12 8 53 78 L4
318-265/12 x-large left 12 8 56 85 L5
Further implant sizes on next page...
60
Implants
Fixed Bearing PS - PE Articulating Surfaces
UHMWPE
H h A B Identi-
Size Side
mm mm mm mm fication
318-251/14 x-small right 14 10 42 62 R1
318-258/14 x-small plus right 14 10 46 69 R1B
318-252/14 small right 14 10 46 69 R2
318-256/14 medium small / medium small plus right 14 10 47 74 R2A / R2B
318-253/14 medium right 14 10 47 74 R3
318-254/14 large right 14 10 53 78 R4
318-255/14 x-large right 14 10 56 85 R5
318-261/14 x-small left 14 10 42 62 L1
318-268/14 x-small plus left 14 10 46 69 L1B
318-262/14 small left 14 10 46 69 L2
318-266/14 medium small / medium small plus left 14 10 47 74 L2A / L2B
318-263/14 medium left 14 10 47 74 L3
318-264/14 large left 14 10 53 78 L4
318-265/14 x-large left 14 10 56 85 L5
318-251/16 x-small right 16 12 42 62 R1
318-258/16 x-small plus right 16 12 46 69 R1B
318-252/16 small right 16 12 46 69 R2
318-256/16 medium small / medium small plus right 16 12 47 74 R2A / R2B
318-253/16 medium right 16 12 47 74 R3
318-254/16 large right 16 12 53 78 R4
318-255/16 x-large right 16 12 56 85 R5
318-261/16 x-small left 16 12 42 62 L1
318-268/16 x-small plus left 16 12 46 69 L1B
318-262/16 small left 16 12 46 69 L2
318-266/16 medium small / medium small plus left 16 12 47 74 L2A / L2B
318-263/16 medium left 16 12 47 74 L3
318-264/16 large left 16 12 53 78 L4
318-265/16 x-large left 16 12 56 85 L5
318-251/18 x-small right 18 14 42 62 R1
318-258/18 x-small plus right 18 14 46 69 R1B
318-252/18 small right 18 14 46 69 R2
318-256/18 medium small / medium small plus right 18 14 47 74 R2A / R2B
318-253/18 medium right 18 14 47 74 R3
318-254/18 large right 18 14 53 78 R4
318-255/18 x-large right 18 14 56 85 R5
318-261/18 x-small left 18 14 42 62 L1
318-268/18 x-small plus left 18 14 46 69 L1B
318-262/18 small left 18 14 46 69 L2
318-266/18 medium small / medium small plus left 18 14 47 74 L2A / L2B
318-263/18 medium left 18 14 47 74 L3
318-264/18 large left 18 14 53 78 L4
318-265/18 x-large left 18 14 56 85 L5
61
Implants
GEMINI SL Tibial Components – Mobile Bearing
to be used with Femoral Components CR (Cruciate Retaining) and Mobile Bearing Tibial Components
Mobile Bearing - PE Articulating Surfaces
UHMWPE
H h A B Identi-
Size Side
mm mm mm mm fication
318-431/12 x-small right 12 5 42 62 R1
318-438/12 x-small plus right 12 5 46 69 R1B
318-432/12 small right 12 5 46 69 R2
318-437/12 medium small / medium small plus right 12 5 47 74 R2A / R2B
318-433/12 medium right 12 5 47 74 R3
318-434/12 large right 12 5 53 78 R4
318-435/12 x-large right 12 5 56 85 R5
318-441/12 x-small left 12 5 42 62 L1
318-448/12 x-small plus left 12 5 46 69 L1B
318-442/12 small left 12 5 46 69 L2
318-447/12 medium small / medium small plus left 12 5 47 74 L2A / L2B
318-443/12 medium left 12 5 47 74 L3
318-444/12 large left 12 5 53 78 L4
318-445/12 x-large left 12 5 56 85 L5
318-431/14 x-small right 14 7 42 62 R1
318-438/14 x-small plus right 14 7 46 69 R1B
318-432/14 small right 14 7 46 69 R2
318-437/14 medium small / medium small plus right 14 7 47 74 R2A / R2B
318-433/14 medium right 14 7 47 74 R3
318-434/14 large right 14 7 53 78 R4
318-435/14 x-large right 14 7 56 85 R5
318-441/14 x-small left 14 7 42 62 L1
318-448/14 x-small plus left 14 7 46 69 L1B
318-442/14 small left 14 7 46 69 L2
318-447/14 medium small / medium small plus left 14 7 47 74 L2A / L2B
318-443/14 medium left 14 7 47 74 L3
318-444/14 large left 14 7 53 78 L4
318-445/14 x-large left 14 7 56 85 L5
318-431/16 x-small right 16 9 42 62 R1
318-438/16 x-small plus right 16 9 46 69 R1B
318-432/16 small right 16 9 46 69 R2
318-437/16 medium small / medium small plus right 16 9 47 74 R2A / R2B
318-433/16 medium right 16 9 47 74 R3
318-434/16 large right 16 9 53 78 R4
318-435/16 x-large right 16 9 56 85 R5
Further implant sizes on next page...
62
Implants
Mobile Bearing - PE Articulating Surfaces
UHMWPE
H h A B Identi-
Size Side
mm mm mm mm fication
318-441/16 x-small left 16 9 42 62 L1
318-448/16 x-small plus left 16 9 46 69 L1B
318-442/16 small left 16 9 46 69 L2
318-447/16 medium small / medium small plus left 16 9 47 74 L2A / L2B
318-443/16 medium left 16 9 47 74 L3
318-444/16 large left 16 9 53 78 L4
318-445/16 x-large left 16 9 56 85 L5
318-431/18 x-small right 18 11 42 62 R1
318-438/18 x-small plus right 18 11 46 69 R1B
318-432/18 small right 18 11 46 69 R2
318-437/18 medium small / medium small plus right 18 11 47 74 R2A / R2B
318-433/18 medium right 18 11 47 74 R3
318-434/18 large right 18 11 53 78 R4
318-435/18 x-large right 18 11 56 85 R5
318-441/18 x-small left 18 11 42 62 L1
318-448/18 x-small plus left 18 11 46 69 L1B
318-442/18 small left 18 11 46 69 L2
318-447/18 medium small / medium small plus left 18 11 47 74 L2A / L2B
318-443/18 medium left 18 11 47 74 L3
318-444/18 large left 18 11 53 78 L4
318-445/18 x-large left 18 11 56 85 L5
Femoral Component
Polyethylene Articula-
ting Surfaces (Mobile)
Tibial Component (Mobile)
63
Implants
GEMINI SL Total Knee Replacement
Taper Caps with LINK PorEx (TiNbN = Titanium Niobium
Nitride) available as custom-made implant on request.
Taper Caps for Tibial Components
cemented The taper of the tibial component
for coupling with modular stems
CoCrMo can be covered with the taper cap,
thus providing a short prosthesis
318-314 Ø 16 mm, length (L) 28 mm
stem (overall length = 49 mm).
The cap is fixed to the taper by
cementless means of a hex screw using a
hex screwdriver (hex. 2.0 mm).
CoCrMo/TiCaP * The cap is provided separately in
sterile packaging.
318-314/01 Ø 16 mm, length (L) 28 mm
Ti6Al4V (Tilastan)
318-315 Ø 16 mm, length (L) 28 mm
* TiCaP double coating: Titan (Ti)/Calcium Phosphate (CaP)
Ø
Patella Resurfacing Components, 3-pegs
H
UHMWPE
Ø Height
Size
mm mm
318-401/25 1 25 7
318-401/28 2 28 8
318-401/31 3 31 9
318-401/34 4 34 10
Instrument Set for Patella Components are available on request.
64
Implants
ØD
Tibial Stems
Ti6Al4V (Tilastan) ØB
cemented
L
ØA ØB ØD L LT
mm mm mm mm mm
318-190/07 7 10 16 50 70
318-190/10 7 10 16 80 100
318-190/14 7 10 16 120 140
ØA
Tibial Stems
Ti6Al4V (Tilastan)
ØD
cementless
ØA ØB ØD L LT ØB
mm mm mm mm mm
L
318-191/07 10 16 16 50 70
318-191/10 10 16 16 80 100
318-191/14 10 16 16 120 140
ØA
LT
65
Instruments
GEMINI SL Instrument Set – Overview
GEMINI SL Instrument Sets
319-011/31 Case 1 SIT-CUT Easy Tibia Resection Alignment
319-020/31 Case 2 Femoral Preparation
319-030/31 Case 3 Spacer
319-040/31 Case 4 Tibia Preparation
445-008/00 Case 5 Femoral Trial Prostheses
319-060/32 Case 6 Femoral PS Preparation
319-070/31 Case 7 General Instruments
319-080/31 Case 8 Final Spacer
340-200/01 Patella Instrument Set
Additional Instrument Sets:
319-010/31 Precision Tibia Resection Alignment
317-800/00 EXTRABONE Instrument Set
66
Instruments
319-011/31 Case 1 – SIT-CUT Easy Tibia Resection Alignment
1 2 3 4 5 6 13 14
7 8 9 12 10 11 15
1 319-611/31 Instrument Tray 1B, empty, 485 x 253 x 100 mm (incl. lid)
2 319-180/01 EM Alignment Rod
3 319-182/00 Foot Clamp, silicon belt connection
4 319-181/01 EM Alignment Guide
5 319-525/00 Alignment Rod, extramedullary
6 319-520/01 Alignment Rod, extramedullary
7 319-125/01 Stylus, Referencer Defect Compartment (2 mm)
8 319-130/01 Stylus, Referencer Intact Compartment (10 mm)
9 319-135/01 Stylus, Referencer Intact Compartment (12 mm)
10 319-185/00 Cutting Block, Tibial Plateau, symmetrical, right and left
11 319-186/00 Cutting Block, Tibial Plateau, asymmetrical, left
12 319-187/00 Cutting Block, Tibial Plateau, asymmetrical, right
13 319-184/00 Stylus Guide
14 319-183/00 Foot Clamp, spring fixation
15 317-538/01 Plastic Connector *
* without illustration
67
Instruments
319-020/31 Case 2 – Femoral Preparation
1 2 3 4 5 11 12 13
17 16 15 6 7 8 9 10 14
1 319-620/31 Instrument Tray 2, empty, 485 x 253 x 100 mm (incl. lid)
2 319-210/00 IM Alignment Instrument, femoral varus/valgus adjustment
3 319-211/01 Guide U-Bolt
4 319-215/00 Cutting Block, distal femur cut „S“
5 319-220/00 Cutting Block, distal femur cut „L“
Cutting Blocks, 4-in-1-Femur Cutting Block
6 319-202/01 for size 1
7 319-202/16 for size 1B
8 319-202/02 for size 2
9 319-202/25 for size 2A
10 319-202/26 for size 2B
11 319-202/03 for size 3
12 319-202/04 for size 4
13 319-202/05 for size 5
14 319-505/00B* Step Drill
15 319-510/00 Guide Rod, intramedullary
16 319-200/02 Femoral Sizing/Rotation Alignment Instrument, femoral sizing and rotation adjustment
17 317-845/00 Femoral Sizer
* with B = Hudson-Fitting
68
Instruments
319-030/31 Case 3 – Spacer
1 2 3 4 5 6
7 8 9 10 11 12 13 14 15 16
1 319-630/31 Instrument Tray 3, empty, 485 x 253 x 100 mm (incl. lid)
Shims
2 319-330/12 12 mm
3 319-330/14 14 mm
4 319-330/16 16 mm
5 319-330/18 18 mm
6 319-300/00 T-Handle
7 319-335/01 Correction Spacer, universal, white
Spacers for flexion gap
8 319-322/01 Size 1, yellow
9 319-322/16 Sizen 1B, grey
10 319-322/02 Size 2+2A, green
11 319-322/26 Size 2B, red
12 319-322/03 Size 3, blue
13 319-322/04 Size 4, black
14 319-322/05 Size 5, brown
Spacers for extension gap
15 319-310/01 Size „S“, white
16 319-315/01 Size „L“, white
69
Instruments
319-040/31 Case 4 – Tibia Preparation
1 2 3 4 5 6 7 8 10 9 11 12
17 16 15 14 13 22 21 20 19 18
23 24 25 26 27 28 29 30 34 33 32 31
1 319-640/31 Instrument Tray 4, empty, 485 x 253 x 100 mm (incl. lid)
Trial Plateaus
2 319-415/01 Size 1
3 319-415/16 Size 1B
4 319-415/02 Size 2
5 319-415/25 Size 2A/2B
6 319-415/03 Size 3
7 319-415/04 Size 4
8 319-415/05 Size 5
9 319-430/01 Guide Sleeve
10 319-401/00 Locking Clip
11 319-570/00 Insertion Forceps
12 319-425/00 Tibial Blade Chisel
Tibial Preparation Plates, right/left
13 319-400/01 Size 1
14 319-400/02 Size 2
15 319-400/03 Size 3
16 319-400/04 Size 4
17 319-400/05 Size 5
Shims
18 319-407/10 10 mm
19 319-407/12 12 mm
20 319-407/14 14 mm
21 319-407/16 16 mm
22 319-407/18 18 mm
Trial Plateaus – Fixed Bearing
23 319-410/01 Size 1
24 319-410/16 Size 1B
25 319-410/02 Size 2
26 319-410/25 Size 2A/2B
27 319-410/03 Size 3
28 319-410/04 Size 4
29 319-410/05 Size 5
30 319-450/00 Impactor Handle for Tibial Component
31 319-437/03B* Tapered Reamer (Taper Cap)
32 319-437/05B* Tapered Reamer (L 50 mm)
33 319-437/08B* Tapered Reamer (L 80 mm)
34 319-437/12B* Tapered Reamer (L 120 mm)
* with B = Hudson-Fitting
70
Instruments
445-008/00 Case 5 – Femoral Trial Prostheses
1 2 3 4 5 6 7 8 9 10 11 12
19 18 17 16 15 14 13
1 445-080/00 Instrument Tray 8, empty, 485 x 253 x 100 mm (incl. lid)
2 317-668/25 Femoral Trial, left, size 2A
3 317-668/01 Femoral Trial, left, size 1
4 317-668/26 Femoral Trial, left, size 2B
5 317-668/16 Femoral Trial, left, size 1B
6 317-668/03 Femoral Trial, left, size 3
7 317-668/02 Femoral Trial, left, size 2
8 317-667/03 Femoral Trial, right, size 3
9 317-667/05 Femoral Trial, right, size 5
10 317-667/26 Femoral Trial, right, size 2B
11 317-667/04 Femoral Trial, right, size 4
12 317-667/25 Femoral Trial, right, size 2A
13 317-667/01 Femoral Trial, right, size 1
14 445-205/00 Drill Template, Ø 5.5 mm
15 317-667/16 Femoral Trial, right, size 1B
16 317-667/02 Femoral Trial, right, size 2
17 317-668/05 Femoral Trial, left, size 5
18 317-668/04 Femoral Trial, left, size 4
19 445-210/00 Femoral Inserter/Extractor
71
Instruments
319-060/32 Case 6 – Femoral PS Preparation
1 12 – 15 20 – 23 24+25 26 27 28
2–8 9 – 11 16 – 19 34 31 32 29 33 30
1 319-660/32 Instrument Tray 6, empty, 485 x 253 x 100 mm (incl. lid)
Trial Plateaus – Fixed Bearing PS
2 319-420/01 Size 1
3 319-420/16 Size 1B
4 319-420/02 Size 2
5 319-420/25 Size 2A/2B
6 319-420/03 Size 3
7 319-420/04 Size 4
8 319-420/05 Size 5
Femoral Trial Boxes
9 317-717/01B* Size 1
10 317-717/16B* Size1B
11 317-717/02B* Size 2
12 317-717/25B* Size 2A/2B
13 317-717/03B* Size 3
14 317-717/04B* Size 4
15 317-717/05B* Size 5
Femur-Probekästen
16 317-669/01 Size 1
17 317-669/16 Size 1B
18 317-669/02 Size 2
19 317-669/25 Size 2A
20 317-669/26 Size 2B
21 317-669/03 Size 3
22 317-669/04 Size 4
23 317-669/05 Size 5
24 317-668 Fixation Screws (4 pieces)
25 319-601/30 Sterilizing Box with base, silicon base and lid
Box Guides
26 317-618/01 Size 1
27 317-618/16 Size 1B
28 317-618/02 Size 2
29 317-618/25 Size 2A
30 317-618/26 Size 2B
31 317-618/03 Size 3
32 317-618/04 Size 4
33 317-618/05 Size 5
34 319-535/00 Screwdriver, hex 2.5 mm
* with B = Hudson-Fitting
72
Instruments
319-070/31 Case 7 – General Instruments
1 2 4 5 6 7 8 9
3 20 19 13–18 12 11 10
1 319-670/31 Instrument Tray 7, empty, 485 x 253 x 100 mm (incl. lid)
2 317-802/10 Impactor for Tibial Component, incl. PE Plateau, Gr. 1, 1B, 2
3 317-802/11 Impactor for Tibial Component, incl. PE Plateau, Gr. 2A, 2B, 3, 4
4 317-802/12 Impactor for Tibial Component, incl. PE Plateau, Gr. 5
5 317-802/53 Cutting Template
6 319-540/00 Screwdriver, hex. 2.0 mm
7 317-586 Inserter/Extraction Forceps for Wire Pins
8 319-260/00 Rasp
9 319-530/01 Screwdriver
10 317-649/08B* Twist Drill, Ø 5.5 mm
11 16-3287/00B* Adapter, for LINK power tool snap lock coupling
12 319-602/30 Sterilizing Box with base, silicon mat and lid
consisting of:
13 319-555/01 Head Pin, 4 pieces
14 319-566/00 Drill Pin, 85 mm, Ø 3.0/3.5 mm, with stop, 4 pieces
15 319-581/00 Drill Pin, 80 mm, Ø 3.0 mm, 4 pieces
16 319-582/00 Drill Pin, 110 mm, Ø 3.0 mm, 4 pieces
17 319-560/02 Thread Pin, 85 mm, Ø 3.0/3.5 mm, with stop, 4 pieces
18 319-212/00 Fixation Adapter, 2 pieces
19 319-500/00 T-Handle
20 317-646 Driver
* with B = Hudson-Fitting
73
Instruments
319-080/31 Case 8 – Final-Spacer
1 2 4 6 8 9–12
3 5 7
1 319-680/31 Instrument Tray 8, empty, 485 x 253 x 100 mm (incl. lid)
Final-Spacer
2 319-341/01 Size 1
3 319-341/16 Size 1B
4 319-341/02 Size 2
5 319-341/25 Size 2A/B
6 319-341/03 Size 3
7 319-341/04 Size 4
8 319-341/05 Size 5
Shim
9 319-330/12 12 mm
10 319-330/14 14 mm
11 319-330/16 16 mm
12 319-330/18 18 mm
74
Instruments
340-200/01 Patella Instrument Set
1 2 3 4 5 6 7 8
26
9-12
25
24
23 22 20 21 19 18 17 16 15 14 13
1 340-015/01 Instrument Tray, empty, 485 x 253 x 100 mm (incl. lid)
2 340-007 Patella Pusher Attachment
3 340-306B Patella Drill, Ø 6.3 mm, with Hudson fitting, for Patella Component 3-pegs
4 optional* Adapter for snap lock chuck, optional (see page 64)
Patella Drill Guides for patella holding clamp, Patella Component 3-pegs
5 340-225 Size 25 for 318-401/25
6 340-228 Size 28 for 318-401/28
7 340-231 Size 31 for 318-401/31
8 340-234 Size 34 for 318-401/34
Patella Reaming Guides for patella holding clamp, Patella Component 3-pegs
9 340-025 Size 25 for 318-401/25
10 340-028 Size 28 for 318-401/28
11 340-031 Size 31 for 318-401/31
12 340-034 Size 34 for 318-401/34
Patella Reamers for patella reaming guides, Patella Component 3-pegs, with Hudson fitting
13 340-134B Size 34 for 318-401/34
14 340-131B Size 31 for 318-401/31
15 340-128B Size 28 for 318-401/28
16 340-125B Size 25 for 318-401/25
Patella Trial Prostheses for Patella Component 3-pegs
17 340-334 Size 34 for 318-401/34
18 340-331 Size 31 for 318-401/31
19 340-328 Size 28 for 318-401/28
20 340-325 Size 25 for 318-401/25
21 340-006 Patella Resection Clamp
22 340-009 Patella Pusher Inserter
23 65-2000 Townley Femur Caliper
24 340-010 Patella Sizing Template
25 15-2042 Inserting Forceps for inserting the manipulating bearings, 215 mm
26 340-005 Patella Holding Clamp
* Adapter optional: 16-3283/00, 16-3284/00, 16-3285/00
75
Instruments
319-010/31 Precision Tibia Resection Alignment
1 2 3 4 5 15 17
7 6 10 9 8 12 11 14 13 16*
1 319-610/31 Instrument Tray 1, empty, 485 x 253 x 100 mm (incl. lid)
2 319-140/01 Tibial Base Guide
3 319-110/01 EM Alignment Rod
4 319-525/00 Alignment Rod, extramedullary (external thread)
5 319-520/01 Alignment Rod, extramedullary (female thread)
6 319-115/01 Fixation Jig, extramedullary
7 319-165/00 Fixation Jig, extramedullary
8 319-125/01 Stylus, Referencer Defect Compartment (2 mm)
9 319-130/01 Stylus, Referencer Intact Compartment (10 mm)
10 319-135/01 Stylus, Referencer Intact Compartment (12 mm)
11 319-145/01 Cutting Block (symmetrical), right and left
12 319-150/01 Cutting Block (asymmetrical), right
13 319-155/01 Cutting Block (asymmetrical), left
14 319-120/01 Guide
15 319-160/00 Foot Clamp, intramedullary, silicon belt connection
16 317-538/01 Plastic Connector *
17 319-183/00 Foot Clamp, spring fixation
* without illustration
76
Instruments
317-800/00 EXTRABONE Instrument Set
1 2 3
9 8 7 6 5 4
1 317-805/00 Instrument Tray, empty, 485 x 253 x 120 mm
2 317-840/00 Alignment Rod
3 317-815/00 Positioner Guide
4 317-825/00 Guide
5 317-820/00 Connector (3 parts)
6 317-810/00 Alignment Instrument
7 317-845/00 Femoral Sizer
8 317-835/00 Axis Alignment Guide
9 317-830/00 Positioner
77
Instruments
Additional Instruments
Hudson-Fitting (B)
Standard tool connection.
Adapter for power tool chuck
Different adapters are available to ensure compatibility
to allow various connections:
Attachment
16-3283/01 Jacobs-Fitting (E)
16-3284/00 AO-Fitting (D)
16-3285/00 Harris-Fitting (C)
A
Sawblade, 317-588/01
without offset teeth, 1.24 mm thick Gauge for ligament tension
Width (A) Width (A)
25 mm 13 mm Fitting
317-654/10 317-656/10 Synthes
317-654/11 317-656/11 Aesculap combi
317-654/12 317-656/12 3M
317-654/13 317-656/13 Zimmer/Hall combi
317-654/14 317-656/14 Stryker system 4
78
Accessories
X-ray Templates
317-684/01 317-683/01
X-ray Templates for GEMINI SL, X-ray Templates for GEMINI SL – PS version,
Set contains femoral and tibial implants Fixed Set contains femoral and tibial implants PS (Posterior
Bearing, 110% actual size, 1 set of 7 sheets Stabilized), 110% actual size, 1 set of 7 sheets
317-658/00
X-ray Template GEMINI SL EXTRABONE,
110% actual size,1 sheet
79
Further Information
PCL Protector GEMIN SL EXTRABONE
Instrument for the protection of the posterior cruciate Extramedullary Femur Referencing
ligament during resection of the tibia.
LINK PorEx (TiNbN = Titan-Niobium-Nitrid) Patella Components
Surface Modification for LINK implants for LINK Knee Prosthesis Systems
For more information please register for our LINK Media Library (linkorthopaedics.com)
80
Important Information
Please note the following regarding the use of our implants:
1. Choosing the right implant is very important.
The size and shape of the human bone determines the size and shape of the implant and also limits the load capacity.
Implants are not designed to withstand unlimited physical stress. Demands should not exceed normal functional loads.
2. Correct handling of the implant is very important.
Under no circumstances should the shape of a finished implant be altered, as this shortens its life span. Our implants
must not be combined with implants from other manufacturers. The instruments indicated in the Surgical Technique
must be used to ensure safe implantation of the components.
3. Implants must not be reused.
Implants are supplied sterile and are intended for single use only. Used implants must not be used again.
4. After-treatment is also very important.
The patient must be informed of the limitations of the implant. The load capacity of an implant cannot compare with
that of healthy bone!
5. Unless otherwise indicated, implants are supplied in sterile packaging.
Note the following conditions for storage of packaged implants:
• Avoid extreme or sudden changes in temperature.
• Sterile implants in their original, intact protective packaging may be stored in permanent buildings up until the
“Use by” date indicated on the packaging.
• They must not be exposed to frost, dampness or direct sunlight, or mechanical damage.
• Implants may be stored in their original packaging for up to 5 years after the date of manufacture. The “Use by” date
is indicated on the product label.
• Do not use an implant if the packaging is damaged.
6. Traceability is important.
Please use the documentation stickers provided to ensure traceability.
7. Further information on the material composition is available on request from the manufacturer.
Follow the instructions for use!
Waldemar Link GmbH & Co. KG, Hamburg
All content in this catalog, including text, pictures and data, is protected by law. Every instance of use, whether in part or in whole and which is not permitted by
law, is subject to our prior consent. In particular, this applies to the reproduction, editing, translation, publishing, saving, processing, or passing on of content
stored in databases or other electronic media and systems, in any manner or form. The information in the catalogs is solely intended to describe the products
and does not constitute a guarantee.
The Surgical Technique described has been written to the best of our knowledge and belief, but it does not relieve the surgeon of his/her responsibility to duly
consider the particularities of each individual case.
Products shown in this document may not be available in your country. The product availability is subject to the approval and/or registration regulations of the
respective country. Please contact Waldemar Link GmbH & Co. KG if you have questions about the availability of LINK products in your country.
Waldemar Link GmbH & Co. KG and/or other corporate affiliated entities own, use or have applied for the following trademarks in many jurisdictions:
LINK, BiMobile, SP II, Modell Lubinus, E-Dur, EndoDur, T.O.P. II, BetaCup, CombiCup PF, CombiCup SC, CombiCup R, MobileLink, C.F.P., LCU, SP-CL, LCP,
MIT-H, Endo-Model, Endo-Model SL, MP, MEGASYSTEM-C, GEMINI SL, SPAR-K, LCK, Link OptiStem, HX, TiCaP, X-LINKed, PorAg, LINK PorEx, BiPorEx,
PorEx-Z, TrabecuLink, Tilastan, customLINK, RescueSleeve, Stactip, VACUCAST.
Other trademarks and trade names may be used in this document to refer to either the entities claiming the marks and/or names or their products and
are the property of their respective owners.
© LINK • 737_GEMINI SL_OP.Impl.Instr_en_2021-03_001
Waldemar Link GmbH & Co. KG
Barkhausenweg 10 · 22339 Hamburg · Germany
Phone +49 40 53995-0 ·
[email protected]
www.linkorthopaedics.com
REVISION KNEE SYSTEM
Surgical Technique
Explanation of Pictograms
Manufacturer Article number
Caution: Federal law restricts this device to sale by or on the order
Material number Rx only
of a physician.
Contents
CCK
02 Overview – Workflow of Surgical Steps:
Surgical Technique:
04 Preoperative Planning
05 Approaches
06 Pins, Pin Instruments and Sawblades
08 Reamers
10 Trial Stems
12 Tibial Preparation:
12 Intramedullary Tibial Guide Assembly
13 Intramedullary Tibial Guide Alignment
16 Tibial Resection
17 Tibial Preparation
19 Tibial Augment Preparation
20 Tibial Male Taper Connection Preparation
21 Tibial Trial Component Assembly
24 Femoral Preparation:
24 Femoral Sizing:
25 Femoral Alignment Guide Assembly
26 Femoral Alignment Guide Positioning
28 Distal Femoral Resection
29 4-in-1 Femoral Cutting Block Assembly
31 4-in-1 Femoral Cutting Block Positioning
33 Femoral 4-in-1 Resection
34 Femoral Box Preparation
36 Femoral Male Taper Connection Preparation
37 Femoral Trial Assembly
40 Trial Reduction
42 Final Preparation
43 Extraction of Femoral Trial Component
44 Patella Preparation (Patella Resurfacing)
47 Final Implantation:
47 Augment Assembly, Modular Stem Assembly
50 Implantation
50 1 Fixed Bearing Modular Tibial Component Implantation
51 2 CCK Femoral Component Implantation
52 3 CCK PE Articulating Surface Implantation
54 4 Patella Implantation
54 Functional Test
55 Appendix 1, Implant Removal
58 Appendix 2, Extraction Stem Adapter and Trial Stems
59 Appendix 3, TrabecuLink Tibial Cones Compatibility
Implants:
60 Implants Combination Overview
62 LinkSymphoKnee Femoral Components
64 LinkSymphoKnee Tibial Components
65 LinkSymphoKnee Articulating Surfaces
67 LinkSymphoKnee Press-Fit Stems
68 LinkSymphoKnee Cemented Stems, Centralizers
69 LinkSymphoKnee Cementless Stems
71 LinkSymphoKnee Replacement Sets for Cone Adapter
72 LinkSymphoKnee Femoral Augments
75 LinkSymphoKnee Tibial Augments
76 LinkSymphoKnee Patella Components
Instruments:
77 LinkSymphoKnee Instrument Sets: Overview
101 Additional Instruments
102 Additional Information
103 Indications/Contraindications
Important Information
NOTE: To assist the reader of this Surgical Technique, the LinkSymphoKnee instruments used in the
illustrated surgical steps are shown again at the bottom of each page (from 08 to 58).
Take care to select the correct size of the instruments according to the individual surgical case.
01
Overview – Workflow of Surgical Steps
CCK
Preoperative Tibial Reaming, Tibial Preparation
Planning IM Tibial Alignment and Resection
(Approach)
CCK Femoral Box Preparation Trial Reduction
02
Overview – Workflow of Surgical Steps
Femoral Reaming and A/P Femoral
Distal Femoral Resection Resection and Chamfer Cuts
Final Tibial Preparation Final Implantation
03
Surgical Technique
Preoperative Planning
The anatomic landmarks in the knee joint are defined preoperatively by imaging
the whole leg on the healthy and the affected side in the standing position. The
angle between the anatomic axis (center of knee joint – intramedullary canal) and
the mechanical axis (center of femoral head – center of knee joint – center of
ankle to the second toe) determines the valgus angle (01).
These angles should be determined for both knees. The valgus angle of a
healthy knee joint is approximately 5°-7°. The appropriate implant size can be
estimated preoperatively with X-ray templates. The necessary resections are
determined by the size of the implant and the deformity corrections required.
ATTENTION: The LinkSymphoKnee CCK has an inbuilt 6° valgus angle in the
femoral stem.
01
01
04
Surgical Technique
Approaches
With the knee in slight flexion, a straight incision is made over the patella, as far as the tibial tuberosity (02).
A medial parapatellar incision is made through the patellar retinaculum, capsule and synovial membrane (03).
When making the parapatellar incision, the patella is pushed to one side to visualize the patellofemoral joint.
Removal of the hypertrophic synovial membrane and parts of the fat pad allow access to the medial, lateral
and intracondylar parts of the joint. Excess synovium should be removed in order to avoid postoperative im-
pingement and adhesions. Some surgeons prefer total synovectomy.
02 03
Component Removal
When removing the components to be revised, great care must be taken to preserve as much of the
remaining bone stock as possible and to avoid the risk of fracture of the residual bone.
Bone preservation can usually be achieved through the use of small flexible osteotomes, saws and
high-speed burring instruments.
05
Surgical Technique
Pins, Pin Instruments and Sawblades
The LinkSymphoKnee Instruments
are designed to be used with:
Drill Pins Headed Drill Pins Bone Nails
The Drill Pins, Headed Drill Pins and Bone Nails may be hammered in using the Universal Pin Inserter/
Extractor (A) or the Universal Pin Inserter (B). The Drill Pins, Headed Drill Pins and Bone Nails may be pulled
out using the Universal Pin Inserter/Extractor (A), the Drill Pins and Headed Drill Pins are also designed to be
drilled in and removed using the Power Driver (C) or the Power Driver with Snap Lock (D). The Bone Nail may
be pulled out using the Universal Pin Inserter/Extractor (A) or the Slaphammer (E).
B C D
A1
E
A
Use the Universal Pin Inserter/Extractor (A) as shown in the picture (A1).
OPTIONAL: ATTENTION: The Thread Pins* are designed
The LinkSymphoKnee Instruments are to be drilled in and removed using the Power
designed to be used also with Thread Pins*, Driver (C) or the Power Driver with Snap Lock
headed or not headed. (D) only.
* only available upon request
The LinkSymphoKnee Instruments are designed for use with sawblades up to a max. thickness of 1.27 mm
(page 101).
06
Surgical Technique
Compatibility
Drill Pins Headed Drill Bone Nails Thread Pins* Headed Thread
Pins Pins*
Universal Pin Inserter/Extractor
Universal Pin Inserter
Power Driver
Power Driver with Snap Lock
Slaphammer
* only available upon request
07
Surgical Technique
Reamers
Cylindrical Reamers:
The Cylindrical Reamers are available in diameters:
10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 22 mm.
Different markings on the Reamers show the depths
for different stems.
ATTENTION: There are different markings for femur
(F) and tibia (T).
ATTENTION: The Cylindrical Reamers can be used
only with the Cylindrical Cementless Stems,
Cylindrical Press-Fit Stems and Conical Cemented
Stems.
Conical Reamers:
The Conical Reamers are available in the following sizes:
• L
ength: 128 mm, with diameter: 12, 13, 14, 15,
16, 17, 18, 19, 20, 21, 22, 23, 24 mm
• L
ength: 158 mm, with diameter: 12, 13, 14, 15,
16, 17, 18, 19, 20, 21, 22, 23, 24 mm
• L
ength: 188 mm, with diameter: 12, 13, 14, 15,
16, 17, 18, 19, 20, 21, 22, 23, 24 mm
ATTENTION: The markings F and T show the depth
of the stem on the Femoral side and Tibial side
respectively.
ATTENTION: The Conical Reamers can be used
only with the Conical Cementless Stems.
ATTENTION: When using the Cementless Modular Stems, ream with increasing diameters until the Reamer
has achieved contact with the cortical bone over a continuous distance of approx. 50 mm. The Implant must
correspond with the last Reamer used in terms of diameter and length.
08
Surgical Technique
Ball Reamers:
The Ball Reamers are available in the diameters:
10, 12, 14, 16, 18, 20, 22, 24, 26 mm
ATTENTION: The Ball Reamers can be used only
with the Conical Cemented Stems.
ATTENTION: In the case of Conical Cemented Stems, the Reamer should be at least 2 mm larger than the
planned stem diameter.
09
Surgical Technique
Trial Stems
Cylindrical Trial Stems, cementless:
ATTENTION: Must be used for the cementless Cylindrical Stems and
Cylindrical Press-Fit Stems only.
Conical Trial Stems, cementless:
ATTENTION: Must be used for the cementless Conical Stems only.
Conical Trial Stems, cemented:
ATTENTION: Must be used for the cemented Conical Stems only.
10
Surgical Technique
When using Cemented Conical Stems, a Centralizer
must be used. The Centralizers are available in diameter:
12, 14, 16, 18, 20, 22, 24 mm.
Trial Centralizer
Every Trial Stem has a hole for inserting a screwdriver
(AF 2.5) to loosen the Trial Stem when the connection is
too tight (04-05).
Hole for the Screwdriver
04
05
ATTENTION: The Trial Stems for the 50 mm Cemented Conical Stems feature an in-built Trial Centralizer.
11
Surgical Technique
Tibial Preparation
Intramedullary Tibial Guide Assembly
Assemble the appropriate Tibial Cutting Block to the CCK IM Tibial Guide and lock it by twisting the knob of
the IM Tibial Guide clockwise until it clicks in the "lock position” (06).
CCK IM Tibial Guide
Tibial Cutting Block
Tibial Cutting Guide
Connector
06
LinkSymphoKnee CCK Instrument Set
includes the following Tibial Cutting Blocks:
left right
Instruments
881-059/00 151-202/00 151-204/00
CCK IM Tibial Guide 3° Tibial Cutting Tibial Cutting Block,
Guide Connector right
12
Surgical Technique
Intramedullary Tibial Guide Alignment
Open the tibial canal using the Step Drill (07).
ATTENTION: In case of primary CCK, the entry point
for opening the tibia can be marked with the
electrocautery, for example. It is usually located at
the point of attachment of the anterior cruciate
ligament.
Assemble the Reamer with the T-handle (08).
Slowly introduce the Reamer into the canal to pre-
vent building up pressure.
Use the Reamer until it achieves a good stability in
the tibial canal.
Slide the CCK IM Tibial Guide assembly onto the
Reamer (09).
ATTENTION: After reaming, it is possible to use
the Trial Stem Rod Intramedullary assembled to the
previously selected Trial Stem. This trial construct 07 08
can be used, instead of the Reamer, as a guide for
all other instruments (09a).
Insert the construct into the bone until reaching the
tibial marking (09b).
The Trial Stem Rod Intramedullary must be used
when it is planned to implant a stem shorter than or
equal to 95 mm.
09 09a 09b
Instruments
319-505-00B
Step Drill
881-059/00 151-202/00 151-204/00 15-6053/00 881-056/00 881-069/16 151-716/16
CCK IM Tibial Tibial Cutting Tibial Cutting T-handle Trial Stem Rod Reamer, cementless Trial Stem, cementless,
Guide 3° Guide Connector Block, right Intramedullary cylindrical, Ø 16 mm cylindrical, Ø 16 mm
13
Surgical Technique
OPTIONAL: To assess tibial alignment, insert the Alignment Rod into the IM-Tibial Guide. Rotation and
alignment may be checked by ensuring that the Alignment Rod remains parallel with the tibial axis (10).
10
Instruments
881-069/16 151-202/00 445-113/20 151-204/00 881-059/00
Reamer, cementless Tibial Cutting Alignment Rod, long Tibial Cutting Block, CCK IM Tibial Guide 3°
cylindrical, Ø 16 mm Guide Connector right
14
Surgical Technique
Insert the foot of the Adjustable Stylus into the slot of the Tibial Cutting Block and adjust it to the appropriate
level (11).
11
The scale on the body of the Adjustable Stylus indicates the quantity of bone and residual cartilage to be
resected.
ATTENTION: LinkSymphoKnee CCK Total Knee System has a minimum tibial component thickness of
10 mm (Tibial baseplate + PE-articulating surface). Take care to restore the joint line.
Set the Adjustable Stylus according to the patient's anatomy, to avoid excessive tibial resection. After the
tibial resection level has been set, pin the Tibial Cutting Block through the anterior parallel "0" holes, using two
Drill Pins (12).
12
Disconnect and remove the IM Tibial Guide and the reamer. The resection level may be adjusted by using the
distal or proximal pin holes, which move the block 2 mm more proximal or 2 mm more distal. If desired, the
Tibial Cutting Block can be fixed with an additional Headed Drill Pin placed through the distal angled hole.
Confirm the tibial resection level using Cutting Template introduced as a free sawblade into the Tibial Cutting
Block.
Instruments
445-124/65 445-125/35 881-059/00 151-202/00 151-204/00 881-069/16 445-111/00
Drill Pin Headed Drill CCK IM Tibial Tibial Cutting Tibial Cutting Reamer, cementless Adjustable Stylus
Pin Guide 3° Guide Connector Block, right cylindrical, Ø 16 mm
15
Surgical Technique
Tibial Resection
Resect the tibia using an oscillating saw and a sawblade max. 1.27 mm thick (13).
13 14
OPTIONAL: It is possible to add the Quick Connect Handle to the Tibial Cutting Block to check the
rotation again (14).
In case of bone loss, use the 5, 10 or 15 mm cutting slot to prepare the bone for Tibial Augments.
ATTENTION: When using only an augment on one side (medial or lateral), use the slot in the middle to per-
form the sagittal resection (15).
15
ATTENTION: When using an Offset Stem, prepare the bone for the Tibial Augment in a later stage of the
surgical procedure, after setting the offset position and magnitude.
Remove the Tibial Cutting Block and Pins.
Instruments
151-204/00 445-124/65 445-125/35 445-112/00 445-113/10
Tibial Cutting Block, Drill Pin Headed Drill Handle, quick connect Alignment Rod, short
right Pin
16
Surgical Technique
Tibial Preparation
Slowly reposition the Reamer into the tibial canal.
Place the Tibial Preparation Plate CCK onto the tibia
over the Reamer. Assess the tibial size to achieve
maximal coverage of the resected surface (16).
Slide the Offset Selector CCK (17) over the Reamer
and assemble it to the Tibial Preparation Plate CCK
(18). 16
Offset Selector CCK:
Offset position indicator Hole for Reamer
Offset magnitude options
0 mm, 3 mm, 6 mm
Turn it clockwise to change the
position of the offset.
Press and push to change the
17 offset magnitude
ATTENTION: There is only one way to assemble the Offset Selector onto the Tibial Preparation Plate CCK.
18
Instruments
881-099/00 881-285/60 881-069/16
Offset Selector CCK Tibial Preparation Plate CCK, Reamer, cementless
Size 6 cylindrical, Ø 16 mm
17
Surgical Technique
Set the appropriate offset magnitude and the position.
Check the rotational alignment by inserting the alignment rod into the handle of the Tibial Preparation Plate
CCK (19).
Fix the Tibial Preparation Plate CCK with 2 Drill Pins (20).
19
20
Record the magnitude and the position of the offset from the Offset Selector CCK.
Instruments
881-099/00 881-285/60 445-113/20 881-069/16 445-124/65
Offset Selector CCK Tibial Preparation Plate CCK, Alignment Rod, long Reamer, cementless Drill Pin
Size 6 cylindrical, Ø 16 mm
18
Surgical Technique
Tibial Augment Preparation
In case of bone loss, Tibial Trial Augments are available corresponding to each final tibial augment size and
side.
For preparing the tibial bone, slide the Tibial Cutting Block with the CCK Tibial Cutting Guide Connector onto
the CCK Tibial Preparation Plate. Remove the Reamer and CCK Offset Selector (21).
Use the medial or lateral (or both) cutting slot at 5, 10 or 15 mm to resect the tibia for the selected augment
(height 5, 10, 15 mm).
21
ATTENTION: When using only half augment (medial or lateral), use the slot in the middle to perform the
sagittal resection (22).
22 23
Assemble the appropriate Trial Augment under Tibial Preparation Plate (23).
Instruments
151-202/00 445-124/65 445-125/35 881-335/11 151-204/00 881-285/60
Tibial Cutting Drill Pin Headed Drill Tibial Trial Augment, Medial-Right/ Tibial Cutting Block, Tibial Preparation Plate CCK,
Guide Connector Pin Lateral-Left, Size 5-6, H = 5mm right Size 6
19
Surgical Technique
Tibial Male Taper Connection Preparation
Attach the Tibial Reamer Guide CCK onto the Tibial Preparation Template (24).
Use the Tapered Reamer CCK until it stops onto the
Tibial Reamer Guide CCK (25).
24
25
ATTENTION: When using an Offset Stem, assemble the Offset Reamer Stopper to the Tapered Reamer CCK
(26 & 27).
26
ATTENTION: This step of the surgical technique can
only be skipped when using straight stems with a
diameter bigger than 16 mm. 27
Instruments
881-285/60 881-065/09 881-065/00 881-067/17 445-124/65
Tibial Preparation Plate CCK, Reamer Stopper, offset Tibial Reamer Guide, CCK Tapered Reamer, CCK Drill Pin
Size 6
20
Surgical Technique
Tibial Trial Component Assembly
In case of bone loss, add the previously selected Tibial Trial Augment to the Tibial Trial Component CCK (28).
Fully assembled
Tibial Trial Augment
1 1
Neutral,
3 mm or 6 mm
Offset Adapter
2 2
28
Trial Stem assembly sequence: Tibial Trial Component CCK + (1) Adapter + (2) Trial Stem (28)
ATTENTION: When using an Offset Stem, select the 3 mm or the 6 mm Offset Adapter according to the
offset magnitude previously selected.
For the correct offset position refer to the setting on the CCK Offset Selector (29).
29
Instruments
881-099/00 881-258/60 151-806/06 151-501/00 151-716/16 881-335/11
Offset Selector CCK Tibial Trial Component, 6 mm Offset Adapter Neutral Stem Adapter Trial Stem, cementless, Tibial Trial Augment, Medial-Right/
CCK, Size 6 cylindrical, Ø 16 mm Lateral-Left, Size 5-6, H = 5mm
21
Surgical Technique
The Offset Adapters for the 3 mm and 6 mm offset feature the same markings as on the CCK Offset Selector
(30 & 31).
Marking on the Offset
Adapter for the offset
position.
Marking on the CCK
Offset Selector for the
offset position.
30 31
On the ventral/anterior side of the CCK Tibial Trial Component there is a marking that shows the correct
reference position for the Trial Offset Adapter (30 & 31a).
31a
Instruments
881-099/00 151-806/06 881-258/60
Offset Selector CCK 6 mm Offset Adapter Tibial Trial Component,
CCK, Size 6
22
Surgical Technique
Once the Trial Stem is assembled, the entire construct is locked and the offset position will not change.
ATTENTION: The length of the assembly (Trial Stem Adapter + Trial Stem) is equal to the length of the final
stem implant (32).
ATTENTION: In case of Cemented Stems, assemble a Trial Centralizer (33)
to the Trial Stem.
Open the handle of the Tibial Inserter and attach it to the Tibial Trial Component.
Use the Tibial Inserter to slowly introduce the Tibial Trial Component
into the tibial bone (34). 32
33
34
It is possible to secure the rotation of the Tibial Trial Component
using two Bone Nails (L = 25 mm or L = 35 mm) (35). 35
Instruments
881-258/60 445-128/35 151-806/06 151-106/16 881-042/00 151-716/16
Tibial Trial Component, Bone Nail 6 mm Offset Adapter Trial Centralizer, 16 mm Tibial Inserter Trial Stem, cementless,
CCK, Size 6 cylindrical, Ø 16 mm
23
Surgical Technique
Femoral Preparation
For femoral preparation, the knee is flexed to 90°. The entry point for opening the femur can be marked with
the electrocautery, for example. It is usually located approx. 3-5 mm medially above the intercondylar fossa.
The medullary canal is opened with the Step Drill (36).
36
Femoral Sizing
The CCK Femoral Sizer Template may be placed over the femur to assess the A/P and M/L size and position
(37).
37
Use the Reamer until it achieves a good stability into the femoral canal. Slowly introduce it into the canal to
prevent building up pressure.
Slide the handle of the CCK Femoral Sizer Template over the reamer and assess the M/L femoral size (38).
The vertical markings on the template handle represent the M/L width of the femoral component.
38 38a 38b
ATTENTION: After reaming, it is possible to use the Trial Stem Rod Intramedullary assembled to the
previously selected Trial Stem. This trial construct can be used, instead of the Reamer, as a guide for all other
instruments (38a).
Insert the construct into the bone until reaching the femoral marking (38b).
The Trial Stem Rod Intramedullary must be used when it is planned to implant stems lengths shorter than or
equal to 95 mm.
Instruments
881-069/16 881-056/00 151-716/16 881-105/60
Reamer, cementless Trial Stem Rod Trial Stem, cementless, Femoral Sizer Template CCK,
cylindrical, Ø 16 mm Intramedullary cylindrical, Ø 16 mm Size 6
24
Surgical Technique
Femoral Alignment Guide Assembly
In order to lock the CCK Distal Femoral Cutting Block (39), twist the knob of the Femoral Alignment Guide,
Varus/Valgus Adjustment 6° clockwise until it clicks in the "lock position” (40).
CCK Femoral Distal Cutting Block
The first cutting slot allows for a clean up
cut of 2 mm.
39
ATTENTION: Ensure that the correct Femoral
Alignment Guide, Varus/Valgus Adjustment 6° has
been selected. The "Right" one or the "Left" one is
selected according to the operated leg.
40
Instruments
881-159/00 881-050/06
Femoral Cutting Block, CCK, Femoral Alignment Guide,
Distal Cut Varus/Valgus Adjustment 6°, right
25
Surgical Technique
Femoral Alignment Guide Positioning
Slide the CCK Femoral Alignment Guide over the Reamer.
ATTENTION: To slide the Alignment Guide onto the Reamer, press the
spring on the side of the CCK Femoral Alignment Guide (41).
ATTENTION: Joint line/patella height corrections noted preoperatively
should be assessed for any proximal/distal (augment/resection) adjust-
ments.
41 Note any augment variations for use on the femoral CCK 4-in-1 Cutting
Block and Femoral Trial (42).
OPTIONAL: To confirm the valgus angle, attach the Quick Connect
Handle to the Distal Cutting Block and then insert the Alignment Rod into
the Quick Connect Handle. Extend the Alignment Rod assembly to the
center of the femoral head (43).
42
43
Instruments
Alignment Rods: 445-113/10 short, 445-113/20 long
881-069/16 317-802/53 881-050/06 881-159/00
Reamer, cementless Cutting Template Femoral Alignment Guide, CCK, 445-112/00 Femoral Cutting Block, CCK,
cylindrical, Ø 16 mm Varus/Valgus Adjustment 6°, right Handle, quick connect Distal Cut
26
Surgical Technique
The rotation can be checked with markings for the Whiteside line and the epicondylar line (44).
44
For a stable resection of the distal cut, the Reamer and the CCK Femoral Alignment Guide may be left in situ.
To secure the CCK Distal Femoral Cutting Block to the femur, two Drill Pins are inserted into the "0" pin holes
(45).
OPTIONAL: To fully secure the CCK
Femoral Cutting Block to the femur,
two Headed Drill Pins are inserted into
oblique holes.
45
Instruments
881-069/16 881-159/00 881-050/06 445-124/65 445-125/35
Reamer, cementless Femoral Cutting Block, CCK, Femoral Alignment Guide, CCK, Drill Pin Headed Drill
cylindrical, Ø 16 mm Distal Cut Varus/Valgus Adjustment 6°, right Pin
27
Surgical Technique
Distal Femoral Resection
The Cutting Template can be used to check the level of the distal resection. Use an oscillating saw
(max. 1.27 mm thick) through the cut slot to resect the distal femur (46).
46
Remove the Pins, the Cutting Block and the CCK Femoral Alignment Guide, leaving the Reamer in situ.
If it is necessary to re-cut the distal femur, reposition the CCK Femoral Cutting Block onto the +2 or +4 holes
for a +2 mm re-cut or a +4 mm re-cut respectively.
Instruments
881-069/16 881-159/00 881-050/06 445-124/65 445-125/35 317-802/53
Reamer, cementless Femoral Cutting Block, CCK, Femoral Alignment Guide, CCK, Drill Pin Headed Drill Cutting Template
cylindrical, Ø 16 mm Distal Cut Varus/Valgus Adjustment 6°, right Pin
28
Surgical Technique
4-in-1 Femoral Cutting Block Assembly
Select the CCK 4-in-1 Femoral Cutting Block that matches the femur A/P size previously determined (47).
47 48
ATTENTION: The LinkSymphoKnee Instrument Set features eleven CCK 4-in-1 Femoral Cutting Blocks,
one for each A/P femoral size. LinkSymphoKnee CCK 4-in-1 Femoral Cutting Block mimics precisely the
M/L size of the standard final femoral components (0 - 10).
The M/L size of the wide femoral components (3+, 4+, 5+) can be assessed later in the surgical technique.
ATTENTION: Joint line/patella height corrections noted preoperatively and at the Distal femoral resection step
should be accounted for in the Femoral Trial Augments selection.
Attach the appropriate Femoral Trial Augment to the posterior face of the CCK 4-in-1 Femoral Cutting Block
(48).
ATTENTION: In order to build a 15 mm Femoral Trial Augment, combine a 10 mm + a 5 mm Femoral Trial
Augment.
Now assemble the CCK Offset Selector (49) onto the CCK 4-in-1 Cutting Block.
Offset position indicator Hole for Reamer
Offset magnitude options
0 mm, 3 mm, 6 mm
Turn it clockwise to change the
position of the offset.
Press and push to change the
49 offset magnitude
Instruments
881-099/00 881-115/60 881-302/91
Offset Selector CCK Femoral Cutting Block, CCK, Femoral Augment Trial, 5 mm
4-in-1 Cut, Size 6
29
Surgical Technique
ATTENTION: The CCK Offset Selector can be used either for the left or the right leg. Ensure that the CCK
Offset Selector is assembled correctly according to the operated leg (50).
Slide the CCK 4-in-1 Cutting Block construct onto the Reamer (51).
50
51
Instruments
881-099/00 881-302/91 881-115/60 881-069/16
Offset Selector CCK Femoral Augment Trial, 5 mm Femoral Cutting Block, CCK, Reamer, cementless
4-in-1 Cut, Size 6 cylindrical, Ø 16 mm
30
Surgical Technique
4-in-1 Femoral Cutting Block Positioning
Assess the A/P and M/L position ensuring the rotation of the CCK 4-in-1 Cutting Block is aligned with the
epicondylar axis.
OPTIONAL: Insert the Cutting Template through the lateral side of the anterior slot of the CCK 4-in-1 Femoral
Cutting Block to verify the correct anterior resection before cutting the femur and to ensure that notching is
unlikely to occur (52).
52
If there is risk of unacceptable notching, it is possible to adjust the offset magnitude or offset position
for prevention.
OPTIONAL: The flexion gap may be checked by using the Flexion Spacer for the 4-in-1 Cut in combination
with the CCK 4-in-1 Femoral Cutting Block. The Flexion Spacer is placed between the CCK 4-in-1 Femoral
Cutting Block (with its stepped side pushed as far as possible under the unresected femoral condyles) and
the resected tibia (53). If needed Shims can be used.
53
ATTENTION: Use the Flexion Spacer for Micro-Sizes with femoral sizes 0, 1 and 2.
Instruments
881-069/16 881-115/60 881-011/02 881-099/00 317-802/53
Reamer, cementless Femoral Cutting Block, CCK, Flexion Spacer 4-in-1 Cut Offset Selector CCK Cutting Template
cylindrical, Ø 16 mm 4-in-1 Cut, Size 6
31
Surgical Technique
OPTIONAL: Use the CCK Flexion Spacer for the 4-in-1 Cut to check the flexion gap between the CCK 4-in-1
Femoral Cutting Block and the top surface of the Tibial Trial Component CCK (54).
54
ATTENTION: All Flexion Spacers are used only to
assess the flexion gap.
ATTENTION: All Flexion Spacers are designed to assess the flexion gap in combination with the CCK 4-in-1
Femoral Cutting Block.
The Flexion Spacer can be connected to a 4 mm or 8 mm Shim to allow evaluation of multiple
thicknesses (55a & 55b):
10 mm + 4 mm shim = 14 mm
12 mm + 4 mm shim = 16 mm
10 mm + 8 mm shim = 18 mm
12 mm + 8 mm shim = 20 mm
10 mm + 4 mm shim + 8 mm shim = 22 mm
12 mm + 4 mm shim + 8 mm shim = 24 mm
55a 55b
Record the magnitude and the position of the offset.
Instruments
881-069/16 881-115/60 881-011/02 881-099/00 881-258/60 881-013/00 881-019/04
Reamer, cementless Femoral Cutting Block, Flexion Spacer Offset Selector Tibial Trial Component, Flexion Spacer CCK Shim, Spacer,
cylindrical, Ø 16 mm CCK, 4-in-1 Cut, Size 6 4-in-1 Cut CCK CCK, Size 6 4-in-1 Cut, H = 10-12 mm H = 4 mm
32
Surgical Technique
Femoral 4-in-1 Resection
After final placement of the CCK 4-in-1 Femoral Cutting Block, insert two Headed Drill Pins into the oblique
pin holes on the medial and lateral aspects of the Cutting Block.
NOTE: when the CCK 4-in-1 Femoral Cutting Block is assembled to a Femoral Augment Trial construct of
15 mm, use the oblique pins located at the opposite side respect to the Femoral Augment Trial.
Protecting the collateral ligaments, use a max. 1.27 mm thick oscillating sawblade to complete anterior,
posterior, posterior chamfer and anterior chamfer resections as well as any 5 mm and 10 mm posterior
augment resections (56).
56
ATTENTION: Use a narrow sawblade with sizes 0, 1 and 2.
Instruments
881-069/16 881-115/60 881-099/00 445-125/35
Reamer, cementless Femoral Cutting Block, Offset Selector CCK Headed Drill
cylindrical, Ø 16 mm CCK, 4-in-1 Cut, Size 6 Pin
33
Surgical Technique
Femoral Box Preparation
Select the appropriate size of the CCK Box Guide that matches the femur A/P size previously determined.
ATTENTION: The CCK Box Guide is symmetrically designed to permit universal use.
The LinkSymphoKnee CCK instruments feature 5 CCK Box Guides (0-1-2, 3-4, 5-6, 7-8 and 9-10)
Assemble the CCK Box Guide to the anterior surface of the CCK 4-in-1 Cutting Block (57).
57
Instruments
445-125/35 881-115/60 881-302/91 881-069/16 881-099/00 881-114/56
Headed Drill Femoral Cutting Block, CCK, Femoral Augment Trial, Reamer, cementless Offset Selector Femoral Box Guide, CCK,
Pin 4-in-1 Cut, Size 6 5 mm cylindrical, Ø 16 mm CCK Size 5-6
34
Surgical Technique
Fix the CCK Box Guide to the bone with two Drill Pins, remove the CCK 4-in-1 Cutting Block, the Reamer
and the Offset Selector (58).
OPTIONAL
OR
58 59
OPTIONAL: The primary PS Box Guide may also be used for the CCK box preparation. The appropriate size
of the primary PS Box Guide can be positioned onto the 2 parallel pins of the CCK Box Guide onto the
anterior femur (59).
Using a narrow sawblade, resect for the M/L walls and the distal wall of the femoral box through the CCK Box
Guide (Anterior/Posterior) (59a).
Slot for the CCK box preparation.
59a 60
ATTENTION: When completing the notch cut, be careful to avoid excessive angulation of the sawblade or
penetration past the posterior femoral cortex to avoid injury to the neurovascular structures.
Avoid undercutting the condyles.
ATTENTION: When using the primary PS Box Guide, use the CCK cutting slot to resect the distal wall of the
femoral box (60).
ATTENTION: When using the primary PS Box Guide, attach the Femoral Trial Augment to increase the
stability of the construct in case of bone loss (60a & 60b).
60a 60b
Instruments
445-124/65 881-302/91 881-114/56 881-113/60
Drill Pin Femoral Augment Trial, Femoral Box Guide, CCK, Femoral PS Box Guide,
5 mm Size 5-6 Size 6
35
Surgical Technique
Femoral Male Taper Connection Preparation
Attach the CCK Femoral Box Reamer Guide onto the CCK Box Guide (61) or to the primary PS Box Guide
(61b).
61 61b
ATTENTION: The CCK Femoral Box Reamer can be used for a LEFT and for a RIGHT knee. Take care to
assemble it properly according to the operated leg (61 & 61b).
Use the Tapered Reamer CCK until it stops onto the CCK Femoral Box Reamer Guide (62).
62
ATTENTION: When using an Offset Stem, assemble the Offset Reamer Stopper to the Tapered Reamer CCK
(63 & 64).
63 64
ATTENTION: This step of the surgical technique can only be skipped when using non-offset stems with
diameter bigger than 16 mm.
Instruments
881-113/60 881-114/56 881-067/17 881-116/00 445-124/65 881-065/09
Femoral PS Box Guide, Femoral Box Guide, Tapered Reamer, CCK Reamer Guide, Drill Pin Reamer Stopper, offset
Size 6 CCK, Size 5-6 Femoral box, CCK
36
Surgical Technique
Femoral Trial Assembly
The LinkSymphoKnee CCK Instrument Set features a
specific femoral trial component for each CCK femoral size.
In case of bone loss, add the previously selected Femoral
Trial Augments to the CCK Femoral Trial Component (65).
(2)
Neutral, (1)
3 mm or 6 mm
Offset Adapter
65
Trial Stem assembly sequence: CCK Femoral Trial Component + (1) Adapter + (2) Trial Stem (65).
ATTENTION: When using an Offset Stem select the 3 mm or the 6 mm Offset Adapter according to the offset
magnitude previously selected.
For the correct position refer to the setting on the CCK Offset Selector (66).
66
Instruments
881-099/00 151-806/06 151-501/00 151-716/16 881-150/60 881-302/91
Offset Selector 6 mm Offset Adapter Neutral Stem Adapter Trial Stem, cementless, Femoral Trial CCK, Femoral Augment Trial,
CCK cylindrical, Ø 16 mm right, size 6 5 mm
37
Surgical Technique
The Trial Offset Adapters for the 3 mm and 6 mm offset have the same markings as the CCK Offset Selector
(67).
Marking on the Offset
Adapter for the offset
position.
Marking on the CCK
Offset Selector for the
offset position.
67
On the lateral side of the box of the CCK Femoral Trial Component there is a marking that shows the correct
reference position for the Trial Offset Adapter (68).
68
Instruments
881-150/60 151-806/06 881-099/00
Femoral Trial CCK, 6 mm Offset Adapter Offset Selector
right, size 6 CCK
38
Surgical Technique
Once the Trial Stem is assembled, the entire construct is locked and the offset position will not change.
ATTENTION: The length of the assembly (Trial Stem Adapter + Trial Stem) is equal to the length of the final
stem (69).
ATTENTION: In case of Cemented Stems, assemble
a Trial Centralizer (70) to the Trial Stem.
69 70
Use the Femoral Inserter/Extractor to slowly introduce the CCK Femoral Trial Component into the femoral
bone (71).
71
OPTIONAL: It is feasible to resect the bone for
femoral augments using an oscillating sawblade
through the CCK Femoral Trial Component (73).
Release the Femoral Inserter/Extractor and use the
Femoral Impactor Tip (assembled to the Impactor/
Extractor Handle) to complete the trial component
impaction (72).
72 73
It is possible to secure the position of the Trial
Femoral Component using pins.
Instruments
151-106/16 151-806/06 881-041/00 151-716/16 881-150/60 445-207/00 881-041/99
Trial Centralizer, 6 mm Offset Adapter Femoral Trial Stem, cementless, Femoral Trial CCK, Impactor/Extractor Femoral Impactor Tip
16 mm Inserter/Extractor cylindrical, Ø 16 mm right, size 6 Handle
39
Surgical Technique
Trial Reduction
Perform a trial reduction with all trial components – CCK Femoral Trial Component, Tibial Trial Component
CCK and CCK Trial Plateau (73).
74a
74
Select the appropriate size of the CCK Trial Plateau. The following table shows the possible size combinations:
CCK Femoral Component
0 1 2 3/3+ 4/4+ 5/5+ 6 7 8 9 10
1 CCK Articulating x x x x x x
CCK Articulating Surface 1-2
2 Surface 1-2up x x x x x x
Tibial Component
3 x CCK Articulating x x x x
CCK Articulating Surface 3-4
4 x Surface 3-4down x x x x
5 x x x x x
CCK Articulating Surface 5-6
6 x x x x x
7 x x x x x
CCK Articulating Surface 7-8
8 x x x x x
9 x x x x x x x
CCK Articulating Surface 9-10
10 x x x x x x x
Symbol Description: x = Combination not permitted
Select the appropriate thickness of the CCK Trial Plateau:
10 mm, 12 mm, 14 mm or 16 mm.
ATTENTION: For adjusting the height of the Trial Plateau beyond 16 mm, use the +4 mm Shim or the +8 mm
Shim (74) in the following combinations:
14 mm Trial Plateau + 4 mm Shim = 18 mm
12 mm Trial Plateau + 8 mm Shim = 20 mm
14 mm Trial Plateau + 8 mm Shim = 22 mm
16 mm Trial Plateau + 8 mm Shim = 24 mm
Instruments
881-250/85 151-716/16 881-258/60 881-255/10 881-150/60 151-806/06
Shim Trial Plateau, CCK Trial Stem, cementless, Tibial Trial Component, Trial Plateau CCK, Femoral Trial CCK, 6 mm Offset Adapter
H = +8 mm, size 5-6 cylindrical, Ø 16 mm CCK, Size 6 Size 5-6, H = 10 mm right, size 6
40
Surgical Technique
ATTENTION: Always use a stem with the Femoral and with the Tibial Component when implanting a
LinkSymphoKnee CCK Articulating Surface.
ATTENTION: For more implant options please refer to the LinkSymphoKnee CR-PS-PS+ Surgical
Technique.
OPTIONAL: It is possible to fix the CCK Trial Plateau to the CCK Tibial Trial Component using a Trial Plateau
Locking Screw. Select the Short Tibial Plateau Locking Screw when using a CCK Trial Plateau without Shim
or with the +4 mm Shim (75), whereas use the Long Trial Plateau Locking Screw when using a CCK Trial
Plateau with +8 mm Shim (76).
Use the Torque Wrench (hex 2.5 mm) to lock the Trial Plateau Locking Screw in place (77).
75 76
77
Instruments
15-2545 Torque Wrench, hex 2.5 mm
Locking Screws Tibial Plateau:
881-052/00 Short, 881-053/00 Long
881-250/85 151-716/16 881-258/60 881-255/10
Shim Trial Plateau, CCK Trial Stem, cementless, Tibial Trial Component, Trial Plateau CCK, 151-806/06
H = +8 mm, size 5-6 cylindrical, Ø 16 mm CCK, Size 6 Size 5-6, H = 10 mm 6 mm Offset Adapter
41
Surgical Technique
Final Preparation
Attach the size-specific CCK Tibial Keel Punch to the Impactor/Extractor Handle.
Choose the appropriate CCK Tibial Keel Punch according to the size of the CCK Tibial Trial Component.
ATTENTION: The LinkSymphoKnee Instrument Set features five CCK Tibial Keel Punches, each for two
tibial sizes: 1-2, 3-4, 5-6, 7-8 and 9-10.
After confirming the rotation of the CCK Tibial Trial Component, carefully impact the CCK Tibial Keel Punch
until it is fully seated onto the CCK Tibial Trial Component (78).
Align Impact Unlock Remove
78
OPTIONAL: It is possible to perform a final trial
reduction (79).
Make sure that no bony structures (e.g. osteophytes)
or local soft tissue interfere with the range of motion.
Remove the Tibial Trial Component assembly.
79
Instruments
445-207/00 881-258/60 881-255/10 881-275/56 151-806/06 151-716/16 881-150/60
Impactor/Extractor Tibial Trial Component, Trial Plateau CCK, Tibial Keel Punch CCK 6 mm Offset Adapter Trial Stem, cementless, Femoral Trial CCK,
Handle CCK, Size 6 Size 5-6, H = 10 mm Size 5-6 cylindrical, Ø 16 mm right, size 6
42
Surgical Technique
Extraction of Femoral Trial Component
OPTIONAL: In order to extract the Femoral Trial Components, it is
possible to use the Slaphammer in combination with the Femoral
Inserter/Extractor (80).
80
Assemble the Femoral Inserter/Extractor on the
CCK Femoral Trial Component (81).
81
Use the Slaphammer to remove the Femoral Trial
assembly (82).
82
Instruments
445-206/00 881-041/00 881-150/60 151-806/06 151-716/16
Slaphammer Femoral Femoral Trial CCK, right, 6 mm Offset Adapter Trial Stem, cementless,
Inserter/Extractor size 6 cylindrical, Ø 16 mm
43
Surgical Technique
Patella Preparation (Patella Resurfacing)
The following instructions describe how to use the LinkSymphoKnee Patella Component and they assume
the use of the instrument set available for this procedure.
Sizing
A Patella Sizing Template corresponding to the implants is available (83). The size of implant is determined by
placing the Sizing Template slightly medial and superior on the patella surface (84).
83 84
ATTENTION: The LinkSymphoKnee system includes 6 sizes of Patella Components.
Each size features a different implant thickness and diameter. In the table below, all Patella Component
dimensions are shown.
LinkSymphoKnee Patella Components – 3-peg
cemented
H
Ø
Ø Height (H)
UHMWPE E-Dur * mm mm
880-501/25 880-511/25 25 6
880-501/28 880-511/28 28 6
880-501/31 880-511/31 31 7
880-501/34 880-511/34 34 8
880-501/37 880-511/37 37 9
880-501/40 880-511/40 40 10
* E-Dur = Highly crosslinked UHMWPE (X-Linked PE) with Vit-E
Instruments
881-509/00
Patella Sizing Template
44
Surgical Technique
Patella Resection
Insert the Resection Guide (A) into the second groove
A from the top of the Patella Clamp Handle (B) (85).
Now, insert the Depth Gauge (C) into the first groove
from the top of the Patella Clamp Handle (B) (86).
B
C
85
86
Set the appropriate resection level corresponding
to the size of the patella component previously
determined (87).
87
Position the Patella Clamp so that the sectional plane lies parallel to the extended patellar tendon.
The Depth Gauge must lie on the bone. While pressing the hand grips, the Patella Clamp firmly fixes the
patella by means of the integrated ratchet (88). If necessary, release the ratchet by pressing the trigger
on the Patella Clamp Handle.
88 89
ATTENTION: It is important to ensure that the remaining patella is sufficiently thick (min. 12 mm).
The resection is carried out using an oscillating saw with a max.1.27 mm thick sawblade. The saw is guided
using the saw slots of the Resection Guide (89).
Instruments
445-902/00 445-903/00 881-500/00
Patella Clamp, Patella Clamp, Patella Clamp,
Handle Resection Guide Depth Gauge
45
Surgical Technique
Drilling the Anchoring Holes
Insert the Drill Guide (E) for the anchoring holes into the first slot from the top of the Patella Clamp Handle (B) (90).
E
B
90
Position the Patella Clamp onto the previously resected patella surface and use the Universal Drill (H) for the
anchoring holes. To prepare the seat for the anchoring holes, push the Universal Drill until stop (91). A Patella
Trial can now be used (91a).
H
91a
91
Instruments
445-902/00 881-012/00 881-511/31 881-501/31
Patella Clamp, Universal Drill, Patella Clamp, Patella Trial,
Handle Hudson Fitting (B) Drill Guide size 31 Ø 31 mm
46
Surgical Technique
Final Implantation
Augment Assembly
Select the appropriate sized Femoral/Tibial Components and the matching sized Femoral/Tibial Augment.
Using the Tibial and Femoral Trials as a guide, assemble the Femoral and/or Tibial Augment with the Torque
Wrench (hex 2.5 mm) (93 & 94).
NOTE: Screws are preassembled to the augments.
93 94
Instruments
15-2545
Torque Wrench, hex 2.5 mm
47
Surgical Technique
Modular Stem Assembly
ATTENTION: Carefully remove the Safety Cap from the taper. The Safety Cap protects the Cone Adapter
during transportation (95).
1. 2.
ATTENTION: The Cone Adapter is pre-assembled to
the Femoral and Tibial implant.
ATTENTION: If the Cone Adapter is not assembled
or is loose, use the Screwdriver for the Cone Adapt-
er to tighten it to the Femoral and Tibial Component
(96 & 97).
1. 2.
95
96
97
Select the final femoral and tibial stem, referencing to
the ones used for the Trial Assembly.
If an Offset Stem is used, insert the male end of the
Femoral or Tibial component into the female Offset
Stem taper. Referencing the Trial Assembly, align the
orientation lines to the appropriate setting (98 & 99).
98 99
Instruments
445-207/00 881-040/99 151-131/00
Impactor/Extractor Tibial Femoral Coupling Base Tip Screwdriver for Cone Adapter
Handle
48
Surgical Technique
ATTENTION: To avoid a metal on metal contact between the Stem and the Hammer, use the Tibial/Femoral
Coupling Base Tip.
Connect the Tibial/Femoral Coupling Base Tip with the Impactor/Extractor Handle (100).
Slide the Tibial/Femoral Coupling Base Tip onto the top of the Stem.
Two Hammer blows are enough (101 & 102).
100
101 102
Secure the Security Screw for the Cone Adapter with the Torque Wrench (hex 2.5 mm) (103 & 104).
ATTENTION: The Security Screw is included in the package of the CCK Femoral Component and of the
Modular Tibial Component.
103 104
ATTENTION: If a Cemented Stem is used, assemble a Centralizer to the final stem. The Centralizer ensures
a central positioning of the stem in the femoral/tibial canal and it avoids stress peaks in the bone in case of
bending load.
Instruments
445-207/00 881-040/99 15-2545
Impactor/Extractor Tibial Femoral Coupling Base Tip Torque Wrench, hex 2.5 mm
Handle
49
Surgical Technique
Implantation
The implantation sequence:
1 Fixed Bearing Modular Tibial Component Implantation
2 CCK Femoral Component Implantation
3 CCK PE Articulating Surface Implantation
4 Patella Implantation
1 Fixed Bearing Tibial Component Implantation
Prepare the sclerotic bone to ensure a continuous cement mantle with good cement interdigitation of 2 mm -
4 mm. This can be done by drilling holes and cleansing the bone with Pulsatile Lavage.
The bone cement is prepared following the specific manufacturer's instructions. Apply a layer of bone cement
to the underside of the Tibial Component, to the bone, or both.
Attach the Tibial Inserter to the fixed bearing tibial implant (105).
Carefully insert the Fixed Bearing Tibial Component, avoiding malrotation and impact into place (105).
If needed, the Tibial Impactor Tip, assembled with the Impactor/Extractor Handle (106), can be used to com-
plete the seating of the component (107).
Complete implantation with several hammer blows to the top of Tibial Impactor assembly.
Then use a Curette to remove all extruded cement.
ATTENTION: Ensure that excess bone cement is completely removed and no loose bone cement particles
remain, especially in the posterior part of the joint.
OPTIONAL:
Attachment for the
Quick Connect Handle
105 106 107
Instruments
881-042/00 445-207/00 881-042/99
Tibial Inserter Impactor/Extractor Tibial Impactor Tip
Handle
50
Surgical Technique
2 CCK Femoral Component Implantation
The bone cement is prepared following the specific manufacturer's instructions. Apply a layer of bone cement
to the backside of the Femoral Component, to the bone, or both. Place the selected Femoral Component
onto the bone by hand or, if preferred, use the Femoral Inserter/Extractor. Deliver several hammer blows to
the Femoral Inserter/Extractor (108).
Release the Femoral Inserter/Extractor and use the Femoral Impactor Tip (assembled to the Impactor/Ex-
tractor Handle (109) to complete the component impaction (110). Then use a Curette to remove all extruded
cement.
ATTENTION: Ensure that excess bone cement is completely removed and no loose bone cement particles
remain, especially in the posterior part of the joint.
108 109
110
Instruments
445-207/00 881-041/00 881-041/99
Impactor/Extractor Femoral Femoral Impactor Tip
Handle Inserter/Extractor
51
Surgical Technique
3 CCK PE Articulating Surface Implantation
Select the appropriate size of the CCK PE Articulating Surface.
The following table shows the possible LinkSymphoKnee size combinations:
CCK Femoral Component
0 1 2 3/3+ 4/4+ 5/5+ 6 7 8 9 10
1 CCK Articulating x x x x x x
CCK Articulating Surface 1-2
2 Surface 1-2up x x x x x x
Tibial Component
3 x CCK Articulating x x x x
CCK Articulating Surface 3-4
4 x Surface 3-4down x x x x
5 x x x x x
CCK Articulating Surface 5-6
6 x x x x x
7 x x x x x
CCK Articulating Surface 7-8
8 x x x x x
9 x x x x x x x
CCK Articulating Surface 9-10
10 x x x x x x x
Symbol Description: x = Combination not permitted
52
Surgical Technique
Select the appropriate thickness of the Fixed Bearing PE Articulating Surface.
The locking mechanism must sit in its resting position before impact. Therefor the following sequences must
be followed.
Please take particular note that all debris and especially bone cement particles are carefully removed from
the surface of the Tibial Component.
First the Fixed Bearing PE Articulating Surface is pushed as far posteriorly as it will go before impaction (111a-c).
The PE posterior lip must rest beneath the posterior dovetail of the Tibial Component.
Then position the Articulating Surface Impactor between 30 and 60 degrees on the insert so that the notch
rests on the anterior edge of the center of the insert (111d). Use a mallet to strike the Articulating Surface Impactor.
Haptic feedback!
The insert is ready for
Push the plateau in the tray (90a, 90b) the impact process (111c)
30°- 60°
111a 111 b 111 c
Confirm seating by circumferential
inspection (111e).
111d
111 e CAUTION:
For appropriate insertion of the PE Articulating Surface for LinkSymphoKnee
configuration PS, PS+ and CCK, the following aspects are to be considered:
Preventing the post-cam mechanism from pushing the PE Articulating Surface
out of position. While pushing the insert to its resting position it only can be pushed
until reaching the post (111f). If so the tibia has to be moved anteriorly. It can be
beneficial to move the knee into extension to help releasing the post cam mechanism. 111f
Wrong Right
The rotation and the M/L position of the tibia
has to be aligned to the femoral component.
It is important that the post is proper aligned
with the femoral box by internally rotating the
tibia (111g). 111 g
Instruments
881-040/01
Articulating Surface Impactor
53
Surgical Technique
4 Patella Implantation
Insert the Clamp Arm (F) into the first slot from the top of the Patella Clamp Handle (B) (112).
F
B
112
The bone cement is prepared following the specific manufacturer's instructions.
Following extensive rinsing and removal of all impeding soft tissue, the bone cement is applied to the
back of the implant, and the implant is placed by hand and pressed on using the Patella Clamp Handle
with the Clamp Arm (113).
ATTENTION: Ensure that excess bone cement is completely removed and no loose bone cement particles
remain in the joint.
113
Functional Test
Perform a final functional test through an entire
range of motion to check that all components are
properly positioned, and also to check for proper
ligament tension and patella tracking (114).
114
Instruments
445-902/00 445-904/00
Patella Clamp, Patella Clamp, Arm
Handle
54
Surgical Technique
Appendix 1, Implant Removal
Articulating Surface removal
Use the Articulating Surface Extractor to disengage the PE Articulating Surface from the Tibial Component
(115).
115
Femoral Component removal
Assemble the Slaphammer to the Femoral Inserter/Extractor and use it to extract the Femoral Component
(116).
ATTENTION: Disrupt cement along the bone
interface with flexible osteotomes or thin sawblades
with an oscillating saw.
NOTE: In the case of a porous component, disrupt
the bone/implant interface by using a Gigli saw, the
thin blade of a reciprocating saw or flexible osteotomes.
116
Instruments
445-206/00 881-019/00 881-041/00
Slaphammer Articulating Surface Extractor Femoral
Inserter/Extractor
55
Surgical Technique
Tibial Component removal
Assemble the Slaphammer to the Tibial Extractor Tip and use it to extract the Tibial Component (117). Make
sure that the Salphammer is fully threaded into the Tibial Extractor Tip, before using the retrograded impactor.
117
Stem removal
If the Stem remains in situ, remove all the cement from the noses of the Stem and from the top of the female
taper (118)
118
Instruments
445-206/00 881-043/00
Slaphammer Tibial Extractor Tip
56
Surgical Technique
Assemble the male taper of the Stem Extractor into the female taper of the stem.
ATTENTION: Clean the female taper before inserting the Stem Extractor.
ATTENTION: The peg of the Stem Extractor (120) must be on the same side as the marking on the top of the
Stem female taper (119).
Marking on top of the Stem
female taper Peg of the Stem Extractor
119
120
Assemble the Slaphammer onto the Stem Extractor and use it to remove the Stem (121).
121
Instruments
445-206/00 151-144/00
Slaphammer Extractor for Modular
Stems/Offset Stems
57
Surgical Technique
Appendix 2, Extraction Stem Adapter and Trial Stems:
OPTIONAL: If the Trial Stem Adapter and the Trial Stem remain inside the bone during the extraction of the
trial components (femoral and/or tibial), it is possible to use the Slaphammer to extract them. Attach the
Modular Extractor Tip to the Slaphammer (122).
Modular Extractor Tip
122
Connect the Modular Extractor Tip with the Stem Adapter (123).
123
Use the Slaphammer to axially remove the Stem Adapter and the Trial Stem.
Instruments
151-132/00 151-716/16 445-206/00 151-501/00
Modular Extractor Tip Trial Stem, cementless, Slaphammer Neutral Stem Adapter
cylindrical, Ø 16 mm
58
Surgical Technique
Appendix 3, TrabecuLink Cones Compatibility
The LinkSymphoKnee CCK Femoral Components and Modular Tibial Components are compatible with the
TrabecuLink Femoral Cones and Tibial Cones, respectively (124).
124
Select the appropriate size of the TrabecuLink Femoral Cone according to the following table:
Modular Stems
LinkSymphoKnee CCK Femoral Component
(no offset only)
(used with centralizer
(with & without slot)
Conical Cemented
max. Ø 18 mm)
cementless
Cylindrical
conical
Femoral
0 1 2 3 3+ 4 4+ 5 5+ 6 7 8 9 10
Cones
variant
(all sizes)
2-Zones & max. max.
2-Zones Ø 18 Ø 17
Proximal mm mm
3-Zones not compatible
Select the appropriate size of the TrabecuLink Tibial Cone according to the following table:
Tibial Components
Tibial Cones 1 2 3 4 5 6 7 8 9 10
all versions
XS x x x x x x x x x x
S x x x x x x x x
M x x x x x x x x
L x x x x x x x x
For all TrabecuLink Cones Implants and Instruments information, refer to the TrabecuLink Cones Surgical Technique.
59
Implants
Implants Combination Overview
Patella
Additional Implants
CR
Femoral Components
CR
Articulating Surfaces
Not modular
Tibial Components
Additional Implants
In the following pages are listed and described all implants and instruments for the CCK configurations.
For all other implants and instruments, refer to the LinkSymphoKnee CR, PS, PS+ Surgical Technique.
60
Implants
TrabecuLink Femoral Cones Modular Stems and Femoral Augments
PS CCK
PS PS+ CCK
PS All-Poly Modular
TrabecuLink Tibial Cones Modular Stems and Tibial Augments
61
Implants
LinkSymphoKnee Femoral Components – CCK Micro-Sizes
cemented
** **
AP
ML
AP ML
CoCrMo CoCrMo Size Side
mm mm
LINK PorEx*
880-030/00 880-080/00 0 right 47 53.5
880-030/10 880-080/10 1 right 50 56
880-030/20 880-080/20 2 right 53 58.5
880-031/00 880-081/00 0 left 47 53.5
880-031/10 880-081/10 1 left 50 56
880-031/20 880-081/20 2 left 53 58.5
LinkSymphoKnee Femoral Components – CCK
cemented
** **
AP
ML
AP ML
CoCrMo CoCrMo Size Side
mm mm
LINK PorEx*
880-030/30 880-080/30 3 right 56 61
880-030/40 880-080/40 4 right 59 63.5
880-030/50 880-080/50 5 right 62 66
880-030/60 880-080/60 6 right 65 69
880-030/70 880-080/70 7 right 68 72
880-030/80 880-080/80 8 right 71 75
880-031/30 880-081/30 3 left 56 61
880-031/40 880-081/40 4 left 59 63.5
880-031/50 880-081/50 5 left 62 66
880-031/60 880-081/60 6 left 65 69
880-031/70 880-081/70 7 left 68 72
880-031/80 880-081/80 8 left 71 75
* LINK PorEx: TiNbN = Titanium Niobium Nitride (gold color).
** Cone Adapter made of Tilastan = Ti6Al4V
62
Implants
LinkSymphoKnee Femoral Components – CCK Macro-Sizes
cemented
** **
AP
ML
AP ML
CoCrMo CoCrMo Size Side
mm mm
LINK PorEx*
880-030/90 880-080/90 9 right 74 78
880-030/X0 880-080/X0 10 right 77 81
880-031/90 880-081/90 9 left 74 78
880-031/X0 880-081/X0 10 left 77 81
LinkSymphoKnee Femoral Components – CCK Wide-Sizes
cemented
** **
AP
ML
AP ML
CoCrMo CoCrMo Size Side
mm mm
LINK PorEx*
880-030/35 880-080/35 3+ right 56 63.5
880-030/45 880-080/45 4+ right 59 66
880-030/55 880-080/55 5+ right 62 69
880-031/35 880-081/35 3+ left 56 63.5
880-031/45 880-081/45 4+ left 59 66
880-031/55 880-081/55 5+ left 62 69
* LINK PorEx: TiNbN = Titanium Niobium Nitride (gold color).
** Cone Adapter made of Tilastan = Ti6Al4V
63
Implants
LinkSymphoKnee Tibial Components – Modular Micro-Sizes
cemented
AP
** ** ML
AP ML
CoCrMo CoCrMo Size
mm mm
LINK PorEx*
880-050/10 880-100/10 1 37.5 59
880-050/20 880-100/20 2 40 62.5
LinkSymphoKnee Tibial Components – Modular
cemented
AP
** ** ML
AP ML
CoCrMo CoCrMo Size
mm mm
LINK PorEx*
880-050/30 880-100/30 3 42.5 66
880-050/40 880-100/40 4 45 69.5
880-050/50 880-100/50 5 47.5 73
880-050/60 880-100/60 6 50 76.5
880-050/70 880-100/70 7 52.5 80
880-050/80 880-100/80 8 55 83.5
LinkSymphoKnee Tibial Components – Modular Macro-Sizes
cemented
AP
** ** ML
AP ML
CoCrMo CoCrMo Size
mm mm
LINK PorEx*
880-050/90 880-100/90 9 57.5 87
880-050/X0 880-100/X0 10 60 90.5
* LINK PorEx: TiNbN = Titanium Niobium Nitride (gold color).
** Cone Adapter made of Tilastan = Ti6Al4V
64
Implants
LinkSymphoKnee Articulating Surfaces – CCK Micro-Sizes
H
AP
ML
AP ML Height (H)
Size
UHMWPE E-Dur * mm mm mm
880-231/10 880-271/10 1-2 37.5 59 10
880-231/12 880-271/12 1-2 37.5 59 12
880-231/14 880-271/14 1-2 37.5 59 14
880-231/16 880-271/16 1-2 37.5 59 16
880-231/18 880-271/18 1-2 37.5 59 18
880-231/20 880-271/20 1-2 37.5 59 20
880-231/22 880-271/22 1-2 37.5 59 22
880-231/24 880-271/24 1-2 37.5 59 24
880-232/10 880-272/10 1-2up 37.5 59 10
880-232/12 880-272/12 1-2up 37.5 59 12
880-232/14 880-272/14 1-2up 37.5 59 14
880-232/16 880-272/16 1-2up 37.5 59 16
880-232/18 880-272/18 1-2up 37.5 59 18
880-232/20 880-272/20 1-2up 37.5 59 20
880-232/22 880-272/22 1-2up 37.5 59 22
880-232/24 880-272/24 1-2up 37.5 59 24
* E-Dur = Highly crosslinked UHMWPE (X-Linked PE) with Vit-E
65
Implants
LinkSymphoKnee Articulating Surfaces – CCK
H
AP
ML
AP ML Height (H)
Size
UHMWPE E-Dur * mm mm mm
880-233/10 880-273/10 3-4 42.5 66 10
880-233/12 880-273/12 3-4 42.5 66 12
880-233/14 880-273/14 3-4 42.5 66 14
880-233/16 880-273/16 3-4 42.5 66 16
880-233/18 880-273/18 3-4 42.5 66 18
880-233/20 880-273/20 3-4 42.5 66 20
880-233/22 880-273/22 3-4 42.5 66 22
880-233/24 880-273/24 3-4 42.5 66 24
880-234/10 880-274/10 3-4down 42.5 66 10
880-234/12 880-274/12 3-4down 42.5 66 12
880-234/14 880-274/14 3-4down 42.5 66 14
880-234/16 880-274/16 3-4down 42.5 66 16
880-234/18 880-274/18 3-4down 42.5 66 18
880-234/20 880-274/20 3-4down 42.5 66 20
880-234/22 880-274/22 3-4down 42.5 66 22
880-234/24 880-274/24 3-4down 42.5 66 24
880-235/10 880-275/10 5-6 47.5 73 10
880-235/12 880-275/12 5-6 47.5 73 12
880-235/14 880-275/14 5-6 47.5 73 14
880-235/16 880-275/16 5-6 47.5 73 16
880-235/18 880-275/18 5-6 47.5 73 18
880-235/20 880-275/20 5-6 47.5 73 20
880-235/22 880-275/22 5-6 47.5 73 22
880-235/24 880-275/24 5-6 47.5 73 24
880-237/10 880-277/10 7-8 52.5 80 10
880-237/12 880-277/12 7-8 52.5 80 12
880-237/14 880-277/14 7-8 52.5 80 14
880-237/16 880-277/16 7-8 52.5 80 16
880-237/18 880-277/18 7-8 52.5 80 18
880-237/20 880-277/20 7-8 52.5 80 20
880-237/22 880-277/22 7-8 52.5 80 22
880-237/24 880-277/24 7-8 52.5 80 24
LinkSymphoKnee Articulating Surfaces – CCK Macro-Sizes
H
AP
ML
AP ML Height (H)
Size
UHMWPE E-Dur * mm mm mm
880-239/10 880-279/10 9-10 57.5 87 10
880-239/12 880-279/12 9-10 57.5 87 12
880-239/14 880-279/14 9-10 57.5 87 14
880-239/16 880-279/16 9-10 57.5 87 16
880-239/18 880-279/18 9-10 57.5 87 18
880-239/20 880-279/20 9-10 57.5 87 20
880-239/22 880-279/22 9-10 57.5 87 22
880-239/24 880-279/24 9-10 57.5 87 24
* E-Dur = Highly crosslinked UHMWPE (X-Linked PE) with Vit-E
66
Implants
LinkSymphoKnee – Cylindrical Press-Fit Stems
D1
L LF LT
D1 L LF LT
Straight 3 mm Offset 6 mm Offset mm mm mm mm
Tilastan* Tilastan* Tilastan*
880-601/10 880-611/10 880-621/10 10 80 115 98
880-601/11 880-611/11 880-621/11 11 80 115 98
880-601/12 880-611/12 880-621/12 12 80 115 98
880-601/13 880-611/13 880-621/13 13 80 115 98
880-601/14 880-611/14 880-621/14 14 80 115 98
880-601/15 880-611/15 880-621/15 15 80 115 98
880-601/16 880-611/16 880-621/16 16 80 115 98
880-601/17 880-611/17 880-621/17 17 80 115 98
880-601/18 880-611/18 880-621/18 18 80 115 98
880-603/10 880-613/10 880-623/10 10 120 155 138
880-603/11 880-613/11 880-623/11 11 120 155 138
880-603/12 880-613/12 880-623/12 12 120 155 138
880-603/13 880-613/13 880-623/13 13 120 155 138
880-603/14 880-613/14 880-623/14 14 120 155 138
880-603/15 880-613/15 880-623/15 15 120 155 138
880-603/16 880-613/16 880-623/16 16 120 155 138
880-603/17 880-613/17 880-623/17 17 120 155 138
880-603/18 880-613/18 880-623/18 18 120 155 138
880-603/19 880-613/19 880-623/19 19 120 155 138
880-603/20 880-613/20 880-623/20 20 120 155 138
880-605/11 880-615/11 880-625/11 11 160 195 178
880-605/12 880-615/12 880-625/12 12 160 195 178
880-605/13 880-615/13 880-625/13 13 160 195 178
880-605/14 880-615/14 880-625/14 14 160 195 178
880-605/15 880-615/15 880-625/15 15 160 195 178
880-605/16 880-615/16 880-625/16 16 160 195 178
880-605/17 880-615/17 880-625/17 17 160 195 178
880-605/18 880-615/18 880-625/18 18 160 195 178
880-605/19 880-615/19 880-625/19 19 160 195 178
880-605/20 880-615/20 880-625/20 20 160 195 178
880-607/12 880-617/12 880-627/12 12 200 235 218
880-607/14 880-617/14 880-627/14 14 200 235 218
880-607/16 880-617/16 880-627/16 16 200 235 218
880-607/18 880-617/18 880-627/18 18 200 235 218
880-607/20 880-617/20 880-627/20 20 200 235 218
880-607/22 880-617/22 880-627/22 22 200 235 218
880-608/12 880-618/12 880-628/12 12 220 255 238
880-608/14 880-618/14 880-628/14 14 220 255 238
880-608/16 880-618/16 880-628/16 16 220 255 238
880-608/18 880-618/18 880-628/18 18 220 255 238
880-608/20 880-618/20 880-628/20 20 220 255 238
880-608/22 880-618/22 880-628/22 22 220 255 238
880-609/12 880-619/12 880-629/12 12 240 275 258
880-609/14 880-619/14 880-629/14 14 240 275 258
880-609/16 880-619/16 880-629/16 16 240 275 258
880-609/18 880-619/18 880-629/18 18 240 275 258
880-609/20 880-619/20 880-629/20 20 240 275 258
880-609/22 880-619/22 880-629/22 22 240 275 258
* Tilastan = Ti6Al4V
67
Implants
LinkSymphoKnee – Conical Cemented Stems
D1
D2
L LF LT
D3
CoCrMo C
oCrMo/ 3 mm Offset 3 mm Offset 6 mm Offset 6 mm Offset D1 D2 D3 L LF LT
LINK PorEx* C
oCrMo C
oCrMo/ C
oCrMo C
oCrMo/ mm mm mm mm mm mm
LINK PorEx* LINK PorEx*
15-2950/17 15-3950/17 16 11 8 50 85 68
15-2950/12 15-3950/12 15-4302/08 15-5302/08 15-4602/08 15-5602/08 16 10 7 80 115 98
15-2950/18 15-3950/18 15-4304/08 15-5304/08 15-4604/08 15-5604/08 16 11 8 80 115 98
15-2950/27 15-3950/27 15-4305/08 15-5305/08 15-4605/08 15-5605/08 16 13 10 80 115 98
15-2950/36 15-3950/36 15-4306/08 15-5306/08 15-4606/08 15-5606/08 16 15 12 80 115 98
15-2950/13 15-3950/13 16 10 7 95 130 113
15-2950/19 15-3950/19 16 11 8 95 130 113
15-2950/28 15-3950/28 16 13 10 95 130 113
15-2950/37 15-3950/37 16 15 12 95 130 113
15-2950/14 15-3950/14 15-4302/12 15-5302/12 15-4602/12 15-5602/12 16 10 7 120 155 138
15-2950/20 15-3950/20 15-4304/12 15-5304/12 15-4604/12 15-5604/12 16 11 8 120 155 138
15-2950/29 15-3950/29 15-4305/12 15-5305/12 15-4605/12 15-5605/12 16 13 10 120 155 138
15-2950/38 15-3950/38 15-4306/12 15-5306/12 15-4606/12 15-5606/12 16 15 12 120 155 138
15-2950/15 15-3950/15 16 10 7 135 170 153
15-2950/21 15-3950/21 16 11 8 135 170 153
15-2950/30 15-3950/30 16 13 10 135 170 153
15-2950/39 15-3950/39 16 15 12 135 170 153
15-2950/16 15-3950/16 15-4302/16 15-5302/16 15-4602/16 15-5602/16 16 10 7 160 195 178
15-2950/22 15-3950/22 15-4304/16 15-5304/16 15-4604/16 15-5604/16 16 11 8 160 195 178
15-2950/31 15-3950/31 15-4305/16 15-5305/16 15-4605/16 15-5605/16 16 13 10 160 195 178
15-2950/40 15-3950/40 15-4306/16 15-5306/16 15-4606/16 15-5606/16 16 15 12 160 195 178
15-2950/44 15-3950/44 15-4304/18 15-5304/18 15-4604/18 15-5604/18 16 11 8 180 215 198
15-2950/45 15-3950/45 15-4305/18 15-5305/18 15-4605/18 15-5605/18 16 13 10 180 215 198
15-2950/46 15-3950/46 15-4306/18 15-5306/18 15-4606/18 15-5606/18 16 15 12 180 215 198
15-2950/23 15-3950/23 16 11 8 200 235 218
15-2950/32 15-3950/32 16 13 10 200 235 218
15-2950/41 15-3950/41 16 15 12 200 235 218
15-2950/24 15-3950/24 15-4304/24 15-5304/24 15-4604/24 15-5604/24 16 11 8 240 275 258
15-2950/33 15-3950/33 15-4305/24 15-5305/24 15-4605/24 15-5605/24 16 13 10 240 275 258
15-2950/42 15-3950/42 15-4306/24 15-5306/24 15-4606/24 15-5606/24 16 15 12 240 275 258
15-2950/25 15-3950/25 16 11 8 280 315 298
15-2950/34 15-3950/34 16 13 10 280 315 298
15-2950/43 15-3950/43 16 15 12 280 315 298
LinkSymphoKnee – Centralizers Ø
L
Ø Length (L)
UHMWPE mm mm
15-2975/12 12 15
15-2975/14 14 15
15-2975/16 16 15
15-2975/18 18 15
15-2975/20 20 15
15-2975/22 22 15
15-2975/24 24 15
* LINK PorEx: TiNbN = Titanium Niobium Nitride (gold color).
68
Implants
LinkSymphoKnee – Cylindrical Cementless Stems
D1
D2 LF
L LT
D3
T ilastan* 3 mm Offset 6 mm Offset D1 D2 D3 L LF LT
T
ilastan* T
ilastan* mm mm mm mm mm mm
15-2951/43 15-4310/08 15-4610/08 16 10 10 80 115 98
15-2951/44 15-4311/08 15-4611/08 16 11 11 80 115 98
15-2951/45 15-4312/08 15-4612/08 16 12 12 80 115 98
15-2951/46 15-4313/08 15-4613/08 16 13 13 80 115 98
15-2951/47 15-4314/08 15-4614/08 16 14 14 80 115 98
15-2951/48 15-4315/08 15-4615/08 16 15 15 80 115 98
15-2951/49 15-4316/08 15-4616/08 16 16 16 80 115 98
15-2951/50 15-4317/08 15-4617/08 17 17 17 80 115 98
15-2951/51 15-4318/08 15-4618/08 18 18 18 80 115 98
15-2951/54 15-4310/12 15-4610/12 16 10 10 120 155 138
15-2951/55 15-4311/12 15-4611/12 16 11 11 120 155 138
15-2951/72 15-4312/12 15-4612/12 16 12 12 120 155 138
15-2951/56 15-4313/12 15-4613/12 16 13 13 120 155 138
15-2951/78 15-4314/12 15-4614/12 16 14 14 120 155 138
15-2951/57 15-4315/12 15-4615/12 16 15 15 120 155 138
15-2951/84 15-4316/12 15-4616/12 16 16 16 120 155 138
15-2951/58 15-4317/12 15-4617/12 17 17 17 120 155 138
15-2951/90 15-4318/12 15-4618/12 18 18 18 120 155 138
15-2951/59 15-4319/12 15-4619/12 19 19 19 120 155 138
15-2951/60 15-4320/12 15-4620/12 20 20 20 120 155 138
15-2951/63 15-4311/16 15-4611/16 16 11 11 160 195 178
15-2951/73 15-4312/16 15-4612/16 16 12 12 160 195 178
15-2951/64 15-4313/16 15-4613/16 16 13 13 160 195 178
15-2951/79 15-4314/16 15-4614/16 16 14 14 160 195 178
15-2951/65 15-4315/16 15-4615/16 16 15 15 160 195 178
15-2951/85 15-4316/16 15-4616/16 16 16 16 160 195 178
15-2951/66 15-4317/16 15-4617/16 17 17 17 160 195 178
15-2951/91 15-4318/16 15-4618/16 18 18 18 160 195 178
15-2951/67 15-4319/16 15-4619/16 19 19 19 160 195 178
15-2951/68 15-4320/16 15-4620/16 20 20 20 160 195 178
15-2951/74 16 12 12 200 235 218
15-2951/80 16 14 14 200 235 218
15-2951/86 16 16 16 200 235 218
15-2951/92 18 18 18 200 235 218
15-2951/75 16 12 12 240 275 258
15-2951/81 16 14 14 240 275 258
15-2951/87 16 16 16 240 275 258
15-2951/93 18 18 18 240 275 258
* Tilastan = Ti6Al4V
69
Implants
LinkSymphoKnee – Conical Cementless Stems
D1
D2
L LF LT
D3
D1 D2 D3 L LF LT
Tilastan* mm mm mm mm mm mm
15-2953/00 16 12 9 128 163 146
15-2953/01 16 13 10 128 163 146
15-2953/02 16 14 11 128 163 146
15-2953/03 16 15 12 128 163 146
15-2953/04 16 16 13 128 163 146
15-2953/05 17 17 14 128 163 146
15-2953/06 18 18 15 128 163 146
15-2953/07 19 19 16 128 163 146
15-2953/08 20 20 17 128 163 146
15-2953/09 21 21 18 128 163 146
15-2953/10 22 22 19 128 163 146
15-2953/11 23 23 20 128 163 146
15-2953/12 24 24 21 128 163 146
15-2954/00 16 12 9 158 193 176
15-2954/01 16 13 10 158 193 176
15-2954/02 16 14 11 158 193 176
15-2954/03 16 15 12 158 193 176
15-2954/04 16 16 13 158 193 176
15-2954/05 17 17 14 158 193 176
15-2954/06 18 18 15 158 193 176
15-2954/07 19 19 16 158 193 176
15-2954/08 20 20 17 158 193 176
15-2954/09 21 21 18 158 193 176
15-2954/10 22 22 19 158 193 176
15-2954/11 23 23 20 158 193 176
15-2954/12 24 24 21 158 193 176
15-2958/00 16 12 9 188 223 206
15-2958/01 16 13 10 188 223 206
15-2958/02 16 14 11 188 223 206
15-2958/03 18 15 12 188 223 206
15-2958/04 16 16 13 188 223 206
15-2958/05 17 17 14 188 223 206
15-2958/06 18 18 15 188 223 206
15-2958/07 19 19 16 188 223 206
15-2958/08 20 20 17 188 223 206
15-2958/09 21 21 18 188 223 206
15-2958/10 22 22 19 188 223 206
15-2958/11 23 23 20 188 223 206
15-2958/12 24 24 21 188 223 206
* Tilastan = Ti6Al4V
70
Implants
Replacement Sets for Cone Adapter
Tilastan* + CoCrMo T ilastan* + CoCrMo/ LINK PorEx**
15-6118/29 15-6118/30
Each package contains:
• One Cone Adapter
• Two Security Screws
* Tilastan = Ti6Al4V
** LINK PorEx: TiNbN = Titanium Niobium Nitride (gold color).
71
Implants
LinkSymphoKnee Femoral Augments – Distal, Micro-Sizes
cemented
H H H
Height
Size Side
Tilastan* mm
880-300/11 0 Medial-Right/Lateral-Left 5
880-300/21 0 Lateral-Right/Medial-Left 5
880-301/11 1-2 Medial-Right/Lateral-Left 5
880-301/21 1-2 Lateral-Right/Medial-Left 5
LinkSymphoKnee Femoral Augments – Distal
cemented
H H H
Height
Size Side
Tilastan* mm
880-303/11 3-4 Medial-Right/Lateral-Left 5
880-303/12 3-4 Medial-Right/Lateral-Left 10
880-303/21 3-4 Lateral-Right/Medial-Left 5
880-303/22 3-4 Lateral-Right/Medial-Left 10
880-305/11 5-6 Medial-Right/Lateral-Left 5
880-305/12 5-6 Medial-Right/Lateral-Left 10
880-305/21 5-6 Lateral-Right/Medial-Left 5
880-305/22 5-6 Lateral-Right/Medial-Left 10
880-307/11 7-8 Medial-Right/Lateral-Left 5
880-307/12 7-8 Medial-Right/Lateral-Left 10
880-307/21 7-8 Lateral-Right/Medial-Left 5
880-307/22 7-8 Lateral-Right/Medial-Left 10
LinkSymphoKnee Femoral Augments – Distal, Macro-Sizes
cemented
H H H
Height
Size Side
Tilastan* mm
880-309/11 9-10 Medial-Right/Lateral-Left 5
880-309/12 9-10 Medial-Right/Lateral-Left 10
880-309/21 9-10 Lateral-Right/Medial-Left 5
880-309/22 9-10 Lateral-Right/Medial-Left 10
* Tilastan = Ti6Al4V
72
Implants
LinkSymphoKnee Femoral Augments – Posterior, Micro-Sizes
cemented
H H H
Height
Size Side
Tilastan* mm
880-310/11 0 Medial-Right/Lateral-Left 5
880-310/21 0 Lateral-Right/Medial-Left 5
880-311/11 1-2 Medial-Right/Lateral-Left 5
880-311/21 1-2 Lateral-Right/Medial-Left 5
LinkSymphoKnee Femoral Augments – Posterior
cemented
H H H
Height
Size Side
Tilastan* mm
880-313/11 3-4 Medial-Right/Lateral-Left 5
880-313/12 3-4 Medial-Right/Lateral-Left 10
880-313/21 3-4 Lateral-Right/Medial-Left 5
880-313/22 3-4 Lateral-Right/Medial-Left 10
880-315/11 5-6 Medial-Right/Lateral-Left 5
880-315/12 5-6 Medial-Right/Lateral-Left 10
880-315/21 5-6 Lateral-Right/Medial-Left 5
880-315/22 5-6 Lateral-Right/Medial-Left 10
880-317/11 7-8 Medial-Right/Lateral-Left 5
880-317/12 7-8 Medial-Right/Lateral-Left 10
880-317/21 7-8 Lateral-Right/Medial-Left 5
880-317/22 7-8 Lateral-Right/Medial-Left 10
LinkSymphoKnee Femoral Augments – Posterior, Macro-Sizes
cemented
H H H
Height
Size Side
Tilastan* mm
880-319/11 9-10 Medial-Right/Lateral-Left 5
880-319/12 9-10 Medial-Right/Lateral-Left 10
880-319/21 9-10 Lateral-Right/Medial-Left 5
880-319/22 9-10 Lateral-Right/Medial-Left 10
* Tilastan = Ti6Al4V
73
Implants
LinkSymphoKnee Femoral Augments – L-Shaped, Micro-Sizes
cemented
H H
Height
Size Side
Tilastan* mm
880-320/12 0 Medial-Right/Lateral-Left 10
880-320/22 0 Lateral-Right/Medial-Left 10
880-321/12 1-2 Medial-Right/Lateral-Left 10
880-321/22 1-2 Lateral-Right/Medial-Left 10
LinkSymphoKnee Femoral Augments – L-Shaped
cemented
H H
Height
Size Side
Tilastan* mm
880-323/13 3-4 Medial-Right/Lateral-Left 15
880-323/23 3-4 Lateral-Right/Medial-Left 15
880-325/13 5-6 Medial-Right/Lateral-Left 15
880-325/23 5-6 Lateral-Right/Medial-Left 15
880-327/13 7-8 Medial-Right/Lateral-Left 15
880-327/23 7-8 Lateral-Right/Medial-Left 15
LinkSymphoKnee Femoral Augments – L-Shaped, Macro-Sizes
cemented
H H
Height
Size Side
Tilastan* mm
880-329/13 9-10 Medial-Right/Lateral-Left 15
880-329/23 9-10 Lateral-Right/Medial-Left 15
* Tilastan = Ti6Al4V
74
Implants
LinkSymphoKnee Tibial Augments – Micro-Sizes
cemented
H
H H
Height
Size Side
Tilastan* mm
880-331/11 1-2 Medial-Right/Lateral-Left 5
880-331/12 1-2 Medial-Right/Lateral-Left 10
880-331/13 1-2 Medial-Right/Lateral-Left 15
880-331/21 1-2 Lateral-Right/Medial-Left 5
880-331/22 1-2 Lateral-Right/Medial-Left 10
880-331/23 1-2 Lateral-Right/Medial-Left 15
LinkSymphoKnee Tibial Augments
cemented
H
H H
Height
Size Side
Tilastan* mm
880-333/11 3-4 Medial-Right/Lateral-Left 5
880-333/12 3-4 Medial-Right/Lateral-Left 10
880-333/13 3-4 Medial-Right/Lateral-Left 15
880-333/21 3-4 Lateral-Right/Medial-Left 5
880-333/22 3-4 Lateral-Right/Medial-Left 10
880-333/23 3-4 Lateral-Right/Medial-Left 15
880-335/11 5-6 Medial-Right/Lateral-Left 5
880-335/12 5-6 Medial-Right/Lateral-Left 10
880-335/13 5-6 Medial-Right/Lateral-Left 15
880-335/21 5-6 Lateral-Right/Medial-Left 5
880-335/22 5-6 Lateral-Right/Medial-Left 10
880-335/23 5-6 Lateral-Right/Medial-Left 15
880-337/11 7-8 Medial-Right/Lateral-Left 5
880-337/12 7-8 Medial-Right/Lateral-Left 10
880-337/13 7-8 Medial-Right/Lateral-Left 15
880-337/21 7-8 Lateral-Right/Medial-Left 5
880-337/22 7-8 Lateral-Right/Medial-Left 10
880-337/23 7-8 Lateral-Right/Medial-Left 15
LinkSymphoKnee Tibial Augments – Macro-Sizes
cemented
H
H H
Height
Size Side
Tilastan* mm
880-339/11 9-10 Medial-Right/Lateral-Left 5
880-339/12 9-10 Medial-Right/Lateral-Left 10
880-339/13 9-10 Medial-Right/Lateral-Left 15
880-339/21 9-10 Lateral-Right/Medial-Left 5
880-339/22 9-10 Lateral-Right/Medial-Left 10
880-339/23 9-10 Lateral-Right/Medial-Left 15
* Tilastan = Ti6Al4V
75
Implants
LinkSymphoKnee Patella Components – 3-peg
cemented
H
Ø
Ø Height (H)
UHMWPE E-Dur * mm mm
880-501/25 880-511/25 25 6
880-501/28 880-511/28 28 6
880-501/31 880-511/31 31 7
880-501/34 880-511/34 34 8
880-501/37 880-511/37 37 9
880-501/40 880-511/40 40 10
* E-Dur = Highly crosslinked UHMWPE ( X-Linked PE) with Vit-E
76
Instruments
LinkSymphoKnee Instrument Sets: Overview
CCK CCK CCK CCK
CCK Instruments Basic Micro- Macro- Wide-
Kit Sizes Sizes Sizes
881-001/00 General Instruments – Blue feet X X X X
881-005/10 CCK Articulating Surface Trials – Yellow feet X X X X
881-006/10 CCK Instruments & Trials – Yellow feet X X X X
881-007/10 CCK Preparation Instruments – Yellow feet X X X X
881-006/00 CCK Instruments & Trials Micro-Sizes – Yellow feet X
881-007/00 CCK Preparation Instruments Micro-Sizes – Yellow feet X
881-006/20 CCK Instruments & Trials Macro-Sizes – Yellow feet X
881-007/20 CCK Preparation Instruments Macro-Sizes – Yellow feet X
881-006/30 CCK Instruments & Trials Wide-Sizes – Yellow feet X
Additional Instrument Sets:
881-008/00 Press-Fit Stem Reamers – Yellow feet
881-008/10 Press-Fit Stem Trials – Yellow feet
151-07/00 Ball Reamers & Trial Centralizers – Green feet
151-08/00 Conical Cemented Stem Trials – Yellow feet
151-09/00 Conical Cementless Stem Instruments & Trials, 128 mm – Yellow feet
151-10/00 Conical Cementless Stem Instruments & Trials, 158 mm – Yellow feet
151-11/00 Conical Cementless Stem Instruments & Trials, 188 mm – Yellow feet
881-009/00 Patella Instruments & Trials – Blue feet
LinkSymphoKnee Instruments upon request
Instruments upon request
445-126/65 Thread Pins, L = 65 mm, Ø 3.0 mm
445-126/95 Thread Pins, L = 95 mm, Ø 3.0 mm
445-127/35 Headed Thread Pins, L = 35 mm, Ø 3.0 mm
445-127/65 Headed Thread Pins, L = 65 mm, Ø 3.0 mm
77
Instruments
881-001/00 General Instruments – Blue feet
1 2 3 4 5 6
16 15 14 13 12 11 10 9 8 7
Z
1 881-010/00 LinkSymphoKnee General Instruments Tray – Upper Tray 1
2 445-206/00 Slaphammer 1
3 16-3283/01 Adapter, Hudson/Jacobs Fitting (E) 1
4 881-012/00 Universal Drill, Hudson Fitting (B) 1
5 445-122/00 Power Driver, Hudson Fitting 1
6 445-122/10 Power Driver with snap lock, Hudson Fitting 1
7 881-011/02 Spacer 4-in-1 Cut, Flexion, H = 10-12 mm 1
8 445-112/00 Handle, Quick Connect 1
9 445-121/00 Pin Inserter, universal 1
10 319-505/00B Step Drill, with Hudson Fitting (B) 1
11 15-2040/02B Twist Drill 3 mm, Hudson Fitting 1
12 317-802/53 Cutting Template 1
13 445-113/10 Alignment Rod, extramedullary, short 1
14 445-113/20 Alignment Rod, extramedullary, long 1
15 445-120/00 Pin Inserter/Extractor, universal 1
16 881-010/02 Spacer Flexion/Extension, H = 10-12 mm 1
78
Instruments
881-001/00 General Instruments – Blue feet
1 2 3 4 5 6 7 8
9
10
11
22 21 20 19 18 17 16 15 14 13 12
Z
1 881-010/00 LinkSymphoKnee General Instruments Tray – Lower Tray 1
2 445-101/00 Intramedullary Rod, Ø 8.0 mm 1
3 881-040/99 Tibial/Femoral Coupling Base Tip 1
4 881-040/01 Articulating Surface Impactor 1
5 15-2545 Torque Wrench, hex 2.5 mm 1
6 881-019/00 Articulating Surface Extractor 1
7 881-043/00 Tibial Extractor Tip 1
8 151-131/00 Screwdriver for Cone Adapter 1
9 881-019/01 Shim, Spacer, H = 1 mm 1
10 881-019/04 Shim, Spacer, H = 4 mm 1
11 881-019/08 Shim, Spacer, H = 8 mm 1
12 445-123/00 Pin Box 1
445-124/65 Drill Pin, L = 65 mm, Ø 3.0 mm 4
445-124/95 Drill Pin, L = 95 mm, Ø 3.0 mm 4
445-125/35 Drill Pin, Headed, L = 35 mm, Ø 3.0 mm 4
445-125/65 Drill Pin, Headed, L = 65 mm, Ø 3.0 mm 4
445-128/25 Bone Nail, L = 25 mm, Ø 3.0 mm 4
445-128/35 Bone Nail, L = 35 mm, Ø 3.0 mm 4
445-128/65 Bone Nail, L = 65 mm, Ø 3.0 mm 4
13 881-042/00 Tibial Inserter 1
14 881-042/90 Tibial Impactor Tip, All Poly 1
15 151-132/00 Modular Extractor Tip 1
16 881-041/99 Femoral Impactor Tip 1
17 881-042/99 Tibial Impactor Tip 1
18 881-053/00 Locking Screw Tibial Plateau, long 1
19 881-052/00 Locking Screw Tibial Plateau, short 1
20 319-601/30 Sterilizing Box, small 1
21 445-207/00 Handle, Impactor/Extractor 1
22 881-041/00 Femoral Inserter/Extractor 1
79
Instruments
881-005/10 CCK Articulating Surface Trials – Yellow feet
1 2-5 6-7 8-11 12-15
16-19
20-21
22-25
26-27
28-31 32-33 34-37 38-39
Z
1 881-050/10 LinkSymphoKnee CCK Articulating Surface Trials Tray 1
2 881-251/10 Trial Plateau, CCK, Size 1-2, H = 10 mm 1
3 881-251/12 Trial Plateau, CCK, Size 1-2, H = 12 mm 1
4 881-251/14 Trial Plateau, CCK, Size 1-2, H = 14 mm 1
5 881-251/16 Trial Plateau, CCK, Size 1-2, H = 16 mm 1
6 881-220/41 Shim, Trial Plateau, H = +4 mm, Size 1-2 1
7 881-250/81 Shim, Trial Plateau, CCK, H = +8 mm, Size 1-2 1
8 881-252/10 Trial Plateau, CCK, Size 1-2up, H = 10 mm 1
9 881-252/12 Trial Plateau, CCK, Size 1-2up, H = 12 mm 1
10 881-252/14 Trial Plateau, CCK, Size 1-2up, H = 14 mm 1
11 881-252/16 Trial Plateau, CCK, Size 1-2up, H = 16 mm 1
12 881-254/10 Trial Plateau, CCK, Size 3-4down, H = 10 mm 1
13 881-254/12 Trial Plateau, CCK, Size 3-4down, H = 12 mm 1
14 881-254/14 Trial Plateau, CCK, Size 3-4down, H = 14 mm 1
15 881-254/16 Trial Plateau, CCK, Size 3-4down, H = 16 mm 1
16 881-253/10 Trial Plateau, CCK, Size 3-4, H = 10 mm 1
17 881-253/12 Trial Plateau, CCK, Size 3-4, H = 12 mm 1
18 881-253/14 Trial Plateau, CCK, Size 3-4, H = 14 mm 1
19 881-253/16 Trial Plateau, CCK, Size 3-4, H = 16 mm 1
20 881-220/43 Shim, Trial Plateau, H= +4 mm, Size 3-4 1
21 881-250/83 Shim, Trial Plateau, CCK, H = +8 mm, Size 3-4 1
22 881-255/10 Trial Plateau, CCK, Size 5-6, H = 10 mm 1
23 881-255/12 Trial Plateau, CCK, Size 5-6, H = 12 mm 1
24 881-255/14 Trial Plateau, CCK, Size 5-6, H = 14 mm 1
25 881-255/16 Trial Plateau, CCK, Size 5-6, H = 16 mm 1
26 881-220/45 Shim, Trial Plateau, H = +4 mm, Size 5-6 1
27 881-250/85 Shim, Trial Plateau, CCK, H = +8 mm, Size 5-6 1
28 881-257/10 Trial Plateau, CCK, Size 7-8, H = 10 mm 1
29 881-257/12 Trial Plateau, CCK, Size 7-8, H = 12 mm 1
30 881-257/14 Trial Plateau, CCK, Size 7-8, H = 14 mm 1
31 881-257/16 Trial Plateau, CCK, Size 7-8, H = 16 mm 1
32 881-220/47 Shim, Trial Plateau, H = +4 mm, Size 7-8 1
33 881-250/87 Shim, Trial Plateau, CCK, H = +8 mm, Size 7-8 1
34 881-259/10 Trial Plateau, CCK, Size 9-10, H = 10 mm 1
35 881-259/12 Trial Plateau, CCK, Size 9-10, H = 12 mm 1
36 881-259/14 Trial Plateau, CCK, Size 9-10, H = 14 mm 1
37 881-259/16 Trial Plateau, CCK, Size 9-10, H = 16 mm 1
38 881-220/49 Shim, Trial Plateau, H = +4 mm, Size 9-10 1
39 881-250/89 Shim, Trial Plateau, CCK, H = +8 mm, Size 9-10 1
80
Instruments
881-006/10 CCK Instruments & Trials – Yellow feet
1 2-8 9 10 11 12 13 14
44
43
15
16
40-42
17
39
36-38 27-35 18-26
Z
1 881-060/10 LinkSymphoKnee CCK Instruments & Trials Tray – Upper Tray 1
2 151-050/12 50 mm Trial Stem Conical Cemented with Ø 12 mm Centralizer 1
3 151-050/14 50 mm Trial Stem Conical Cemented with Ø 14 mm Centralizer 1
4 151-050/16 50 mm Trial Stem Conical Cemented with Ø 16 mm Centralizer 1
5 151-050/18 50 mm Trial Stem Conical Cemented with Ø 18 mm Centralizer 1
6 151-050/20 50 mm Trial Stem Conical Cemented with Ø 20 mm Centralizer 1
7 151-050/22 50 mm Trial Stem Conical Cemented with Ø 22 mm Centralizer 1
8 151-050/24 50 mm Trial Stem Conical Cemented with Ø 24 mm Centralizer 1
9 881-285/30 Tibial Preparation Plate, CCK, Size 3 1
10 881-285/40 Tibial Preparation Plate, CCK, Size 4 1
11 881-285/50 Tibial Preparation Plate, CCK, Size 5 1
12 881-285/60 Tibial Preparation Plate, CCK, Size 6 1
13 881-285/70 Tibial Preparation Plate, CCK, Size 7 1
14 881-285/80 Tibial Preparation Plate, CCK, Size 8 1
15 151-501/00 Neutral Stem Adapter 1
16 881-070/08 Sleeve, Assignment for Intramedullary Rod, Ø 8 mm 1
17 881-067/17 Tapered Reamer, CCK 1
18 881-333/11 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 3-4, H = 5 mm 1
19 881-333/12 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 3-4, H = 10 mm 1
20 881-333/13 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 3-4, H = 15 mm 1
21 881-335/11 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 5-6, H = 5 mm 1
22 881-335/12 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 5-6, H = 10 mm 1
23 881-335/13 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 5-6, H = 15 mm 1
24 881-337/11 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 7-8, H = 5 mm 1
25 881-337/12 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 7-8, H = 10 mm 1
26 881-337/13 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 7-8, H = 15 mm 1
27 881-333/21 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 3-4, H = 5 mm 1
28 881-333/22 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 3-4, H = 10 mm 1
29 881-333/23 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 3-4, H = 15 mm 1
30 881-335/21 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 5-6, H = 5 mm 1
31 881-335/22 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 5-6, H = 10 mm 1
32 881-335/23 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 5-6, H = 15 mm 1
33 881-337/21 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 7-8, H = 5 mm 1
34 881-337/22 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 7-8, H = 10 mm 1
35 881-337/23 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 7-8, H = 15 mm 1
36 151-080/15 80 mm Trial Stem Conical Cemented, Ø 15 mm 1
37 151-095/15 95 mm Trial Stem Conical Cemented, Ø 15 mm 1
38 151-120/15 120 mm Trial Stem Conical Cemented, Ø 15 mm 1
39 881-065/09 Reamer Stopper, Offset 1
40 881-275/34 Tibial Keel Punch, CCK, Size 3-4 1
41 881-275/56 Tibial Keel Punch, CCK, Size 5-6 1
42 881-275/78 Tibial Keel Punch, CCK, Size 7-8 1
43 881-055/00 Neutral Alignment Guide, CCK, 0 mm offset 1
44 881-065/00 Tibial Reamer Guide, CCK 1
81
Instruments
881-006/10 CCK Instruments & Trials – Yellow feet
1 2-7
8
9
10
21-26
11
12
13
15-20 14
Z
1 881-060/10 LinkSymphoKnee CCK Instruments & Trials Tray – Lower Tray 1
2 881-150/30 Femoral Trial, CCK, right, Size 3 1
3 881-150/40 Femoral Trial, CCK, right, Size 4 1
4 881-150/50 Femoral Trial, CCK, right, Size 5 1
5 881-150/60 Femoral Trial, CCK, right, Size 6 1
6 881-150/70 Femoral Trial, CCK, right, Size 7 1
7 881-150/80 Femoral Trial, CCK, right, Size 8 1
8 881-302/91 Femoral Trial Augment, H = 5 mm 4
9 881-302/92 Femoral Trial Augment, H = 10 mm 4
10 881-323/93 Femoral Trial Augment, L-Shaped, Size 3-4, H = 15 mm 2
11 881-325/93 Femoral Trial Augment, L-Shaped, Size 5-6, H = 15 mm 2
12 881-327/93 Femoral Trial Augment, L-Shaped, Size 7-8, H = 15 mm 2
13 319-603/30 Sterilizing Box, medium high 1
14 881-116/00 Reamer Guide, Femoral Box, CCK 1
15 881-151/30 Femoral Trial, CCK, left, Size 3 1
16 881-151/40 Femoral Trial, CCK, left, Size 4 1
17 881-151/50 Femoral Trial, CCK, left, Size 5 1
18 881-151/60 Femoral Trial, CCK, left, Size 6 1
19 881-151/70 Femoral Trial, CCK, left, Size 7 1
20 881-151/80 Femoral Trial, CCK, left, Size 8 1
21 881-258/30 Tibial Trial Component, CCK, Size 3 1
22 881-258/40 Tibial Trial Component, CCK, Size 4 1
23 881-258/50 Tibial Trial Component, CCK, Size 5 1
24 881-258/60 Tibial Trial Component, CCK, Size 6 1
25 881-258/70 Tibial Trial Component, CCK, Size 7 1
26 881-258/80 Tibial Trial Component, CCK, Size 8 1
82
Instruments
881-007/10 CCK Preparation Instruments – Yellow feet
1 2 3 4 5 6 7 8 9 10-11 12-17
18-23
33-35
32
31 30 29 28 27 26 25 24
Z
1 881-070/10 LinkSymphoKnee CCK Preparation Instruments Tray 1
2 151-201/00 Tibial Guide, Intramedullary, 0° 1
3 881-059/00 Tibial Guide, Intramedullary, 3°, CCK 1
4 445-111/00 Stylus, adjustable 1
5 15-6053/00 T-Handle 1
6 151-501/00 Neutral Stem Adapter 2
7 151-803/03 3 mm Offset Adapter 2
8 151-806/06 6 mm Offset Adapter 2
9 881-159/00 Femoral Cutting Block, CCK, Distal Cut 1
10 881-013/16 Shim, Spacer 4-in-1 Cut CCK, Flexion, H = 4 mm 1
11 881-013/20 Shim, Spacer 4-in-1 Cut CCK, Flexion, H = 8 mm 1
12 881-105/30 Femoral Sizer, Template, Size 3, CCK 1
13 881-105/40 Femoral Sizer, Template, Size 4, CCK 1
14 881-105/50 Femoral Sizer, Template, Size 5, CCK 1
15 881-105/60 Femoral Sizer, Template, Size 6, CCK 1
16 881-105/70 Femoral Sizer, Template, Size 7, CCK 1
17 881-105/80 Femoral Sizer, Template, Size 8, CCK 1
18 881-115/30 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 3 1
19 881-115/40 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 4 1
20 881-115/50 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 5 1
21 881-115/60 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 6 1
22 881-115/70 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 7 1
23 881-115/80 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 8 1
24 881-050/06 Femoral Alignment Guide, CCK, Varus/Valgus Adjustment 6°, right 1
25 881-051/06 Femoral Alignment Guide, CCK, Varus/Valgus Adjustment 6°, left 1
26 881-056/00 Trial Stem Rod, Intramedullary 1
27 881-013/00 Spacer 4-in-1 Cut, CCK, Flexion, H = 10-12 mm 1
28 151-144/00 Extractor for Modular Stems/Offset Stems 1
29 881-099/00 Offset Selector, CCK 1
30 151-204/00 Tibial Cutting Block, right 1
31 151-203/00 Tibial Cutting Block, left 1
32 151-202/00 Tibial Guide Connector 1
33 881-114/34 Femoral Box Guide, CCK, Size 3-4 1
34 881-114/56 Femoral Box Guide, CCK, Size 5-6 1
35 881-114/78 Femoral Box Guide, CCK, Size 7-8 1
83
Instruments
881-006/00 CCK Instruments & Trials Micro-Sizes – Yellow feet
1 2 3 4-5 6 7-9 10-12 13-15
16
17
18
19
20
21-23
Z
1 881-060/00 LinkSymphoKnee CCK Instruments & Trials Micro-Sizes Tray 1
2 881-285/10 Tibial Preparation Plate, CCK, Size 1 1
3 881-285/20 Tibial Preparation Plate, CCK, Size 2 1
4 881-258/10 Tibial Trial Component, CCK, Size 1 1
5 881-258/20 Tibial Trial Component, CCK, Size 2 1
6 881-275/12 Tibial Keel Punch, CCK, Size 1-2 1
7 881-150/00 Femoral Trial, CCK, right, Size 0 1
8 881-150/10 Femoral Trial, CCK, right, Size 1 1
9 881-150/20 Femoral Trial, CCK, right, Size 2 1
10 881-331/11 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 1-2, H = 5 mm 1
11 881-331/12 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 1-2, H = 10 mm 1
12 881-331/13 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 1-2, H = 15 mm 1
13 881-331/21 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 1-2, H = 5 mm 1
14 881-331/22 Tibial Trial Augment, Lateral-Right/Lateral-Left, Size 1-2, H = 10 mm 1
15 881-331/23 Tibial Trial Augment, Lateral-Right/Lateral-Left, Size 1-2, H = 15 mm 1
16 881-301/91 Femoral Trial Augment, Micro-Sizes, H = 5 mm 4
17 881-301/92 Femoral Trial Augment, Micro-Sizes, H = 10 mm 4
18 881-320/92 Femoral Trial Augment, L-Shaped, Size 0, H = 10 mm 2
19 881-321/92 Femoral Trial Augment, L-Shaped, Size 1-2, H = 10 mm 2
20 319-603/30 Sterilizing Box, medium high 1
21 881-151/00 Femoral Trial, CCK, left, Size 0 1
22 881-151/10 Femoral Trial, CCK, left, Size 1 1
23 881-151/20 Femoral Trial, CCK, left, Size 2 1
84
Instruments
881-007/00 CCK Preparation Instruments Micro-Sizes – Yellow feet
1 2 3 4
5
10-11 9 6-8
Z
1 881-070/00 LinkSymphoKnee CCK Preparation Instruments Micro-Sizes Tray 1
2 881-105/00 Femoral Sizer, Template, Size 0, CCK 1
3 881-105/10 Femoral Sizer, Template, Size 1, CCK 1
4 881-105/20 Femoral Sizer, Template, Size 2, CCK 1
5 881-010/01 Spacer, Micro-Sizes, Flexion/ Extension, H = 10-12 mm 1
6 881-115/00 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 0 1
7 881-115/10 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 1 1
8 881-115/20 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 2 1
9 881-114/02 Femoral Box Guide, CCK, Size 0-1-2 1
10 881-019/94 Shim, Spacer, H = 4 mm, Micro-Sizes 1
11 881-019/98 Shim, Spacer, H = 8 mm, Micro-Sizes 1
85
Instruments
881-006/20 CCK Instruments & Trials Macro-Sizes – Yellow feet
1 2 3 4-5 6 7 8 9-11 12-14
15
16
17
18
19 20
Z
1 881-060/20 LinkSymphoKnee CCK Instruments & Trials Macro-Sizes Tray 1
2 881-285/90 Tibial Preparation Plate, CCK, Size 9 1
3 881-285/X0 Tibial Preparation Plate, CCK, Size 10 1
4 881-258/90 Tibial Trial Component, CCK, Size 9 1
5 881-258/X0 Tibial Trial Component, CCK, Size 10 1
6 881-275/9X Tibial Keel Punch, CCK, Size 9-10 1
7 881-150/90 Femoral Trial, CCK, right, Size 9 1
8 881-150/X0 Femoral Trial, CCK, right, Size 10 1
9 881-339/11 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 9-10, H = 5 mm 1
10 881-339/12 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 9-10, H = 10 mm 1
11 881-339/13 Tibial Trial Augment, Medial-Right/Lateral-Left, Size 9-10, H = 15 mm 1
12 881-339/21 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 9-10, H = 5 mm 1
13 881-339/22 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 9-10, H = 10 mm 1
14 881-339/23 Tibial Trial Augment, Lateral-Right/Medial-Left, Size 9-10, H = 15 mm 1
15 881-303/91 Femoral Trial Augment, Macro-Sizes, H = 5 mm 4
16 881-303/92 Femoral Trial Augment, Macro-Sizes, H = 10 mm 4
17 881-329/93 Femoral Trial Augment, L-Shaped, Size 9-10, H = 15 mm 2
18 319-603/30 Sterilizing Box, medium high 1
19 881-151/90 Femoral Trial, CCK, left, Size 9 1
20 881-151/X0 Femoral Trial, CCK, left, Size 10 1
86
Instruments
881-007/20 CCK Preparation Instruments Macro-Sizes – Yellow feet
1 2 3 4
5
6
Z
1 881-070/20 LinkSymphoKnee CCK Preparation Instruments Macro-Sizes Tray 1
2 881-105/90 Femoral Sizer, Template, Size 9, CCK 1
3 881-105/X0 Femoral Sizer, Template, Size 10, CCK 1
4 881-115/90 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 9 1
5 881-115/X0 Femoral Cutting Block, CCK, 4-in-1 Cut, Size 10 1
6 881-114/9X Femoral Box Guide, CCK, Size 9-10 1
87
Instruments
881-006/30 CCK Instruments & Trials Wide-Sizes – Yellow feet
1 2 3 4
5 6 7
Z
1 881-060/30 LinkSymphoKnee CCK Instruments & Trials Wide-Sizes Tray 1
2 881-150/35 Femoral Trial, CCK, right, Size 3+ 1
3 881-150/45 Femoral Trial, CCK, right, Size 4+ 1
4 881-150/55 Femoral Trial, CCK, right, Size 5+ 1
5 881-151/35 Femoral Trial, CCK, left, Size 3+ 1
6 881-151/45 Femoral Trial, CCK, left, Size 4+ 1
7 881-151/55 Femoral Trial, CCK, left, Size 5+ 1
88
Instruments
881-008/00 Press-Fit Stem Reamers – Yellow feet
1 2 3 4 5 6 7 8 9 10 11
13 12
Z
1 881-080/00 LinkSymphoKnee Press-Fit Stem Reamers 1
2 881-069/10 Reamer, cementless, cylindrical, Ø 10 mm 1
3 881-069/11 Reamer, cementless, cylindrical, Ø 11 mm 1
4 881-069/12 Reamer, cementless, cylindrical, Ø 12 mm 1
5 881-069/13 Reamer, cementless, cylindrical, Ø 13 mm 1
6 881-069/14 Reamer, cementless, cylindrical, Ø 14 mm 1
7 881-069/15 Reamer, cementless, cylindrical, Ø 15 mm 1
8 881-069/16 Reamer, cementless, cylindrical, Ø 16 mm 1
9 881-069/17 Reamer, cementless, cylindrical, Ø 17 mm 1
10 881-069/18 Reamer, cementless, cylindrical, Ø 18 mm 1
11 881-069/19 Reamer, cementless, cylindrical, Ø 19 mm 1
12 881-069/20 Reamer, cementless, cylindrical, Ø 20 mm 1
13 881-069/22 Reamer, cementless, cylindrical, Ø 22 mm 1
89
Instruments
881-008/10 Press-Fit Stem Trials – Yellow feet
1 2-10 11-21
22-31
Z
1 881-080/10 LinkSymphoKnee Press-Fit Stem Trials – Upper Tray 1
2 151-780/10 Trial Stem, cementless, cylindrical, Ø 10 mm, Stem.L 80 mm 1
3 151-780/11 Trial Stem, cementless, cylindrical, Ø 11 mm, Stem.L 80 mm 1
4 151-780/12 Trial Stem, cementless, cylindrical, Ø 12 mm, Stem.L 80 mm 1
5 151-780/13 Trial Stem, cementless, cylindrical, Ø 13 mm, Stem.L 80 mm 1
6 151-780/14 Trial Stem, cementless, cylindrical, Ø 14 mm, Stem.L 80 mm 1
7 151-780/15 Trial Stem, cementless, cylindrical, Ø 15 mm, Stem.L 80 mm 1
8 151-780/16 Trial Stem, cementless, cylindrical, Ø 16 mm, Stem.L 80 mm 1
9 151-780/17 Trial Stem, cementless, cylindrical, Ø 17 mm, Stem.L 80 mm 1
10 151-780/18 Trial Stem, cementless, cylindrical, Ø 18 mm, Stem.L 80 mm 1
11 151-712/10 Trial Stem, cementless, cylindrical, Ø 10 mm, Stem.L 120 mm 1
12 151-712/11 Trial Stem, cementless, cylindrical, Ø 11 mm, Stem.L 120 mm 1
13 151-712/12 Trial Stem, cementless, cylindrical, Ø 12 mm, Stem.L 120 mm 1
14 151-712/13 Trial Stem, cementless, cylindrical, Ø 13 mm, Stem.L 120 mm 1
15 151-712/14 Trial Stem, cementless, cylindrical, Ø 14 mm, Stem.L 120 mm 1
16 151-712/15 Trial Stem, cementless, cylindrical, Ø 15 mm, Stem.L 120 mm 1
17 151-712/16 Trial Stem, cementless, cylindrical, Ø 16 mm, Stem.L 120 mm 1
18 151-712/17 Trial Stem, cementless, cylindrical, Ø 17 mm, Stem.L 120 mm 1
19 151-712/18 Trial Stem, cementless, cylindrical, Ø 18 mm, Stem.L 120 mm 1
20 151-712/19 Trial Stem, cementless, cylindrical, Ø 19 mm, Stem.L 120 mm 1
21 151-712/20 Trial Stem, cementless, cylindrical, Ø 20 mm, Stem.L 120 mm 1
22 151-716/11 Trial Stem, cementless, cylindrical, Ø 11 mm, Stem.L 160 mm 1
23 151-716/12 Trial Stem, cementless, cylindrical, Ø 12 mm, Stem.L 160 mm 1
24 151-716/13 Trial Stem, cementless, cylindrical, Ø 13 mm, Stem.L 160 mm 1
25 151-716/14 Trial Stem, cementless, cylindrical, Ø 14 mm, Stem.L 160 mm 1
26 151-716/15 Trial Stem, cementless, cylindrical, Ø 15 mm, Stem.L 160 mm 1
27 151-716/16 Trial Stem, cementless, cylindrical, Ø 16 mm, Stem.L 160 mm 1
28 151-716/17 Trial Stem, cementless, cylindrical, Ø 17 mm, Stem.L 160 mm 1
29 151-716/18 Trial Stem, cementless, cylindrical, Ø 18 mm, Stem.L 160 mm 1
30 151-716/19 Trial Stem, cementless, cylindrical, Ø 19 mm, Stem.L 160 mm 1
31 151-716/20 Trial Stem, cementless, cylindrical, Ø 20 mm, Stem.L 160 mm 1
90
Instruments
881-008/10 Press-Fit Stem Trials – Yellow feet
1 2-7 8-13 14-19
Z
1 881-080/10 LinkSymphoKnee Press-Fit Stem Trials – Lower Tray 1
2 151-720/12 Trial Stem, cementless, cylindrical, Ø 12 mm, Stem.L 200 mm 1
3 151-720/14 Trial Stem, cementless, cylindrical, Ø 14 mm, Stem.L 200 mm 1
4 151-720/16 Trial Stem, cementless, cylindrical, Ø 16 mm, Stem.L 200 mm 1
5 151-720/18 Trial Stem, cementless, cylindrical, Ø 18 mm, Stem.L 200 mm 1
6 151-720/20 Trial Stem, cementless, cylindrical, Ø 20 mm, Stem.L 200 mm 1
7 151-720/22 Trial Stem, cementless, cylindrical, Ø 22 mm, Stem.L 200 mm 1
8 151-722/12 Trial Stem, cementless, cylindrical, Ø 12 mm, Stem.L 220 mm 1
9 151-722/14 Trial Stem, cementless, cylindrical, Ø 14 mm, Stem.L 220 mm 1
10 151-722/16 Trial Stem, cementless, cylindrical, Ø 16 mm, Stem.L 220 mm 1
11 151-722/18 Trial Stem, cementless, cylindrical, Ø 18 mm, Stem.L 220 mm 1
12 151-722/20 Trial Stem, cementless, cylindrical, Ø 20 mm, Stem.L 220 mm 1
13 151-722/22 Trial Stem, cementless, cylindrical, Ø 22 mm, Stem.L 220 mm 1
14 151-724/12 Trial Stem, cementless, cylindrical, Ø 12 mm, Stem.L 240 mm 1
15 151-724/14 Trial Stem, cementless, cylindrical, Ø 14 mm, Stem.L 240 mm 1
16 151-724/16 Trial Stem, cementless, cylindrical, Ø 16 mm, Stem.L 240 mm 1
17 151-724/18 Trial Stem, cementless, cylindrical, Ø 18 mm, Stem.L 240 mm 1
18 151-724/20 Trial Stem, cementless, cylindrical, Ø 20 mm, Stem.L 240 mm 1
19 151-724/22 Trial Stem, cementless, cylindrical, Ø 22 mm, Stem.L 240 mm 1
91
Instruments
151-07/00 Ball Reamers & Trial Centralizers – Green feet
1 2-8 9 10 11 12 13 14 15 16 17
Z
1 151-07/07 Ball Reamers & Trial Centralizers Tray 1
2 151-106/12 Trial Centralizer, 12 mm 2
3 151-106/14 Trial Centralizer, 14 mm 2
4 151-106/16 Trial Centralizer, 16 mm 2
5 151-106/18 Trial Centralizer, 18 mm 2
6 151-106/20 Trial Centralizer, 20 mm 2
7 151-106/22 Trial Centralizer, 22 mm 2
8 151-106/24 Trial Centralizer, 24 mm 2
9 15-1133/02B Ball Reamer, 10 mm 1
10 15-1133/03B Ball Reamer, 12 mm 1
11 15-1133/04B Ball Reamer, 14 mm 1
12 15-1133/05B Ball Reamer, 16 mm 1
13 15-1133/06B Ball Reamer, 18 mm 1
14 15-1133/07B Ball Reamer, 20 mm 1
15 15-1133/08B Ball Reamer, 22 mm 1
16 15-1133/09B Ball Reamer, 24 mm 1
17 15-1133/10B Ball Reamer, 26 mm 1
92
Instruments
151-08/00 Conical Cemented Stem Trials – Yellow feet
1 2-5 6-9 10-13 14-17 18-21 28-30
34-40 21-24 25-27 31-33
Z
1 151-08/08 Conical Cemented Stem Trials Tray 1
2 151-080/10 Trial Stem Conical Cemented, Ø 10 mm, 80 mm 1
3 151-080/11 Trial Stem Conical Cemented, Ø 11 mm, 80 mm 1
4 151-080/13 Trial Stem Conical Cemented, Ø 13 mm, 80 mm 1
5 151-080/15 Trial Stem Conical Cemented, Ø 15 mm, 80 mm 1
6 151-095/10 Trial Stem Conical Cemented, Ø 10 mm, 95 mm 1
7 151-095/11 Trial Stem Conical Cemented, Ø 11 mm, 95 mm 1
8 151-095/13 Trial Stem Conical Cemented, Ø 13 mm, 95 mm 1
9 151-095/15 Trial Stem Conical Cemented, Ø 15 mm, 95 mm 1
10 151-120/10 Trial Stem Conical Cemented, Ø 10 mm, 120 mm 1
11 151-120/11 Trial Stem Conical Cemented, Ø 11 mm, 120 mm 1
12 151-120/13 Trial Stem Conical Cemented, Ø 13 mm, 120 mm 1
13 151-120/15 Trial Stem Conical Cemented, Ø 15 mm, 120 mm 1
14 151-135/10 Trial Stem Conical Cemented, Ø 10 mm, 135 mm 1
15 151-135/11 Trial Stem Conical Cemented, Ø 11 mm, 135 mm 1
16 151-135/13 Trial Stem Conical Cemented, Ø 13 mm, 135 mm 1
17 151-135/15 Trial Stem Conical Cemented, Ø 15 mm, 135 mm 1
18 151-160/10 Trial Stem Conical Cemented, Ø 10 mm, 160 mm 1
19 151-160/11 Trial Stem Conical Cemented, Ø 11 mm, 160 mm 1
20 151-160/13 Trial Stem Conical Cemented, Ø 13 mm, 160 mm 1
21 151-160/15 Trial Stem Conical Cemented, Ø 15 mm, 160 mm 1
22 151-180/11 Trial Stem Conical Cemented, Ø 11 mm, 180 mm 1
23 151-180/13 Trial Stem Conical Cemented, Ø 13 mm, 180 mm 1
24 151-180/15 Trial Stem Conical Cemented, Ø 15 mm, 180 mm 1
25 151-200/11 Trial Stem Conical Cemented, Ø 11 mm, 200 mm 1
26 151-200/13 Trial Stem Conical Cemented, Ø 13 mm, 200 mm 1
27 151-200/15 Trial Stem Conical Cemented, Ø 15 mm, 200 mm 1
28 151-240/11 Trial Stem Conical Cemented, Ø 11 mm, 240 mm 1
29 151-240/13 Trial Stem Conical Cemented, Ø 13 mm, 240 mm 1
30 151-240/15 Trial Stem Conical Cemented, Ø 15 mm, 240 mm 1
31 151-280/11 Trial Stem Conical Cemented, Ø 11 mm, 280 mm 1
32 151-280/13 Trial Stem Conical Cemented, Ø 13 mm, 280 mm 1
33 151-280/15 Trial Stem Conical Cemented, Ø 15 mm, 280 mm 1
34 151-050/12 Trial Stem Conical Cemented, Ø 11 mm with Ø 12 mm Centralizer, 50 mm 1
35 151-050/14 Trial Stem Conical Cemented, Ø 11 mm with Ø 14 mm Centralizer, 50 mm 1
36 151-050/16 Trial Stem Conical Cemented, Ø 11 mm with Ø 16 mm Centralizer, 50 mm 1
37 151-050/18 Trial Stem Conical Cemented, Ø 11 mm with Ø 18 mm Centralizer, 50 mm 1
38 151-050/20 Trial Stem Conical Cemented, Ø 11 mm with Ø 20 mm Centralizer, 50 mm 1
39 151-050/22 Trial Stem Conical Cemented, Ø 11 mm with Ø 22 mm Centralizer, 50 mm 1
40 151-050/24 Trial Stem Conical Cemented, Ø 11 mm with Ø 24 mm Centralizer, 50 mm 1
93
Instruments
151-09/00 Conical Cementless Stem Instruments & Trials, 128 mm – Yellow feet
1 2 3 4 5 6 7 9 11 8 10 12
13 15 19 17 14 16 18
Z
1 151-09/09 Conical Cementless Stem Instruments & Trials, 128 mm Tray – Upper Tray 1
2 151-718/00 Reamer Stopper, Tibia, XS-L 2
3 151-719/00 Reamer Stopper, Femur, XS 2
4 151-719/01 Reamer Stopper, Femur, S 2
5 151-719/02 Reamer Stopper, Femur, M 2
6 151-719/03 Reamer Stopper, Femur, L 2
7 151-100/12 Trial Stem Conical, Cementless, 128 mm, Ø 12 mm 1
8 151-100/13 Trial Stem Conical, Cementless, 128 mm, Ø 13 mm 1
9 151-100/14 Trial Stem Conical, Cementless, 128 mm, Ø 14 mm 1
10 151-100/15 Trial Stem Conical, Cementless, 128 mm, Ø 15 mm 1
11 151-100/16 Trial Stem Conical, Cementless, 128 mm, Ø 16 mm 1
12 151-100/17 Trial Stem Conical, Cementless, 128 mm, Ø 17 mm 1
13 151-100/18 Trial Stem Conical, Cementless, 128 mm, Ø 18 mm 1
14 151-100/19 Trial Stem Conical, Cementless, 128 mm, Ø 19 mm 1
15 151-100/20 Trial Stem Conical, Cementless, 128 mm, Ø 20 mm 1
16 151-100/21 Trial Stem Conical, Cementless, 128 mm, Ø 21 mm 1
17 151-100/22 Trial Stem Conical, Cementless, 128 mm, Ø 22 mm 1
18 151-100/23 Trial Stem Conical, Cementless, 128 mm, Ø 23 mm 1
19 151-100/24 Trial Stem Conical, Cementless, 128 mm, Ø 24 mm 1
94
Instruments
151-09/00 Conical Cementless Stem Instruments & Trials, 128 mm – Yellow feet
1 2 4 6 8 10 3 5 7 9 11
12 14 13
Z
1 151-09/09 Conical Cementless Stem Instruments & Trials, 128 mm Tray – Lower Tray 1
2 151-610/12 Reamer Conical, 128 mm, Ø 12 mm 1
3 151-610/13 Reamer Conical, 128 mm, Ø 13 mm 1
4 151-610/14 Reamer Conical, 128 mm, Ø 14 mm 1
5 151-610/15 Reamer Conical, 128 mm, Ø 15 mm 1
6 151-610/16 Reamer Conical, 128 mm, Ø 16 mm 1
7 151-610/17 Reamer Conical, 128 mm, Ø 17 mm 1
8 151-610/18 Reamer Conical, 128 mm, Ø 18 mm 1
9 151-610/19 Reamer Conical, 128 mm, Ø 19 mm 1
10 151-610/20 Reamer Conical, 128 mm, Ø 20 mm 1
11 151-610/21 Reamer Conical, 128 mm, Ø 21 mm 1
12 151-610/22 Reamer Conical, 128 mm, Ø 22 mm 1
13 151-610/23 Reamer Conical, 128 mm, Ø 23 mm 1
14 151-610/24 Reamer Conical, 128 mm, Ø 24 mm 1
95
Instruments
151-10/00 Conical Cementless Stem Instruments & Trials, 158 mm – Yellow feet
1 2 3 4 5 6 7 9 11 8 10 12
13 15 19 17 14 16 18
Z
1 151-10/10 Conical Cementless Stem Instruments & Trials, 158 mm Tray – Upper Tray 1
2 151-718/00 Reamer Stopper, Tibia, XS-L 2
3 151-719/00 Reamer Stopper, Femur, XS 2
4 151-719/01 Reamer Stopper, Femur, S 2
5 151-719/02 Reamer Stopper, Femur, M 2
6 151-719/03 Reamer Stopper, Femur, L 2
7 151-130/12 Trial Stem Conical, Cementless, 158 mm, Ø 12 mm 1
8 151-130/13 Trial Stem Conical, Cementless, 158 mm, Ø 13 mm 1
9 151-130/14 Trial Stem Conical, Cementless, 158 mm, Ø 14 mm 1
10 151-130/15 Trial Stem Conical, Cementless, 158 mm, Ø 15 mm 1
11 151-130/16 Trial Stem Conical, Cementless, 158 mm, Ø 16 mm 1
12 151-130/17 Trial Stem Conical, Cementless, 158 mm, Ø 17 mm 1
13 151-130/18 Trial Stem Conical, Cementless, 158 mm, Ø 18 mm 1
14 151-130/19 Trial Stem Conical, Cementless, 158 mm, Ø 19 mm 1
15 151-130/20 Trial Stem Conical, Cementless, 158 mm, Ø 20 mm 1
16 151-130/21 Trial Stem Conical, Cementless, 158 mm, Ø 21 mm 1
17 151-130/22 Trial Stem Conical, Cementless, 158 mm, Ø 22 mm 1
18 151-130/23 Trial Stem Conical, Cementless, 158 mm, Ø 23 mm 1
19 151-130/24 Trial Stem Conical, Cementless, 158 mm, Ø 24 mm 1
96
Instruments
151-10/00 Conical Cementless Stem Instruments & Trials, 158 mm – Yellow feet
1 2 4 6 8 10 3 5 7 9 11
12 14 13
Z
1 151-10/10 Conical Cementless Stem Instruments & Trials, 158 mm Tray – Lower Tray 1
2 151-613/12 Reamer Conical, 158 mm, Ø 12 mm 1
3 151-613/13 Reamer Conical, 158 mm, Ø 13 mm 1
4 151-613/14 Reamer Conical, 158 mm, Ø 14 mm 1
5 151-613/15 Reamer Conical, 158 mm, Ø 15 mm 1
6 151-613/16 Reamer Conical, 158 mm, Ø 16 mm 1
7 151-613/17 Reamer Conical, 158 mm, Ø 17 mm 1
8 151-613/18 Reamer Conical, 158 mm, Ø 18 mm 1
9 151-613/19 Reamer Conical, 158 mm, Ø 19 mm 1
10 151-613/20 Reamer Conical, 158 mm, Ø 20 mm 1
11 151-613/21 Reamer Conical, 158 mm, Ø 21 mm 1
12 151-613/22 Reamer Conical, 158 mm, Ø 22 mm 1
13 151-613/23 Reamer Conical, 158 mm, Ø 23 mm 1
14 151-613/24 Reamer Conical, 158 mm, Ø 24 mm 1
97
Instruments
151-11/00 Conical Cementless Stem Instruments & Trials, 188 mm – Yellow feet
1 2 3 4 5 6 7 9 11 8 10 12
13 15 17 19 14 16 18
Z
1 151-11/11 Conical Cementless Stem Instruments & Trials, 188 mm Tray – Upper Tray 1
2 151-718/00 Reamer Stopper, Tibia, XS-L 2
3 151-719/00 Reamer Stopper, Femur, XS 2
4 151-719/01 Reamer Stopper, Femur, S 2
5 151-719/02 Reamer Stopper, Femur, M 2
6 151-719/03 Reamer Stopper, Femur, L 2
7 151-160/12 Trial Stem Conical, Cementless, 188 mm, Ø 12 mm 1
8 151-016/13 Trial Stem Conical, Cementless, 188 mm, Ø 13 mm 1
9 151-160/14 Trial Stem Conical, Cementless, 188 mm, Ø 14 mm 1
10 151-016/15 Trial Stem Conical, Cementless, 188 mm, Ø 15 mm 1
11 151-160/16 Trial Stem Conical, Cementless, 188 mm, Ø 16 mm 1
12 151-160/17 Trial Stem Conical, Cementless, 188 mm, Ø 17 mm 1
13 151-160/18 Trial Stem Conical, Cementless, 188 mm, Ø 18 mm 1
14 151-160/19 Trial Stem Conical, Cementless, 188 mm, Ø 19 mm 1
15 151-160/20 Trial Stem Conical, Cementless, 188 mm, Ø 20 mm 1
16 151-160/21 Trial Stem Conical, Cementless, 188 mm, Ø 21 mm 1
17 151-160/22 Trial Stem Conical, Cementless, 188 mm, Ø 22 mm 1
18 151-160/23 Trial Stem Conical, Cementless, 188 mm, Ø 23 mm 1
19 151-160/24 Trial Stem Conical, Cementless, 188 mm, Ø 24 mm 1
98
Instruments
151-11/00 Conical Cementless Stem Instruments & Trials, 188 mm – Yellow feet
1 2 4 6 8 10 3 5 7 9 11
12 14 13
Z
1 151-11/11 Conical Cementless Stem Instruments & Trials, 188 mm Tray – Lower Tray 1
2 151-616/12 Reamer Conical, 188 mm, Ø 12 mm 1
3 151-616/13 Reamer Conical, 188 mm, Ø 13 mm 1
4 151-616/14 Reamer Conical, 188 mm, Ø 14 mm 1
5 151-616/15 Reamer Conical, 188 mm, Ø 15 mm 1
6 151-616/16 Reamer Conical, 188 mm, Ø 16 mm 1
7 151-616/17 Reamer Conical, 188 mm, Ø 17 mm 1
8 151-616/18 Reamer Conical, 188 mm, Ø 18 mm 1
9 151-616/19 Reamer Conical, 188 mm, Ø 19 mm 1
10 151-616/20 Reamer Conical, 188 mm, Ø 20 mm 1
11 151-616/21 Reamer Conical, 188 mm, Ø 21 mm 1
12 151-616/22 Reamer Conical, 188 mm, Ø 22 mm 1
13 151-616/23 Reamer Conical, 188 mm, Ø 23 mm 1
14 151-616/24 Reamer Conical, 188 mm, Ø 24 mm 1
99
Instruments
881-009/00 Patella Instruments & Trials – Blue feet
2 3-5 6-8 9 10-15
1
16
17
19 18
Z
1 881-090/00 LinkSymphoKnee Patella Instruments & Trials Tray 1
2 881-012/00 Universal Drill, Hudson Fitting (B) 1
3 881-501/25 Patella Trial, Ø 25 mm, H = 6 mm 1
4 881-501/28 Patella Trial, Ø 28 mm, H = 6 mm 1
5 881-501/31 Patella Trial, Ø 31 mm, H = 7 mm 1
6 881-501/34 Patella Trial, Ø 34 mm, H = 8 mm 1
7 881-501/37 Patella Trial, Ø 37 mm, H = 9 mm 1
8 881-501/40 Patella Trial, Ø 40 mm, H = 10 mm 1
9 445-904/00 Patella Clamp, Clamp Arm 1
10 881-511/25 Patella Drill Guide, Ø 25 mm 1
11 881-511/28 Patella Drill Guide, Ø 28 mm 1
12 881-511/31 Patella Drill Guide, Ø 31 mm 1
13 881-511/34 Patella Drill Guide, Ø 34 mm 1
14 881-511/37 Patella Drill Guide, Ø 37 mm 1
15 881-511/40 Patella Drill Guide, Ø 40 mm 1
16 881-509/00 Patella Sizing Template 1
17 445-903/00 Patella Clamp, Resection Guide 1
18 445-902/00 Patella Clamp, Handle 1
19 881-500/00 Patella Depth Gauge 1
100
Instruments
Additional Instruments
Hudson Fitting (B)
Standard tool connection.
Adapter for power tool chuck
Different adapters are available to ensure compatibility
to allow various connections:
Attachment
16-3283/01 Jacobs Fitting (E)
16-3284/00 AO Fitting (D)
16-3285/00 Harris Fitting (C)
A
Sawblades,
without offset teeth, 1.24 mm thick
Width (A) Width (A)
Fitting
25 mm 13 mm
317-654/10 317-656/10 Synthes
317-654/11 317-656/11 Aesculap Combi
Zimmer / Hall
317-654/13 317-656/13
Combi
317-654/14 317-656/14 Stryker System 4
X-ray Templates
X-ray Templates for LinkSymphoKnee CCK
881-750/00 To be used with LinkSymphoKnee CCK, 110% actual size
101
Additional Information
For more information please register for our LINK Media Library (linkorthopaedics.com)
102
Indications/Contraindications
Specified Indications and Contraindications: LinkSymphoKnee System
General Indications:
The LinkSymphoKnee System is intended for primary and revision total knee replacement in skeletally mature patients with the
following conditions.
This device is intended for cemented use only unless a cementless modular stem is indicated for use.
Indications:
Primary degenerative arthritis/osteoarthritis
Secondary arthritis resulting from rheumatoid arthritis
Fracture
Contraindications (absolute):
All LinkSymphoKnee Designs
Acute and chronic infections, local and systemic, insofar as they may compromise the successful implantation
Any bone defect that will result in insufficient implant fixation (based on the fact, that using stems, bone grafts and metal bone
substitutes like cones, a minimum bone stock for implant fixation cannot be defined)
Severe insufficiency or loss of extensor mechanism
LinkSymphoKnee Cruciate Retaining Fixed Bearing (CR FB)
Moderate or severe instability or complete loss of the medial or collateral ligament
Instability or loss of the posterior cruciate ligament
LinkSymphoKnee Posterior Stabilized Fixed Bearing and All-Poly (PS FB and PS All-Poly)
Moderate or severe instability or complete loss of the medial or collateral ligament
LinkSymphoKnee Posterior Stabilized Plus Fixed Bearing (PS+ FB)
Severe instability or complete loss of the medial or collateral ligament
LinkSymphoKnee Condylar Constrained Knee Fixed Bearing (CCK FB)
Complete loss of the medial or lateral collateral ligament
Contraindications (relative):
All LinkSymphoKnee Designs
Allergy to one of the implant materials
LinkSymphoKnee Cruciate Retaining Fixed Bearing (CR FB)
Extension deficit >30°
Varus or valgus deformity >30°
LinkSymphoKnee Posterior Stabilized Plus Fixed Bearing (PS+ FB)
Situations in which an overall leg alignment will result, that is outside a range of 5° varus or valgus in reference to the mechanical axis.
In such a situation the PS+ mechanism may fail over time as a result of the shear forces.
LinkSymphoKnee Condylar Constrained Knee Fixed Bearing (CCK FB)
Situations in which an overall leg alignment will result, that is outside a range of 5° varus or valgus in reference to the mechanical axis.
In such a situation the CC mechanism may fail over time as a result of the shear forces
Please note:
These indications/contraindications refer to standard cases. The ultimate decision on whether or not an implant is suitable for a
patient must be made by the surgeon based on his/her individual analysis and his/her experience.
103
Notes
104
Important Information
Please note the following regarding the use of our implants:
1. Choosing the right implant is very important.
The size and shape of the human bone determines the size and shape of the implant and also limits the load capacity.
Implants are not designed to withstand unlimited physical stress. Demands should not exceed normal functional loads.
2. Correct handling of the implant is very important.
Under no circumstances should the shape of a finished implant be altered, as this shortens its life span. Our implants
must not be combined with implants from other manufacturers. The instruments indicated in the Surgical Technique
must be used to ensure safe implantation of the components.
3. Implants must not be reused.
Implants are supplied sterile and are intended for single use only. Used implants must not be used again.
4. After-treatment is also very important.
The patient must be informed of the limitations of the implant. The load capacity of an implant cannot compare with
that of healthy bone!
5. Unless otherwise indicated, implants are supplied in sterile packaging.
Note the following conditions for storage of packaged implants:
• Avoid extreme or sudden changes in temperature.
• Sterile implants in their original, intact protective packaging may be stored in permanent buildings up until the
“Use by” date indicated on the packaging.
• They must not be exposed to frost, dampness or direct sunlight, or mechanical damage.
• Implants may be stored in their original packaging for up to 5 years after the date of manufacture. The “Use by” date
is indicated on the product label.
• Do not use an implant if the packaging is damaged.
6. Traceability is important.
Please use the documentation stickers provided to ensure traceability.
7. Further information on the material composition is available on request from the manufacturer.
Follow the instructions for use!
Waldemar Link GmbH & Co. KG, Hamburg
All content in this catalog, including text, pictures and data, is protected by law. Every instance of use, whether in part or in whole and which
is not permitted by law, is subject to our prior consent. In particular, this applies to the reproduction, editing, translation, publishing, saving,
processing, or passing on of content stored in databases or other electronic media and systems, in any manner or form. The information
in the catalogs is solely intended to describe the products and does not constitute a guarantee.
The Surgical Technique described has been written to the best of our knowledge and belief, but it does not relieve the surgeon of his/her
responsibility to duly consider the particularities of each individual case.
Products shown in this document may not be available in your country. The product availability is subject to the approval and/or registration
regulations of the respective country. Please contact Waldemar Link GmbH & Co. KG if you have questions about the availability of LINK
products in your country.
Waldemar Link GmbH & Co. KG and/or other corporate affiliated entities own, use or have applied for the following trademarks in many
jurisdictions: LINK, BiMobile, SP II, Modell Lubinus, E-Dur, EndoDur, T.O.P. II, BetaCup, CombiCup PF, CombiCup SC, CombiCup R,
MobileLink, C.F.P., LCU, SP-CL, LCP, MIT-H, Endo-Model, Endo-Model SL, MP, MEGASYSTEM-C, GEMINI SL, SPAR-K, LCK, HX, TiCaP,
X-LINKed, PorAg, LINK PorEx, BiPorEx, PorEx-Z, TrabecuLink, Tilastan, customLINK, RescueSleeve, Stactip, VACUCAST.
Other trademarks and trade names may be used in this document to refer to either the entities claiming the marks and/or names or their
products and are the property of their respective owners.
© LINK • 7501_LSK-CCK_OP-Impl-Instr_us_2022-10_003 B MAR-02836 3.0
This publication is intended for distribution in the U.S. only.
Waldemar Link GmbH & Co. KG
Barkhausenweg 10 · 22339 Hamburg · Germany
Phone +49 40 53995-0 ·
[email protected]
www.linkorthopaedics.com
OXINIUM Technology in Knees
THE MOMENT F CHANGE
Changing times,
changing needs
It’s time to change everything
As the number of cases and the age of patients broadens,1-3
orthopaedic applications demand a high performance implant
material that can deliver proven clinical performance in hip and
knee arthroplasty.4-8
But what does real change look like?
At Smith+Nephew, we know you want the best possible
clinical outcomes for your knee replacement patients. In
order to do that, you need to have confidence that the
implant material you use provides both performance
and value.
The problem is there are a range of implant materials
available that may create uncertainty regarding which is
the most suitable to meet the expectations of your patients.
We understand that corrosion, strength, wear and metal
composition can all impact implant survivorship and quality
of life for the patient.9,10 You need an implant material that
can address these concerns through material science.4-8,11-19
OXINIUM◊ Technology
THE MOMENT F CHANGE
Does your knee
Exclusive to Smith+Nephew, OXINIUM◊ Technology is an implant have
award-winning,18 advanced implant material available for
hip and knee arthroplasty. OXINIUM?
With more than 20 years of clinical experience across
120 countries, OXINIUM Technology brings unrivalled
material properties to a portfolio which contains
best-in-class implant designs.11,14,15,19,20
OXINIUM Technology has shown strong clinical and
health economic outcomes,delivering value for
patients, payers and providers. 4,8,21,22
Minimize wear, corrosion and
nickel/cobalt/chromium with
OXINIUM Technology15,20,26
What is OXINIUM◊
Oxidized Zirconium?
• An OXINIUM implant is composed of an award-winning18 oxidized
zirconium (OxZr) alloy. The original metal is transformed through heating
in air at over 500°C to create a 5 μm thick ceramicised oxide.11 Oxidation
process
• The unique manufacturing process means that, in contrast to other
materials used in TJA, the ceramicised metal is not an externally applied
coating. The result is a truly uniform surface transformation that provides
the implant with the durability of metal, the wear of a ceramic bearing,
and fretting/corrosion resistance that is better than both.11-13
• Combined with Smith+Nephew’s implant designs, OXINIUM Technology
provides unique material properties to support excellent clinical and
economic outcomes for patients.21,23,24
Dig deeper into the value
that OXINIUM can provide
Unrivalled Established
material economic
science11-15,19 benefits 21,25
Proven clinical Ideal for
performance8,24 revisions26,27,33
Unrivalled material science
45 million is the number of simulated wear cycles tested in knee simulators without any
measurable loss in oxide thickness – 9 times the industry standard which is estimated to be 30 years of use*15
2x the surface hardness of cobalt chrome, contributing to wear performance 12
Mean (+/- Std. Dev.) oxide thickness of the OXINIUM femoral
components during laboratory wear testing15
4900x more abrasion resistant than cobalt chrome after 10M cycles of pin-on-disc lab testing
using bone cement28 Comparison of metal content between
OXINIUM◊ alloys and CoCr alloys26,29
160x smoother than a cobalt chrome surface following 10M cycles of pin-on-disc testing 28
Virtually zero levels of nickel, cobalt and chrome, 26,29
common metal sensitizers detected in
10-15% of the population 34
OXINIUM◊ may also have a reduced impact on the inflammatory response as demonstrated by lower
pro-inflammatory cytokine expression in cells exposed to OXINIUM versus CoCrMo and Ti alloy†19
Unrivalled material science addresses your concerns regarding wear,
strength and virtually undetectable nickel, cobalt and chrome –
Why wouldn’t you have OXINIUM?
*The LEGION◊ Primary CR Knee System completed 45 million cycles of in vitro simulated wear testing, which is an estimate of 30 years of
activity. Other LEGION VERILAST Primary Knee Systems underwent similar lab testing comparable to industry standards. The results of in
vitro wear simulation testing have not been proven to quantitatively predict clinical wear performance. Also, a reduction in total polyethylene
wear volume or wear rate alone may not result in improved clinical outcomes as wear particle size and morphology are also critical factors in
the evaluation of the potential for wear mediated osteolysis and associated aseptic implant loosening. Particle size and morphology were not *All metal implants contain residual levels of metal elements.
evaluated as part of the testing. May contain<0.01% wt.
†The results of in vitro cytokine expression analyses have not been proven to quantitatively predict clinical cytokine expression.
3B 5B 7B 10B
St
Longer-term evidence Orthopaedic D
13A*
GENESIS II
3A*
Stronger evidence
Total Knee System
Proven clinical performance 3A
CR OXINIUM
Dished PE Insert and Biconvex Patella
13A* 5A* 7A* 3B
13 A*
Longer-term e
GENESIS◊ II JOURNEY UNI LEGION
Total Knee System Knee System Total Knee System
Multiple studies demonstrate excellent 10 year survivorship GENESIS II
Knee ODEP ratings 32
CR OXINIUM OXINIUM PS OXINIUM
Dished PE Insert and Biconvex Patella JOURNEY UNI tibia, JOURNEY UNI Insert and Total Knee System
GENESIS II PS High Flex PE Insert and
• 95.4% (n=44) in patients <50 years old§30
no patella. GENESIS II Round and oval patella. CR OXINIUM
Dished PE Insert and Biconv
• 97.1% (100% survivorship for aseptic loosening; n=303)¶31
13A* 5A*A
15 75A*
13
A* A* 7A*5A* 7A*
• 97.8% (revision aseptic loosening; n=98) ¶8
GENESIS◊ II JOURNEY
GENESIS IIUNI LEGION
GENESIS
JOURNEY ◊
II
II BCS
◊ LEGION
JOURNEY
◊
UNI LEGION
Total Knee System Knee
Total System
Knee System Total KneeKnee
Total
OXINIUM System System TotalKnee
KneeSystem
System Total Knee System
CR OXINIUM OXINIUM
PS OXINIUM PS OXINIUM
CRII OXINIUM
JOURNEY BCS CR OXINIUM
OXINIUM PS OXINIUM
Dished PE Insert and Biconvex Patella JOURNEY UNI IItibia, JOURNEY UNI Insert
II PS and OXINIUM with PE
Dished JOURNEY
Insert tibia,
and JOURNEY
Biconvex II with GENESIS
JOURNEY II Baseplate,
UNI tibia,GENESIS
JOURNEY II PS
UNI Insert and
with GENESIS Baseplate, GENESIS GENESIS II PS High Flex PE Insert and Patella GENESIS II PS High Flex PE
no
PE patella.
High Flex Insert & GENESIS II Round XLPE Insert and JOURNEY Resurfacing Patella. PE Highno
Flex Insert & GENESIS II Round
patella.
GENESIS II Round and oval patella. GENESIS II Round and oval
A*/A Independent assessment by the Orthopaedic Data Evaluation Panel (ODEP)
and Oval Resurfacing Patella and Oval Resurfacing Patella
Latest ODEP ratings can be found at www.odep.org.uk (last accessed on 24th Aug 2020).
◊
15Atop levels of survivorship
shows that OXINIUM knee implants perform at or above 5A*13
up A* 7A*15
5A*A 7
5A*
13 A*
A*
7
5A*
A*
to 15 years post-implantation.32
GENESIS II JOURNEY
GENESIS ◊
II BCS
◊
II LEGION ◊
JOURNEY
GENESIS II UNI LEGION ◊◊ II BCS
JOURNEY LEGION◊
GENESIS II JOURNEY UNI
Total Knee System OXINIUM
Total Knee System TotalTotal
KneeKnee
Knee System
System System Total Knee System
OXINIUM Total Knee System
Total Knee System Knee System
PS OXINIUM CRII BCS
JOURNEY OXINIUM CR OXINIUM
OXINIUM
PS OXINIUM PS OXINIUM
JOURNEY II BCS CR OXINIUM
with GENESIS II Baseplate, GENESIS II PS OXINIUM with JOURNEY
Dished tibia,
PE Insert and JOURNEY
Biconvex II
Patella with GENESIS
JOURNEY
with II Baseplate,
GENESIS GENESIS
UNIIItibia, JOURNEY
Baseplate, II PS
UNI Insert
GENESIS II PS and
CR OXINIUM
OXINIUM with JOURNEY tibia, JOURNEY II OXINIUM
with GENESIS II Baseplate, G
XLPE Insert and JOURNEY Resurfacing Patella. GENESISPEIIInsert
PS High Flex PE Insert and JOURNEY UNI tibia,
PE High Flex Insert & GENESIS II Round PE HighPE
Flex
no Insert
Flex&Insert
patella.
High GENESIS II RoundII Round
& GENESIS Dished
XLPE Insert and Biconvex
and JOURNEY Patella
Resurfacing
GENESIS II Round and oval patella.
Patella. PE High Flex Insert &JOURN
GENES
2X Lower In one analysis of registry data, the use of OXINIUM Technology
and Oval Resurfacing Patella and Oval Resurfacing
and Patella Patella
Oval Resurfacing no patella.
and Oval Resurfacing Patell
Latest ODEP ratings can be found at www.odep.org.uk (last accessed on 24th Aug 2020). Latest ODEP ratings can be found at www.odep.org.uk (last accessed on 24th Aug 2020).
resulted in a revision rate that was two times lower than cobalt chrome in patients 15A 7A*A
younger than 65 years old,23 typically a challenging population for implant survivorship.7 13A*5A* 5A*15 7A* 5A*
GENESIS II JOURNEY◊ II BCS LEGION◊
GENESIS◊ II GENESIS
JOURNEY UNIII System JOURNEY◊ II BCS
LEGION
Total Knee System OXINIUM Total Knee
TotalJOURNEY
Knee System Total
KneeCR
System Knee System TotalOXINIUM
Knee System
Latest Hip and PS OXINIUM
with GENESIS II Baseplate, GENESIS II PS
PE High Flex Insert & GENESIS II Round
CR OXINIUM
XLPE
Dished PE
II BCS
Insert
Insert andand JOURNEY
Biconvex Patella
with
OXINIUM
PS OXINIUM
OXINIUM with JOURNEY tibia, JOURNEY II OXINIUM GENESIS II Baseplate, GENESIS II PS
Resurfacing Patella.JOURNEY
with
PE GENESIS
High
UNI tibia, II Baseplate,
Flex JOURNEY
Insert UNIGENESIS
& GENESIS II PS
II Round
Insert and
PS OXINIUM
JOURNEY II BCS
OXINIUM with JOURNEY tib
Knee ODEP ratings
GENESIS II PSInsert
High Flex
and PE Insert a
OXINIUM Technology delivers proven clinical
PE High
and OvalFlex Insert & GENESIS
Resurfacing Patella II Round XLPE JOURNEY
and Oval Resurfacing Patella no patella.
and Oval Resurfacing Patella GENESIS II Round and oval patella.
Latest ODEP ratings can be found at www.odep.org.uk (last accessed on 24th Aug 2020).
performance in patients across a range of age and
Latest ODEP ratings can be found at www.odep.org.uk (last accessed on 24th Au
Orthopaedic Data Evaluation Panel (ODEP) Rating Chart
15 A
3A* 5A* 7A* 10A* 5A*
13 15A* 7A*
activity expectations emphasizing the importance A*
Stronger evidence
3A II 5A 7A 10AJOURNEY 15A
13A ◊ II BCS
of having OXINIUM. GENESIS
Total Knee System OXINIUM
LEGION◊
Total Knee System
3B
PS OXINIUM 5B 7B 10BJOURNEY
13II BCS
B 15B CR OXINIUM
with GENESIS II Baseplate, GENESIS II PS OXINIUM with JOURNEY tibia, JOURNEY II with GENESIS II Baseplate, GENESIS
PE High Flex Insert & GENESIS II Round XLPE Insert and JOURNEY Resurfacing Patella. PE High Flex Insert & GENESIS II Rou
Longer-term evidence and Oval Resurfacing Patella
and Oval Resurfacing Patella
Latest ODEP ratings can be found at www.odep.org.uk (last accessed on 24th Aug 2020).
§ using OXINIUM PROFIX implants, ¶ using OXINIUM GENESIS II implants
13A*
Established economic benefit
Length of stay Discharge to home Discharge to SNF
JOURNEY◊ II BCS implants JOURNEY II BCS patients were 35% more likely JOURNEY II BCS patients were 41% less likely
with OXINIUM◊ significantly to be discharged to home/home health care to be discharged to a Skilled Nursing Facility
reduced hospital stays (p<0.0001)21 versus other total knee systems (p<0.0001)21
versus other total knee
systems (p<0.0001)21 26% more likely versus TriathlonTM Knee System 28% less likely versus TriathlonTM Knee System
(p=0.003)25 (p=0.0003)25
Readmission
Patients receiving JOURNEY II BCS with OXINIUM were 51% less likely to be readmitted within 30
days compared to other TKA systems (p=0.0037)21
27% less likely to be readmitted within 1-year compared to TriathlonTM Knee System (p=0.0264)25
$1,690 OXINIUM powered implant designs, like
The average total hospital cost
savings comparing JOURNEY
JOURNEY II BCS, allow the patient to
II BCS, powered by OXINIUM, potentially get home faster, return to life
compared to all other TKA with a lower chance of complications and
systems using US CMS/
Medicare data, a significant save money across the episode of care -
saving including the cost of Can you afford not to have OXINIUM?
implant (p<0.0001)21
Ideal for revisions
60% The proportion of patients with poorly performing joint
replacements who show reaction to nickel, cobalt and/or chromium34
Virtually zero levels of nickel, cobalt and chrome 26,29
and
chromium in OXINIUM Oxidized Zirconium
OXINIUM◊ Technology in knee applications has been
shown to minimize metallic artifacts with magnetic resonance imaging to
confirm placement during follow-up compared to cobalt chrome33
Lower 13-year revision rate of LEGION Revision, powered by OXINIUM,
◊
versus the class average for revision TKA in the National Joint Registry for England,
Wales and Northern Ireland‡27 (11.9% vs. 12.7%)
LEGION Revision, powered by OXINIUM, demonstrates significantly fewer
re-revisions due to aseptic loosening of the femur compared to the class
average (p=0.01)‡27
The same unrivalled material science that makes OXINIUM
Technology great for primary procedures can provide benefit
to revisions as well – Why wouldn’t you have OXINIUM?
‡ We thank the patients and staff of all the hospitals in England, Wales and Northern Ireland who have contributed data to the National Joint Registry. We are grateful to the Healthcare Quality Improvement Partnership
(HQIP), the NJR Steering Committee and staff at the NJR Centre for facilitating this work. The views expressed represent those of Smith+Nephew and do not necessarily reflect those of the National Joint Registry
Steering Committee or the Health Quality Improvement Partnership (HQIP) who do not vouch for how the information is presented.
The data used for this analysis was obtained from the National Joint Registry (“NJR”), part of the Healthcare Quality Improvement Partnership (“HQIP”). HQIP, the NJR and/or its contractor, Northgate Public Services (UK)
Limited (“NPS”) take no responsibility (except as prohibited by law) for the accuracy, currency, reliability and correctness of any data used or referred to in this report, nor for the accuracy, currency, reliability and correctness
of links or references to other information sources and disclaims all warranties in relation to such data, links and references to the maximum extent permitted by legislation including any duty of care to third party readers
of the data analysis.
Smith & Nephew, Inc. www.smith-nephew.com ◊
Trademark of Smith+Nephew
1450 Brooks Road All Trademarks acknowledged
Memphis, Tennessee 38116 ©2021 Smith & Nephew, Inc.
USA 30007 V1 08/21
References
1. Klug A, Gramlich Y, Rudert M, et al. The projected volume of primary and revision total knee arthroplasty will place an immense burden on future health care systems over the next 30 years. Knee Surgery, Sports
Traumatology, Arthroscopy. 2020;15:1-12. 2. Sloan M, Premkumar A, Sheth NP. Projected Volume of Primary Total Joint Arthroplasty in the U.S., 2014 to 2030. J Bone Joint Surg Am. 2018;100:1455-1460. 3.
Ackerman IN, Bohensky MA, Zomer E, et al. The projected burden of primary total knee and hip replacement for osteoarthritis in Australia to the year 2030. BMC Musculoskeletal Disorders. 2019;23(1):90. 4.
Davis ET, Pagkalos J, Kopjar B. Bearing surface and survival of cementless and hybrid total hip arthroplasty in the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man. Journal of Bone
Joint Surgery. 2020;5(2):pe0075. 5. Peters RM, Van Steenbergen LN, Stevens M, et al. The effect of bearing type on the outcome of total hip arthroplasty. Acta Orthopaedica. 2018; 89(2):163-169. 6. Atrey
A, Ancarani C, Fitch D, Bordini B. Impact of bearing couple on long-term component survivorship for primary cementless total hip replacement in a large arthroplasty registry. Poster presented at: Canadian
Orthopedic Association; June 20–23, 2018; Victoria, British Columbia, Canada. 7. Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) Hip, Knee & Shoulder Arthroplasty: 2020
Annual Report. 8. Innocenti M, Matassi F, Carulli C, Nistri L, Civinini C. Oxidized zirconium femoral component for TKA: A follow-up note of a previous report at a minimum of 10 years. The Knee. 2014;21:858–
861. 9. Fernandez-Fernandez R, Cruz-Pardos A, Garcia-Rey E. Revision Total Hip Arthroplasty: Epidemiology and Causes. In: Rodríguez-Merchán E. Revision Total Joint Arthoplasty. Springer, 2020. 10. Lewis PL,
Robertsson O, Graves SE, et al. Variation and trends in reasons for knee replacement revision: a multi-registry study of revision burden. Acta Orthopaedica. 2020. DOI: 10.1080/17453674.2020.1853340. 11.
Hunter G, Dickinson J, Herb B, et al. Creation of oxidized zirconium orthopaedic implants. Journal of ASTM International. 2005;2:1-14. 12. Long M, Riester L, Hunter G. Nano-hardness Measurements of Oxidized
Zr-2.5Nb and Various Orthopaedic Materials. Abstract presented at: 24th Annual Meeting of the Society for Biomaterials. April 22-26, 1998, San Diego, California. 13. Parikh A, Hill P, Hines G, Pawar V. Wear of
conventional and highly crosslinked polyethylene liners during simulated fast walking/jogging. Poster presented at: 55th Annual Meeting of the Orthopaedic Research Society, 2009. Poster no. 2340. 14. Parikh A,
Hill P, Pawar V, Sprague J. Long-term Simulator Wear Performance of an Advanced Bearing Technology for THA. Poster presented at: 2013 Annual Meeting of the Orthopaedic Research Society. Poster no. 1028.
15. Papannagari R, Hines G, Sprague J, Morrison M. Long-term wear performance of an advanced bearing technology for TKA. Poster presented at: 2011 Annual Meeting of the Orthopaedic Research Society.
Poster no. 1141. 16. Aldinger P, Williams T, Woodard E. Accelerated Fretting Corrosion Testing of Zirconia Toughened Alumina Composite Ceramic and a New Composition of Ceramicised Metal Femoral Heads.
Poster presented at: 2017 Annual Meeting of the Orthopaedic Research Society. Poster no. 1037. 17. Smith+Nephew 2016. OR-16-127. 18. 2005 ASM International Engineering Materials Achievement Award.
19. Dalal A, Pawar V, McAllister K, Weaver C, Hallab NJ. Orthopedic implant cobalt-alloy particles produce greater toxicity and inflammatory cytokines than titanium alloy and zirconium alloy-based particles in
vitro, in human osteoblasts, fibroblasts, and macrophages. J Biomed Mater Res Part A. 2012;100A:2147-2158. 20. Pawar V, Jones B, Sprague J, Salehi A, Hunter G. Acidic Fretting Tests of Oxidized Zr-2.5Nb, CoCr
and SS Femoral Heads. Paper presented at: Materials & Processes for Medical Devices Conference, ASMI, St. Paul, Minnesota, 2004. 21. Mayman DJ, Patel AR, Carroll KM. Hospital related clinical and economic
outcomes of a bicruciate knee system in total knee arthroplasty patients. Poster presented at: ISPOR Symposium; May 19-23, 2018; Baltimore, Maryland, USA. 22. Patrick C, Delhougne G, Patel A. Retrospective
Analysis of Oxidized Zirconium Bearing Surface in Hip Replacement 90-Day Episode Claims. ISPOR EU Poster, 2019. 23. Civinini R, Carulli C, Matassi F, et al. The Survival of Total Knee Arthroplasty: Current
Data from Registries on Tribology. HSS Journal. 2017;13:28-31. 24. Civinini R, Matassi F, Carulli C, et al. Clinical Results of Oxidized Zirconium Femoral Component in TKA. A Review of Long-Term Survival. HSS
Journal. 2017;13:32-34. 25. Patel AR, Delhougne G. Hospital related clinical and economic outcomes of two premium knee system in total knee arthroplasty patients. Poster presented at: ISPOR Symposium;
May 22, 2019; New Orleans, LA, USA. 26. ASTM International Standard Specification for Wrought Zirconium-2.5 Niobium Alloy for Surgical Implant Applications (UNS R60901) Designation: F 2384 – 10. 27.
National Joint Registry for England, Wales and Northern Ireland: LEGION revision OXINIUM (with revision tibial) implant summary report. 18 May 2022. Available at: https://bit.ly/3w8TEl9. 28. Hunter G, Long M.
Abrasive wear of oxidized Zr-2. 5 Nb, CoCrMo, and Ti-6 Al-4 V against bone cement. Abstract presented at: Sixth World Biomaterials Congress, 2000. 29. ASTM International Standard Specification for Cobalt-28
Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075): Designation: F 75 – 12. 30. Glover AW, Santini AJA, Davidson JS, Pope JA. Mid- to long-term survivorship of
oxidised zirconium total knee replacements performed in patients under 50 years of age. The Knee. 2018;25:617-622. 31. Ahmed I, Salmon LJ, Waller A, et al. Total knee arthroplasty with an oxidized zirconium
femoral component: Ten Year Survivorship Analysis. Bone Joint J. 2016;98-B:58–64. 32. The Orthopaedic Data Evaluation Panel (ODEP). www.odep.org.uk. Accessed April 12th, 2021. 33. Lee KY, Slavinsky JP,
Ries MD, Blumenkrantz G, Majumdar S. Magnetic Resonance Imaging of In Vivo Kinematics After Total Knee Arthroplasty. Journal of Resonance Imaging. 2005;21:172–178. 34. Hallab N, Merritt K, Jacobs JJ.
Metal Sensitivity in Patients with Orthopaedic Implants. JBJS. 2001;83(3):428-436.
Surgical Technique
LEGION TKS
Distal Cut First Preparation with Universal Instruments
LEGION™ Total Knee System
Distal Cut First Preparation
Primary Total Knee Arthroplasty
Contents
Introduction................................................................1
Technique highlights..................................................2
Preoperative planning................................................6
Instrument assembly..................................................7
DCF femoral resection
Intramedullary alignment.....................................9
Distal resection....................................................10
Sizing guide procedure.......................................11
A/P resection.......................................................15
Extramedullary (EM) tibial resection
EM tibial preparation............................................17
Tibial resection.....................................................19
Tibial sizing..........................................................21
Component trialing..............................................23
Posterior stabilized resection.....................................24
Component implantation............................................29
Options/Alternatives
Patella preparation...............................................33
Intramedullary (IM) tibial preparation...................38
Appendix A
Implant size interchangeability............................43
Catalog Information....................................................44
Implant Compatibilities...............................................50
Nota Bene
The following technique is for informational and educational purposes only. It
is not intended to serve as medical advice. It is the responsibility of treating
physicians to determine and utilize the appropriate products and techniques
according to their own clinical judgment for each of their patients. For more
information on the product, including its indications for use, contraindications,
and product safety information, please refer to the product’s label and the
Instructions for Use (IFU).
Additional LEGION Total Knee System surgical technique brochures are available
for the other LEGION components and sets. LEGION instruments listed within
reflect examples in the LEGION Universal Instrument Set.
Introduction
The LEGION™ Total Knee System has been designed
to offer the orthopaedic surgeon solutions to address
intraoperative situations. Implant function is directly related
to accurate surgical technique. The instrumentation can be
used in minimally invasive or standard exposures.
1
Technique highlights
Distal femoral resection
Use the 9.5mm drill to open up the femoral canal
and slide the valgus alignment assembly until at
least one side contacts the distal femur.
After the assembly is placed in neutral rotation,
impact the floating spikes into the distal femur
and secure the distal block with pins.
Remove the IM rod, unlock the lever on
the valgus alignment guide and remove
the valgus alignment assembly using the
universal extractor.
Resect the distal femur.
Position the sizing guide flush against the
distal femur, while ensuring that the posterior
paddles are contacting the underside of both
posterior condyles.
To set rotation
Pin through lateral pivot pinhole, located on the
lower lateral corner of the sizing guide. Adjust
external rotation of the sizing guide by turning
the rotational adjustment knob clockwise (0-6°).
Fixed posterior referencing
Ensure the sizing guide is set in the ‘0’ position.
Drill and insert two pins through the locator
holes of the sizing guide. Determine the size of
the component by the graduations on the stylus.
If the femur is in-between two sizes, choose the
larger size.
2
Adjustable anterior referencing
Position the sizing guide stylus so that it contacts
the lateral ridge of the anterior cortex and
determine the size from the graduations on the
shaft of the stylus.
If the indicated size is in-between two sizes,
turn the upper hex screw clockwise to the next
lower size for anterior referencing or turn the
upper hex screw counterclockwise to the next
larger size. To lock in position, tighten the locking
hex screw. Drill to mark the locator holes for the
A/P cutting block.
Place the correctly sized A/P cutting block on
the distal femur and make anterior, posterior
and chamfer cuts.
3
Tibial resection
Place the extramedullary tibial guide with the
non-spiked (shown) or spiked rod and place on
tibia. Align guide over medial third of the tibial
tubercle and parallel to the tibia.
Attach the tibial stylus to the tibial cutting block
and lower the cutting block until the stylus
touches the low point on the least affected
side of the tibia. Once the resection level is
determined, insert pins to secure and remove
alignment assembly.
Resect the proximal tibia.
Size the tibia.
Posterior stabilized housing resection
Pin trial through anterior flange. Select the
Housing Resection Collet matching the femoral
trial size (either 1-2 or 3-8).
Ream through the collet in the anterior position
until the depth stop makes contact. Remove
the reamer and move the collet to the posterior
position. Ream through the collet once more until
contact is made.
Impact the Housing Box Chisel anteriorly
and then posteriorly through the Housing
Resection Collet to square the corners of the
PS box resection.
Select the appropriate size femoral trial cam
module, insert the arms of the cam module into
the anterior aspect of the femoral trial box and
rotate posteriorly.
4
Final preparation
After trial range of motion and alignment checks,
select the appropriate trial fin punch and punch
through the trial.
Place the femoral implant on the femur and use
the femoral impactor to fully seat the implant.
Apply tibial baseplate cover to protect
tibial baseplate.
Seat the tibial implant with the tibial impactor.
Insert the articular insert by placing the insert
assembly tool into the center notch of the
anterior lock detail (handle up) and engage the
two tabs of the tool into the two recesses on the
anterior periphery of the insert. Squeeze the
tool handle until the insert is fully seated within
the tibial component.
5
Preoperative planning
Determine the angle between the anatomical M V M A
and the mechanical axes. This measurement 6º
will be used intraoperatively to select the
appropriate valgus angle so that correct limb
3º 3º
alignment is restored. (Beware of misleading
angles in knees with a flexion contracture or
rotated lower extremities.)
Tip Many surgeons prefer to simply select a
standard angle for the distal femoral cut
(ie, 5º, 6º or 7º) based on the patient and
surgical experience.
Recommended sawblades
T
1.27 Sawblade (for standard blocks)
Cutting thickness and blade thickness
should be 0.050" or 1.27mm.
3º 3º
1.35 Sawblade (for MIS blocks)
Cutting thickness and blade thickness
should be 1.35mm.
M = Mechanical Axis
A = Anatomical Axis
T = Transverse Axis
V = Vertical Axis
6
Instrument assembly
IM femoral assembly
1 Attach the selected valgus angle bushing (5°, 6°
or 7°) to the valgus alignment guide. Check the
bushing position to make sure that ‘left’ is facing
anteriorly when operating on a left knee and
‘right’ is facing anteriorly when operating
on a right knee.
2 Attach a modular T-handle to the IM rod and
insert through the alignment assembly (Figure 1).
3 Assemble the distal femoral cutting block onto
the valgus alignment guide. Positioning the block
at the ‘primary’ resection level will ensure the
cut will equal the distal thickness of the femoral
prosthesis. Lock by pressing the lever in a
horizontal position toward the medial side.
Figure 1
Extramedullary tibial alignment guide
Insert the ankle clamp into the distal end of the
alignment tube and thread the locking pin into
the ankle clamp (Figure 2).
Gold Knob
After the ankle clamp is moved into the proper
position, lock into place with the gold knob.
Choose the correct left or right tibial cutting block.
Select the spiked or non-spiked fixation rod.
Figure 2
Valgus Bushing Alignment Guide T-handle IM Rod Distal Cutting Ankle Clamp
5° 71440014 71441144 71110080 Long 71512040 Block 71440444
6° 71440016 Short 71512035 71441147
7° 71440018 Alignment Tube
71440448
7
Instrument assembly
Central Knob
Non-spiked fixation rod
Place the appropriate left or right tibial cutting
block on top of the disc on the non-spiked
fixation rod (Figure 3). Tighten the central knob to
lock the block into position.
Introduce the rod into the extramedullary
assembly and adjust and lock the cam in
the assembly.
Spiked fixation rod
Place the spiked fixation rod through the hole
in the tibial cutting guide; adjust the block
and tighten the central knob to lock the block
into position.
Locking Cam
Introduce the spiked fixation rod into the proximal
end of the alignment assembly and adjust and
Figure 4
lock the cam on the assembly (Figure 4).
Figure 3
Tibial Cutting Block Non-spiked Spiked Fixation Rod
Left 71441136 Fixation Rod 71440198
Right 71441137 71440446
8
DCF Femoral resection
Intramedullary alignment
1 Open the femoral canal with a 9.5mm
intramedullary drill (Figure 5).
Tip If desired, the distal femoral cutting block
may be set to resect an additional +2, +5 or
+7mm of bone.
Figure 5
2 Slide the intramedullary rod of the assembly
into the femoral canal until the alignment guide
Locked
contacts the distal femur (Figure 6).
Tip There may be times when only one side of
the guide will touch bone.
3 Orient rotation of the assembly neutral to the
posterior condyles (Figure 7) and impact one or
both of the floating spikes into the distal femur.
Instrument assembly
Figure 6
Figure 7
Valgus Bushing Alignment Guide T-handle IM Rod Distal Cutting IM Drill
5° 71440014 71441144 71110080 Long 71512040 Block 74012111
6° 71440016 Short 71512035 71441147
7° 71440018
9
Distal resection
1 Using non-headed pins, pin the distal femoral
cutting block to the anterior femur using the
holes marked ‘0’. Once adequate distal femoral
resection is noted, an additional headed or
non-headed pin should be placed obliquely
to provide additional stability (Figure 8).
2 Unlock the lever on the valgus alignment
guide, remove the intramedullary rod and the
valgus alignment assembly using the universal
extractor (Figure 9). Only the distal femoral
cutting block should remain on the femur.
Figure 8
3 Resect the distal femur (Figure 10) then remove
the distal femoral cutting block.
Unlocked
Tip If the distal femoral resection is not
adequate, remove the oblique pin, and
reposition the block through the pin holes
marked +2 or +4mm for the desired level of
resection and re-insert the oblique pin.
Figure 9
Figure 10
Valgus Bushing Alignment Guide Universal Extractor IM Rod Distal Cutting
5° 71440014 71441144 71440366 Long 71512040 Block
6° 71440016 Short 71512035 71441147
7° 71440018
10
Instrument assembly
Sizing guide procedure A/P Axis
The sizing guide allows for external rotation to
be set from 0-6º based on surgeon preference
and patient anatomy. Rotational alignment
may be checked by aligning the A/P axis with
the pointer on the sizing guide or by ensuring
that the laser marked lines on the face of the Epicondylar Axis
guide are parallel with the epicondylar axis. The
rotational adjustment knob on the lower portion Rotational Adjustment
Knob
of the guide is turned to dial in rotation (Figure
11).
Figure 11
The guide can be used for fixed posterior
referencing or can be adjusted anteriorly or
posteriorly for fine tuning. When in-between
sizes, the surgeon can choose to adjust sizing
up to 4mm anteriorly, thereby taking up to an
additional 4mm off the posterior condyles, or up
to 2mm posteriorly, taking up to an additional
2mm off of the anterior cortex.
If the anterior surface of the guide is in-between
two sizes when it is at the zero position, the
upper hex screw can be rotated to shift the
Upper Hex Screw
anterior face of the sizing guide up to the next
smaller size or down to the next larger size on Locking Hex Screw
the stylus. As a result, the locator holes for the
A/P cutting block are shifted either anteriorly
or posteriorly to align with the next implant size
(Figure 12).
Tip The gap between the top of the sizing Figure 12 Locator Holes
guide and the stylus graduation line indicates
how much bone will be removed from either the
anterior cortex or posterior condyles by choosing
the next larger size (Figure 13).
Figure 13
Sizing Guide Sizing Stylus Hex Screwdriver
Left 71440007 71441140 115035
Right 71440008
11
Sizing guide procedure: rotation
1 Flex the knee, approximately 90º so the
posterior condyles are accessible.
2 Choose appropriate sizing guide, ‘Left’ for a left
knee and ‘Right’ for a right knee.
3 Position the femoral sizing guide flush against
the distal femur, while ensuring the posterior Lateral Pivot Pin
paddles are contacting the underside of both
posterior condyles. Once correct position of Figure 14
sizing guide is established, place a pin through
lateral pivot pinhole located in the posterior/
lateral corner on the face of the sizing guide
(Figure 14). A/P Axis
4 Adjust the external rotation of the sizing guide
to be aligned anatomically with the epicondylar
and/or A/P axis. This can be achieved by
turning the rotational adjustment knob (0-6º)
using a hex screwdriver (Figure 15).
Epicondylar Axis
5 Once rotation is set, sizing can be established
either by fixed posterior referencing or
adjustable referencing.
Rotational Adjustment Knob
Figure 15
Sizing Guide Sizing Stylus Hex Screwdriver
Left 71440007 71441140 115035
Right 71440008
12
Sizing guide procedure: fixed
posterior referencing
1 Ensure that the anterior surface of the sizing
guide is set in the ‘0’ position.
2 Drill and insert two pins through the locator
holes of the sizing guide to secure the guide.
3 Position the sizing guide stylus so that it contacts
Figure 16
the lateral ridge of the anterior femoral cortex
(highest point on the anterior cortex of the femur)
(Figure 16).
4 Determine the size of the component from the
graduations on the shaft of the stylus.
5 If the femur is between sizes, chose the
larger size.
6 Remove the pins and the sizing guide.
Sizing Guide Sizing Stylus Hex Screwdriver
Left 71440007 71441140 115035
Right 71440008
13
Sizing guide procedure: adjustable
anterior referencing
1 Ensure that the anterior surface of the sizing
guide is set in the ‘0’ position.
Upper Hex Screw
2 Drill and insert two pins through the oblique
holes of the sizing guide to secure the guide. Locking Hex Screw
3 Position the sizing guide stylus so that it
contacts the lateral ridge of the anterior femoral Locator Holes
cortex (highest point on the anterior cortex of
the femur) (Figure 17).
Figure 17
4 Determine the size of the component from the
graduations on the shaft of the stylus.
5 If the indicated size is in-between sizes, you
can turn the upper hex screw to shift the
anterior surface up to an additional 4mm to the
next smaller size or down an additional 2mm to
the next larger size (Figures 18a and b). Once
the appropriate size is selected, turn the locking
hex screw to lock the anterior surface and
locator holes into position (Figure 17).
6 Drill the locator holes to set the position for the
cutting block. Anterior Reference Shift of 4mm
Figure 18a
7 Remove the pins and sizing guide.
Posterior Reference Shift of 2mm
Figure 18b
Sizing Guide Sizing Stylus Hex Screwdriver
Left 71440007 71441140 115035
Right 71440008
14
Instrument assembly
A/P resection
1 Position the fixed spikes on the A/P cutting
block into the predrilled holes.
Tip It is not necessary that the block be
centered M/L on the distal femur.
2 Ensure that the cutting block is flush with the Figure 19
resected distal femur. Several holes in the A/P
block allow fixation of the block. Place one
pin centrally through one of the middle holes
just medial or lateral to the quick-connect
attachment. For additional stability, a headed
pin may be placed through the holes on the
medial or lateral side of the block (Figure 19).
3 Complete the anterior, posterior and chamfer Figure 20
cuts (Figures 20-23). The block is designed
to allow for angling of the sawblade during
the cuts.
Tip To maintain block stability, the anterior
chamfer cut should be completed last.
Figure 21
Figure 22
Figure 23
A/P Cutting Block
sz 5 71441153
15
Downsizing the femoral component
1 Attach the downsizing drill guide to the cut
femur, placing the spikes on the back of the
plate into the same location holes used for the
A/P cutting block (Figure 24).
2 Drill new location holes through the downsizing
drill guide (shifted 2mm anterior).
3 Place the smaller A/P cutting block into the new
location holes. Redo the posterior, anterior and
chamfer cuts.
Tip It is useful to mark the original pin track Figure 24
holes with a marking pen in order to properly
identify the new holes.
Downsizing Drill
Guide
71440860
16
Extramedullary (EM) tibial resection
EM tibial preparation
When using the extramedullary tibial alignment,
the surgeon may use a non-spiked or spiked
fixation rod.
Non-spiked fixation
1 Place the arms of the extramedullary alignment
clamp around the ankle, and adjust the distal
M/L slide directly over the middle of the tibiotalar
joint, which is also approximated by the second
ray of the foot proximal to the malleoli (Figure 25).
The cutting block on the proximal end of the
assembly should be proximal to the tibial
tubercle (Figure 26).
2 Assess rotation of the alignment guide and slope
Figure 25 Figure 26
of the cutting plane. The goal is to align the
extramedullary alignment assembly rotationally
so that it aligns over the medial third of the tibial
tubercle and over the second toe (Figure 27).
3 Rotational alignment is critical due to the 3°
posterior sloped cut. The slope can be adjusted
according to the patient’s anatomy (Figure 28).
Note 3-5° of slope is built into the articular
insert (depending on which insert is chosen) and
3° of slope is built into the tibial cutting block.
A neutral or slightly sloped alignment should
usually be chosen.
Tip Neutral or minimally sloped alignment may
be achieved by palpating the fibula followed by
aligning the alignment guide parallel to the fibula.
Tibial bowing and soft tissue bulk may make
external tibial referencing unreliable.
Figure 27 Figure 28
Ankle Clamp Alignment Tube Tibial Cutting Block Non-spiked
71440444 71440448 Left 71441136 Fixation Rod
Right 71441137 71440446
17
Spiked fixation
1 Place the arms of the extramedullary alignment
clamp around the ankle, and adjust the distal
M/L slide directly over the middle of the
tibiotalar joint, which is also approximated
by the second ray of the foot proximal to the
malleoli (Figure 29).
The cutting block on the proximal end of the
assembly should be proximal to the tibial
Figure 30
tubercle (Figure 30).
2 Impact the longer spike of the spiked fixation
rod into the proximal tibia (Figure 31).
Figure 29
3 Assess rotation of the alignment guide and
slope of the cutting plane. The goal is to
align the extramedullary alignment assembly
rotationally so that it aligns over the medial
third of the tibial tubercle and over the second
toe (Figure 32).
4 Rotational alignment is critical due to the
3° posterior sloped cut. The slope can be
adjusted according to the patient’s anatomy Figure 31
(Figure 33). Impact the second spike to secure
the assembly (Figure 34).
Note 3-5° of slope is built into the articular
insert (depending on which insert is chosen)
and 3° of slope is built into the tibial cutting
block. A neutral or slightly sloped alignment
should usually be chosen.
Figure 32
Tip Neutral or minimally sloped alignment may
be achieved by palpating the fibula followed
by aligning the alignment guide parallel to the
fibula. Tibial bowing and soft tissue bulk may
make external tibial referencing unreliable. Figure 33
Figure 34
Ankle Clamp Alignment Tube Tibial Cutting Block Spiked Fixation Rod
71440444 71440448 Left 71441136 71440198
Right 71441137
18
Tibial resection
1 Attach the tibial stylus to the tibial cutting block
by inserting the stylus foot into the cutting slot.
2 Lower the cutting block until the stylus touches
the low point on the least affected side of the
tibia (Figure 35). The stylus can be adjusted for
a 1-13mm tibial resection by twisting the knob
on top of the stylus. If the affected side of the
tibia is to be used as a reference, the stylus
may be adjusted for a 1-9mm resection level. Figure 35
3 Pin the tibial cutting block to the tibia by
inserting pins first through the central holes;
then the oblique hole.
Tip Pinning through the central holes marked
0mm with smooth pins will allow the block to
be moved +2mm should additional resection
be required (Figure 36).
Tip A 9mm resection is recommended since
9mm of metal and plastic is the thinnest
available component.
Tip To do an extramedullary alignment check,
Figure 36
place the extramedullary alignment rod through
the tibial cutting block.
4 To remove the assembly:
a For the assembly with spiked rod, release the
cam at the top of the alignment tube and use
the universal extractor to remove the spiked
fixation rod (Figure 37) after loosening the
thumbscrew.
b The assembly with the non-spiked rod may
be left in place or removed by loosening the
thumbscrew and lowering the non-spiked rod
to disengage from the tibial cutting block.
Knob (unlocked)
Figure 37
Tibial Stylus Alignment Rod Tibial Cutting Block
71441143 71441148 Left 71441136
Right 71441137
19
5 Cut the tibia by first directing the blade in the
posterior direction and then laterally (Figure 38).
Check Alignment and Balance
1 Assemble the Quick Connect Handle to the
Flexion/Extension block. Attach desired
thickness of Flexion/Extension Spacer onto the
Figure 38
Flexion/Extension block.
2 Insert the Flexion/Extension block into the
extension gap. Evaluate alignment, balance, and
extension space (Figure 39)
3 Adjust the thickness of the spacer as needed to
determine the extension space. Spacer Block
4 Remove the block and reassemble desired
thickness of spacer for flexion gap evaluation.
5 With the knee flexed to 90 degrees, place the
Flexion/Extension Block into the joint space. Figure 39
6 Apply a varus/valgus force and assess the
medial and lateral compartment laxity levels of
the flexion space. Then adjust the thickness of
the spacer as needed to determine the flexion
space.
7 When the flexion space is determined, compare
the thickness selected relative to the extension
space previously determined.
Note: Remember any difference between the
Extension and Flexion Gap Assessments as this
will affect how the femoral implant is positioned
in the steps ahead (e.g. 10mm Ext - 11mm Flex =
-1mm Flex Imbalance).
Universal Extractor Tibial Cutting Block Flexion/Extension Flexion/Extension Quick Connect
71440366 Left 71441136 Block Standard Spacers Handle
Right 71441137 74018603 9mm 74018608 71440044
11mm 74018611
13mm 74018613
15mm 74018615
18mm 74018618
21mm 74018621
20
Tibial sizing
Option A – stemless tibial trials
1 Attach a quick-connect handle to a stemless
trial one size below the femoral component size
and place on the cut tibia to assess coverage
(Figure 40). As needed, additional sizes should
be templated using the stemless trials.
2 Once the appropriate size is determined, pin the
medial side of the selected stemless trial with a
short headed pin.
3 Place a trial insert into the stemless tibial trial
tray and perform a trial range of motion to allow
the baseplate to center on the femoral trial. Quick-connect Handle
(As a secondary check, the surgeon may pass
the alignment rod through the quick-connect
handle to assess alignment) (Figure 41). Pin the Figure 40
lateral side of the trial.
Tip After putting the knee through a trial
range of motion, the surgeon should note the
proper rotation of the trial tibial component on
the proximal tibia and mark the tibia for
future reference.
Tip The center-line marks on the femoral and
tibial trial components should line up. Drop Rod
4 Using the tibial fin/stem punch, rotational
alignment may be set now or at the time of
trial placement.
Tip If the tibial bone is sclerotic, first drill for the
stem using the 11mm tibial drill. Begin the fin
Figure 41
slot with a burr or thin sawblade before using the
fin punch to prevent tibial fracture.
Quick-connect Stemless Trials Fin/Stem Punch 11mm Tibial Drill
Handle Left sz 3-4 71449993 71440040
71440044 sz 4 71430167
Right
sz 4 71430183
21
Option B – stemmed tibial trials
(NOT AVAILABLE IN UNIVERSAL INSTRUMENT
SETS)
1 Place a tibial drill guide one size below the
femoral component size on the cut tibia to
assess coverage. As needed, additional sizes
should be templated (Figure 42).
2 Once the tibial drill guide has been centralized
Figure 42
on the proximal tibia, pin the drill guide in place.
Retract the gold collar on the drill guide handle
and insert the 11mm tibial collet.
3 With the 11mm tibial collet in place, drill with the
11mm tibial drill (Figure 43) and punch with the
11mm tibial punch (Figure 44).
4 Remove the tibial drill guide.
5 Place the stemmed tibial trial into the
prepared hole. Figure 43
6 Using the tibial fin punch, rotational alignment
may be set now or at the time of trial placement.
Tip After putting the knee through a trial
range of motion, the surgeon should note the
proper rotation of the trial tibial component on
the proximal tibia and mark the tibia for
future reference.
Tip The center-line marks on the femoral and
tibial trial components should line up.
Figure 44
11mm Tibial Punch 11mm Tibial Drill Stem Trials Fin/Stem Punch
71440262 71440040 Left sz 3-4 71440266
sz 4 71430166
Right
sz 4 71430182
22
Component trialing
Femoral component trialing
1 Flex the knee to 90° and insert the femoral trial
using the femoral trial impactor (Figure 45).
2 Perform a trial range of motion to assess patellar
tracking. With cruciate-retaining knees, medial/ Figure 45
lateral placement of the femoral trial can be
adjusted to optimize patellar tracking
(Figure 46).
3 For cruciate-retaining femorals, prepare the
femoral lug holes through the femoral trial with
the femoral lug punch (Figure 47).
Note: This step is also required for cemented
posterior-stabilizing femorals using femoral lugs
or Flex-Lok pegs. Figure 46
Note: It is recommended to leave the femoral
trial on at this point to perform full component
trialing. However, if desired, you may attach the
end of the universal extractor to the femoral trial
and remove the femoral trial (Figure 48).
Lug Punch
Figure 47
Extractor
Figure 48
Femoral Trial Femoral Trial Lug Punch Universal Extractor
Left Impactor 71440183 71440366
sz 5 71433345 71440009
Right
sz 5 71433355
23
Posterior stabilized resection
Femoral housing box resection
1 Pin the femoral trial through the anterior flange
(Figure 49).
2 Choose the housing resection Collet matching
the femoral trial size (either 1-2 or 3-8). Attach
the Collet to the femoral trial by sliding the
Housing Collet (anterior to posterior) into the
slots on the distal face of the femoral trial and
threading the two posts into the femoral trial. The
Housing Collet should be secured in the anterior
Figure 49
position first and then shifted to the posterior
position
and screwed (Figure 50).
Housing Collet
Figure 50
Femoral Trial Housing Collet
Left Resection
sz 5 71433345 sz 3-8 71434424
Right
sz 5 71433355
24
3 Attach the Housing Reamer Dome and the PS
Reamer Sleeve to the Patellar Reamer Shaft
(Figure 51).
Ream through the housing resection Collet in
both the anterior and posterior positions until
the depth stop contacts the Collet (Figure 52).
4 Impact the Housing Box Chisel through the
Housing Resection Collet to square the corners
of the housing. The Housing Box Chisel
should be used anteriorly and then posteriorly
to ensure that the full length of the box is
Housing Reamer Dome
prepared (Figure 53).
Figure 51
Figure 52
Figure 53
Housing Reamer Housing Box Chisel Patellar Reamer Reamer Sleeve
Dome 71440373 Shaft 71440464
71440145 71440324
25
Femoral trial cam module assembly
1 Select the appropriate sized femoral trial
cam module (matching the femoral trial
size selected).
2 Insert the arms of the femoral cam module
into the anterior aspect of the femoral trial box
and rotate posteriorly until seated (Figures 54
and 55).
Figure 54
Figure 55
Cam Modules Femoral Trial
sz 5 71433365 Left
sz 5 71433345
Right
sz 5 71433355
26
Tibial component trialing
1 Use the appropriate insert trial (begin with a
9mm trial) to determine stability and alignment.
2 Perform a trial range of motion. The alignment
marks on the front of the femoral and tibial
trials should line up (Figure 56). The quick-
connect handle may be attached to the tibial
trial and used to set the appropriate rotational
alignment.
Optional Extend the knee fully with the
Figure 56
handle attached to the tibial trial. Pass the
extramedullary rod through the handle to
assess full leg alignment (Figure 57).
Figure 57
PS High Flex Insert PS Insert Trials Constrained Insert Insert Trial Spacers
Trials 9mm Trials 13mm
9mm sz 3-4 71430815 9mm sz 3-4 74033634
sz 3-4 71430408 11mm sz 3-4 71440473 15mm
10mm sz 3-4 71430817 11mm sz 3-4 74033635
sz 3-4 71433453 sz 3-4 71430524 18mm
11mm CR Insert Trials Deep Dished sz 3-4 74033636
sz 3-4 71430409 9mm Insert Trials 21mm
12mm sz 3-4 71430490 9mm sz 3-4 74033637
sz 3-4 71433454 11mm sz 3-4 71430766
sz 3-4 71430492 11mm
sz 3-4 71430768
CR Deep Flex
Insert Trials
9mm
sz 3-4 71430444
10mm
sz 3-4 71433443
11mm
sz 3-4 71430445
12mm Note: 10 and 12mm
sz 3-4 71433444 Insert Trials are an
Add-On Set
27
3 Mark correct tibial rotational alignment on the
anterior tibia using a cautery knife (Figure 58).
4 If not previously performed, select the
appropriate tibial fin punch to prepare the fins
and punch through the tibial trial (Figure 59).
Tip If the tibial bone is sclerotic, begin the fin
slot with a burr or thin sawblade before using
the fin punch to prevent tibial fracture.
Figure 58
6 Remove the femoral and tibial trials.
Figure 59
28
Component implantation
Femoral implantation
Cemented
1 Mix and prepare bone cement for femoral
component and distal femur. Apply to the femoral
component or prepared bone, based on the
surgeon’s preference.
Tip Many surgeons put cement on the bone
rather than, or supplemental to, cement on the
underside of the implant.
Note: If using femoral lugs or Flex-Lok pegs with
cemented posterior-stabilizing femorals, screw
those components into the femoral prior to
cement application.
2 Place the femoral implant onto the femur and
use the femoral impactor to fully seat the implant Figure 60
(Figure 60).
3 Remove excess cement. Extend the knee to
remove cement anteriorly without retracting the
proximal soft tissue.
Tip After tibial implant is implanted, place the tibial
insert trial onto the tibial implant and extend the
leg to pressurize the cement.
Tip Place the CR tibial trial in the tibial implant
tray to assist with aligning the femoral component
during implantation.
Porous
1 Ensure that flat, clean cuts are made to all of the
femoral resection cuts. This will help to achieve an
optimal press-fit.
Tip Lavaging during resection helps ensure flat,
clean cuts.
2 Place the femoral implant onto the femur and use
the femoral impactor to fully seat the implant.
Tip If extraction of the femoral component is
needed, attach the locking impactor and move in
side-to-side motions to leverage off, then adjust
and reimpact.
OXINIUM™ CR Femoral Impactor PS Femoral Impactor
Impactor (Universal 71440190 71440005
Impactor)
71440890
29
Tibial implantation
Apply cement on the proximal tibia and/or the
implant, tibial insert covers on the articular surface
of the baseplate, and seat the tibial implant with the
tibial impactor (Figure 61). Remove excess cement.
Recommended Insert Placement Method
1 Clear any debris from the locking mechanism and
manually slide the insert into the tibial baseplate
engaging the locking mechanism until the insert
periphery is within 1-2mm of the Tibial Figure 61
Component periphery.
2 Insert the tip of the JOURNEY™ Articular Insert Tool
into the center notch of the anterior lock detail
(handle up) and engage the two tabs of the Tool
into the two recesses on the anterior periphery of
the insert (Figure 62).
Note: Make sure the tool is level with the plane of
the baseplate.
3 Squeeze the tool handle until the insert is fully
seated within the Tibial Component. The insert Figure 62
should not move under any pressure in flexion
or extension.
Cruciate-retaining, dished and posterior
stabilized insert placement
1 Determine the correct articular insert thickness.
2 Clear any debris from the locking mechanism
and manually slide the insert into the tibial
baseplate engaging the locking mechanism.
For the PS insert, begin insertion in flexion
and extend the leg to engage the locking
mechanism.
3 Attach the articular inserter/extractor to the
tibial tray. Lift the inserter superiorly until the
anterior lip of the articular insert is fully seated
(Figure 63). Figure 63
Tibial Base Articular Inserter/ JOURNEY Articular
Impactor Extractor Inserter Tool
74018901 71440194 74018911
30
PS and CR high flex insert placement
1 Attach the appropriately sized impactor head
(either 1-2 or 3-8) to the impactor handle.
2 Position the knee in approximately 90° flexion.
3 Align the articular insert with the locking
Figure 64
mechanism of the tibial baseplate.
4 Push the insert posteriorly until the top of the
anterior rail of the baseplate is visible.
5 Place the impactor head on the anterior chamfer
of the insert. The mating surfaces should be very
conforming (Figures 64 and 65).
6 Impact the handle until the insert is fully seated.
Figure 65
High-flex Impactor High-flex Impactor
Head Handle
sz 1-2 71441553 71441552
sz 3-8 71441554
31
MIS PS high flex insert placement
When using the PS high flex insert in a minimally
invasive procedure, the femoral cam mechanism
is likely to prevent the insert from fully seating
into the locking mechanism while the knee is
in flexion. To use the PS high flex insert in a
MIS case:
1 Flex the knee to 90° and push the insert as far
Figure 66
back as it will go posteriorly with the knee in
flexion (Figure 66).
Tip Lift the distal femur to prevent scratching of
the posterior condyle of the component.
2 Placing your thumb on the anterior of the insert
to hold it on the baseplate (Figure 67), move the
knee into extension.
3 Use the impactor handle with the appropriately Figure 67
sized impactor head to fully seat the insert
and engage the anterior portion of the dovetail
locking mechanism (Figure 68).
Figure 68
32
Options/Alternatives
Patella preparation
Locking Screw
Instrument assembly
Patellar Reamer Guide
Determine the appropriate diameter patellar
implant, and select the correctly-sized Patellar
Reamer Collet and slide it into place on the
Reamer Collet
Patellar Reamer Guide (Figure 69).
Depth gauge and reamer assembly
1 Attach the appropriate patellar depth gauge Figure 80
Figure 69 Reamer Guide
(red = resurfacing, black = large resurfacing/
round) to the Reamer Guide (Figure 70).
2 Attach the matching sized patellar reamer dome
and patellar depth stop to the patellar reamer
shaft (Figures 71 and 72). Lower the assembly
through the patellar Reamer Guide until the
reamer dome contacts the patella.
Figure 70
Reamer Dome
Figure 71 Figure 72
Reamer Collet Resurfacing Depth Patellar Depth Reamer Shaft
26mm 71440512 Gauge Stop 71440324
29mm 71440514 71440330 71440326 Reamer Dome
32mm 71440516 26mm 71440348
35mm 71440518 Large Resurfacing/ Large Resurfacing/ 29mm 71440342
Oval Depth Gauge Oval Depth Stop 32mm 71440344
Reamer Guide 71440431 71440427 35mm 71440346
71440311
33
Resurfacing patellar preparation
The surgeon can choose from a freehand
cutting technique with towel clips, or if
desired, they can choose one of the
following instrumented techniques.
Reaming technique
The objective of this technique is to resurface the
articular surface of the patella with the precision
of a reaming technique. The reamed patellar
surface can accommodate an oval or round
resurfacing patellar component.
1 Trim tissue surrounding the patella using
electrocautery (bovie) (Figure 73).
Figure 73
2 Use a rongeur to remove osteophytes and
reduce the patella to its true size (Figure 74). It is
recommended to leave the superior rim of bone
intact. The bovie should also be used to release
soft tissue attachments to the estimated level of
resection.
3 Place the collet over the patella so that it fits
snugly around the patellar diameter (Figure 75).
The goal is to reduce the patella to its smallest Figure 74
diameter so that the smallest possible collet
will fit around the entire patella. Use the patellar
reamer collet as a sizing template to select the
appropriately sized collet and reamer.
Tip The collet should be resting on the soft
tissue surrounding the patella. If the patella does
not enter the collet evenly but instead enters
at an angle, the collet may not be completely
surrounding the patella, but instead resting on
part of the bone. If the collet is only slightly smaller
than the patella, you may trim 1-2mm of the
medial and lateral edges of the patella to ensure a Figure 75
snug fit. If the collet is far smaller than the patella,
choose the next size up and assess fit.
Resurfacing Drill Drill Patellar Calipers Reamer Collet
Guide 71440360 11-4943 26mm 71440512
26mm 71440402 29mm 71440514
29mm 71440403 32mm 71440516
32mm 71440404 35mm 71440518
35mm 71440405
34
4 Measure patella thickness with the patellar
calipers (Figure 76).
Tip The patella should measure a minimum of
19mm before reaming to use this resurfacing
Collet
technique.
Determine the design and diameter of the
patellar implant to be used. The round
resurfacing patella is 9mm thick, and the Figure 76
depth stop for this technique prepares for
9mm of resection.
Tip Minor adjustments may be necessary at the
time of resection to accommodate the largest Depth Stop
diameter oval patellar implants. (Please see chart
on page 44).
5 Rotate the appropriate resurfacing
patellar depth gauge (red = round) Depth Gauge
around so that the hooked end or ‘claw’
surrounds the patellar reamer shaft (Figure 77).
Lower the depth stop by compressing the button
until it meets the depth gauge (Figure 78).
Remove the depth gauge from the assembly. Clamp
Ream the patella until the depth stop engages
Figure 77
the patellar Reamer Guide (Figure 79).
Tip Excessive force on the reamer shaft may alter
the depth of resection, causing overreaming.
Figure 78 Figure 79
Resurfacing Depth Reamer Dome Reamer Guide
Stop 26mm 71440348 71440311
71440326 29mm 71440342
32mm 71440344 Resurfacing Depth
Reamer Shaft 35mm 71440346 Gauge
71440324 71440330
35
6 After reaming, the patella should have a
completely flat articular surface (Figure 80).
Measure the resected patella to ensure adequate
resection (the resected patella should measure
its original depth minus 9mm).
7 Drill the appropriate fixation holes for the
Target Depth
resurfacing patellar implant using the correctly
Original Depth - 9mm
sized drill guide and resurfacing drill (Figure 81).
8 Place the patellar trial into the prepared patella.
Collet
If desired, use the calipers to remeasure the
composite thickness of bone and trial (Figure 82). Figure 80
Drill Guide
Drill
Figure 81
Figure 82
Resurfacing Drill Drill Patellar Calipers
Guide 71440360 114943
26mm 71440402
29mm 71440403
32mm 71440404
35mm 71440405
36
Component trialing
1 Place the patellar trial into the prepared
patella (Figure 83).
2 Perform a trial range of motion to assess
patellar tracking. Medial/lateral placement of
the femoral trial can be adjusted to optimize
patellar tracking (Figure 84).
Figure 83
Figure 84
Patellar implantation
1 Assemble the patellar cement clamp to the Clamp
patellar Reamer Guide.
2 Apply bone cement to the patella.
3 Place the patellar implant onto the patella
and clamp into the bone (Figure 85). Remove
excess cement.
Figure 85
Resurfacing Reamer Guide Cement Clamp
Patellar Trial 71440311 71440322
26mm 71430580
29mm 71430574
32mm 71430576
35mm 71430578
37
Intramedullary (IM) tibial preparation
Technique highlights
Place the IM alignment assembly on the tibia.
The alignment rod should align with the medial
third of the tibial tubercle. Impact assembly.
Attach the tibial stylus to the tibial cutting block
and lower the cutting block until the stylus
touches the low point on the least affected
side of the tibia. Once the resection level is
determined, insert pins to secure and remove
alignment assembly.
Resect the proximal tibia.
Size the tibia.
Final preparation
After trial range of motion and alignment checks,
select the appropriate trial fin punch and punch
through the trial.
Seat the tibial implant with the tibial impactor.
Insert the articular insert by placing the insert
assembly tool into the center notch of the
anterior lock detail (handle up) and engage the
two tabs of the tool into the two recesses on the
anterior periphery of the insert. Squeeze the tool
handle until the insert is fully seated within the
tibial component.
38
Instrument assembly
Intramedullary Tibial Alignment Guide
1 Insert the external rod of the Intramedullary
Tibial Alignment Guide through the hole on the
correct left or right Tibial Cutting Block and lock Alignment Sleeve
the cam (Figure 86).
2 Attach the T-handle to the IM Rod and pass it
through the cannulated alignment sleeve on
the alignment assembly (Figure 87).
Alignment Rod
Figure 86
Figure 87
T-handle Tibial Cutting Block IM Alignment IM Rod
71110080 Left 71441136 Guide Short 71512035
Right 71441137 71440200 Long 71512040
39
IM tibial preparation
1 Make a 9.5mm pilot hole into the tibial canal
(Figure 88). A preliminary resection of the tibial
spine may facilitate seating of the tibial drill guide
onto the proximal tibia.
2 Slowly insert the IM rod into the tibial canal.
3 Assess rotation of the intramedullary tibial Figure 88
alignment guide. Rotational alignment is critical
due to the 3º posterior sloped cut. The
alignment rod of the intramedullary tibial
alignment assembly should align with the
medial third of the tibial tubercle (Figure 89).
4 Impact the proximal end of the cannulated
alignment sleeve to drive the distal spikes into
the proximal tibia to lock rotational alignment
(Figure 90).
Figure 89
Figure 90
Tibial Cutting Block IM Alignment IM Rod
IM Drill
Left 71441136 Guide Short 71512035
74012111
Right 71441137 71440200 Long 71512040
40
Tibial resection
1 Attach the Tibial Stylus to the Tibial Cutting
Block by inserting the stylus foot into the
cutting slot.
2 Lower the cutting block until the stylus touches
the low point on the least affected side of the
tibia (Figure 91). The stylus can be adjusted for
a 1-13mm tibial resection by twisting the knob
on top of the stylus. If the affected side of the
tibia is to be used as a reference, the stylus
may be adjusted for a 1-9mm resection level.
Figure 91
3 Pin the Tibial Cutting Block to the tibia by
inserting pins first through the central holes;
then the oblique hole.
Tip Pinning through the central holes marked
0mm with smooth pins will allow the block to
be moved +2mm should additional resection
be required (Figure 92).
Tip A 9mm resection is recommended since
9mm of metal and plastic is the thinnest
available component.
Figure 92
Tip To do an extramedullary alignment check,
place the extramedullary Alignment Rod
through the tibial cutting block.
Tibial Stylus Tibial Cutting Block Alignment Rod
71441143 Left 71441136 71441148
Right 71441137
41
4 To remove the assembly:
Use the universal extractor leaving the cutting
block on the anterior tibia (Figure 93) after
loosening the thumbscrew.
5 Cut the tibia by first directing the blade in
the posterior direction and then laterally
(Figure 94).
6 Check alignment and balance with spacer
block and rod. Balance ligaments in
standard fashion.
Figure 93
Figure 94
Universal Extractor Flexion / Extension Quick Connect Alignment Rod
(Slap Hammer) Spacers Handle 71441148
71440366 9mm 74018608 71440044
11mm 74018611
Flexion / Extension 13mm 74018613 Tibial Cutting Block
Block Standard 15mm 74018615 Left 71441136
74018603 18mm 74018618 Right 71441137
21mm 74018621
42
Appendix A
Implant size interchangeability
Femoral Size
Insert Size 1 2 3 4 5 6 7 8
1-2 PS, DD, Con
1-2 PSHF, CRHF
3-4 PS, DD, Con
3-4 PSHF, CRHF
5-6 PS, DD, Con
5-6 PSHF, CRHF
7-8 PS, DD, Con
7-8 PSHF, CRHF
PS – Standard posterior stabilized
DD – Deep dish
Con – Constrained
PSHF – Posterior stabilized high flex
CRHF – Cruciate retaining high flex
Note: CR – Standard cruciate retaining inserts are interchangeable with all size femoral components
43
Catalog Information
LEGION™ PS OXINIUM™ Femorals LEGION Narrow CR OXINIUM Femorals
Cat. No. Description Cat. No. Description
71421202 LEGION PS OXINIUM Femoral Size 2 Right 71421243 LEGION Narrow CR OXINIUM Size 3N Left
71421203 LEGION PS OXINIUM Femoral Size 3 Right 71421244 LEGION Narrow CR OXINIUM Size 4N Left
71421204 LEGION PS OXINIUM Femoral Size 4 Right 71421245 LEGION Narrow CR OXINIUM Size 5N Left
71421205 LEGION PS OXINIUM Femoral Size 5 Right 71421246 LEGION Narrow CR OXINIUM Size 6N Left
71421206 LEGION PS OXINIUM Femoral Size 6 Right 71421253 LEGION Narrow CR OXINIUM Size 3N Right
71421207 LEGION PS OXINIUM Femoral Size 7 Right 71421254 LEGION Narrow CR OXINIUM Size 4N Right
71421208 LEGION PS OXINIUM Femoral Size 8 Right 71421255 LEGION Narrow CR OXINIUM Size 5N Right
71421212 LEGION PS OXINIUM Femoral Size 2 Left 71421256 LEGION Narrow CR OXINIUM Size 6N Right
71421213 LEGION PS OXINIUM Femoral Size 3 Left
71421214 LEGION PS OXINIUM Femoral Size 4 Left
LEGION CR Non-Porous Femorals
71421215 LEGION PS OXINIUM Femoral Size 5 Left
Cat. No. Description
71421216 LEGION PS OXINIUM Femoral Size 6 Left
71423202 LEGION CR Non-Porous Femoral Size 2 Left
71421217 LEGION PS OXINIUM Femoral Size 7 Left
71423203 LEGION CR Non-Porous Femoral Size 3 Left
71421218 LEGION PS OXINIUM Femoral Size 8 Left
71423204 LEGION CR Non-Porous Femoral Size 4 Left
71423205 LEGION CR Non-Porous Femoral Size 5 Left
LEGION CR OXINIUM Femorals 71423206 LEGION CR Non-Porous Femoral Size 6 Left
Cat. No. Description
71423207 LEGION CR Non-Porous Femoral Size 7 Left
71421222 LEGION CR OXINIUM Femoral Size 2 Right
71423208 LEGION CR Non-Porous Femoral Size 8 Left
71421223 LEGION CR OXINIUM Femoral Size 3 Right
71423212 LEGION CR Non-Porous Femoral Size 2 Right
71421224 LEGION CR OXINIUM Femoral Size 4 Right
71423213 LEGION CR Non-Porous Femoral Size 3 Right
71421225 LEGION CR OXINIUM Femoral Size 5 Right
71423214 LEGION CR Non-Porous Femoral Size 4 Right
71421226 LEGION CR OXINIUM Femoral Size 6 Right
71423215 LEGION CR Non-Porous Femoral Size 5 Right
71421227 LEGION CR OXINIUM Femoral Size 7 Right
71423216 LEGION CR Non-Porous Femoral Size 6 Right
71421228 LEGION CR OXINIUM Femoral Size 8 Right
71423217 LEGION CR Non-Porous Femoral Size 7 Right
71421232 LEGION CR OXINIUM Femoral Size 2 Left
71423218 LEGION CR Non-Porous Femoral Size 8 Right
71421233 LEGION CR OXINIUM Femoral Size 3 Left
71421234 LEGION CR OXINIUM Femoral Size 4 Left
71421235 LEGION CR OXINIUM Femoral Size 5 Left
LEGION PS Non-Porous Femorals
71421236 LEGION CR OXINIUM Femoral Size 6 Left
Cat. No. Description
71421237 LEGION CR OXINIUM Femoral Size 7 Left
71423222 LEGION PS Non-Porous Femoral Size 2 Left
71421238 LEGION CR OXINIUM Femoral Size 8 Left
71423223 LEGION PS Non-Porous Femoral Size 3 Left
71423224 LEGION PS Non-Porous Femoral Size 4 Left
71423225 LEGION PS Non-Porous Femoral Size 5 Left
LEGION Narrow PS OXINIUM Femorals 71423226 LEGION PS Non-Porous Femoral Size 6 Left
Cat. No. Description
71423227 LEGION PS Non-Porous Femoral Size 7 Left
71421263 LEGION Narrow PS OXINIUM Size 3N Left
71423228 LEGION PS Non-Porous Femoral Size 8 Left
71421264 LEGION Narrow PS OXINIUM Size 4N Left
71423232 LEGION PS Non-Porous Femoral Size 2 Right
71421265 LEGION Narrow PS OXINIUM Size 5N Left
71423233 LEGION PS Non-Porous Femoral Size 3 Right
71421266 LEGION Narrow PS OXINIUM Size 6N Left
71423234 LEGION PS Non-Porous Femoral Size 4 Right
71421273 LEGION Narrow PS OXINIUM Size 3N Right
71423235 LEGION PS Non-Porous Femoral Size 5 Right
71421274 LEGION Narrow PS OXINIUM Size 4N Right
71423236 LEGION PS Non-Porous Femoral Size 6 Right
71421275 LEGION Narrow PS OXINIUM Size 5N Right
71423237 LEGION PS Non-Porous Femoral Size 7 Right
71421276 LEGION Narrow PS OXINIUM Size 6N Right
71423238 LEGION PS Non-Porous Femoral Size 8 Right
44
LEGION™ Porous CR Femorals LEGION CR Non-Porous Narrow Femorals
Cat. No. Description Cat. No. Description
71423242 LEGION Porous CR Femoral Size 2 Left 71933640 LEGION CR Non-Porous Narrow Femoral Size 3 Left
71423243 LEGION Porous CR Femoral Size 3 Left 71933641 LEGION CR Non-Porous Narrow Femoral Size 4 Left
71423244 LEGION Porous CR Femoral Size 4 Left 71933642 LEGION CR Non-Porous Narrow Femoral Size 5 Left
71423245 LEGION Porous CR Femoral Size 5 Left 71933643 LEGION CR Non-Porous Narrow Femoral Size 6 Left
71423246 LEGION Porous CR Femoral Size 6 Left 71933644 LEGION CR Non-Porous Narrow Femoral Size 3 Right
71423247 LEGION Porous CR Femoral Size 7 Left 71933645 LEGION CR Non-Porous Narrow Femoral Size 4 Right
71423248 LEGION Porous CR Femoral Size 8 Left 71933646 LEGION CR Non-Porous Narrow Femoral Size 5 Right
71423252 LEGION Porous CR Femoral Size 2 Right 71933647 LEGION CR Non-Porous Narrow Femoral Size 6 Right
71423253 LEGION Porous CR Femoral Size 3 Right
71423254 LEGION Porous CR Femoral Size 4 Right
71423255 LEGION Porous CR Femoral Size 5 Right
71423256 LEGION Porous CR Femoral Size 6 Right
LEGION PS Non-Porous Narrow Femorals
Cat. No. Description
71423257 LEGION Porous CR Femoral Size 7 Right
71933648 LEGION PS Non-Porous Narrow Femoral Size 3 Left
71423258 LEGION Porous CR Femoral Size 8 Right
71933649 LEGION PS Non-Porous Narrow Femoral Size 4 Left
71933650 LEGION PS Non-Porous Narrow Femoral Size 5 Left
71933651 LEGION PS Non-Porous Narrow Femoral Size 6 Left
LEGION Porous CR HA Femorals 71933652 LEGION PS Non-Porous Narrow Femoral Size 3 Right
Cat. No. Description 71933653 LEGION PS Non-Porous Narrow Femoral Size 4 Right
71425202 LEGION Porous CR HA Femoral Size 2 Left 71933654 LEGION PS Non-Porous Narrow Femoral Size 5 Right
71425203 LEGION Porous CR HA Femoral Size 3 Left 71933655 LEGION PS Non-Porous Narrow Femoral Size 6 Right
71425204 LEGION Porous CR HA Femoral Size 4 Left
71425205 LEGION Porous CR HA Femoral Size 5 Left
71425206 LEGION Porous CR HA Femoral Size 6 Left
71425207 LEGION Porous CR HA Femoral Size 7 Left
71425208 LEGION Porous CR HA Femoral Size 8 Left
71425212 LEGION Porous CR HA Femoral Size 2 Right
71425213 LEGION Porous CR HA Femoral Size 3 Right
71425214 LEGION Porous CR HA Femoral Size 4 Right
71425215 LEGION Porous CR HA Femoral Size 5 Right
71425216 LEGION Porous CR HA Femoral Size 6 Right
71425217 LEGION Porous CR HA Femoral Size 7 Right
71425218 LEGION Porous CR HA Femoral Size 8 Right
45
LEGION™ CR High Flex XLPE Inserts LEGION PS High Flex XLPE Inserts
Cat. No. Description Cat. No. Description
71453101 LEGION CR High Flex XLPE Inserts Size 1-2 9mm 71453171 LEGION PS High Flex XLPE Inserts Size 1-2 10mm
71453102 LEGION CR High Flex XLPE Inserts Size 1-2 11mm 71453172 LEGION PS High Flex XLPE Inserts Size 1-2 12mm
71453103 LEGION CR High Flex XLPE Inserts Size 1-2 13mm 71453173 LEGION PS High Flex XLPE Inserts Size 3-4 10mm
71453104 LEGION CR High Flex XLPE Inserts Size 1-2 15mm 71453174 LEGION PS High Flex XLPE Inserts Size 3-4 12mm
71453105 LEGION CR High Flex XLPE Inserts Size 1-2 18mm 71453175 LEGION PS High Flex XLPE Inserts Size 5-6 10mm
71453111 LEGION CR High Flex XLPE Inserts Size 3-4 9mm 71453176 LEGION PS High Flex XLPE Inserts Size 5-6 12mm
71453112 LEGION CR High Flex XLPE Inserts Size 3-4 11mm 71453177 LEGION PS High Flex XLPE Inserts Size 7-8 10mm
71453113 LEGION CR High Flex XLPE Inserts Size 3-4 13mm 71453178 LEGION PS High Flex XLPE Inserts Size 7-8 12mm
71453114 LEGION CR High Flex XLPE Inserts Size 3-4 15mm 71453201 LEGION PS High Flex XLPE Inserts Size 1-2 9mm
71453115 LEGION CR High Flex XLPE Inserts Size 3-4 18mm 71453202 LEGION PS High Flex XLPE Inserts Size 1-2 11mm
71453121 LEGION CR High Flex XLPE Inserts Size 5-6 9mm 71453203 LEGION PS High Flex XLPE Inserts Size 1-2 13mm
71453122 LEGION CR High Flex XLPE Inserts Size 5-6 11mm 71453204 LEGION PS High Flex XLPE Inserts Size 1-2 15mm
71453123 LEGION CR High Flex XLPE Inserts Size 5-6 13mm 71453205 LEGION PS High Flex XLPE Inserts Size 1-2 18mm
71453124 LEGION CR High Flex XLPE Inserts Size 5-6 15mm 71453206 LEGION PS High Flex XLPE Inserts Size 1-2 21mm
71453125 LEGION CR High Flex XLPE Inserts Size 5-6 18mm 71453211 LEGION PS High Flex XLPE Inserts Size 3-4 9mm
71453131 LEGION CR High Flex XLPE Inserts Size 7-8 9mm 71453212 LEGION PS High Flex XLPE Inserts Size 3-4 11mm
71453132 LEGION CR High Flex XLPE Inserts Size 7-8 11mm 71453213 LEGION PS High Flex XLPE Inserts Size 3-4 13mm
71453133 LEGION CR High Flex XLPE Inserts Size 7-8 13mm 71453214 LEGION PS High Flex XLPE Inserts Size 3-4 15mm
71453134 LEGION CR High Flex XLPE Inserts Size 7-8 15mm 71453215 LEGION PS High Flex XLPE Inserts Size 3-4 18mm
71453135 LEGION CR High Flex XLPE Inserts Size 7-8 18mm 71453216 LEGION PS High Flex XLPE Inserts Size 3-4 21mm
71453181 LEGION CR High Flex XLPE Inserts Size 1-2 10mm 71453221 LEGION PS High Flex XLPE Inserts Size 5-6 9mm
71453182 LEGION CR High Flex XLPE Inserts Size 1-2 12mm 71453222 LEGION PS High Flex XLPE Inserts Size 5-6 11mm
71453183 LEGION CR High Flex XLPE Inserts Size 3-4 10mm 71453223 LEGION PS High Flex XLPE Inserts Size 5-6 13mm
71453184 LEGION CR High Flex XLPE Inserts Size 3-4 12mm 71453224 LEGION PS High Flex XLPE Inserts Size 5-6 15mm
71453185 LEGION CR High Flex XLPE Inserts Size 5-6 10mm 71453225 LEGION PS High Flex XLPE Inserts Size 5-6 18mm
71453186 LEGION CR High Flex XLPE Inserts Size 5-6 12mm 71453226 LEGION PS High Flex XLPE Inserts Size 5-6 21mm
71453187 LEGION CR High Flex XLPE Inserts Size 7-8 10mm 71453231 LEGION PS High Flex XLPE Inserts Size 7-8 9mm
71453188 LEGION CR High Flex XLPE Inserts Size 7-8 12mm 71453232 LEGION PS High Flex XLPE Inserts Size 7-8 11mm
71453233 LEGION PS High Flex XLPE Inserts Size 7-8 13mm
71453234 LEGION PS High Flex XLPE Inserts Size 7-8 15mm
71453235 LEGION PS High Flex XLPE Inserts Size 7-8 18mm
71453236 LEGION PS High Flex XLPE Inserts Size 7-8 21mm
46
LEGION™ XLPE Dished Inserts GENESIS™ II Tibial Baseplates (Cemented)
Cat. No. Description Cat. No. Description
71453271 LEGION XLPE Dished Insert Size 1-2 9mm 71420160 GENESIS II Cemented Tibial Baseplate Size 1 Left
71453272 LEGION XLPE Dished Insert Size 1-2 11mm 71420162 GENESIS II Cemented Tibial Baseplate Size 2 Left
71453273 LEGION XLPE Dished Insert Size 1-2 13mm 71420164 GENESIS II Cemented Tibial Baseplate Size 3 Left
71453274 LEGION XLPE Dished Insert Size 1-2 15mm 71420166 GENESIS II Cemented Tibial Baseplate Size 4 Left
71453275 LEGION XLPE Dished Insert Size 1-2 18mm 71420168 GENESIS II Cemented Tibial Baseplate Size 5 Left
71253276 LEGION XLPE Dished Insert Size 1-2 21mm 71420170 GENESIS II Cemented Tibial Baseplate Size 6 Left
71453277 LEGION XLPE Dished Insert Size 3-4 9mm 71420172 GENESIS II Cemented Tibial Baseplate Size 7 Left
71453278 LEGION XLPE Dished Insert Size 3-4 11mm 71420174 GENESIS II Cemented Tibial Baseplate Size 8 Left
71453279 LEGION XLPE Dished Insert Size 3-4 13mm 71420176 GENESIS II Cemented Tibial Baseplate Size 1 Right
71453281 LEGION XLPE Dished Insert Size 3-4 15mm 71420180 GENESIS II Cemented Tibial Baseplate Size 2 Right
71453282 LEGION XLPE Dished Insert Size 3-4 18mm 71420182 GENESIS II Cemented Tibial Baseplate Size 3 Right
71453283 LEGION XLPE Dished Insert Size 3-4 21mm 71420184 GENESIS II Cemented Tibial Baseplate Size 4 Right
71453284 LEGION XLPE Dished Insert Size 5-6 9mm 71420186 GENESIS II Cemented Tibial Baseplate Size 5 Right
71453285 LEGION XLPE Dished Insert Size 5-6 11mm 71420188 GENESIS II Cemented Tibial Baseplate Size 6 Right
71453286 LEGION XLPE Dished Insert Size 5-6 13mm 71420191 GENESIS II Cemented Tibial Baseplate Size 8 Right
71453287 LEGION XLPE Dished Insert Size 5-6 15mm 71931716 GENESIS II Cemented Tibial Baseplate Size 9 Right
71453288 LEGION XLPE Dished Insert Size 5-6 18mm 71931923 GENESIS II Cemented Tibial Baseplate Size 9 Left
71453289 LEGION XLPE Dished Insert Size 5-6 21mm
71453291 LEGION XLPE Dished Insert Size 7-8 9mm
71453292 LEGION XLPE Dished Insert Size 7-8 11mm
71453293 LEGION XLPE Dished Insert Size 7-8 13mm
71453294 LEGION XLPE Dished Insert Size 7-8 15mm
71453295 LEGION XLPE Dished Insert Size 7-8 18mm
71453296 LEGION XLPE Dished Insert Size 7-8 21mm
47
LEGION™ Porous HA Tibial Baseplate with Holes GENESIS™ II Round Resurfacing Patellae
Cat. No. Description Cat. No. Description
71934054 LEGION Porous HA Tibial Baseplate Size 2 Left 71420574 GENESIS II Round Resurfacing Patella 29mm
71934055 LEGION Porous HA Tibial Baseplate Size 3 Left 71420576 GENESIS II Round Resurfacing Patella 32mm
71934056 LEGION Porous HA Tibial Baseplate Size 4 Left 71420578 GENESIS II Round Resurfacing Patella 35mm
71934057 LEGION Porous HA Tibial Baseplate Size 5 Left 71420580 GENESIS II Round Resurfacing Patella 26mm
71934058 LEGION Porous HA Tibial Baseplate Size 6 Left 71926225 GENESIS II Round Resurfacing Patella 38mm
71934059 LEGION Porous HA Tibial Baseplate Size 7 Left 71926226 GENESIS II Round Resurfacing Patella 41mm
71934060 LEGION Porous HA Tibial Baseplate Size 8 Left
71934061 LEGION Porous HA Tibial Baseplate Size 2 Right
71934062 LEGION Porous HA Tibial Baseplate Size 3 Right
71934063 LEGION Porous HA Tibial Baseplate Size 4 Right
GENESIS II Oval Resurfacing Patellae
71934064 LEGION Porous HA Tibial Baseplate Size 5 Right
Cat. No. Description
71934065 LEGION Porous HA Tibial Baseplate Size 6 Right
71421029 GENESIS II Oval Resurfacing Patella 29mm
71934066 LEGION Porous HA Tibial Baseplate Size 7 Right
71421032 GENESIS II Oval Resurfacing Patella 32mm
71934067 LEGION Porous HA Tibial Baseplate Size 8 Right
71421035 GENESIS II Oval Resurfacing Patella 35mm
71421038 GENESIS II Oval Resurfacing Patella 38mm
71421041 GENESIS II Oval Resurfacing Patella 41mm
LEGION Porous HA Tibial Baseplate without Holes
Cat. No. Description
71934161 LEGION Porous HA Tibial Baseplate without Holes Size 2 Left
71934162 LEGION Porous HA Tibial Baseplate without Holes Size 3 Left GENESIS II Biconvex Resurfacing Patellae
71934163 LEGION Porous HA Tibial Baseplate without Holes Size 4 Left Cat. No. Description
71934164 LEGION Porous HA Tibial Baseplate without Holes Size 5 Left 71420566 GENESIS II Biconvex Resurfacing Patella 23mm
71934165 LEGION Porous HA Tibial Baseplate without Holes Size 6 Left 71420568 GENESIS II Biconvex Resurfacing Patella 26mm
71934166 LEGION Porous HA Tibial Baseplate without Holes Size 7 Left 71420570 GENESIS II Biconvex Resurfacing Patella 29mm
71934167 LEGION Porous HA Tibial Baseplate without Holes Size 8 Left 71420572 GENESIS II Biconvex Resurfacing Patella 32mm
71934168 LEGION Porous HA Tibial Baseplate without Holes Size 2 Right
71934169 LEGION Porous HA Tibial Baseplate without Holes Size 3 Right
71934170 LEGION Porous HA Tibial Baseplate without Holes Size 4 Right
71934171 LEGION Porous HA Tibial Baseplate without Holes Size 5 Right
71934172 LEGION Porous HA Tibial Baseplate without Holes Size 6 Right
71934173 LEGION Porous HA Tibial Baseplate without Holes Size 7 Right
71934174 LEGION Porous HA Tibial Baseplate without Holes Size 8 Right
48
PROFIX Metaphyseal Tibial Stems GENESIS II Femoral Flex-Lok Pegs
Cat. No. Description Cat. No. Description
71501009 PROFIX Metaphyseal Tibial Stem 16mm 71420063 GENESIS II Femoral Flex-Lok Peg
71501010 PROFIX Metaphyseal Tibial Stem 18mm
71501020 PROFIX Metaphyseal Tibial Stem 20mm
GENESIS II Primary Femoral Lug
Cat. No. Description
GENESIS™ II Tibial Bone Screws
71420999 GENESIS II Primary Femoral Lug
Cat. No. Description
71420974 GENESIS II 6.5 mm Screw 15 mm Length
71420976 GENESIS II 6.5 mm Screw 20 mm Length
71420978 GENESIS II 6.5 mm Screw 25 mm Length
71420980 GENESIS II 6.5 mm Screw 30 mm Length
71420982 GENESIS II 6.5 mm Screw 35 mm Length GENESIS™ II Long Stems
71420984 GENESIS II 6.5 mm Screw 40 mm Length Cat. No. Description
71420986 GENESIS II 6.5 mm Screw 45 mm Length 71420628 GENESIS II Long Stem 10mm x 100mm
71420988 GENESIS II 6.5 mm Screw 50 mm Length 71420630 GENESIS II Long Stem 12mm x 100mm
71420632 GENESIS II Long Stem 14mm x 100mm
71420634 GENESIS II Long Stem 16mm x 100mm
71420636 GENESIS II Long Stem 18mm x 100mm
71420638 GENESIS II Long Stem 20mm x 100mm
71420640 GENESIS II Long Stem 22mm x 100mm
71420642 GENESIS II Long Stem 24mm x 100mm
71420647 GENESIS II Long Stem 10mm x 150mm with Slot
71420648 GENESIS II Long Stem 14mm x 150mm
71420649 GENESIS II Long Stem 12mm x 150mm with Slot
71420650 GENESIS II Long Stem 16mm x 150mm
49
LEGION™ CR TKS Implant Constructs
Tibial
Femoral Insert Patella
Baseplate
LEGION CR LEGION LEGION GENESIS™ II
Non-Porous Femoral (CoCr) XLPE Dished Insert Revision Tibial Baseplate Round Resurfacing Patella
(Cemented)
LEGION CR LEGION CR GENESIS II
Narrow Non-Porous High Flex XLPE Insert LEGION Oval Resurfacing Patella
Femoral (CoCr) Porous HA Tibial Baseplate
with Holes GENESIS II
LEGION CR Biconvex Resurfacing Patella
Porous Femoral (CoCr) LEGION
Porous HA Tibial Baseplate
LEGION CR without Holes
Porous + HA Femoral (CoCr)
LEGION CR
Femoral (OXINIUM™)
LEGION CR
Narrow Femoral (OXINIUM)
50
LEGION™ CR Compatibility
LEGION CR Femoral (OXINIUM™) LEGION XLPE Deep Dish Insert
LEGION CR Narrow Femoral (OXINIUM) LEGION CR High Flex XLPE Insert
LEGION CR Non-Porous Femoral (CoCr)
LEGION CR Porous Femoral (CoCr)
LEGION CR Porous HA Femoral (CoCr)
LEGION CR Narrow Non-Porous Femoral (CoCr)
Patella Inserts Tibial Baseplate Tibial Baseplate Femoral
GENESIS™ II LEGION GENESIS II CR LEGION LEGION CR
Round Resurfacing Patella XLPE Dished Insert All-Poly Tibial Baseplate Revision Tibial Baseplate Non-Porous Femoral
(Cemented) (CoCr)
GENESIS II LEGION CR
Oval Resurfacing Patella High Flex XLPE Insert LEGION LEGION CR
Porous HA Tibial Narrow Non-Porous
GENESIS II GENESIS II CR Baseplate Femoral (CoCr)
Biconvex Resurfacing Patella Insert with Holes
LEGION CR
GENESIS II CR LEGION Porous Femoral (CoCr)
Deep Dish Insert Porous HA Tibial
Baseplate LEGION CR
GENESIS II CR without Holes Porous + HA Femoral
High Flex (Deep Flex) (CoCr)
Insert GENESIS II
Tibial Baseplate LEGION CR
(Cemented) Femoral (OXINIUM™)
LEGION CR
Narrow Femoral
(OXINIUM)
GENESIS II CR
Femoral (CoCr)
GENESIS II CR
Femoral (OXINIUM)
51
Optional Compatibility for LEGION™ CR Femoral Components and Baseplates
LEGION CR LEGION LEGION LEGION
LEGION CR
Narrow Non- Revision Tibial Porous HA Tibial Porous HA
Non-Porous
Porous Femoral Baseplate Baseplate with Tibial Baseplate
Femoral (CoCr)
(CoCr) (Cemented) Holes without Holes
LEGION RK GENESIS II LEGION GENESIS II PROFIX
Distal Femoral Wedge Femoral Flex-Lok Peg Finned Tibial Wedge Tibial Bone Screw Metaphyseal Tibial Stem
(71501009, 71501010,
LEGION RK LEGION RK/HK PROFIX 71501020)
Posterior Femoral Wedge Hemi-Stepped Tibial Wedge Metaphyseal Tibial Stem
(71501009, 71501010, GENESIS II
LEGION RK LEGION RK/HK 71501020) Long Stem
“L” Femoral Wedge Full-Stepped Tibial Wedge
GENESIS II
LEGION LEGION Long Stem
Distal Wedge Screw Tibial Wedge Screw
LEGION LEGION
Posterior Wedge Screw Locking Set Screw
GENESIS™ II LEGION
Femoral Flex-Lok Peg Tibial Wedge Lug
LEGION
Tibial Cone
LEGION
Press-fit Stem
LEGION
Cemented Stem
LEGION
Stem Extension
LEGION
Offset Coupler
LEGION
Offset Coupler Angled
LEGION
Male to Male Mini Coupler
52
LEGION™ CR Component Size Compatibility
LEGION
Compatible Component Size
Component
LEGION CR Non-Porous LEGION XLPE Dish Insert Size 1-8, 9-21 mm
Femoral (CoCr)
LEGION CR High Flex XLPE Insert Size 1-8, 9-18 mm
LEGION CR Narrow Non-
Porous Femoral (CoCr) GENESIS™ II CR Insert Size 1-8, 9-18 mm
LEGION CR Porous GENESIS II CR Deep Dish Insert Size 1-8, 9-21 mm
Femoral (CoCr)
GENESIS II CR High Flex (Deep Flex) Insert Size 1-8, 9-18 mm
LEGION CR Porous + HA
Femoral (CoCr) GENESIS II Round Resurfacing Patella 26-41 mm
LEGION CR Femoral GENESIS II Oval Resurfacing Patella 29-41 mm
(OXINIUM™)
GENESIS II Biconvex Resurfacing Patella 23-32 mm
LEGION CR Narrow
Femoral (OXINIUM) GENESIS II CR All-Poly Tibial Baseplate Size 1-8, 9-15 mm, LT/RT
LEGION Revision Knee Distal Femoral Wedge Size 2-8, 5-15 mm
LEGION Revision Knee Posterior Femoral Wedge Size 1-8
LEGION Revision Knee “L” Femoral Wedge Size 2-8, 5-15 mm Distal
LEGION CR Non-Porous x 5-10 mm Posterior
Femoral (CoCr)
LEGION Distal Wedge Screw 5-15 mm
LEGION Posterior Wedge Screw 5-10 mm
GENESIS II Femoral Flex-Lok Peg N/A
LEGION CR Narrow Non-
GENESIS II Femoral Flex-Lok Peg N/A
Porous Femoral (CoCr)
LEGION CR Non-Porous Femoral (CoCr) Size 2-8 LT/RT
LEGION CR Narrow Non-Porous Femoral (CoCr) Size 3-6 LT/RT
LEGION CR Porous Femoral (CoCr) Size 2-8 LT/RT
LEGION CR Porous + HA Femoral (CoCr) Size 2-8 LT/RT
LEGION LEGION CR Femoral (OXINIUM) Size 2-8 LT/RT
XLPE Dished Insert
LEGION CR Narrow Femoral (OXINIUM) Size 3-6 LT/RT
LEGION CR
High Flex XLPE Insert GENESIS II CR Femoral (CoCr) Size 1-8 LT/RT
GENESIS II CR Femoral (OXINIUM) Size 2-9 LT/RT
LEGION Revision Tibial Baseplate (Cemented) Size 1-8, LT/RT
LEGION Porous HA Tibial Baseplate with Holes Size 2-8, LT/RT
LEGION Porous HA Tibial Baseplate without Holes Size 2-8, LT/RT
GENESIS II Tibial Baseplate (Cemented) Size 1-9, LT/RT
53
Compatibility Table (continued)
LEGION™
Compatible Component Size
Component
LEGION Revision Tibial LEGION XLPE Dished Insert Size 1-8, 9-21 mm
Baseplate (Cemented) LEGION CR High Flex Insert Size 1-8, 9-18 mm
LEGION Porous HA Tibial
GENESIS™ II CR Insert Size 1-8, 9-18 mm
Baseplate with Holes
LEGION Porous HA Tibial GENESIS II CR Deep Dish Insert Size 1-8, 9-21 mm
Baseplate without Holes GENESIS II CR High Flex (Deep Flex) Insert Size 1-8, 9-18 mm
LEGION Porous HA Tibial GENESIS II Tibial Bone Screw 15-50 mm
Baseplate with Holes
LEGION Porous HA Tibial PROFIX Metaphyseal Tibial Stem 16, 18, 20 mm
Baseplate with Holes
GENESIS II Long Stem 100, 150mm
LEGION Porous HA Tibial
Baseplate without Holes
LEGION Finned Tibial Wedge Size 2-8, 10-20 mm LT/RT
LEGION RK/HK Hemi-Stepped Tibial Wedge Size 1-8, 5-15 mm LL/RM LM/RL
LEGION RK/HK Full-Stepped Tibial Wedge Size 1-8, 10-15 mm
LEGION Tibial Wedge Screw N/A
LEGION Locking Set Screw N/A
LEGION Tibial Wedge Lug 10, 15 mm
LEGION Tibial Cone 18-30 mm, Short/Long
LEGION Revision Tibial LEGION Press-fit Stem 9-24 mm x 120 mm
Baseplate (Cemented) 9-24 mm x 160 mm
Straight - 10-24 mm x 220mm
Bowed - 10-24 mm x 220mm
Bowed - 10-24 mm x 280 mm
LEGION Cemented Stem 10-20 mm x 120 mm
10-20 mm x 160 mm
LEGION Stem Extension 10-14 mm x 80 mm
LEGION Offset Coupler 2-6 mm
LEGION Offset Coupler Angled N/A
LEGION Male to Male Mini Coupler 2-6 mm
54
LEGION™ PS TKS Implant Constructs
Tibial
Femoral Insert Patella
Baseplate
LEGION PS LEGION PS LEGION GENESIS™ II
Non-Porous Femoral (CoCr) High Flex XLPE Insert Revision Tibial Baseplate Round Resurfacing Patella
(Cemented)
LEGION PS LEGION GENESIS II
Narrow Non-Porous Constrained Insert LEGION Oval Resurfacing Patella
Femoral (CoCr) Porous HA Tibial Baseplate
with Holes GENESIS II
LEGION PS Biconvex Resurfacing Patella
Femoral (OXINIUM™) LEGION
Porous HA Tibial Baseplate
LEGION PS without Holes
Narrow Femoral (OXINIUM)
LEGION RK
Constrained Femoral (CoCr)
LEGION RK
Constrained Femoral
(OXINIUM)
55
LEGION™ PS Compatibility
LEGION PS Femoral LEGION LEGION PS LEGION PS LEGION
(OXINIUM™) PS Femoral High Flex XLPE High Flex XLPE Constrained
(OXINIUM) Insert Insert Insert*
LEGION PS Narrow
Femoral (OXINIUM) LEGION PS LEGION
Narrow Femoral Constrained
LEGION PS Non-Porous (OXINIUM)
Femoral (CoCr) Insert*
LEGION PS
LEGION PS Narrow Non-Porous
Non-Porous Femoral Femoral (CoCr)
(CoCr)
LEGION PS
LEGION RK Constrained Narrow
Femoral (CoCr) Non-Porous
LEGION RK Constrained Femoral (CoCr)
Femoral (OXINIUM)
Patella Inserts Tibial Baseplate Tibial Baseplate Femoral Femoral
GENESIS™ II LEGION PS GENESIS II LEGION RK LEGION PS LEGION PS
Round Resurfacing High Flex XLPE PS All-Poly Tibial Baseplate Non-Porous Femoral (CoCr) Femoral (OXINIUM™)
Patella Insert Tibial Baseplate Cemented)
LEGION PS LEGION PS
GENESIS II LEGION PS LEGION Narrow Non-Porous Narrow Femoral (OXINIUM)
Oval Resurfacing Constrained Insert* Porous HA Tibial Baseplate Femoral (CoCr)
Patella with Holes**
LEGION PS
GENESIS II PS LEGION PS Non-Porous Femoral (CoCr)
GENESIS II Insert LEGION Femoral (OXINIUM™)
Biconvex Porous HA Tibial Baseplate
Resurfacing Patella GENESIS II PS without Holes** LEGION PS
LEGION PS Narrow Non-Porous
High Flex Insert Narrow Femoral (OXINIUM)
GENESIS II Femoral (CoCr)
GENESIS II Tibial Baseplate (Cemented)
LEGION RK LEGION RK
Constrained Insert*
Constrained Femoral(CoCr) Constrained Femoral
(CoCr)
LEGION RK
Constrained Femoral LEGION RK
(OXINIUM) Constrained Femoral
(OXINIUM)
GENESIS II CR
Non-Porous Femoral
(CoCr)**
GENESIS II CR
Non-Porous Femoral
(OXINIUM)**
* Component must be used with the LEGION Femorals (CoCr and OXINIUM) and Tibial Stem
** Not compatible with LEGION and GENESIS II Constrained Insert
56
Optional Compatibility with LEGION™ PS
LEGION PS LEGION RK
LEGION PS Narrow Non-Porous Constrained Femoral
Non-Porous Femoral Femoral (CoCr) (CoCr)
(CoCr)
LEGION PS LEGION RK
LEGION PS Narrow Constrained Femoral
Femoral (OXINIUM™) Femoral (OXINIUM) (OXINIUM)
LEGION RK GENESIS II LEGION RK
Distal Femoral Wedge Femoral Flex-Lok Peg Distal Femoral Wedge
LEGION RK GENESIS II LEGION RK
Posterior Femoral Wedge Primary Femoral Lug Posterior Femoral Wedge
LEGION RK LEGION RK
“L” Femoral Wedge “L” Femoral Wedge
LEGION LEGION
Distal Wedge Screw Distal Wedge Screw
LEGION LEGION
Posterior Wedge Screw Posterior Wedge Screw
GENESIS™ II LEGION
Femoral Flex-Lok Peg Locking Set Screw
GENESIS II LEGION
Primary Femoral Lug Femoral Cone
LEGION
Press-fit Stem
LEGION
Cemented Stem Straight
LEGION
Short Stem Extension
LEGION
Offset Coupler Angled
LEGION
Offset Coupler
LEGION
Male to Male Coupler
57
Optional Compatibility with LEGION™ PS (continued)
LEGION™ RK LEGION LEGION
Tibial Baseplate Porous HA Tibial Porous HA Tibial
(Cemented) Baseplate with Holes Baseplate without Holes
LEGION GENESIS II PROFIX
Finned Tibial Wedge Tibial Bone Screw Metaphyseal Tibial Stem
(71501009, 71501010, 71501020)
LEGION RK/HK PROFIX
Hemi-Stepped Tibial Wedge Metaphyseal Tibial Stem GENESIS II
(71501009, 71501010, 71501020) Long Stem
LEGION RK/HK
Full-Stepped Tibial Wedge GENESIS II
Long Stem
LEGION
Tibial Wedge Screw
LEGION
Locking Set Screw
LEGION
Tibial Wedge Lug
LEGION
Tibial Cone
GENESIS™ II
Long Stem
LEGION
Press-fit Stem
LEGION
Cemented Stem Straight
LEGION
Short Stem Extension
LEGION
Offset Coupler Angled
LEGION
Offset Coupler
LEGION
Male to Male Coupler
58
LEGION™ PS Component Size Compatibility
LEGION
Compatible Component Size
Component
LEGION PS Non-Porous LEGION PS High Flex XLPE Insert Size 1-8, 9-21 mm
Femoral (CoCr)
LEGION Constrained Insert* Size 1-8, 9 mm
LEGION PS Narrow Non-
Porous Femoral (CoCr) GENESIS™ II Constrained Insert* Size 1-8, 11-30 mm
LEGION PS GENESIS II PS Insert Size 1-8, 9-25 mm
Femoral (CoCr)
GENESIS II PS High Flex Insert Size 1-8, 9-25 mm
LEGION PS Narrow
Femoral (OXINIUM™) GENESIS II Round Resurfacing Patella 26-41 mm
LEGION RK Constrained GENESIS II Oval Resurfacing Patella 29-41 mm
Femoral (CoCr)
GENESIS II Biconvex Resurfacing Patella 23-32 mm
LEGION RK Constrained
Femoral (OXINIUM)
LEGION PS GENESIS II PS All-Poly Tibial Baseplate Size 1-8, 9-15 mm LT/RT
Femoral (OXINIUM)
LEGION PS Narrow
Femoral (OXINIUM™)
LEGION PS Non-Porous GENESIS II Primary Femoral Lug N/A
Femoral (CoCr)
LEGION PS Narrow Non-
Porous Femoral (CoCr) GENESIS II Femoral Flex-Lok Peg N/A
LEGION PS Non-Porous LEGION Revision Knee Distal Femoral Wedge Size 2-8, 5-15 mm
Femoral (CoCr)
LEGION Revision Knee Posterior Femoral Wedge Size 1-8, 5-10 mm
LEGION PS
Femoral (OXINIUM) LEGION Revision Knee “L” Femoral Wedge Size 2-8, 5-15 mm Distal x
5-10 mm Posterior
LEGION RK Constrained
Femoral (CoCr) LEGION Distal Wedge Screw 5-15 mm
LEGION RK Constrained LEGION Posterior Wedge Screw 5-10 mm
Femoral (OXINIUM)
LEGION Locking Set Screw N/A
LEGION Femoral Cone 18-30 mm, LT/RT
LEGION Press-fit Stem 9-24 mm x 120 mm
9-24 mm x 160 mm
Straight - 10-24 mm x 220mm
LEGION RK Constrained Bowed - 10-24 mm x 220mm
Femoral (CoCr) Bowed - 10-24 mm x 280 mm
LEGION RK Constrained
LEGION Cemented Stem 10-20 mm x 160 mm
Femoral (OXINIUM)
LEGION Short Stem Extension 10-14 mm x 80 mm
LEGION Offset Coupler Angled N/A
LEGION Offset Coupler 2, 4, 6 mm
LEGION Male to Male Mini Coupler 2, 4, 6 mm
* Component must be used with the LEGION Femorals (CoCr and OXINIUM) and Tibial Stem
59
Compatibility Table (continued)
LEGION™
Compatible Component Size
Component
LEGION PS Non-Porous Femoral (CoCr) Size 2-8 LT/RT
LEGION PS Narrow Non-Porous Femoral (CoCr) Size 3-6 LT/RT
LEGION PS Femoral (OXINIUM™) Size 2-8 LT/RT
LEGION PS LEGION PS Narrow Femoral (OXINIUM) Size 3-6 LT/RT
High Flex XLPE Insert
LEGION RK Constrained Femoral (CoCr) Size 2-8 LT/RT
LEGION
Constrained Insert* LEGION RK Constrained Femoral (OXINIUM) Size 2-8 LT/RT
GENESIS II Tibial Baseplate (Cemented) Size 1-9, LT/RT
LEGION Revision Tibial Baseplate (Cemented) Size 1-8, LT/RT
LEGION Porous HA Tibial Baseplate With Holes Size 2-8, LT/RT
LEGION Porous HA Tibial Baseplate without Holes Size 2-8, LT/RT
LEGION PS GENESIS II PS Femoral Component (CoCr)** Size 1-8 LT/RT
High Flex XLPE Insert GENESIS II PS Femoral (OXINIUM)** Size 2-9 LT/RT
* Component must be used with the LEGION Femorals (CoCr and OXINIUM) and Tibial Stem
** Not compatible with LEGION and GENESIS II Constrained Insert
60
Compatibility Table (continued)
LEGION™
Compatible Component Size
Component
LEGION Revision Tibial LEGION PS High Flex XLPE Insert Size 1-8, 10-21 mm
Baseplate (Cemented) LEGION Constrained Insert* Size 1-8, 9 mm
LEGION Porous HA Tibial
GENESIS™ II PS Insert Size 1-8, 9-25 mm
Baseplate with Holes
LEGION Porous HA Tibial GENESIS II PS High Flex Insert Size 1-8, 9-25 mm
Baseplate without Holes GENESIS II Constrained Insert* Size 1-8, 11-30 mm
LEGION Finned Tibial Wedge Size 2-8, 10-20 mm LT/RT
LEGION RK/HK Hemi-Stepped Tibial Wedge Size 1-8, 5-15 mm LL/RM LM/RL
LEGION RK/HK Full-Stepped Tibial Wedge Size 1-8, 10-15 mm
LEGION Tibial Wedge Screw N/A
LEGION Locking Set Screw N/A
LEGION Tibial Wedge Lug 10, 15 mm
LEGION Tibial Cone 18-30 mm, Short/Long
LEGION Revision Tibial 9-24 mm x 120 mm
Baseplate (Cemented) 9-24 mm x 160 mm
LEGION Press-fit Stem Straight - 10-24 mm x 220mm
Bowed - 10-24 mm x 220mm
Bowed - 10-24 mm x 280 mm
10-20 mm x 120 mm
LEGION Cemented Stem Straight
10-20 mm x 160 mm
LEGION Short Stem Extension 10-14 mm x 80 mm
LEGION Offset Coupler 2-6 mm
LEGION Offset Coupler Angled N/A
LEGION Male to Male Mini Coupler 2-6 mm
LEGION Porous HA Tibial GENESIS II Tibial Bone Screw 15 – 50 mm
Baseplate with Holes
LEGION Porous HA Tibial PROFIX Metaphyseal Tibial Stem 16, 18, 20 mm
Baseplate with Holes
LEGION Porous HA Tibial GENESIS II Long Stem 100, 150mm
Baseplate without Holes
* Component must be used with the LEGION Femorals (CoCr and OXINIUM™) and Tibial Stem
61
Products may not be available in all markets because product availability is subject to the regulatory and/or medical practices in individual markets.
Please contact your Smith & Nephew representative or distributor if you have questions about the availability of Smith & Nephew products in your area.
Smith & Nephew, Inc. www.smith-nephew.com
7135 Goodlett Farms Parkway
Cordova, TN 38016
USA
Telephone: 1-901-396-2121
Information: 1-800-821-5700
Orders and Inquiries: 1-800-238-7538
©2022 Smith & Nephew, Inc.
™Trademark of Smith & Nephew. All rights reserved.
All Trademarks acknowledged 02860 V3 71281978 REVB 11/22
Surgical Technique
Table of contents
Introduction........................................................................................................................................1
Precautions regarding surgical technique.........................................................................................2
Preoperative planning.......................................................................................................................3
Surgical technique.............................................................................................................................4
Explantation of the stem..................................................................................................................14
Postoperative treatment...................................................................................................................15
Sterilization.......................................................................................................................................15
Implants............................................................................................................................................16
Dimensions...................................................................................................................................... 17
Instruments......................................................................................................................................19
Note Bena
The surgical technique described in this brochure is the procedure suggested by the authors for
uncomplicated surgery. The surgeon must, however, decide which procedure is the most suitable
and effective for each individual patient. e-IFU homepage:ifu.smith-nephew.com
The following technique is for informational and educational purposes only. It is not intended to
serve as medical advice. It is the responsibility of treating physicians to determine and utilize the
appropriate products and techniques according to their own clinical judgment for each of their
patients. For more information on the product, including its indications for use, contraindications,
and product safety information, please refer to the product’s label and the Instructions for Use.
Introduction
The SL-PLUS™ MIA INTEGRATION-PLUS™ stem, also referred to as SL-PLUS MIA, has been in clinical use since
2007. The innovative design of the SL-PLUS MIA is based on the SL-PLUS by elimination of the trochanteric
wing and retaining the successful biological meta-diaphyseal anchoring concept.1-7 The dual-tapered straight
stem with its rectangular cross section helps to establish excellent primary and rotational stability in the
cortical bone.3,6
The SL-PLUS MIA stem is designed to implant using minimally invasive surgical techniques as it preserves and
protects the greater trochanter bone.* This may minimise the risk of fracture as a result of better trochanteric
bone stability.4-5,8-10
The cementless SL-PLUS MIA range includes 14 standard stems with a CCD angle of 131° and 12 lateral stems
with a CCD angle of 123°. All stems are coated with the INTEGRATION-PLUS open-pored titanium plasma/HA
coating in the proximal area leading to accelerated and increased bone on-/in-growth.**6-7,11-12
*Compared to SL-Plus Primary
**Compared to HA only coated stems and non-HA coated stems
1
Precautions regarding surgical technique
Reliable fit of femoral ball heads on stem tapers
The taper connection can only be reliably and firmly seated if the surface of the ball head cone and the surface
and structure of the hip stem taper are completely intact. The disposable plastic cap protecting the stem taper
from damage shall not be removed until the trial ball is attached. To ensure that the ball head performs as
required, it is essential to take great care when attaching it to the stem taper: Clean and dry the neck taper with a
clean, sterile cloth. Place the prosthetic femoral head on the neck taper and firmly impact with the femoral ball
head impactor and a mallet several times. Never reuse a femoral ball head that has been impacted onto a stem
cone and then removed.
Restrictions on head/insert combinations
• BIOLOX®forte/delta ceramic inserts must only be combined with
BIOLOX®forte/delta ceramic ball heads.
Compatible Devices/Devices used with:
The surgeon must always ensure that the individual components of the total/partial hip arthroplasty are
compatible. SL-PLUS™ MIA must not be combined with other manufacturer implants, unless the implant
combination is described within this document, the SL-PLUS MIA IFU, or is listed in the compatibility matrix
(lit. no. 04758) available on ifu.smithnephew.com. The compatibility matrix provides all femoral head combination
options for SL- PLUS MIA relating sizes and material. Ball heads should only be combined with hip stems of
identical taper dimensions. SL-PLUS MIA can be combined with all acetabular cup systems manufactured by
Smith & Nephew Orthopaedics AG and Smith & Nephew Inc. taken into account any restrictions given by the
relevant femoral head.
Further, SL-PLUS MIA should always be implanted using Smith & Nephew Orthopaedics AG surgical instruments,
unless these are commonly used in the operating room and/or described in this surgical technique. All liability is
excluded for the unauthorised use of third-party products.
Note:
- SL-PLUS MIA must not be combined with femoral ball heads made of stainless steel
- Size 01 SL-PLUS MIA standard stem must not be used in combination with femoral ball heads
with neck length +16
Femoral ball head revisions
For the revision of a femoral ball head, apply the following:
• The corresponding insert has to be revised as well.
• Replacement only by a metal ball head or a special ceramic revision femoral ball head featuring a metallic taper
adaptor (BIOLOX® OPTION). Do not assemble a standard ceramic head on a used taper; the ceramic head may
fracture from irregularities on the femoral component taper.
• In the case of a revision involving a fractured ceramic component, femoral ball head and insert have to be
revised. Remove all loose identifiable fragments and thoroughly irrigate and lavage the operative site. A special
ceramic revision femoral ball head coupled with a metallic taper adaptor should be used if the femoral taper
is intact. Metal or OXINIUM◊ femoral ball heads with a polyethylene insert shall not be used for revising
fractured ceramic components. In case the taper is damaged or no appropriate ceramic revision femoral ball
head is available, the femoral stem must be revised to provide a suitable femoral taper to attach a new ceramic
ball head.
2
Preoperative planning
Preoperative planning is essential to determine the correct
orientation and size of implant. AP and axial X-rays are required
for this purpose. The offset and neck length to be achieved with
the prosthesis are determined with the aid of X-ray templates.
Preoperative planning requires:
x
• X-rays
x
• Templates for the acetabular component
and the stem or
• Digital templates for the acetabular component
and the stem
The following acetate X-ray templates are available
for SL-PLUS™ MIA:
Item No. Description
2202 SL-PLUS MIA X-ray templates Standard / Lateral 1.15
Today, digital templating is the new standard in many
institutions. Smith & Nephew is supporting a number of
third party digital pre-operative planning tools.
To determine the entry point in the medullary canal, the surgeon
is advised to draw the femoral axis on the AP radiograph
and extend this proximally. This line indicates how far laterally
the box chisel needs to be placed to open up the medullary
canal, thus facilitating identification of the lateral entry point
at operation.
It is also helpful to define the position of the stem within the
canal. The distance x from the shoulder of the stem to the
greater and lesser trochanter is measured and can serve as
an additional leg length check.
Note
X-ray templating is not as reliable for determining implant
sizing in revision cases as they are in planning for primary
implantations and can therefore only provide an approximate
estimate of the actual stem size required.
Warnings and precautions
SL-PLUS MIA implants must not be implanted with cement.
Please contact your local Smith & Nephew representative or
distributor to order X-ray templates or in case of any other
product related question.
3
Surgical technique
Position of the patient and approach
For the purpose of this surgical technique, the anterolateral,
minimally invasive approach is chosen with the patient in the
supine position.
Whether an anterolateral, lateral, posterolateral or posterior
approach is used is at the surgeon’s discretion. The skin
incision and muscle detachment depend on the selected
approach. All implants and instruments suit conventional or
minimally invasive surgical techniques.
Capsular incision and dissection
Perform a blunt dissection between the tensor fasciae latae and
the gluteus medius/minimus from the lateral side to expose the
femoral neck.
One sharp and one blunt Hohmann retractor are placed laterally
and medially respectively. The curved rectus tendon is exposed,
undermined, divided and released from its attachment to
the capsule.
The third retractor is placed between the ventral joint capsule
and the overlying muscles (curved rectus tendon and the
ileopsoas muscle).
The femoral neck is then exposed via an H-shaped incision of
the joint capsule:
• longitudinal incision, as far medially as possible, from the
acetabular rim to the intertrochanteric line
• proximal transverse incision of the acetabular labrum from
approx. the nine o’clock to the three o’clock position
• distal transverse incision extending along the
intertrochanteric line
4
After the wing-like opening of the joint capsule, the dissection
of the capsule can be continued by extending the distal incision
along the intertrochanteric line in the direction of the lesser
trochanter and the proximal incision medially and/or laterally.
Two blunt Hohmann retractors are positioned inside the joint.
Problematic osteophytes on the acetabular rim are removed.
The technique selected by the surgeon for the neck resection
(single or double osteotomy) depends on the patient
(coxa vara/valga).
Capsular release
In order to facilitate alignment of the stem, the surgeon switches
the leg to the “Figure of 4” position and performs an additional
release of the posterior capsule.
Now a retractor is placed between the capsule and the gluteus
muscle holding the leg in a 10-20° abducted position. This
facilitates the placement of the retractor and gives access to the
lateral and cranial part of the capsule. In that way this part of the
capsule should be released or resected.
The leg is hyperextended, adducted approximately 30–40° and
externally rotated by 90°. The leg being operated on is placed
under the other leg in a “Figure of 4” position. Alternatively, the
other leg is lowered and the leg being operated on is placed
on top.
The proximal femur is mobilized with two retractors, a lateral
trochanter retractor and a second retractor on the medial side of
the femoral neck.
Capsular releases must be performed in the direction of both
the lesser trochanter and the trochanteric fossa to the tip of the
trochanter. Optionally, it can also be performed on the caudal
rim of the acetabulum.
5
The cup is then implanted according to the corresponding
surgical technique.
Stem Preparation
In very muscular or obese patients, patients with a valgus
femoral neck, or in cases where the proximal femur sits deep to
the skin surface, further release of the posterior capsule or
release of the piriformis tendon may be necessary to allow
Inferior gemellus m.
adequate mobilization of the femur prior to preparation of the
Internal obturator m.
implantation site.
Superior gemellus m.
Piriformis m.
Access to the medullary canal
The box chisel (75006419/41000029) is placed close to the
posterior cortex at the resection level. The box chisel is
introduced in alignment with the femoral axis and a small
square block of bone is removed. If, after the neck resection, the
box chisel is not used to expose the canal, splitting of the
trochanter may occur during rasp insertion.
Do not impact the box chisel too deeply.
The MIA curved rasp facilitates opening of the diaphyseal
medullary cavity. Removal of the intramedullary fat by suction to
avoid fat embolism.
Further opening of the diaphyseal medullary cavity and probing
of the diaphysis with a corresponding awl are recommended.
6
Use of the MIA guide rasp
The aim of the MIA guide rasps is to ensure easier alignment
Medial
of the rasping bed, thereby preventing a varus implant seating.
The MIA guide rasp should be introduced with the desired
degree of anteversion, matching the planned rotation of
the stem.
The rasp depth can be checked by means of the line markings
123456 on the stem. These markings correspond to the shoulder level
Markings on implant shoulder of the respective SL-PLUS™ MIA implant size. During the rasping
operation, care should be taken to restrict the depth of insertion
of the MIA guide rasp to around one or two sizes smaller than
the planned implant size. As soon as cortical contact is
achieved the guide rasp should not be driven any further. An
optional MIA guide rasp, size 01-3, is available for small stem
sizes of 01 or 0.
When mounting the rasp on the slap hammer or rasping
machine, ensure that the side marked “MEDIAL” is correctly
oriented. If the medial and lateral sides are inadvertently
reversed, the rasp handle may strike the medial aspect of
the greater trochanter, forcing the rasp out of its planned
neutral alignment.
Broaching of the femur
The rasp cuts longitudinal grooves in the femoral cortex. The
goal is to achieve extensive surface area contact to provide
cortical support for the implant. By gradually increasing the
depth of rasping within the medullary cavity, the area of cortical
contact increases, along with the resistance to the advance of
the rasp. As soon as the rasp is fully engaged within cortical
bone, the pitch of the hammer blows rises. The critical resource
for checking the correct size is the preoperative planning with
the X-ray.
Preparation of the bone bed up to implants of size 4 starts with
a size 01 rasp. For implants of size 5 and above, a size 1 rasp
can be used initially.
Important
Since the first rasp determines the position of all the following
rasps, its orientation is critical for exact positioning of the stem.
7
For correct alignment, the rasp is inserted along the anatomical
axis of the stem. If the operator deviates from this axis there is a
risk of subsequent varus positioning of the stem in the femur.
The mountable MIA detachable rasps are used to create an
implantation site of the correct size and alignment for the
femoral implant, as described below.
Adapters with differing offsets are available to accommodate the
selected surgical approach and/or patient size. Please refer to
the instrument sets on page 21 and 22.
All modular adapters can be connected to the slap hammer or
the MIA Rasping Machine (IMT Woodpecker) or the modular
knock plate (75000642/21000378).
At the start of the rasping process, the depth of the rasp must
be checked to ensure that it is not inserted too deeply. It is
Bone loss in calcar arch
extremely important to understand that the osteotomy is
unrelated to the final position of the implant. Because of the
initially reduced rasp resistance, there is a tendency for
surgeons to insert the smaller rasps too deeply into the femur.
This will result in excessive enlargement of the implantation site
and lead to gaps on the medial aspect between the bone and
the final implant.
8
The subsequent rasp is introduced into the cavity until
resistance is felt. The rasp is then driven laterally and distally
into the femur using the (slap) hammer or the rasping machine.
This process is repeated with sequential rasp sizes until an
adequate depth is achieved for the desired implant size. If the
double offset adapters are used, care should be taken to ensure
that these instruments are used on the corresponding side of
the patient.
The rasp is introduced into the cavity along a slightly arc-shaped
path.
Attention should be paid to the anteversion and varus/valgus
alignment of the slap hammer or the rasping machine with
respect to the femoral axis. Insertion of the rasps in a varus
inclination involves the risk of perforation and/or fracture of the
lateral cortex of the femur.
If rasping is carried out using the slap hammer or the MIA
Rasping Machine, the weight of these instruments helps ensure
the precise longitudinal alignment of the rasp within the femur. It
is important that lateral pressure is continuously applied to the
rasping machine to ensure that the rasp moves in line with the
axis of the canal and does not seat in a varus position.
9
The shoulder of the rasp corresponds to the shoulder height of
the implant and should correspond to the preoperatively
determined distance to the greater trochanter (marked x). If the
greater trochanter is not present anymore, the lesser trochanter
x
can be taken instead.
Only in rare cases does the planned prosthesis size not
correspond to the size rasped at operation. If there is a
discrepancy of two or more sizes, the rasp may not have
reached the necessary depth because of incorrect angulation or
the presence of an obstacle in the canal. In these cases the
prosthesis would be too small and stable cortical anchorage
would not be ensured in the long term. If doubt exists,
intraoperative radiography or fluoroscopy is necessary so that
the obstruction can be assessed.
If the preplanned implant size goes too deep into the femoral
canal, a fracture (fissure) should be excluded by intraoperative
radiography or fluoroscopy.
Once the optimum size and stability of the rasp and its height
position are achieved, the offset adapter is removed from the
mounted detachable rasp.
10
Trial Reposition
The neck module is attached to the detachable rasp by hand.
Standard neck modules are available for detachable rasp sizes
01–0, 1–6 and 7–12. Matching “lateral” neck modules are
available for detachable rasp sizes 1–6 and 7–12.
It is important to ensure that the neck module fits flush with the
detachable rasp and that it clicks into position.
Trial Reposition
The selected trial ball head can either be fitted to the neck
module in advance or in situ. Ideally fitting of a M/+4 or L/+8
trial ball head is recommended to leave room for up-/down-
sizing on the final stem.
The joint is reduced and the leg length, soft tissue tension and
range of motion are checked. During the initial operations, it is
recommended that the surgeon obtain AP and lateral
intraoperative radiographs (image intensifier) to verify the
position and size of the rasp in both planes.
If necessary the trial ball head and/or the neck module
(standard or lateral) should be replaced until a satisfactory
result is achieved.
If luxation is still possible even though the lateralized modular
neck is used and the rasp is securely seated, check if the next
sized stem should be used. This then generally protrudes
further proximally out of the femoral bone compared to the
smaller size and can therefore lead to leg lengthening. The
patient should be made aware of this risk before the surgery.
11
The neck module can either be removed manually from the
detachable rasp or with a Kocher retractor.
The rasp adapter is connected to the detachable rasp. The
connected rasping machine or the slap hammer is used to
remove the detachable rasp.
Removal of the rasp, as with its introduction, must be performed
along a curved arc to minimize disturbance of the bone bed and
to avoid fractures in the vicinity of the trochanter.
Implantation of the stem
The stem of the correct size is introduced manually as deep as
possible into the canal. The SL-PLUS™ MIA stem is then seated
with the MIA stem impactor (75007255/600621), using
appropriately measured strokes to minimize the risk of
femoral fractures.
The Ti/HA coating leads to a total oversize of 0.7 mm and,
particularly if hard bone is present, can prevent the implant from
being inserted as deeply as the rasp. In this case, it is advisable
to recheck the leg length, soft tissue tension and range of
motion with the trial ball head. This oversize is not usually
relevant for clinical purposes and can be offset with a shorter
ball head. Otherwise, the implant bed must be enlarged
accordingly with the last rasp.
The protective cap remains mounted on the cone
during impaction.
Please note
It is not sufficient to press the stem in solely by hand.
It is not permitted to impact the stem deeper than the prepared
bony bed or to correct the position once the stem is in the
cone bed as this would inevitably split the femoral stem. It is
recommended to recheck the leg length, soft tissue tension
and range of motion with the trial ball head.
12
Ball head introduction
Remove the plastic taper protection. Before positioning the ball
head, carefully wash the stem cone with water and then dry.
The ball head is then fitted by screwing it lightly into place.
The ball head is firmly impacted using the ball head impactor
(75002160/110242). The joint is then reduced, manipulated and
retested to ensure proper functioning.
Note
Please carefully read the chapter “Reliable fit of femoral ball
heads on stem tapers” on page 2. Ceramic heads must never
be impacted using a metal instrument.
Each femoral stem has a standard 12/14 cone for coupling with
OXINIUM™, ceramic or CoCr ball heads supplied by
Smith & Nephew Orthopaedics AG or Smith & Nephew Inc.
Wound closure
Insert Redon drains and close the wound. Position the leg in slight abduction.
13
Explantation of the stem
The SL-PLUS™ MIA stem can be extracted using the M6
extraction screw (75002165/110249).
In case of difficulty, an extractor block secured with the M8
extraction screw is also available.
Both extraction screws, M6 and M8, can be used in connection
with either the slap hammer or the rasping machine.
Please note
Ensure that the extraction screw is inserted axially.
The removal of a well osseointegrated stem should be done
by the following four steps:
1. reak the bone bridges in the proximal (Ti-/HA-coated) part
B
of the stem by using a thin flat chisel.
2. Make a short cut in the cortical bone with a thin saw blade
at the distal end of the stem. Both ends of this cut should be
drilled with a Ø4.0mm drill to avoid fissure following this cut.
This cut should be placed in the ventral - lateral part of the
femur corticalis.
3. Spread the bone edges of this cut carefully from the implant
using a chisel with gentle wedge action. Doing so, you will
break the bone bridges between the distal end of the
implant and the surrounding bone.
4. Try to remove the implant stem with the appropriate extrac-
tion tools as described above.
Note
For additional stem removal tools, consider Smith & Nephew’s
RENOVATION™ Implant Removal System, lit no 7138-0701.
14
Postoperative treatment
The postoperative treatment depends on the patient’s age and general state of health. The operated leg may be
immediately weight bearing. For 48 hours a splint (foam) in slight abduction is recommended.
The use of crutches can be helpful during the first days.
The use of antibiotics and thrombosis prophylaxis as well as suture removal are at the surgeon’s discretion.
Serious incidents related to any Smith & Nephew medical device should be reported to Smith & Nephew
(
[email protected]) and to the competent authority of the country in which the user and/or patient
is established.
Sterilization
Implants
All the implants described in this Surgical Technique are sterile when they are delivered by the manufacturer.
Resterilization is not allowed.
Instruments
Surgical instruments and trials are not sterile when they are delivered. They are to be cleaned, inspected and
sterilized before use as described in the guide “Processing (cleaning, disinfection and sterilization) of instruments
from Smith & Nephew Orthopaedics AG” (Lit. No. 03389). Sterilization must also be conducted in accordance
with the legal regulations and guidelines applicable in the country of use.
A separate guide is available for the MIA Rasping Machine (Woodpecker, 75004683/800-09-001) from its
manufacturer IMT: please, refer to “Operation Manual Woodpecker” on https://www.imt-medical.com/handling-
and-maintenance/.
15
Implants
SL-PLUS™ MIA INTEGRATION-PLUS™ Cementless Stem
Material: Ti-6Al-7Nb ISO 5832-11, Cone 12/14
With Ti-plasma/hydroxyapatite coating proximally:
Standard Stem CCD 131° Lateral Stem CCD 123°
SAP No. Art. No. Size SAP No. Art. No. Size
75000172 11000422 01 - - -
75000173 11000423 0 - - -
75000174 11000424 1 75000186 11000436 1
75000175 11000425 2 75000187 11000437 2
75000176 11000426 3 75000188 11000438 3
75000177 11000427 4 75000189 11000439 4
75000178 11000428 5 75000190 11000440 5
75000179 11000429 6 75000191 11000441 6
75000180 11000430 7 75000192 11000442 7
75000181 11000431 8 75000193 11000443 8
75000182 11000432 9 75000194 11000444 9
75000183 11000433 10 75000195 11000445 10
75000184* 11000434 11 75000196* 11000446 11
75000185* 11000435 12 75000197* 11000447 12
*Outlier size (optional) * Outlier size (optional)
Demo Implants
SAP No. Art. No. Size Description
75023398 91000428 5 SL-PLUS™ MIA INTEGRATION-PLUS Standard
75023399 91000440 5 SL-PLUS MIA INTEGRATION-PLUS Lateral
16
Dimensions
All dimensions in mm
Specifications Neck Height Standard
Size Stem Stem M/L CCD CCD Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16
length I length II width angle angle
Standard Lateral
01 128 109 26 131 01 17 19 21 24 27 32
0 132 113 27 131 0 18 19 22 25 27 33
1 137 117 28 131 123 1 18 20 23 25 28 33
2 141 121 29 131 123 2 19 21 23 26 29 34
3 145 124 30 131 123 3 20 21 24 27 29 35
4 150 128 32 131 123 4 20 22 25 27 30 35
5 154 132 33 131 123 5 21 23 26 28 31 36
6 159 136 34 131 123 6 22 24 26 29 32 37
7 163 140 35 131 123 7 23 25 27 30 32 38
8 168 144 37 131 123 8 24 25 28 31 33 39
9 173 148 39 131 123 9 24 26 29 32 34 39
10 178 152 40 131 123 10 25 27 30 33 35 40
11 183 157 42 131 123 11 26 28 31 34 36 41
12 188 162 44 131 123 12 27 29 32 35 37 42
Offset
Standard Lateral
Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16
01 28 30 33 36 39 42 1 35 38 41 44 48 51
0 29 31 34 37 40 43 2 37 39 42 46 49 52
1 30 32 35 38 41 44 3 38 40 44 47 50 54
2 31 33 36 39 42 45 4 39 42 45 48 52 55
3 32 34 37 40 43 46 5 41 43 47 50 53 57
4 33 35 38 41 44 47 6 42 45 48 51 55 58
5 34 37 40 43 46 49 7 44 46 50 53 56 60
6 36 38 41 44 47 50 8 45 48 51 55 58 61
7 37 39 42 45 48 51 9 47 50 53 56 60 63
8 38 41 44 47 50 53 10 49 51 55 58 61 65
9 40 42 45 48 51 54 11 51 53 56 60 63 66
10 41 43 46 49 53 56 12 52 55 58 61 65 68
11 43 45 48 51 54 57
12 44 46 50 53 56 59
Neck length
Standard Lateral
Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16
01 17 20 24 28 32 36 1 26 29 33 37 41 45
0 18 21 25 29 33 37 2 27 30 34 38 42 46
1 19 22 26 30 34 38 3 28 31 35 39 43 47
2 20 23 27 31 35 39 4 29 32 36 40 44 48
3 21 23 27 31 35 39 5 30 33 37 41 45 49
4 22 24 28 32 36 40 6 32 34 38 42 46 50
5 22 25 29 33 37 41 7 33 36 40 44 48 52
6 23 26 30 34 38 42 8 34 37 41 45 49 53
7 24 27 31 35 39 43 9 35 38 42 46 50 54
8 25 28 32 36 40 44 10 37 40 44 47 52 56
9 26 29 33 37 41 45 11 38 41 45 49 53 57
10 27 30 34 38 42 46 12 39 42 46 50 54 58
11 29 31 35 39 43 47
12 30 33 37 41 45 49
17
Offset Lateral
XXL/+16
Offset Standard
XL/+12
M/L width
L/+8
M/+4
Neck height S/+0
XS/-3
Neck length
Standard
Neck length
Stem length I CCD angle
Lateral
Stem length II
18
Instrument Set
SL-PLUS™ MIA Basic Set Set No. SAP/Art.
75200838/0942010
SAP No. Art. No. Description Size
75002198 110450 Case Basic Instruments 1/4
75007661 990019 Easy Tray Lid Plastic
1 75002319 110901 Slap Hammer
2 75002320 110902 Extractor Block
3 75002325 110911 Extraction Screw M8
4 75002165 110249 Extraction Screw M6
5 75002160 110242 Ball Head Impactor
6 75007255 600621 MIA Stem Impactor
7 75004495 21000138 MIA Curved Rasp
8 75006420 41000030 MIA Guide Rasp 1-6
9 75100839* 75100839 Trial Ball Head 22 S/+0
75100840 75100840 Trial Ball Head 22 M/+4
75100841 75100841 Trial Ball Head 22 L/+8
75100842* 75100842 Trial Ball Head 22 XL/+12
75100843* 75100843 Trial Ball Head 28 XS/-3
75100844 75100844 Trial Ball Head 28 S/+0
75100845 75100845 Trial Ball Head 28 M/+4
75100846 75100846 Trial Ball Head 28 L/+8
75100847 75100847 Trial Ball Head 28 XL/+12
75100848 75100848 Trial Ball Head 28 XXL/+16
75100849* 75100849 Trial Ball Head 32 XS/-3
75100850 75100850 Trial Ball Head 32 S/+0
75100851 75100851 Trial Ball Head 32 M/+4
75100852 75100852 Trial Ball Head 32 L/+8
75100853 75100853 Trial Ball Head 32 XL/+12
75100854 75100854 Trial Ball Head 32 XXL/+16
75100855* 75100855 Trial Ball Head 36 XS/-3
75100856 75100856 Trial Ball Head 36 S/+0
75100857 75100857 Trial Ball Head 36 M/+4
75100858 75100858 Trial Ball Head 36 L/+8
75100859 75100859 Trial Ball Head 36 XL/+12
* Optional Instruments
Optional:
SAP No. Art. No. Description Size
75210292 75210292 SET 40 mm Trial Ball Head XS/-4 to L/+8
75210293 75210293 SET 44 mm Trial Ball Head XS/ 1-6 to L/+8
75000645 21000397 MIA Curved Rasp 45°
75004615 21000265 MIA Start Rasp
75004683 800-09-001 MIA Rasping Machine (IMT Woodpecker)*
*Manufacturer:
IMT Integral Medizintechnik AG,
Haldenstrasse 5, 6006 Luzern,
Switzerland
19
8
7
1
4
3
9
2 5
6
20
SL-PLUS™ MIA Trial Rasps Set Set No. SAP
75210409
Set No. SAP No. Description Size
75007312 600930 MIA Instrument Case
75007661 990019 Easy Tray Lid Plastic
1 75004481 21000123 MIA Detachable Rasp 01
75004482 21000124 MIA Detachable Rasp 0
75004483 21000125 MIA Detachable Rasp 1
75004484 21000126 MIA Detachable Rasp 2
75004485 21000127 MIA Detachable Rasp 3
75004486 21000128 MIA Detachable Rasp 4
75004487 21000129 MIA Detachable Rasp 5
75004488 21000130 MIA Detachable Rasp 6
75004489 21000131 MIA Detachable Rasp 7
75004490 21000132 MIA Detachable Rasp 8
75004491 21000133 MIA Detachable Rasp 9
75004492 21000134 MIA Detachable Rasp 10
75004493* 21000135 MIA Detachable Rasp 11
75004494* 21000136 MIA Detachable Rasp 12
2 75004603 21000253 MIA Modular Neck for Detachable Rasp Std 01–0
75004604 21000254 MIA Modular Neck for Detachable Rasp Std 1-6
75004605 21000255 MIA Modular Neck for Detachable Rasp Std 7-12
3 75004606 21000256 MIA Modular Neck Detachable Rasp lat. 1–6
75004607 21000257 MIA Modular Neck Detachable Rasp lat. 7-12
4 75006419 41000029 MIA Box Chisel
5 75009352* SYS251374 Trochanter Retractor
*Optional Instruments
Optional Rasp Adapters
Set No. SAP No. Art No. Description Size
75200166 75007307 600920 MIA Offset Adapter 25mm
75200168 75007308 600921 MIA Offset Adapter 40mm
6 75200169 75007309 600922 MIA Offset Adapter 10mm
7 75007310 600923 MIA Double Offset Adapter Left 17/13mm
75200171
8 75007311 600924 MIA Double Offset Adapter Right 17/13mm
75004612 21000262 MIA Double Offset Adapter Left 60/25mm
75210202
75004613 21000263 MIA Double Offset Adapter Right 60/25mm
75210372 75004614 21000264 MIA Offset Adapter 45°
75210406 75004468 21000107 MIA Offset Adapter with Plate 25mm
75000642 21000378 Knock Plate
21
4
5
OPTIONAL
1
OPTIONAL
OPTIONAL
3 2
7
6
8
22
Notes
References
1. Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) 2021. Automated Industry Report (AIRS) ID 4082 -
SL-PLUS INTEGRATION PLUS MIA Stem (Combined Standard and Lateral). Adelaide.
2. Affatato S, Comitini S, Fosco M, Toni A, Tigani D. Radiological identification of Zweymüller-type femoral stem prosthesis in revision
cases. Int Orthop. 2016;40(11):2261-2269.
3. Smith+Nephew 2019. Clinical Obserservation Study of the Hydroxylapatite-Coated SL-PLUS™ MIA Hip Stem. Internal Report. Study
D10080-2.
4. Swiss National Joint Registry (SIRIS) 2021. SIRIS Industry Report for Smith+Nephew SL-PLUS INTEGRATION PLUS MIA Stem. Bern.
5. Endoprothesenregister Deutschland (EPRD) 2021. Data Summary for SL-PLUS INTEGRATION MIA Stem (H1.5 Femoral Component).
Berlin.
6. Reiner T, Sonntag R, Kretzer JP, et al. The migration pattern of a cementless hydroxyapatite-coated titanium stem under immediate full
weight-bearing—a randomized controlled trial using model-based rsa. J Clin Med. 2020;9(7):1-10.
7. Tanaka A, Kaku N, Tabata T, Tagomori H, Tsumura H. Comparison of early femoral bone remodeling and functional outcome after total
hip arthroplasty using the SL-PLUS MIA stem with and without hydroxyapatite coating. Musculoskelet Surg. 2019;104(3):313-320.
8. Junk-Jantsch S, Pflüger G. MIS and the Demands on Bearing Couples. In: Bioceramics and Alternative Bearings in Joint Arthroplasty:
Springer; 2007.
9. Pflüger G, Junk-Jantsch S, Schöll V. Adaptation of the Zweymüller stem (SL-PLUS® stem) for minimally invasive hip surgery: Experiences
after 1½ year. EFORT-Poster. 2007.
10. Brandenberg J, Steiger U. Der SL-PLUS-MIA-Schaft-ergebnisse nach 1000 Implantationen; The SL-Plus-Mia-stem-results after 1000
implantations. OUP. 2012;1(5):208.
11. Strnad Z, Strnad J, Povysil C, Urban K. Effect of plasma-sprayed hydroxyapatite coating on the osteoconductivity of commercially pure
titanium implants. Int J Oral Maxillofac Surg. 2000;15(4):483-490.
12. Zheng X, Huang M, Ding C. Bond strength of plasma-sprayed hydroxyapatite/Ti composite coatings. Biomaterials. 2000;21(8):841-849.
Products may not be available in all markets because product availability is subject to the regulatory and/or medical practices in individual markets.
Please contact your Smith+Nephew representative or distributor if you have questions about the availability of Smith+Nephew products in your area.
www.smith-nephew.com
Manufacturer
Smith & Nephew Orthopaedics AG ™Trademark of Smith & Nephew.
Theilerstrasse 1A All trademarks acknowledged
6300 Zug ©2024 Smith & Nephew Supporting healthcare professionals for over 150 years
Switzerland 31156-en V2 11/24
0123
Surgical Technique
INTENDED FOR NON-US USE ONLY
Design surgeon list
Smith & Nephew thanks the following surgeons for their participation
as part of the R3™ system design team:
Robert Barrack, MD John L. Masonis, MD
St. Louis, Missouri OrthoCarolina
Hip & Knee Center
Robert Bourne, MD Charlotte, North Carolina
London Health Sciences Center
London, Ontario, Canada Henrik Malchau, MD
Associate Professor, Harvard Medical School
Jonathan Garino, MD Co-Director, The Harris Orthopaedic Biomechanics
University of Pennsylvania and Biomaterials Laboratory
School of Medicine Massachusetts General Hospital
Philadelphia, Pennsylvania Boston, Massachusetts
Wayne M. Goldstein, MD Michael Ries, MD
Clinical Professor of Orthopaedics University of California
University of Illinois at Chicago San Francisco, California
Illinois Bone and Joint Institute
Chicago, Illinois Cecil Rorabeck, MD
Professor of Orthopaedic Surgery
Richard Kyle, MD University of Western Ontario
Minneapolis, Minnesota London, Ontario, Canada
Stephen J. McMahon, MB, BS, FRACS(Orth), FA(Orth)A
Senior Lecturer Monash University
Malabar Orthopaedic Clinic
Melbourne, Australia
Nota Bene: The technique description herein is made available to the healthcare professional to
illustrate the authors’ suggested treatment for the uncomplicated procedure. In the final analysis,
the preferred treatment is that which addresses the needs of the patient. For more information on
the products in this surgical technique, including indications for use, contraindications, effects,
precautions and warnings, please consult the products’ Instructions for Use (IFU).
Short technique
1. Preoperative planning 4. Shell insertion
2.Acetabular reaming 5. Acetabular screw insertion
3.Acetabular trialing 6. Acetabular poly liner insertion
Preoperative planning
Preoperative X-Rays should include an A/P of the
pelvis centered over the symphysis and an A/P
and lateral of the affected hip.
Templating can be done on the affected side, but
it is important that the contralateral hip also be
templated to verify the size.
To ensure a congruent fit, the acetabular
component should be medialized to the medial
aspect of the acetabulum, as indicated by the
teardrop.
The center of rotation also should be marked for
subsequent reference.
Surgical tips:
• To minimize the need of assistance,
each of the acetabular retractors can
be tied directly to a Charnley retractor.
• Dividing the transverse acetabular
Acetabular exposure ligament will allow reaming to begin
inferiorly, preventing the tendency of
the reamer to migrate superiorly.
Complete exposure of the acetabulum is required, • Removal of soft tissue and
overhanging osteophytes from the
regardless of the type of approach. Use the
foveal notch aids visualization of the
approach with which you are most familiar and quadralateral plate and the depth that
achieve the best surgical results. the acetabulum should be reamed.
First, resect the acetabular labrum and place a
blunt retractor anteriorly.
After identifying the transverse acetabular
ligament, place a blunt retractor around the inferior
margin of the acetabulum.
Depending on the exposure, a third retractor can
be placed posteriorly following the excision of the
labrum.
Remove all overhanging soft tissue and
osteophytes in order to visualize the entire
acetabular socket.
The acetabulum should be medialized to restore
the normal center of hip rotation.
4
Acetabular reaming
Select an acetabular reamer that is considerably Surgical tips:
smaller than the templated size of the cup. • Each successive reamer must be fully
Generally, reaming 6-8mm lower than the seated within the acetabulum. Failure
templated size is suitable. to do so will result in lateralization of
the trial and exposure of the porous
Position the initial reamer in a vertical direction coating. If lateralization occurs, go back
(1) to ensure the reamer is taken down to the to a smaller reamer and begin again,
medial wall. checking each size to ensure that the
reamers are fully seated.
Direct the second reamer and all subsequent • Increasing the reamer size by 2mm
reamers in approximately 45° of abduction is recommended, although in smaller
and 20° of anteversion for final position of the patients 1mm increments may be
acetabular component (2). preferred.
• Mark the medial wall with an
Preserve subchondral bone to provide good electrocautery prior to using the last
support for the prosthesis. This might mean reamer. If the last reamer does not
the reamer will not be medialized all the way to remove the mark, repeat reaming,
the inner wall. One might suggest leaving some dropping back a size if necessary.
remaining subchondral bone and removing the
medial bone that is osteophyte and is covering
fatty tissue.
Instrument tips:
Frequently palpate the posterior and anterior
walls of the acetabulum during the reaming • The acetabular reamer has an open
process as these walls will determine the largest back, which helps visualize reaming and
acetabular size that can be accommodated. Avoid allows easy access to bone chips. This
style of reamer is hemispherical and
allowing the reamer to drift posteriorly where the when fully seated it should be covered
bone might be less dense and the path of least by the rim of the acetabulum.
resistance for the reamer.
• Gently rock reamer handle back and
To press-fit an R3™ acetabular shell the acetabulum forth approximately 5° for last size used
can either be under-reamed by 1mm or may be only to ensure rim is accurate for the
desired press-fit.
reamed line-to-line depending on the bone quality
and size of the acetabulum.
5
Acetabular trialing
After preparing the acetabulum, the trial shell Surgical tip:
should be inserted to verify size and position of • The bone at the edge of the trial shell
the cup. The surgeon should note the appropriate can be marked with an electrocautery to
orientation of the acetabular trial to position the help in final component positioning.
cup correctly.
A trial liner insert cannot be inserted into a trial Instrument tip:
shell for trial reduction.
• The trial shells are the exact size
If trial reduction using a trial insert is desired specified. They can be used to assess
the accuracy of reaming or can be
at this time, then the preparation of the femur
press-fit into the acetabulum if using a
should occur up until the trial reduction stage. larger size than the final reamer.
The surgeon then has the option of inserting a
trial acetabular liner (preferred) in the acetabular
implant for subsequent leg length, offset and
stability assessments or the real acetabular insert.
6
Acetabular shell insertion
Select the appropriate acetabular implant, attach
the shell to the cup positioner/impactor and insert
it into the acetabulum.
Surgical tips:
Rotate the X-Bar shaft so that it is in line with the
liner removal slot. For the THREE HOLE cup, this • The change in pitch that occurs as the
shell is seated against the medial wall
positions the three holes in the superior direction.
is often audible. A depth gauge can be
When implanting a MULTI HOLE shell, future access inserted through the screw holes and
to the liner removal slot should be considered. apex hole to determine the adequacy of
shell seating.
Position the X-Bar so that the vertical bar is
• Close attention should be paid to initial
perpendicular to the long axis of the body and the
positioning of the R3 shell. However, the
appropriate crossbar (left or right) aligns with the use of the slap hammer may be helpful
long axis of the body. in extracting the shell for repositioning.
Firmly tap the inserter with a mallet until the cup is
fully seated.
Gently toggle the impactor handle to assess the
stability and contact of the shell.
Instrument tips:
Remove the X-Bar, then disengage the impactor • The plastic tip on the cup impactor is
handle and look through the impactor hole to removable for cleaning, or replacement if
judge the distance between the medial wall and damaged.
the shell.
• The X-Bar references 45° of abduction
and 20° of anteversion .
If the cup is firmly seated, there should be no gap
between the shell and the medial wall and no
apparent movement in the component.
Specific to shells for R3™ acetabular
ceramic liners:
Proper range of motion is critical for implant
longevity. If any repositioning of the shell is
required, it should only be performed using the
shell positioner. Any use of a punch, osteotome
or other instrument on the shell’s rim could result
in damage to the taper section and compromise
the integrity of the shell and ceramic liner mating
and lead to liner fracture. It is important to protect
the shell’s rim and inner taper from any damage
during implantation.
7
Acetabular screw insertion
Screw fixation is simple, fast and the most Surgical tip:
common method of assuring additional fixation. • Screws have been shown to be a
Acetabular screws work in compression, which reliable method of assuring fixation;
allows the shell to fully seat in the acetabular however, it is important to avoid
cavity. neurovascular complications by proper
screw placement, avoiding the anterior/
For screw fixation, each screw hole must be superior or anterior/inferior quadrants.
predrilled. Using the variable angle drill guide,
adjust the angle of the tip to align with the
selected screw hole and press firmly in the shell.
After drilling the hole, use the depth gauge to
verify appropriate screw length(s).
Use the screw forceps to hold the screw. Attach
the ball-joint or flexible screwdriver shaft to the
end of the screw. Then introduce the screw
into the hole and screw it into place using the
ratcheting screwdriver handle. Make sure the
screw is fully seated within the screw hole so that
it will not impinge on the acetabular shell/liner.
8
R3™ Acetabular Liner insertion
A trial reduction should be performed with the final Surgical tips:
shell and broach in place to appropriately assess • Running a finger around the
head length, stem offset, liner style and position. circumference of the shell and a visual
With XLPE liners, use of ‘skirted’ modular heads check will help determine if the liner is
should be avoided when possible to maximize flush with the shell face.
range of motion. • The XLPE liner requires an impaction
force between 60 and 120 pounds,
Before inserting the R3 acetabular liner, lavage increasing with the diameter of the shell.
any unused holes and insert the hole covers, if
desired. Using the angled hole cover inserter, • The XLPE liner can be removed and
repositioned once without compromising
place screw hole covers over any remaining screw
the locking mechanism of the liner. To
holes and then impact with the peg impactor. remove R3 liners, insert the liner removal
Cover the apex hole with the threaded hole cover. tool fully into the removal slot and pry or
Using the straight screwdriver, screw in the hole impact the liner loose.
cover until it stops and is flush with the inner
diameter of the shell.
For XLPE liner insertion, screw the appropriate
Instrument tips:
sized liner impactor head on the end of the cup
impactor handle and ensure that the tabs on the • The liner trials are designed with flexible
liner are aligned with the indentions in the shell. locking tabs around the periphery that
Ensure all soft tissue and osteophytes have been is a quick-snap design. Align tabs with
indentions and snap into position.
removed from the periphery of the shell to avoid
Do not force trial. The trial liners are
interference with the liner lock. removed with the trial liner removal tool
via the removal slot at the apex of the
Wipe the shell ID with a lap sponge or gauze trial liner and a clockwise twist of the
until clean and dry. Press the liner impactor removal tool.
firmly until liner is partially locked. Then use light,
repetitive impacts with the mallet until the liner is
fully seated.
Inspect the liner/shell interface for proper seating.
The liner should sit flush with the face of the shell.
9
R3™ Acetabular Liner insertion continued
R3 Hard Bearing insertion
R3 Hard Bearings come preassembled with a
disposable single-use Hard Bearing Alignment
Guide. The liner/alignment guide assembly is then
introduced by hand and sits flush on the face of
the shell. The liner must be checked for proper
orientation. Verification of proper liner seating in
the shell should be confirmed by both a visual
check to see that the insertion ring is sitting on the
shell face and a manual check with the fingers to
feel that the ring does not rock on the face of the
shell. Do not impact the liner if it is not oriented Surgical tip:
properly, as this can damage the shell and/or • It may prove helpful to rotate the liner/
locking mechanism. Once orientation has been guide slightly to ensure soft tissues and
osteophytes are clear.
confirmed, impact the liner into place using the
appropriate sized liner impactor head placed on
the shell positioner/impactor. Once impacted,
the alignment guide will disengage onto the shell
positioner/impactor and should be removed at that
time.*
* Cautionary Statement
Be sure to remove the disposable Hard
Bearing Alignment Guide. It is not intended for
implantation.
In the event that the Hard Bearing Alignment
Guide is disengaged from the liner, the alignment
guide should be reassembled to the liner before
implantation. This is accomplished by taking the
disposable alignment guide and placing it upside
down on the back table. The liner can then be
placed upside down on the alignment guide such
that the peripheral rim is sitting on the alignment
guide. Simply push the liner onto the guide until
the insertion ring locks snugly on the liner. The
assembly is ready for placement in the shell.
10
Specific to R3™ ceramic liners
Use extreme care in handling and storage
of ceramic implant components. Damage to
components may induce internal stresses that
are not obvious to the observer, and it may lead
to premature failure of the component. Before
use of ceramic implants, carefully examine each
component for indications of damage that may
have occurred during shipping or prior in-hospital
handling. All surfaces should be smooth without
pitting, scratches or other surface irregularities.
Only Smith & Nephew ceramic femoral heads can Surgical tip:
be used with the R3 ceramic acetabular liners. • Should a correction or revision of a R3
Do not mix the ceramic liner or ceramic head ceramic liner be necessary, a new R3
with any other manufacturer’s acetabular shell or ceramic insert must be used.
stem, respectively. Use the appropriate size head
• The ceramic liner can be removed by
and liner only. A sizing mismatch may result in placing the liner removal tool in the
premature implant failure. Once the liner or the removal slot and prying (or impacting if
head are impacted, the ridges machined into the necessary) the liner loose.
metal taper deform. If, for any reason, the ceramic
femoral head is removed, the metal stem taper
cannot be reused with a ceramic component. If
the R3 ceramic liner is removed, a new R3 ceramic
liner must be used.
Head compatibility with liners
XLPE Liners may be used with CoCr, OXINIUM™
and ceramic heads. Constrained Liners may be
used with CoCr, and OXINIUM heads. Ceramic
liners may only be used with ceramic heads.Use
the appropriate size head for the liner as a sizing
mismatch may result in premature implant failure.
11
Hard Bearing Liner insertion
1
2
3
4
12
Shell and liner offerings
XLPE Ceramic
Cups 22 28 32 36 40 44 32 36
40
42
44
46
48
50
52
54
56
58
60
62
64
66
68
70
72
74
76
78
80
Range of motion1,2†*
(SPECTRON™ stem [size 3], +4 head offset)
REFLECTION™ Acetabular System with XLPE liner
22mm 28mm 32mm 36mm
0º 133º 142º 145º 148º
20º 112º 122º 126º 128º
R3™ Acetabular System with XLPE liner
22mm 28mm 32mm 36mm 40mm 44mm
0º 140º 150º 154º 157º 157º 157º
20º 132º 134º 136º 138º 138º 138º
* Individual results in patients may vary.
† Based on computer modelling.
13
Poly thickness chart
Poly Poly
Poly Thickness Poly Thickness
Thickness Load- Thickness Load-
Taper Bearing Taper Bearing
Region Region Region Region
Shell OD Poly OD mm mm Shell OD Poly OD mm mm
40 22 5.5 6.1 62 32 11.3 12.1
42 22 6.5 7.1 62 36 9.3 10.1
44 22 7.5 8.1 62 40 7.5 8.0
46 28 5.4 6.1 62 44 5.3 6.0
48 28 6.4 7.1 64 36 10.3 11.1
48 32 4.3 5.1 64 40 8.4 9.0
50 28 7.3 8.1 64 44 6.4 7.0
50 32 5.3 6.1 66-70 36 11.3 12.1
52 28 8.3 9.1 66-70 40 9.3 10.0
52 32 6.3 7.1 66-70 44 7.2 8.0
52 36 4.3 5.1 72-74 36 13.8 14.0
54 28 9.3 10.1 72-74 40 11.8 12.0
54 32 7.3 8.1 72-74 44 9.8 10.0
54 36 5.3 6.1 76-80 36 15.8 16.0
56 28 10.3 11.1 76-80 40 13.8 14.0
56 32 8.3 9.1 76-80 44 11.8 12.0
56 36 6.3 7.1
56 40 4.6 5.0
58 28 11.3 12.1
58 32 9.3 10.1 Taper region
58 36 7.3 8.1
58 40 5.3 6.0
60 28 12.3 13.1
60 32 10.3 11.1
60 36 8.3 9.1
60 40 6.5 7.0 Load-bearing
60 44 4.3 5.0
14
Catalog
R3™ NO HOLE Acetabular Shells
Standard size shells Small size shells
Cat. no. ODmm Cat. no. ODmm
71331846 46 71331840 40
71331848 48 71331842 42
71331850 50 71331844 44
71331852 52
71331854 54 Large size shells
71331856 56 Cat. no. ODmm
71331858 58 71331866 66
71331860 60 71331868 68
71331862 62
71331864 64
R3 NO HOLE HA Acetabular Shells
Standard size shells Small size shells
Cat. no. ODmm Cat. no. ODmm
71332246 46 71332240 40
71332248 48 71332242 42
71332250 50 71332244 44
71332252 52
71332254 54 Large size shells
71332256 56 Cat. no. ODmm
71332258 58 71332266 66
71332260 60 71332268 68
71332262 62
71332264 64
R3 THREE HOLE Acetabular Shells
Standard size shells Small size shells
Cat. no. ODmm Cat. no. ODmm
71335546 46 71335540 40
71335548 48 71335542 42
71335550 50 71335544 44
71335552 52
71335554 54 Large size shells
71335556 56 Cat. no. ODmm
71335558 58 71335566 66
71335560 60 71335568 68
71335562 62
71335564 64
15
R3™ THREE HOLE HA Acetabular Shells
Standard size shells Small size shells
Cat. no. ODmm Cat. no. ODmm
71331946 46 71331940 40
71331948 48 71331942 42
71331950 50 71331944 44
71331952 52
71331954 54 Large size shells
71331956 56 Cat. no. ODmm
71331958 58 71331966 66
71331960 60 71331968 68
71331962 62
71331964 64
R3 MULTI HOLE Acetabular Shells
Standard size shells Large size shells
Cat. no. ODmm Cat. no. ODmm
71338663 48 71338673 66
71338664 50 71338674 68
71338665 52 71338675 70
71338666 54
71338667 56 Jumbo size shells
71338668 58 Cat. no. ODmm
71338669 60 71338676 72
71338671 62 71338677 74
71338672 64 71338678 76
71330009 78
71330010 80
R3 HA MULTI HOLE Acetabular Shells
Standard size shells Large size shells
Cat. no. ODmm Cat. no. ODmm
71338733 48 71338742 66
71338734 50 71338743 68
71338735 52 71338744 70
71338736 54
71338737 56 Jumbo size shells
71338738 58 Cat. no. ODmm
71338739 60 71338745 72
71338740 62 71338747 74
71338741 64 71338749 76
71331138 78
71331139 80
16
R3™ XLPE Acetabular Liners
0° XLPE liner 20° XLPE liner 0° +4 XLPE liner 20°+4 XLPE liner
ID OD Cat. no. Cat. no. Cat. no. Cat. no.
22 40 71334840 71334940 71335840 71337140
22 42 71334842 71334942 71335842 71337142
22 44 71334844 71334944 71335844 71337144
28 46 71337546 71334946 71335946 71337746
28 48 71337548 71334948 71335948 71337748
28 50 71337550 71334950 71335950 71337750
28 52 71337552 71334952 71335952 71337752
28 54 71337554 71334954 71335954 71337754
28 56 71337556 71334956 71335956 71337756
28 58 71337558 71334958 71335958 71337758
28 60 71337560 71334960 71335960 71337760
32 48 71339548 71337648 71336648 71337948
32 50 71339550 71337650 71336650 71337950
32 52 71339552 71337652 71336652 71337952
32 54 71339554 71337654 71336654 71337954
32 56 71339556 71337656 71336656 71337956
32 58 71339558 71337658 71336658 71337958
32 60 71339560 71337660 71336660 71337960
32 62 71339562 71337662 71336662 71337962
36 52 71332752 71335752 71336952 71338552
36 54 71332754 71335754 71336954 71338554
36 56 71332756 71335756 71336956 71338556
36 58 71332758 71335758 71336958 71338558
36 60 71332760 71335760 71336960 71338560
36 62 71332762 71335762 71336962 71338562
36 64 71332764 71335764 71336964 71338564
36 66-70 71330766 71331266 71331566 71332666
36 72-74 71338686 71338694 71338703 71338712
36 76-80 71331103 71331112 71331114 71338946
17
R3™ XLPE Acetabular Liners (continued)
0° XLPE liner 20° XLPE liner 0° +4 XLPE liner 20°+4 XLPE liner
ID OD Cat. no. Cat. no. Cat. no. Cat. no.
40 56 71338679 71338687 71338695 71338704
40 58 71338680 71338688 71338696 71338705
40 60 71338681 71338689 71338697 71338706
40 62 71338682 71338690 71338698 71338707
40 64 71338683 71338691 71338699 71338708
40 66-70 71338684 71338692 71338701 71338709
40 72-74 71338685 71338693 71338702 71338711
40 76-80 71331094 71331104 71331113 71331116
44 60 71331096 71331106 71330011 71331118
44 62 71331097 71331107 71330012 71331119
44 64 71331098 71331108 71330013 71331121
44 66-70 71331099 71331109 71330014 71331122
44 72-74 71331101 71331110 71330016 71331123
44 76-80 71331102 71331111 71330017 71331124
18
R3™ INTL Delta Ceramic Liners*
ID OD Cat. no.
32 48 71331748
32 50 71331750
36 52 71331752
36 54 71331754
36 56 71331756
36 58 71331758
36 60 71331760
36 62 71331762
36 64 71331764
36 66/68 71331766
* For Use with Smith & Nephew BIOLOX® Delta and BIOLOX Forte Ceramic Heads only.
19
R3™ Trial Shells
Standard size trial shells Small size trial shells
Cat. no. ODmm Cat. no. ODmm
71360745 45 71360739 39
71360746 46 71360740 40
71360747 47 71360741 41
71360748 48 71360742 42
71360749 49 71360743 43
71360750 50 71360744 44
71360751 51
71360752 52 Large size trial shells
71360753 53 Cat. no. ODmm
71360754 54 71360765 65
71360755 55 71360766 66
71360756 56 71360767 67
71360757 57 71360768 68
71360758 58 71366524 69
71360759 59 71366525 70
71360760 60
71360761 61 Jumbo size trial shells
71360762 62 Cat. no. ODmm
71360763 63 71366526 71
71360764 64 71366527 72
71366528 73
71366529 74
71366530 75
71366531 76
71362019 77
71362020 78
71362021 79
71362022 80
20
Catalog
R3™ Poly Trial Liners
20° XLPE trial
0° XLPE trial liner liner 0° +4 XLPE trial 20°+4 XLPE trial
ID OD Cat. no. Cat. no. liner Cat. no. liner Cat. no.
22 40 71360540 71365340 71366140 71368640
22 42 71360542 71365342 71366142 71368642
22 44 71360544 71365344 71366144 71368644
28 46 71360546 71366446 71368346 71368746
28 48 71360548 71366448 71368348 71368748
28 50 71360550 71366450 71368350 71368750
28 52 71360552 71366452 71368352 71368752
28 54 71360554 71366454 71368354 71368754
28 56 71360556 71366456 71368356 71368756
28 58 71360558 71366458 71368358 71368758
28 60 71360560 71366460 71368360 71368760
32 48 71365148 71366548 71368448 71368848
32 50 71365150 71366550 71368450 71368850
32 52 71365152 71366552 71368452 71368852
32 54 71365154 71366554 71368454 71368854
32 56 71365156 71366556 71368456 71368856
32 58 71365158 71366558 71368458 71368858
32 60 71365160 71366560 71368460 71368860
32 62 71365162 71366562 71368462 71368862
36 52 71365252 71367952 71368552 71369152
36 54 71365254 71367954 71368554 71369154
36 56 71365256 71367956 71368556 71369156
36 58 71365258 71367958 71368558 71369158
36 60 71365260 71367960 71368560 71369160
36 62 71365262 71367962 71368562 71369162
36 64 71365264 71367964 71368564 71369164
36 66-70 71365266 71367966 71368566 71369166
36 72-74 71366571 71366574 71366577 71366580
36 76-80 71362312 71362314 71362316 71362318
21
R3™ Poly Trial Liners
20° XLPE trial
0° XLPE trial liner liner 0° +4 XLPE trial 20°+4 XLPE trial
ID OD Cat. no. Cat. no. liner Cat. no. liner Cat. no.
40 56 71363420 71363422 71362030 71362035
40 58 71362023 71362026 71362031 71362036
40 60 71362024 71362027 71362032 71362037
40 62 71363421 71362028 71362033 71362038
40 64 71362025 71362029 71362034 71362039
40 66-70 71366569 71366572 71366575 71366578
40 72-74 71366570 71366573 71366576 71366579
40 76-80 71362311 71362313 71362315 71362317
44 60 71366081 71366094 71366087 71366101
44 62 71366082 71366095 71366088 71366102
44 64 71366083 71366096 71366089 71366103
44 66-70 71366084 71366097 71366091 71366104
44 72-74 71366085 71366098 71366092 71366105
44 76-80 71366086 71366099 71366093 71366106
22
R3™ Ceramic Snap in Trial Liners
ID OD Cat. no.
32 48 71369748
32 50 71369758
36 52 71369752
36 54 71369754
36 56 71369756
36 58 71369758
36 60 71369760
36 62 71369762
36 64 71369764
36 66/68 71369766
R3 Liner Impactor Heads
Cat. no. Size mm
71366428* 28
71366432* 32
71366436* 36
71366438* 38-42
71366444* 44-48
71366451* 50-54
*Exclusively for liner impaction
Note: When using a 22mm Liner, you will need to order a Reflection 22mm Impactor Head (71362222) separately.
R3 MIS Instruments
Cat. no. Description
71368569 Offset Shell Impactor
71366052 Offset X-Bar
71363077 Offset Impactor Tip
71364073 Offset Reamer Handle
23
Catalog
R3™ Straight Shell Impactor
Cat. no. 71364450
R3 Impactor Replacement Tip
Cat. no. 71368570
R3 Depth Gauge
Cat. no. 71364451
X-Bar
Cat. no. MT-2201
Screw Forceps
Cat. no. 71362298
Ball Joint Screwdriver
Cat. no. 71362295
R3 Variable Angle Drill Guide
Cat. no. 71364477
Reamer Handle
Cat. no. 71362279
Flexible Screw Drills
Cat. no. Length mm
71362915 15
71362925 25
71362935 35
71362950 50
Captured Flexible
Screwdriver Shaft
Cat. no. 71362291
Captured U-Joint
Screwdriver Shaft
Cat. no. 71362292
R3 Surgical Templates
sizes 40-68 (not shown)
Cat. no. 71380666
R3 Surgical Templates
sizes 70-80 (not shown)
Cat. no. 71381508
24
R3™ Trial Liner Removal Tool
Cat. no. 71364455
R3 Liner Removal Tool
Cat. no. 71366021
Hole Cover Impactor
Cat. no. 73-2117
Trial Shell Handle
Cat. no. 71362297
Flexible Screwdriver
Cat. no. 71362290
Ratchet Handle
Cat. no. 71362294
Small Slap Hammer
Cat. no. 71367541
REFLECTION™ Mallet
Cat. no. 71362106
Hole Cover Inserter
Cat. no. 73-2133
Straight Screwdriver Shaft
Cat. no. 71362293
Power Adaptors (not shown)
Cat. no. 71362781 Synthes
71362782 Asculap
71362783 Hudson
25
Catalog
Reamer Domes
Standard size Small size
Cat. no. Size mm Cat. no. Size mm
71362742 42 71362738 38
71362743 43 71362739 39
71362744 44 71362740 40
71362745 45 71362741 41
71362746 46
71362747 47 Large size
71362748 48 Cat. no. Size mm
71362749 49 71362765 65
71362750 50 71362766 66
71362751 51 71362767 67
71362752 52 71362768 68
71362753 53 71362769 69
71362754 54 71362770 70
71362755 55 71362771 71
71362756 56 71362772 72
71362757 57 71362773 73
71362758 58 71362774 74
71362759 59 71362775 75
71362760 60 71362776 76
71362761 61 71362777 77
71362762 62 71362778 78
71362763 63 71362779 79
71362764 64 71362780 80
26
R3™/REFLECTION™ Threaded Hole
Cover
Cat. no. 71336500
Spherical Head Screws
Cat. no. Length mm
71332515 15
71332520 20
71332525 25
71332530 30
71332535 35
71332540 40
71332545 45
71332550 50
71332555 55
71332560 60
71332565 65
71332570 70
R3 Screw Hole Cover
Cat. no. 71369894
Small Outer Case
Cat. no. 71129401 (not shown)
Lid for Outer Case
Cat. no. 71129402 (not shown)
R3 Trial Shell Tray
Cat. no. 71362213 (not shown)
R3 Main Instrument Tray
Cat. no. 71362211 (not shown)
R3 MIS Instrument Tray
Cat. no. 71362219 (not shown)
R3 Primary Reamer Dome Tray
Cat. no. 71362212 (not shown)
R3 CDH Trial Tray
Cat. no. 71361077 (not shown)
27
Catalog
R3™ Jumbo Trial Shell & Reamer Set
Cat. no. 71362230
R3 XL Revision Reamer Set
Cat. no. 71362160
R3 XL Revision Trial Shell Set
Cat. no. 71362170
R3 Jumbo Trial Liner Set 66-70mm
Cat. no. 71362090
R3 XL 0 Deg. Trial Liner Set 71-80mm
Cat. no. 71362180
R3 XL 20 Deg. Trial Liner Set 71-80mm
Cat. no. 71362190
R3 0 Deg. Multi ID Trial Liner Set 46-64mm
Cat. no. 71360675
R3 20 Deg. Multi ID Trial Liner Set 46-64mm
Cat. no. 71360681
R3 0 Deg. +4 Multi ID Trial Liner Set 46-64mm
Cat. no. 71360676
R3 20 Deg. +4 Multi ID Trial Liner Set 46-64mm
Cat. no. 71360683
R3 Ceramic Trial Liner Set 48-64mm
Cat. no. 71360685
R3 0° 40mm Trial Liner Tray
Cat. no. 71360782
R3 20° 40mm Trial Liner Tray
Cat. no. 71360783
R3 0° +4 40mm Trial Liner Tray
Cat. no. 71360784
R3 20° +4 40mm Trial Liner Tray
Cat. no. 71360786
R3 0° and 20° 44mm Permanent Trial Liner Tray 60-70mm
Cat. no. 71366981
R3 0° and 20° +4 44mm Permanent Trial Liner Tray 60-70mm
Cat. no. 71366982
R3 0° and 20° +4 44mm Permanent Trial Liner Tray 72-80mm
Cat. no. 71366983
28
Notes
Notes
References
1. Barrack RL, Thornberry RL, Ries MD, Lavernia C, Tozakoglou E. The Effect of Component Design on Range of Motion to Impingement in Total Hip Arthroplasty. AAOS
Instructional Course Lectures. 2001;50:275 - 280.
2. Thornberry RL, Lavernia CJ, Barrack RL, Tozakoglou E. The Effects of Neck Geometry in Acetabular Design on Motion to Impingement. Paper presented at: AAHKS
Eighth Annual Meeting; 1998.
Products may not be available in all markets because product availability is subject to the regulatory and/or medical practices in individual markets.
Please contact your Smith & Nephew representative or distributor if you have questions about the availability of Smith & Nephew products in your area.
Smith & Nephew, Inc. www.smith-nephew.com
7135 Goodlett Farms Parkway
Cordova, TN 38016
USA
Telephone: 1-901-396-2121
Information: 1-800-821-5700
Orders and Inquiries: 1-800-238-7538
©2022 Smith & Nephew
™ Trademark of Smith & Nephew. All rights reserved.
All Trademarks acknowledged. 01069 V4 71381560 REVC 12/22
SP-CL HX
Anatomically Adapted Cementless Hip System
Explanation of Pictograms
Manufacturer Article number
Product meets the applicable requirements, which are regulated in
Material (number)
the EU harmonization legislation for the affixing of the CE marking.
Contents
SP-CL HX
Anatomically Adapted Cementless Hip System
Surgical Technique
02 Preoperative Planning
05 Surgical Approaches
06 Surgical Technique
Implants
10 SP-CL Hip Stems, standard, CCD 126°
11 SP-CL Hip Stems, standard, CCD 135°
12 SP-CL Hip Stems, lateralized, CCD 126°
Instruments
13 Instrument Set for SP-CL Hip System
17 Handle for Compressor Stems
18 Additional Instruments
Accessories
18 Instructions for Cleaning and Maintenance, X-ray Templates
19 Literature
20 Indications / Contraindications
Important Information
01
Surgical Technique
Preoperative Planning
Preoperative planning is conducive towards optimal When evaluating the X-rays, it is important to factor in
surgical outcomes by ensuring the most appropriate any magnification incurred (Abb. 1). Two factors are
implants are selected for the patient. The key ob- decisive:
jectives involved are the correct positioning of the
central rotational point of the hip, correct leg length 1) Focal distance
and finally preservation or restoration of sufficient Focal spot X-ray tube x Film cassette
soft tissue tension by avoiding medialisation of the A focal distance of 100 cm gives magnification of
femur. about 10%.
Achieving anatomically appropriate CCD or neck 2) Object film distance
angle and head-neck length are of paramount im- Femoral axis x Film cassette
portance. The SP-CL system offers the CCD angles
126° and 135° as well as a 126° lateralized version.
This combined with femoral heads with up to four Focal spot X-ray tube
head-neck lengths allows the surgeon great
flexibility.
Planning should ideally be based on two X-rays:
Focal distance
an AP film of the pelvis and a mediolateral X-ray
of the hip in question. When performing the pelvic
2
X-ray it is important to ensure that:
Object film distance
1
1. Both femurs are shown in their entirety.
Film level
2. The femurs are straight and parallel and, if
(Cassette)
possible, internally rotated approximately 5° in
that position.
Fig. 1
3. Key landmarks needed for planning are visible:
the inferior margins of the obturator foramen and
of the acetabular teardrop.
02
Surgical Technique
Practical steps
First, geometrical measurements are taken on the This provides an overview and landmarks for orien-
basis of the pelvic radiograph. This can be done tation during surgery, e.g. transfer of dimensional
on the X-ray directly (Fig. 2), but it is better to trace reference to the bone. It must always be remem-
the skeletal contours onto tracing paper (Fig. 3). bered that the measurements on the radiograph
include a magnification effect that must be allowed
A horizontal reference line is drawn along the inferior for if the measurements are transferred to bone.
margins of the obturator foramen, followed by a If the magnification is 10%, measurements taken
vertical reference line along the sacral crest, ideally from the radiograph must be divided by 1.1. So, for
passing through the center of the pubic symphysis. example, 60 mm apparent ÷ 1.1 = 54.5 mm actual
measurement. The same applies for other magnifica-
From these two lines, the center of rotation, tions: e.g. at 15% magnification a 60 mm apparent
difference in leg length, left/right femoral distance, measurement gives 60 mm ÷ 1.15 = 52.2 mm actual
distance between the left/right muscle T lever arms, measurement.
etc. are defined and marked on the tracing paper.
Fig. 2 Fig. 3
Once the dimensions have been entered, the tem- In addition to pelvic radiograph, the mediolateral
plates are used to select the best implant com- radiograph is used to determine the stem shape
ponents for the particular case. The template is and size of the femoral prosthesis as seen from the
positioned on the radiograph such that the center lateral view.
of rotation coincides with the anatomical center of
rotation as determined in the drawing.
The implant components selected should correct
any anatomical insufficiencies derived from the
measurements.
03
Surgical Technique
The planned result becomes clearer when the Materials required:
transparent sheet with the outlined skeletal contours, 1. Tracing paper
measurements, and sketched-in position of the 2. Transparent ruler, 1:1
acetabular cup is placed on top of the radiograph and 3. Transparent protractor
adjusted so that the femur in the radiograph is in the 4. Transparent radius/hole template
desired outcome position in relation to the drawing Ø 24 to 58 mm, in 2 mm increments
of the pelvis. This position is then traced onto the
tracing paper, preferably in a different color (Fig. 4).
The differences on the tracing paper, e.g. actual and INFORMATION:
planned positions of the femur, provide the visual Preoperative planning may be time-consuming
overview required for surgical planning and precise but it provides intraoperative guidance and can
selection of the implant components using the X-ray enhance the final result.
templates or, if necessary, for custom-design
implants (Fig. 5).
Fig. 4 Fig. 5
04
Surgical Technique
Surgical Approaches
The choice of the approach depends on the
surgeon‘s experience and his/her decision based on
the individual situation.
The following approaches are usual:
- antero-lateral - Watson Jones (Fig. A)
- direct lateral - Hardinge (Fig. B)
- postero-lateral - Moore (Fig. C)
Fig. A: Watson Jones
Fig. B: Hardinge
Fig. C: Moore
05
Surgical Technique
Surgical Technique
Implantation of the acetabular cup
Avoid damaging the acetabular metal casing when
implanting the cup prior to the implantation of the
stem (Fig. 1).
Resection of the femoral neck
The hip is dislocated in the usual way. The standard
osteotomy plane is normally at 45° to the axis of the
Fig. 1
femoral stem.
For purposes of orientation the resection guide can
be placed on the lesser trochanter parallel to the
longitudinal axis of the femur (Fig. 2). Resection can
then be carried out along the slit corresponding to
the level selected in preoperative planning. The guide
indicates both the level and the angle of resection.
Care should be taken to ensure that resection is
also carried out at 90° to the axis of the femoral
Fig. 2
neck in the a-p plane.
A second vertical resection may be necessary in
certain cases (Fig. 3). A compressor can also be
used to check that the orientation of the resection
plane is correct.
The acetabular cup is almost always prepared before
the stem.
Fig. 3
Preparation of the femur
The medullary cavity is opened with a gouge. This
should be done as far laterally as possible to avoid
varus positioning of the stem (Fig. 4 + 5). This also
makes it easier to insert the compressor and drasti-
cally reduces the risk of fracturing the femoral shaft
when driving in the prosthesis.
Fig. 4 Fig. 5
06
Surgical Technique
The opening compressor can be used to prepare
the cancellous bone in the lateral area of the great
trochanter to prevent varus positioning of the com-
pressor and the implant.
After the medullary cavity has been opened, it is pre-
pared with the smallest compressor size. Attention
should also be paid to the axial orientation
during this process.
To fit the compressor, the lever on the handle is opened
Fig. 6 and the compressor is inserted with its medial side
facing the lever. The lever is then closed again.
The anatomical form of the compressors means that
the anteversion usually adjusts itself automatically
when they are driven in. The femoral canal is prepared
with compressor of increasing size until maximum
stability is achieved. The process is complete as
soon as the compressor is positioned in the femo-
ral canal so that it is rotationally stable, central and
aligned with the axis.
INFORMATION:
The compressor size is not necessarily the same
as in preoperative planning, but can vary up to
Fig. 7 two sizes.
SP-CL stems are fixed in a bed of compacted can-
cellous bone. The compressors used ensure that the
bone substance is compacted without removing it
from the bone. When driving the compressors in and
out, unintentional turning movements of the handle/
compressor unit should be avoided to prevent damage
to the implant bed.
Fig. 8
INFORMATION:
The SP-CL compressors have a marking line a
few millimeters below the handle attachment
point. The height of this line on the compressor
is exactly at the level of the transition between
the modified surface and the blasted neck sec-
tion of the SP-CL stem. The final prosthesis stem
is driven in with reference to this line up to the
corresponding position of the last compressor.
The position required for this component can
Fig. 9 thus be reproduced exactly.
07
Surgical Technique
Trial reduction
The acetabular component is usually implanted before
the stem so that trial reduction is possible.
This is carried out with the final size of compressor in
situ. The handle is removed and the trial neck segment
selected in preoperative planning (CCD angle, type) is
placed on the compressor. The different trial heads
are then used to check for optimal offset and correct
leg length and to test whether stability is adequate.
Fig. 10 The range of movement is also checked to avoid
impingement of bone and implant and rule out any
instability.
The combination of the prosthetic head with the trial
neck segments is not permissable.
When all checks have been made, the trial head and
trial neck segments are removed by hand. The handle
is then used to remove the compressor, which is still in
situ, from the femoral canal.
Fig. 11
Implanting the final stem
The stem to be used (size, CCD angle, type) has
been determined by performing trial reduction. The
appropriate implant is now removed from the sterile
packaging and inserted manually as far into the
medullary canal as possible. This helps to protect
the prepared implant bed.
The tip of the impactor is placed in the lateral
depression on the upper surface of the prosthesis.
The design of the impactor ensures that any lateral
forces occurring during implantation do not reach
Fig. 12 the stem (Fig. 12).
08
Surgical Technique
Using steady hammer blows the stem is now care-
fully driven in until it reaches its final position. This is
indicated when the transition line on the prosthesis
is at the same height as that on the compressor last
used (Figure 13).
Trial reduction can then be repeated. The selected
trial head is placed on the implant and trial reduction
is carried out.
The trial head is then removed again.
Abb. 13
Fig. 13
Attaching the final prosthesis head
The taper of the stem is cleaned and dried thoroughly.
This is particularly important with ceramic heads.
The head is then attached by hand with axial pres-
sure and a turning movement.
The head is hit lightly with the driver (Fig. 14 + 15).
Once the joint surfaces have been cleaned the joint
is reduced with the final implant components.
The wound is closed in layers.
Abb. 14
Fig. 14
Removing the components
The different components can be removed if required.
The prosthesis head can be removed separately in
an
axial direction using a plunger/stem/rod which is
placed at the base of the head.
The femoral component can be driven out with the
extraction instrument (130-252/00), which is con-
nected to the compressor handle with the adapter.
CAUTION:
If a ceramic head has to be replaced with ano-
ther ceramic head, only ceramic revision heads
with a metal inner taper must be used.
Fig. 15
09
Implants
SP-CL Hip Stem, standard, CCD 126°
Tilastan-S (Ti6Al4V), HX-coated
Offset A A
D
L
C B B
Offset L A B C D
RIGHT LEFT Stem size mm mm mm mm mm mm
177-200/26* 177-220/26* 1 34.7 115 14.4 5.4 6.4 22.5
177-201/26* 177-221/26* 2 36.1 120 15.0 5.6 6.7 23.4
177-202/26* 177-222/26* 3 37.4 125 15.5 5.8 6.9 24.3
177-203/26 177-223/26 4 38.8 130 16.1 6.0 7.2 25.2
177-204/26 177-224/26 5 40.2 135 16.7 6.2 7.4 26.1
177-205/26 177-225/26 6 41.6 140 17.3 6.5 7.7 27.0
177-206/26 177-226/26 7 43.0 145 17.8 6.7 7.9 27.9
177-207/26 177-227/26 8 44.4 150 18.4 6.9 8.2 28.8
177-208/26 177-228/26 9 45.8 155 19.0 7.1 8.4 29.7
177-209/26 177-229/26 10 47.2 160 19.6 7.3 8.7 30.6
177-210/26 177-230/26 11 48.6 165 20.1 7.5 8.9 31.5
177-211/26 177-231/26 12 49.6 165 20.7 7.8 9.2 32.2
177-212/26 177-232/26 13 50.5 165 21.3 8.0 9.5 33.2
177-213/26 177-233/26 14 51.5 165 21.8 8.2 9.7 34.0
177-214/26 177-234/26 15 52.5 165 22.4 8.5 10.0 34.8
177-215/26 177-235/26 16 53.5 165 23.0 8.7 10.3 35.6
177-241/26 177-251/26 18 53.5 165 24.2 9.1 10.8 37.2
177-243/26 177-253/26 20 53.5 165 25.4 9.6 11.3 38.8
177-245/26 177-255/26 22 53.5 165 26.5 10.0 11.9 40.4
* on request, not part of the standard set configuration
10
Implants
SP-CL Hip Stem, standard, CCD 135°
Tilastan-S (Ti6Al4V), HX-coated
Offset A A
D
L
C B B
Offset L A B C D
RIGHT LEFT Stem size mm mm mm mm mm mm
177-400/35* 177-420/35* 1 33.2 115 14.4 5.4 6.4 22.5
177-401/35* 177-421/35* 2 34.5 120 15.0 5.6 6.7 23.4
177-402/35* 177-422/35* 3 35.7 125 15.5 5.8 6.9 24.3
177-403/35 177-423/35 4 37.0 130 16.1 6.0 7.2 25.2
177-404/35 177-424/35 5 38.3 135 16.7 6.2 7.4 26.1
177-405/35 177-425/35 6 39.6 140 17.3 6.5 7.7 27.0
177-406/35 177-426/35 7 40.9 145 17.8 6.7 7.9 27.9
177-407/35 177-427/35 8 42.1 150 18.4 6.9 8.2 28.8
177-408/35 177-428/35 9 43.4 155 19.0 7.1 8.4 29.7
177-409/35 177-429/35 10 44.7 160 19.6 7.3 8.7 30.6
177-410/35 177-430/35 11 46.0 165 20.1 7.5 8.9 31.5
177-411/35 177-431/35 12 46.8 165 20.7 7.8 9.2 32.2
177-412/35 177-432/35 13 47.7 165 21.3 8.0 9.5 33.2
177-413/35 177-433/35 14 48.6 165 21.8 8.2 9.7 34.0
177-414/35 177-434/35 15 49.6 165 22.4 8.5 10.0 34.8
177-415/35 177-435/35 16 50.3 165 23.0 8.7 10.3 35.6
177-441/35 177-451/35 18 50.3 165 24.2 9.1 10.8 37.2
177-443/35 177-453/35 20 50.3 165 25.4 9.6 11.3 38.8
177-445/35 177-455/35 22 50.3 165 26.5 10.0 11.9 40.4
* on request, not part of the standard set configuration
11
Implants
SP-CL Hip Stem, lateralized, CCD 126°
Tilastan-S (Ti6Al4V), HX-coated
Offset A A
D
L
C B B
Offset L A B C D
RIGHT LEFT Stem size mm mm mm mm mm mm
177-603/26 177-623/26 4 45.3 130 16.1 6.0 7.2 25.2
177-604/26 177-624/26 5 46.7 135 16.7 6.2 7.4 26.1
177-605/26 177-625/26 6 48.1 140 17.3 6.5 7.7 27.0
177-606/26 177-626/26 7 49.5 145 17.8 6.7 7.9 27.9
177-607/26 177-627/26 8 50.8 150 18.4 6.9 8.2 28.8
177-608/26 177-628/26 9 52.2 155 19.0 7.1 8.4 29.7
177-609/26 177-629/26 10 53.6 160 19.6 7.3 8.7 30.6
177-610/26 177-630/26 11 55.0 165 20.1 7.5 8.9 31.5
177-611/26 177-631/26 12 56.0 165 20.7 7.8 9.2 32.2
177-612/26 177-632/26 13 57.0 165 21.3 8.0 9.5 33.2
177-613/26 177-633/26 14 58.0 165 21.8 8.2 9.7 34.0
177-614/26 177-634/26 15 59.0 165 22.4 8.5 10.0 34.8
177-615/26 177-635/26 16 60.0 165 23.0 8.7 10.3 35.6
177-641/26 177-651/26 18 60.0 165 24.2 9.1 10.8 37.2
177-643/26 177-653/26 20 60.0 165 25.4 9.6 11.3 38.8
177-645/26 177-655/26 22 60.0 165 26.5 10.0 11.9 40.4
12
Instruments
Instrument set for LINK SP-CL Hip Stems
Compressors, right Compressors, left
Compressors, extra large sizes, right and left
Instrument Set, complete
175-110/00 Instrument set Instruments, complete
175-120/00 Instrument set Compressors, right, complete
175-130/00 Instrument set Compressors, left, complete
175-140/00 Instrument set Compressors, right and left, complete, extra large sizes
13
Instruments
175-110/00 Instrument set, complete
1 2 3 4 5-8 9-12 13-16 17-20
21 22 23
Description Qty.
1 175-111/00 Tray, only 1
2 175-380 Gouge 1
3 175-360 Impactor for prosthesis heads with exchangeable plastic head, 280 mm 1
4 130-622 Impactor 1
5 175-928/11 Plastic trial head 28 S 1
6 175-928/12 Plastic trial head 28 M 1
7 175-928/13 Plastic trial head 28 L 1
8 175-928/14 Plastic trial head 28 XL 1
9 175-932/11 Plastic trial head 32 S 1
10 175-932/12 Plastic trial head 32 M 1
11 175-932/13 Plastic trial head 32 L 1
12 175-932/14 Plastic trial head 32 XL 1
13 175-936/11 Plastic trial head 36 S 1
14 175-936/12 Plastic trial head 36 M 1
15 175-936/13 Plastic trial head 36 L 1
16 175-936/14 Plastic trial head 36 XL 1
17 175-940/11* Plastic trial head 40 S 1
18 175-940/12* Plastic trial head 40 M 1
19 175-940/13* Plastic trial head 40 L 1
20 175-940/14* Plastic trial head 40 XL 1
21 130-394/01 Handle for compressors with quick coupling stems 1
22 175-390 Compressor to start, curved 1
23 175-310/05 Resection guide, 72 mm 1
* on request, not part of the standard set configuration
14
Instruments
175-120/00 Instrument set, Compressors, RIGHT, complete
175-130/00 Instrument set, Compressors, LEFT, complete (NOT ILLUSTRATED)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
18 19 20 21
Version Size Version Size
1 175-121/00 Compressor tray, RIGHT, empty 175-131/00 Compressor tray, LEFT, empty
2 175-200/01* right 1 175-201/01* left 1
3 175-200/02* right 2 175-201/02* left 2
4 175-200/03* right 3 175-201/03* left 3
5 175-200/04 right 4 175-201/04 left 4
6 175-200/05 right 5 175-201/05 left 5
7 175-200/06 right 6 175-201/06 left 6
8 175-200/07 right 7 175-201/07 left 7
9 175-200/08 right 8 175-201/08 left 8
10 175-200/09 right 9 175-201/09 left 9
11 175-200/10 right 10 175-201/10 left 10
12 175-200/11 right 11 175-201/11 left 11
13 175-200/12 right 12 175-201/12 left 12
14 175-200/13 right 13 175-201/13 left 13
15 175-200/14 right 14 175-201/14 left 14
16 175-200/15 right 15 175-201/15 left 15
17 175-200/16 right 16 175-201/16 left 16
Single angled Handle Single angled Handle
18 130-394/02* 130-394/03*
for compressors, right for compressors, left
* on request, not part of the standard set configuration
Neck segment for fitting to compressor with taper 12/14 mm
Description Type CCD Version for compressor- & stem size
19 175-195/26 Neck segment std. 126° right 1 - 16
20 175-197/26 Neck segment lat. 126° right 4 - 16
21 175-195/35 Neck segment std. 135° right 1 - 16
175-196/26 Neck segment std. 126° left 1 - 16
175-198/26 Neck segment lat. 126° left 4 - 16
175-196/35 Neck segment std. 135° left 1 - 16
15
Instruments
175-140/00 Instrument set, Compressors, complete, extra large sizes, right and left
1 2 3 4 5 6 7
Version Size
1 175-141/00 Compressor tray, empty
2 175-201/18 left 18
3 175-201/20 left 20
4 175-201/22 left 22
5 175-200/22 right 22
6 175-200/20 right 20
7 175-200/18 right 18
16
Instruments
130-394/01 Universal Handle for rasp stems and compressors
130-394/02 Universal Handle for rasp stems and compressors, angled, right
130-394/03 Universal Handle for rasp stems and compressors, angled, left
Coupling of the compressor
Fig. 1
To couple compressor and handle, the catch is retracted fully (left arrow). Then, the compressor fitting is
inserted into the mount on the front of the handle (right arrow, Fig. 1).
Fig. 2
To secure the connection between compressor and handle, the catch is pushed forwards (arrow) (Fig. 2).
Fig. 3
To disengage open the handle again (arrow) (Fig. 3).
The handle for compressors can then be detached from the compressor.
17
Instruments / Accessories
Additional Instruments
(not included in instrument set)
179-122 Insertion Forceps, lockable 179-122/01 Exchangeable Taper Cap
for Insertion Forceps 179-122
130-252/00 Instrument set for extraction
(not illustrated)
Instructions for Cleaning and Maintenance
Specific instructions for instruments are available on request: E-mail
[email protected]
X-ray Templates
for SP-CL Hip Prosthesis Stems, cementless
Stem Ti6Al4V, HX Coating, taper 12/14 mm, sizes 1-16, set with 16 sheets
CCD angle Actual size
175-801/02 126° 110%
175-801/04 135° 110%
175-801/03 126° lateralized 110%
18
Literature
Additional Information
Catalogs on request: E-mail
[email protected]
For more information please register for our LINK Media Library (linkorthopaedics.com)
19
Indications/Contraindications
SP-CL cementless Hip Prosthesis System (all types)
General Indications
Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur which cannot be treated by
conservative or osteosynthetic procedures
Indications
Primary and secondary osteoarthritis
Rheumatoid arthritis
Correction of functional deformities
Avascular necrosis
Femoral neck fractures
Revision surgeries
Contraindications
Acute and chronic infections, local and systemic insofar as they compromise the successful implantation of
a total hip prosthesis (preoperative microbiological analysis recommended)
Allergies to (implant) materials
Insufficient / inadequate bone mass- or quality, which prevents a stable anchorage of the prosthesis
The stems are indicated for cementless use only.
INFORMATION:
These indications/contraindications refer to standard cases. The ultimate decision on whether or
not an implant is suitable for a patient must be made by the surgeon based on his/her individual
analysis and his/her experience.
The SP-CL hip prosthesis stem can only be combined with prosthesis heads that do not exceed
+10.5 mm additional neck length.
20
Important Information
Please note the following regarding the use of our implants:
1. Choosing the right implant is very important.
The size and shape of the human bone determines the size and shape of the implant and also limits the load capacity.
Implants are not designed to withstand unlimited physical stress. Demands should not exceed normal functional loads.
2. Correct handling of the implant is very important.
Under no circumstances should the shape of a finished implant be altered, as this shortens its life span. Our implants
must not be combined with implants from other manufacturers. The instruments indicated in the Surgical Technique
must be used to ensure safe implantation of the components.
3. Implants must not be reused.
Implants are supplied sterile and are intended for single use only. Used implants must not be used again.
4. After-treatment is also very important.
The patient must be informed of the limitations of the implant. The load capacity of an implant cannot compare with
that of healthy bone!
5. Unless otherwise indicated, implants are supplied in sterile packaging.
Note the following conditions for storage of packaged implants:
• Avoid extreme or sudden changes in temperature.
• Sterile implants in their original, intact protective packaging may be stored in permanent buildings up until the
“Use by” date indicated on the packaging.
• They must not be exposed to frost, dampness or direct sunlight, or mechanical damage.
• Implants may be stored in their original packaging for up to 5 years after the date of manufacture. The “Use by” date
is indicated on the product label.
• Do not use an implant if the packaging is damaged.
6. Traceability is important.
Please use the documentation stickers provided to ensure traceability.
7. Further information on the material composition is available on request from the manufacturer.
Follow the instructions for use!
Waldemar Link GmbH & Co. KG, Hamburg
All content in this catalog, including text, pictures and data, is protected by law. Every instance of use, whether in part or in whole and which
is not permitted by law, is subject to our prior consent. In particular, this applies to the reproduction, editing, translation, publishing, saving,
processing, or passing on of content stored in databases or other electronic media and systems, in any manner or form. The information
in the catalogs is solely intended to describe the products and does not constitute a guarantee.
The Surgical Technique described has been written to the best of our knowledge and belief, but it does not relieve the surgeon of his/her
responsibility to duly consider the particularities of each individual case.
Products shown in this document may not be available in your country. The product availability is subject to the approval and/or registration
regulations of the respective country. Please contact Waldemar Link GmbH & Co. KG if you have questions about the availability of LINK
products in your country.
Waldemar Link GmbH & Co. KG and/or other corporate affiliated entities own, use or have applied for the following trademarks in many
jurisdictions: LINK, BiMobile, SP II, Modell Lubinus, E-Dur, EndoDur, T.O.P. II, BetaCup, CombiCup PF, CombiCup SC, CombiCup R,
MobileLink, C.F.P., LCU, SP-CL, LCP, MIT-H, Endo-Model, Endo-Model SL, MP, MEGASYSTEM-C, GEMINI SL, SPAR-K, LCK, Link OptiStem,
HX, TiCaP, X-LINKed, PorAg, LINK PorEx, BiPorEx, PorEx-Z, TrabecuLink, Tilastan, customLINK, RescueSleeve, Stactip, VACUCAST.
Other trademarks and trade names may be used in this document to refer to either the entities claiming the marks and/or names or their
products and are the property of their respective owners.
21
© LINK 6462_SP-CL_HX_OP_EN_2021-12_003 MAR-01475 3.0
N Waldemar Link GmbH & Co. KG
Barkhausenweg 10 · 22339 Hamburg · Germany
Phone +49 40 53995-0 ·
[email protected]
www.linkorthopaedics.com
SP-CL ®
Anatomically Adapted Cementless Hip System
“The most important advancement in total hip arthroplasty in the last 50 years has
been the adaption of femoral components to the anatomy of the femur.“*
Being one of the inventors of the anatomical
prostheses design, LINK has more than three
decades of experience in the development of
this type of hip stems. The new anatomical
cementless hip prostheses system SP-CL®
from LINK represents a consistent progres-
sion of the cemented Lubinus SP II® hip
system towards cementless fixation – the
SP II® concept that has proven its success
in the Swedish Hip registry**.
During the development of the SP-CL® the
anatomical design features were subjected
to various biomechanical tests.
The result is a harmonically aligned hip
prosthesis system, that provides a stable
and long term solution for a large variety
of patients with hip problems. It supports
surgeons in their aim to operate gentle and
less invasive and high reproducibility of
clinical results is expected.
* W.T. Stillwell. The Art of the Total Arthroplasty. Grune & Stratton,
Inc. 1987; pp. 296
** Annual Report 2012; Swedish Hip Arthroplasty Register;
www.shpr.se
Made in Germany
Features of the SP-CL® Hip System:
• Complementary to the cemented SP® family,
backed by decades of successful use
• Cementless metaphyseal fixation
• Bioharmonic left and right stem types,
each available in 13 standard sizes
• Two CCD angles - 126° and 135°
• Size and offset increase following stem growths
from one size to the next
Neutralising of torsional forces*
acting on the proximal femur
Photoelastic study of stresses
Stress-load analysis proves that the true
adaption to the femur of the LINK anatomi-
cally shaped hip stems result in the most
natural stress distribution, eliminating the
harmful pinpoint stress concentration at
the bone / implant interface.
Straight stems produce S-shaped stems result
stress risers in an in even stress distribution
anatomically S-shaped
medullar canal
Anatomic design
and integrated antetorsion
5°
The stems in the SP-CL® system have a 12/14
taper and can be combined with every modular
ceramic or metal prosthesis head made by LINK.
The stem has an anatomic design with an
S-shaped curvature in M/L view. An antetorsion
of 5° is integrated in the stem.
Ribbed profile for excellent
primary stability
Designed with deep grooves and from
LINK® Tilastan-S® alloy the SP-CL® results
in a stem with constructive and material
elasticity to act against stress shielding.
The pronounced ribs provide primary
stability.
Trochanter protecting
Protects the greater trochanter due
to flattened proximal lateral profile.
LINK® HX® (CaP) coating
The surface promotes osteoconduction
and allows stable osseointegration for
secondary stability.
HX®-Coating (calcium phosphate) The HX®-Coating is an osteoconductive coating approxi-
vertical, cristalline coating structure mately 15 µm thick. The proprietary electrochemical coating
IIIIIIIII I approx. process results in extraordinary mechanical strength of the
I I
15 µm coating, that endures the stress of implantation. The porous
I
I
I I
I cell structure of approx. 160 µm cell diameter of the substrate‘s
Substrate surface is retained open, due to the thin overcoat and provides
an optimal structure for osteoconduction.
Polished distal stem
For easy and safe insertion the polished surface
allows distal gliding to counteract thigh pain.
Anatomical shape of the compressors
Anatomically shaped stems require anatomically shaped instruments.
The compressors of the SP-CL® system strictly follow the anatomical
stem design and prepare a bony bed for the SP-CL® stem following
the natural shape of the intramedullar canal of the proximal femur.
The terraced profile
of the compressors
The SP-CL® stems are anchored in a bed
of compacted cancellous bone. The teeth
of the SP-CL® compressors ensure that the
cancellous substance is firmly compacted.
High fixation zones
Toothless bilateral isles for maximum grip
of the proximal ribs.
Press-fit
The compressors and implants are
matched to provide the optimal press-fit.
Rounded distal tip
of compressor for safe guidance
and to avoid via falsa.
© LINK 644_SP CL_Flyer_en_2017-08_004
Waldemar Link GmbH & Co. KG
Barkhausenweg 10 · 22339 Hamburg, Germany
P.O. Box 63 05 52 · 22315 Hamburg, Germany
Tel.: +49 40 53995-0 · Fax: +49 40 5386929
E-mail:
[email protected] · www.linkorthopaedics.com
!" #$%& '
() * +,
-./0- -12342 5 -6678 9:;<8:=>?@; 5 43ABB4CD/ 2-B-
!""# $%& '() *+,- ."#
/012/345
6"7," 7 89:; < =>??, @ ABC 89:; DC :.."?7, "& 89:; 7# .?7.-
1E/0/0330/2502E333 F*
--0- -12342 5 -6678 9E:FG;<8:H 5 414IAJA-CKD/ 2-B-
*:G:FHHI*F
6"7," 7 89:; @ ABC 89:; :.."?7, "& 89:;
12/0/03300J452K0//
H< L7 9M 7 < #++ 9M N ' O(PB C ( )$A 97Q.-R+ S7! 9M 7 7# ##++
FG3T :H H9 U:6 I*V 3 I:WW /E0/0 HI:
!+ +? 87?S, < +X7#; < =>??, YP O CZ 8[ $ O$(OA\D; DC :"?7, "Q L-7, 87 7?S+ 7# ]"#+; 7#
.?7.-
/310&00 8^!+ ,?_, !!++# !_!+!^7# &00; F*
H>,?+ `( A ,>+ "Q L-7,
*>X= =#>?+,?+ +,
a,7?S ( $ AAb' *Q7. "&
3200013/3K
I7?+,+? ` C (NAPZ M"++"7 Q I7?+,+?!"7," < =>??, YBC ) O Cb (NAPZ DC
0&00 M"++"7 Q .."?7, 7# .?7.-
(-1L- .A11ABM/ 5 EN<O8PQN86?8 R;6@; 5 -1.D/ JS1IAC3-BAS1
V!>7# 7,7,L7!9?+7++ 7,7, S7!
*J2/EEK4431 //&0K&303E /4c/Kc32
:H H9 U7! I"7d 3 /E0/0 I>>77 +, !""# /00043E3
3/3 JEJ
11.06.2025
VAIDLUSTUS
Riigihange: „Traumatoloogia metallosteosünteesi vahendid ja liigeste
endoproteesid“
viitenumber: 295377
Vaidlustaja: OneMed OÜ
registrikood: 10891247
aadress: Veerenni 38, 10138 Tallinn
Esindaja: Erki Fels, vandeadvokaat
Gregor Saluveer, vandeadvokaadi abi
Advokaadibüroo TGS Baltic
e-post:
[email protected]
Hankija: sihtasutus Tartu Ülikooli Kliinikum
registrikood: 90001478
aadress: L. Puusepa 1a, Tartu 50406
e-post:
[email protected]
TAOTLUSED
1. Peatada riigihanke „Traumatoloogia metallosteosünteesi vahendid ja liigeste
endoproteesid“ (vr 295377) menetlus kõikides riigihanke osades.
2. Kohustada sihtasutust Tartu Ülikooli Kliinikum viima õigusaktidega kooskõlla järgmised
riigihanke alusdokumentide tingimused:
2.1. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ punkt 3;
2.2. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ punkt 4.1;
2.3. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ punkt 5.1;
2.4. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ punkt 6.1;
2.5 riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 6.1;
2.6. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 8.1;
2.7. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 8.2;
2.8. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 9.2;
2.9. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 9.3;
2.10. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ punkt 9.6;
Advokaadibüroo TGS Baltic AS Ahtri 6a, 10151 Tallinn Kaluri 2, 51004 Tartu
[email protected] Telefon +372 626 4300 Telefon +372 730 1610
2.11. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 1.5;
2.12. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 1.6;
2.13. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 2.5;
2.14. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 2.8;
2.15. riigihanke osa 10 alusdokumendi „Lisa 10 osa 10 - Mittetsemeneteeritav
puusaprotees (põhiproteesina)“ punkt 3.1;
2.16. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 1.1;
2.17. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 1.2;
2.18. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 1.6;
2.19. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 1.7;
2.20. riigihanke osa 11 alusdokumendi „Lisa 11 osa 11 - Mittetsemeneteeritav
puusaprotees (alternatiiv)“ punkt 2.4;
2.21. riigihanke osa 1 alusdokumendis „Lisa 1 osa 1 - tsemeneteeritav standardne
puusaprotees (põhi)“ sätestatud nõuded, sealhulgas peatükk „Muud täiendavad
komponendid“;
2.22. riigihanke osa 2 alusdokumendis „Lisa 2 osa 2 - tsemeneteeritav standardne
puusaprotees (alternatiiv)“ sätestatud nõuded, sealhulgas peatükk „Muud
täiendavad komponendid“;
2.23. riigihanke osa 3 alusdokumendis „Lisa 3 osa 3 - puusaliigese eriprotees“
sätestatud nõuded, sealhulgas peatükid „Muud täiendavad komponendid“ ja
„Lisavahendid“;
2.24. riigihanke osa 5 alusdokumendis „Lisa 5 osa 5 - Tsemeneteeritav põlveprotees
tagumise ristatisideme stabiliseerimisega“ sätestatud nõuded, sealhulgas
peatükid „Lisavõimalused“ ja „Teised lisavõimalused“;
2.25. riigihanke osa 7 alusdokumendis „Lisa 7 osa 7 - Põlveliigese eriprotees“ sätestatud
nõuded, sealhulgas punkt 8;
2.26. riigihanke osa 13 alusdokumendis „Lisa 13 osa 13 - Megaprotees“ sätestatud
nõuded;
2.27. riigihanke alusdokumendi „Lisa_24_pakkumuse_maksumus“ punkt 2.3;
2.28. riigihanke osa 4 alusdokumendi „Lisa 4 osa 4 - Põlveliigese kondülaarne protees
tagumise ristatisideme säilitamisega“ peatükid „Muud täiendavad komponendid“,
„Lisavahendid“ ja „Teised lisavõimalused“;
2.29. riigihanke osa 6 alusdokumendi „Lisa 6 osa 6 - Erikonstruktsiooniga
tsemeneteeritav põlveprotees“ peatükid „Lisavõimalused“, „Teised
lisavõimalused“ ja „Muud täiendavad komponendid ja võimalused“;
2.30. riigihanke osa 7 alusdokumendi „Lisa 7 osa 7 - Põlveliigese eriprotees“ peatükid
„Lisavõimalused“, „Teised lisavõimalused“ ja „Muud täiendavad komponendid ja
võimalused“;
2.31. riigihanke osa 8 alusdokumendi „Lisa 8 osa 8 - Õlaliigese protees“ peatükk „Muud
täiendavad komponendid ja võimalused“;
2/13
2.32. riigihanke osa 9 alusdokumendi „Lisa 9 osa 9 - Küünarliigese protees“ peatükk
„Muud täiendavad komponendid ja võimalused“;
2.33. riigihanke osade 1—14 tehniliste kirjelduste tingimused, mis algavad sõnadega
„Endoproteesi komplekti baashinna arvutamiseks /…/“;
2.34 riigihanke osade 1—23 tehniliste kirjelduste tingimused, mis algavad sõnadega
„Pakkuja peab hankijale tagama spetsiifiliste instrumentide komplektid /…/“;
2.35. riigihanke osade 1—23 tehniliste kirjelduste tingimused, mis algavad sõnadega
„Garanteeritud instrumentide komplektide arv arvestatakse /…/“ ja „Lisaks
garanteeritud instrumentide komplektidele /…/“;
2.36. riigihanke alusdokumendi „Juhised pakkujale“ punktid 6.18—6.18.2 ja 7.2.
3. Jätta OneMed OÜ menetluskulud sihtasutuse Tartu Ülikooli Kliinikum kanda.
1. ASJAOLUD
1.1. 19.05.2025 avaldas sihtasutus Tartu Ülikooli Kliinikum (hankija) avatud hankemenetlusena
läbiviidava riigihanke „Traumatoloogia metallosteosünteesi vahendid ja liigeste endoproteesid“ (vr
295377) (riigihange) hanketeate ja tegi kättesaadavaks riigihanke alusdokumendid (RHAD).
Tehnilise kirjelduse (TK) järgi on hankelepingu esemeks traumatoloogia metallosteosünteesi
vahendite ja liigeste endoproteeside ost.
1.2. Pakkumuste esitamise tähtpäev on 19.06.2025.
1.3. TK on vastuolus riigihangete seaduse (RHS) § 88 lg-ga 7 ja RHS § 3 p-dega 1—3 ja 5. Riigihanke
osades 1, 2, 3, 4, 5, 6, 7, 10, 11 ja 13 on hankija koostanud TK nii, et neile vastab kindel toode.
OneMed OÜ (vaidlustaja) osalemine riigihankes on põhjendamatult takistatud. Ülejäänud
riigihanke osades on tingimused ebaselged, mistõttu pole vaidlustajal võimalik aru saada, kas
tema pakutavad tooted tingimustele vastavad või mitte, või ebaproportsionaalsed.
1.4. Vaidlustaja taotleb riigihangete vaidlustuskomisjonilt (VAKO) kohustada hankijat viima TK
õigusaktidega kooskõlla (RHS § 185 lg 2 p 1).
2. RIIGIHANKE PEATAMISE TAOTLUS
2.1. VAKO võib vaidlustaja põhjendatud taotluse alusel teha igas vaidlustusmenetluse staadiumis
otsuse riigihanke peatamise kohta (RHS § 193 lg 1). Vaidlustaja taotleb riigihanke peatamist
kõikides riigihanke osades, kuna kõikides osades on õigusvastaseid tingimusi, mis vaidlustaja
õiguseid riivavad. Riigihanke peatamine on põhjendatud, kuna pakkumuste esitamise tähtpäev
saabub enne VAKO otsuse tegemist (RHS § 200 lg 1).1
2.2. Vaidlustuse eesmärk on saavutada olukord, kus hankija kehtestaks õigusvastaste
hanketingimuste asemel õiguspärased tingimused. Sellise eesmärgi saavutamine ei ole enam
võimalik, kui pakkumused on esitatud. Kui pakkumuste esitamise tähtaeg möödub, on RHADis
sätestatud tingimused kohustuslikud nii hankijale kui pakkujatele, kellel tuleb edasises
menetluses nendest tingimustest lähtuda. VAKO ega kohus ei saa kohustada hankijat pärast
pakkumuste esitamise tähtpäeva möödumist RHADi muutma.2 Vaidlustaja on valmis taotlusest
loobuma, kui hankija soostub ise pakkumuste esitamise tähtpäeva edasi lükkama nii, et tagatud
oleks VAKO otsuse tegemine ja ka kaebetähtaeg.
1
VAKO 93-18/194960, p 4.
2
TlnRKm 3-15-1817, p 10; VAKO 166-17/190812, p-d 8, 11—14; VAKO 187-17/192426, p-d 8, 11—14.
3/13
3. PÕHJENDUSED
3.1. Vaidlustaja vaidlustab tingimusi kahel põhjusel. Esiteks, TK on koostatud kindlate toodete järgi
(ptk A). Teiseks, TK on ebaselge (ptk 3.6).
A. TEHNILINE KIRJELDUS ON KOOSTATUD KINDLATE TOODETE JÄRGI
3.1. Hankija on koostanud TK-d vaidlustatud osades kindlate toodete spetsifikatsioonide järgi
(täpsemalt toodud allolevas tabelis). TK-d tervikuna lugedes on selge, et hankija on lihtsalt
kopeerinud TK-sse nende toodete omadused.
3.2. Riigihangete korraldamise üheks üldpõhimõtteks on konkurentsi efektiivne ärakasutamine (RHS
§ 3 p 3). Hankija on vaidlusaluse riigihanke kirjutanud n-ö lukku nendele pakkujatele, kes pakuvad
neid tooteid, mille omadusi on TK-s kirjeldatud. TK ei tohi aga tekitada objektiivselt
põhjendamatuid takistusi riigihangete avamisel konkurentsile (RHS § 88 lg 7). TK ei tohi olla
koostatud nii, et tingimustele vastaks üksnes üks toode või pakkuja.3
3.3. Hanketingimused peavad olema põhjendatud. Praegune TK, millega suunatakse hankeleping
kindlate toodete pakkujatele, kindlasti põhjendatud ei ole. Hankija ei tohi seada hankelepingu eset
silmas pidades põhjendamatuid piiranguid, mis välistavad ausa ja läbipaistva konkurentsi või on
mõistliku põhjenduseta piiravad. 4 Hanketingimus ei tohi minna kaugemale, kui riigihankes
taotletava eesmärgi saavutamiseks on mõistlikult vajalik.5
3.4. Vaidlustaja juhib tähelepanu, et 2021. aastal korraldas hankija endoproteeside riigihanke
(238201) sarnase ülesehitusega ja tulemuseks oli olukord, kus hankija maksis endoproteeside
eest 4 aasta jooksul ebamõistlikult kõrget hinda võrreldes samade või paremate tulemustega
Eestis kasutatavate endoproteesidega. Hankija tegevus ei vasta seega ka üldpõhimõttele
kasutada rahalisi vahendeid säästlikult ja otstarbekalt (RHS § 3 p 5).
3.5. Allolevas tabelis on välja toodud tingimused, mis takistavad vaidlustajal pakkumust esitamast.
Paremalt teises lahtris on viide dokumendile, millest nähtub, et hankija valitud tootel on olemas
täpselt kehtestatud nõudele vastav omadus. Parempoolses lahtris selgitab vaidlustaja, miks ei
ole nõuded sellised kujul põhjendatud.
TK Viide hankija valitud Miks ei ole tingimus
TK punkti sisu
punkt tootele põhjendatud
Osa 4: Põlveliigese kondülaarne protees tagumise ristatisideme säilitamisega
Hankija nõuded kirjeldavad OÜ Wimberg poolt müüdava Waldemar Link tooteid
„SymphoKnee“ ja „Gemini SL“ kombinatsioonis veel mõne Waldemar Linki tootega,
mille edasimüügiõigus on vaid OÜ-l Wimberg.
3 „Reieluu ja sääreluu Dokument „SymphoKnee Nõue on objektiivselt
komponentide väljavõte“ (lisa 1) põhjendamatu ja valitud kindla
suuruse ühilduvus toote järgi. Kõigil tootjatel on
vähemalt +/- 2 omad kombineeritavuse tabelid,
suurust.“ mis täidavad samaväärselt
hankija eesmärki.
3
TlnRnKo 3-10-3225, p 18.
4
TlnHKo 3-20-582, p 12.
5
VAKO 34-19/206183, p 15.
4/13
4.1 “mobiilse ja Dokument „Gemini SL Nõue on objektiivselt
toestatud kande surgical technique“ (lisa põhjendamatu ja valitud kindla
komponendid;” 2), lk 2 toote järgi. Mobiilne kanne võib
käia roteeruva tibia platoo
mõistena, mitte reieosa
komponendi nõudena.
5.1 “mobiiline kanne ja Dokument „Gemini SL Nõue on objektiivselt
fikseeritud kanne;” surgical technique“ (lisa põhjendamatu ja valitud kindla
2), lk 2 toote järgi. Mobiilse kande nõude
eesmärk on füsioloogiline liigese
rotatsioon, eri tootjad lahendavad
seda erinevalt süsteemipõhiselt,
näiteks mediaalselt roteeruv
protees. See täidab samaväärselt
hankija eesmärki.
6.1 „Ülikõrge Dokument „Link Nõue on objektiivselt
molekulmassiga Symphoknee Surgical põhjendamatu ja valitud kindla
polüetüleenist technique“ (lisa 3), lk 65 toote järgi. Laialtlevinud standard
(UHMWPE) on HXLPE või tootjaspetsiifiline
liugpinnad.“ variant, mis täidab samaväärselt
või paremini hankija eesmärki.
UHMWPE materjal antud nõude
kombineerimisel ülejäänud
nõuetega on konkurentsi välistav.
Osa 6: Erikonstruktsiooniga tsemeneteeritav põlveprotees
Hankija nõuded kirjeldavad OÜ Ortolink Invest poolt müüdavat Smith & Nephew toodet
„Primary Knee System“.
6.1 „reisosa materjal Dokument „OXINIUM- “Keramiseeritud metall” on üks
keramiseeritud Knee-Brochure“ (lisa 4), võimalustest lahendada
metall (ZrNb), et lk 4 probleemi, mis esineb 1-l
sobiks ka patsiendil 100st (metalliallergia).
metalliallergiaga Tegemist on Smith & Nephew
patsientidele“ patenteeritud tehnoloogiaga
(OXINIUM). On olemas ka
alternatiive, nt Ti-Nidium, mis
täidavad samaväärselt hankija
eesmärki.
8.1 “plastikosa igal Dokument „Legion TKS „High flex insert“ on nimetus, mida
mõõdul vähemalt 9 surgical technique“ (lisa kasutab ainult Smith & Nephew
erinevat kõrgust, 5), lk 46 oma plastikosade kirjeldamiseks.
high flex ja Teised tootjad võivad selle
constrained tingimuse eesmärki täita näiteks
variandid;” erilise disainiga reie
komponendiga.
5/13
8.2 “vähemalt 6 Dokument „Legion TKS Nõue on objektiivselt
erinevat surgical technique“ (lisa põhjendamatu ja valitud kindla
polüetüleenist 5), lk-d 59 ja 61 toote järgi. Enamik süsteeme
liigestuva pinna algab suurusest 10mm, mis
suurust, CCK täidab samaväärselt hankija
versioonil eesmärki.
paksusega alates 9
mm;”
9.2 „Monteeritavad Dokument „Legion TKS Nõue objektiivselt põhjendamatu
intramedullaarsed surgical technique“ (lisa ja valitud kindla toote järgi. Igal
pikendused: varre 5), lk 59 tootjal on oma suuruste ja
pikendused alates pikkustega varred, mis vastavad
diameetrist 10 mm, süsteemile ja täidavad
pikkusega vähemalt samaväärselt hankija eesmärki.
vahemikus 80 mm –
275 mm, sirged.
Tsemendivabad ja
tsemendiga
versioonid.“
9.3 “Revisjonvarred: Dokument „Legion TKS Nõue on objektiivselt
pikendused koos surgical technique“ (lisa põhjendamatu ja valitud kindla
mõõdetava Offset- 5), lk 59 toote järgi. Igal tootjal on oma
ga , alates suuruste ja pikkustega varred, mis
diameetrist 9 mm ja vastavad süsteemile ja täidavad
pikkusega vähemalt samaväärselt hankija eesmärki.
vahemikus 120 –
270 mm. Nii
tsementeeritavad
kui tsemendivabad.”
9.6 „Põlvekedra Dokument „Legion TKS Nõue on objektiivselt
protees: vähemalt surgical technique“ (lisa põhjendamatu ja valitud kindla
kolme erineva 5), lk 48 toote järgi. Igal tootjal on oma
kujuga ja vähemalt suuruste ja variatsioonidega
6 erineva põlvekedra proteesid, mis
paksusega.“ vastavad süsteemile ja täidavad
samaväärselt hankija eesmärki.
Osa 10: Mittetsemeneteeritav puusaprotees (põhiproteesina)
Hankija nõuded kirjeldavad OÜ Ortolink Invest poolt müüdavaid Smith & Nephew
tooteid „SL Plus MIA“ ja „R3“ kombinatsioonis veel mõne Smith & Nephew tootega,
mille edasimüügiõigus on vaid OÜ-l Ortolink Invest.
1.5 “reieosa 2 erineva Dokument „SL PLUS Nõue on objektiivselt
kaela nurgaga- MIA Surgical technique“ põhjendamatu ja valitud kindla
standard (131°) ja (lisa 6), lk 16 toote järgi (täpselt 131°).
Vaidlustaja toote puhul on reieosa
6/13
lateraliseeritud 2 erineva kaela nurgaga: standard
(123°);” (133°) ja lateraliseeritud (123°).
See täidab hankija eesmärki
samaväärselt.
1.6 ”reieosa kaela Dokument „SL PLUS Nõue on objektiivselt
pikkus standard MIA Surgical technique“ põhjendamatu ja valitud kindla
varrel 17-49mm ja (lisa 6), lk 17 toote järgi (täpselt 17-49mm ja
lateraliseeritud 26-58mm). Vaidlustaja toote puhul
variandil 26-58mm” on reieosa kaela pikkus standard
varrel 28,3-49,4mm ja
lateraliseeritud variandil 33,6-
54,6mm. See täidab hankija
eesmärki samaväärselt.
2.5 ”napa sisemine - Nõue on objektiivselt
vooder saadaval nii põhjendamatu. PE vooder on
PE kui XLPE vananenud tehnoloogia. Kuna PE
variandis;” voodriga implantaatide kliinilised
tulemused on kehvemad kui
näiteks XLPE versioonil, siis ei
paku enamik vastutustundlikuid
tootjad enam PE versiooni. Kuna
hankija nõuab nii üht kui teist,
teeb hankija võimatuks XLPE
variandis tooteid pakkuda (mis ei
täida hankija eesmärki isegi
samaväärselt, vaid paremini).
2.8 „Pressfit napaosa Dokument „R3 Surgical Nõue on objektiivselt
suurused 40mm – technique“ (lisa 7) lk-d 13 põhjendamatu ja valitud kindla
80 mm millele ja 17. toote järgi. 0° ja 10°
sobivad XLPE kõrgendusega on saavutatav
plastikosad 0° ja samaväärne tulemus, mis 0° ja
20° kõrgendusega;“ 20° kõrgendusega.
3.1 „reiepea materjal Dokument „OXINIUM- „Keramiseeritud metall” on Smith
keramiseeritud Knee-Brochure“ (lisa 4), & Nephew patenteeritud
metall (ZrNb);“ lk 4 tehnoloogia (OXINIUM). On
olemas ka alternatiive, nt Ti-
Nidium, mis täidavad
samaväärselt hankija eesmärki.
Osa 11: Mittetsemeneteeritav puusaprotees (alternatiiv)
Hankija nõuded kirjeldavad OÜ Wimberg poolt müüdavat Waldemar Link toodet „SP-
CL“ kombinatsioonis veel mõne Waldemar Link tootega, mille edasimüügiõigus on
vaid OÜ-l Wimberg.
7/13
1.1 ”bioloogiliselt/anato Dokument „SP-CL Nõue on objektiivselt
omiliselt sobilik, Product Rationale“ (lisa põhjendamatu. S-kõverus on
mille S-kõverus 8), lk 3 ainult Waldemar Linki toodetele
järgib reieluu omane omadus. Samaväärseid
loomulikku või paremaid kliinilisi tulemusi on
anatoomilist kuju;” võimalik saavutada ka teiste
kujudega proteesidega.
1.2 ”Vasaku ja parema Dokument „SP-CL Vasak ja parem versioon on
poole versioonid” Anatomically Adapted vajalikud anatoomilise disaini
Cementless Hip System“ puhul. Samaväärsete või
(lisa 9), lk 3 paremate tulemustega teise
disainiga vartel ei ole vaja vasakut
ja paremat versiooni.
1.6 ”Kahe CCD nurga Dokument „SP-CL Nõue on objektiivselt
võimalus – 126 ja Anatomically Adapted põhjendamatu ja valitud kindla
135 kraadi” Cementless Hip System“ toote järgi (täpselt 126 ja 135
(lisa 9), lk 3 kraadi, nagu Waldemar Link
tootel).
1.7 ”Reieosal vähemalt Dokument „SP-CL Vasak ja parem versioon on
18 erinevat suurust Anatomically Adapted vajalikud anatoomilise disaini
vasakule ja Cementless Hip System“ puhul. Samaväärsete või
paremale” (lisa 9), lk 3 paremate tulemustega teise
disainiga vartel ei ole vaja vasakut
ja paremat versiooni.
2.4 ”napaosa võimalik - Nõue on objektiivselt
modulaarsete põhjendamatu ja valitud kindla
komponentidega toote järgi. 0° ja 10°
10° ja 20° serva kõrgendusega on saavutatav
kõrgendada” samaväärne tulemus, mis 0° ja
20° kõrgendusega.
3.6. Lisaks eelnevale vaidlustab vaidlustaja TK-sid tervikuna osades 1, 2, 3, 5, 7 ja 13. Riigikohus on
selgitanud, et tingimuste kogum on õigusvastane, kui selle eesmärk on ebakohastel eesmärkidel
konkreetse ettevõtja soosimine.6 Viidatud osade TK-d on kogumis koostatud nii, et neile vastavad
vaid ühe tootja tooted:
a. Osa 1 („tsementeeritav standardne puusaprotees (põhiproteesina)“) nõuded kirjeldavad
OÜ Wimberg poolt müüdavat Waldemar Link toodet „Lubinus SPII“ kombinatsioonis veel
mõne Waldemar Link tootega, mille edasimüügiõigus on vaid OÜ-l Wimberg.
b. Osa 2 („tsementeeritav standardne puusaprotees (alternatiivse proteesina)“) nõuded
kirjeldavad OÜ Ortolink Invest poolt müüdavaid Smith & Nephew tooteid „Polar Stem“ ja
„Polar Cup“ kombinatsioonis veel mõne Smith & Nephew tootega, mille edasimüügiõigus
on vaid OÜ-l Ortolink Invest.
c. Osa 3 („puusaliigese eriprotees“) nõuded kirjeldavad OÜ Wimberg poolt müüdavat
Waldemar Link toodet „Link MP Reconstruction“ kombinatsioonis veel mõne Waldemar
6
RKHKo 3-20-718, p 18.
8/13
Link tootega, mille edasimüügiõigus on vaid OÜ-l Wimberg.
d. Osa 5 („tsemeneteeritav põlveprotees tagumise ristatisideme stabiliseerimisega“) nõuded
kirjeldavad Johnson&Johnson AB Eesti filiaali poolt müüdava DePuy Synthese toodet
„Attune“.
e. Osa 7 („põlveliigese eriprotees“) nõuded kirjeldavad OÜ Wimberg poolt müüdava
Waldemar Link põlveproteesi kombinatsioonis veel mõne Waldemar Link tootega, mille
edasimüügiõigus on vaid OÜ-l Wimberg.
f. Osa 13 („megaprotees“) nõuded kirjeldavad OÜ Wimberg poolt müüdava Waldemar Link
tooteid. Eriti ilmselt avaldub see tõsiasi TK punktis 5, kus nõuab hankija, et pakutav toode
peab olema kombineeritav põlveliigese Endo-model põlve implantaadi ning puusaliigese
MP rekonstruktsiooni proteesiga. Tegemist on Waldemar Link toodetega. Lisaks, Eestis on
erinevatel tootjatel oluliselt kaasaegsemaid lahendusi, mille kliinilised tulemused on
paremad kui mainitud Endo-model proteesil.
B. TEHNILINE KIRJELDUS ON EBASELGE
3.7. Teatud TK tingimusi vaidlustab vaidlustaja ka põhjusel, et nende sisu on ebaselge ning seega
vastuolus läbipaistvuse põhimõttega (RHS § 3 p 1). Vaidlustajal ei ole nende tingimuste sisu
ebaselguse tõttu võimalik aru saada, kas tema pakutavad tooted tingimustele vastavad või mitte.
3.8. Vaidlustaja hinnangul on hankija teadlikult koostanud segaste ning asjatundmatute mõistetega,
tihti omavahel vastuolevad tingimused, et varjata soovi osta kindlaid tooteid kindlatelt
edasimüüjatelt.
TK punkt TK punkti sisu Miks on tingimus ebaselge
Osa 7 TK punkt „Peab ühilduma Pakkujal pole võimalik aru saada, millega
8 megaproteesi (osa 13) peab osas 7 hangitav põlveliigese
komponentidega.“ eriprotees ühilduma. See, millist toodet
hankija osa 13 („megaprotees“) raames
hangib, sõltub sellest, milline pakkumus
osutub edukaks. Pakkumuse esitamise
hetkel ei ole olemas „megaproteesi (osa
13) komponente“, millega oleks üldse
ühilduda võimalik. Jääb mulje, nagu
hakkab hankija osas 7 kontrollima
pakkumuste vastavust selle alusel,
milline pakkumus osutus edukaks osas
13. Selline teguviis on ilmselgelt
õigusvastane.
Lisa 24 Esitada tuleb „reiepea Osas 2, millega pakkumuse maksumuse
(pakkumuse komplekti“ hind. vormi (lisa 24) punkt 2.3 korreleerub, ei
maksumuse ole sellist asja nagu „reiepea komplekt“.
tabel), punkt 2.3 Tsemendilise puusaproteesi varre külge
ei paigaldata komplekte, vaid reiepäid.
Terminist “reiepea komplekt” võib aru
saada kui mitmest osast koosnevast
lahendusest, mis on oluliselt kallim
tavapärasest reiepeast. Hankija on
teabevahetuses kinnitanud, et tegemist
9/13
on „inimliku eksimusega“ ja ta peab
silmas reiepead, kuid ei ole
hankedokumente parandanud. Hankijal
ei ole võimalik teabevahetuse teel
hankedokumente muuta. Maksumuse
vormi ebaselgus võib põhjustada
pakkumuste tagasi lükkamise või
maksumuste võrreldamatuse.
Osade 1—4 ja TK-des on osad nõuded Ebaselgeks jääb, mida need nõuded
10 TK-de liigitatud kui „Muud endast kujutavad – kas tegemist on
peatükid „Muud täiendavad komponendid“, kohustusliku nõuetega või mitte.
täiendavad „Lisavahendid“, „Teised Dokumendi „Juhised pakkujale“ p-s 1.4
komponendid“ lisavõimalused“, on kirjutatud, et mistahes dokumendis
„Lisavõimalused“, „Muud sätestatud kohustus, tingimus või nõue
Osade 3–4 TK-
täiendavad komponendid ja on pakkujale kohustuslik. Sama
de peatükid
võimalused“. dokumendi p-s 6.9 on kirjas, et
„Lisavahendid“
pakutavad tooted peavad vastama
Osade 4—7 TK- kõikidele TK-s esitatud nõuetele. See on
de peatükid ka loogiline, kuna tehniline kirjeldus on
„Teised „asjade omaduste ja neile esitatavate
lisavõimalused“ nõuete loetelu“ (RHS § 87 lg 1 p 1).
Osade 5—7 TK- Teabevahetuses (ID 972062) on hankija
de peatükid aga andnud teada, et n-ö lisanõuete
„Lisavõimalused“ täitmine pole siiski kohustuslik: „Need ei
ole kohustuslikud, kuid kui pakkujal on
Osade 6—9 TK- neid pakkuda, peavad need olema
de peatükid sobivad ja ühilduvad nõutud
„Muud implantaatidega.“ Hankija ei saa
täiendavad teabevahetuse kaudu muuta
komponendid ja kohustuslikku tingimust vabatahtlikuks
võimalused“ (RHS § 46 lg 3). Hankija ei ole lubanud
alternatiivsete pakkumuste esitamist
(„Juhised pakkujale“ p 3.4). Ka praktilises
mõttes jääb täiesti arusaamatuks, kuidas
saavad olla teatud TK nõuded
iseloomuga „kui pakkujal on pakkuda, siis
paku“. Hankija peab hanketingimusi
muutma, et sellest tuleneks üheselt
selgelt, kuidas pakkujad peavad n-ö
lisanõuete osas käituma. Ebaselgus võib
põhjustada pakkumuste tagasi lükkamise
või maksumuste võrreldamatuse.
Seejuures on ilmselgelt asjakohatu
hankija põhjendus, justkui võimaldavad
lisavõimalused „saada pakkumusi
laiemalt“, arvestades et hankija on TK-d
kirjutanud lukku kindlatele toodetele.
Osade 1—14 „Endoproteesi komplekti Jääb arusaamatuks, miks on vaja
TK-de baashinna arvutamiseks endoproteesi komplekti baashinda. Mitte
tingimused, mis liidetakse kõige enam kusagil mujal hanketingimustes ei ole
10/13
algavad kasutatavate mõõtudega reie, viidet endoproteesi komplekti
sõnadega sääre ja liigespinna baashinnale. Millise metoodika järgi
„Endoproteesi komponentide hinnad.“ liidetake kõige enam kasutatavate
komplekti mõõtudega reie, napa ja reiepea
baashinna komponentide hinnad ja miks seda vaja
arvutamiseks on? Kust tuleb sisend „enim kasutatavate
/…/“ mõõtude jaoks“? Pole aru saada, millele
pakkuja selles tingimuses kinnituse
annab.
Osade 1—23 Jääb arusaamatuks, mida tähendab
TK-de instrumentide komplekti “tagamine”.
„Pakkuja peab hankijale
tingimused, mis Endoproteeside valdkonnas on
tagama spetsiifiliste
algavad tavapärane, et hanke võitja annab
instrumentide komplektid,
sõnadega instrumentide komplekti(d) tasuta
mille maksumus on arvutatud
„Pakkuja peab kasutamiseks. Kui tegemist on väga vähe
implantaatide osana (nn
hankijale kasutatavate implantaatidega, siis on ka
garanteeritud instrumentide
tagama võimalus, et hankija ostab instrumentide
komplektid).“
spetsiifiliste komplekti välja. Sõnakasutus „tagama“
instrumentide põhjustab ebaselgust, millele pakkuja
komplektid /…/“. oma kinnituse andma peab. „Tagama“
viitab sellele, et pakkujal peab olema
teoreetiline võimekus komplekte
hankijale tarnida, mitte ei pea ta neid
kindlasti tarnima.
Osade 1—23 „Garanteeritud instrumentide Jääb arusaamatuks, mida tähendab
TK-de komplektide arv arvestatakse “garanteeritud instrumentide komplekt”.
tingimused, mis 1 komplekt/100 endoproteesi Mujal hanketingimused eristata
algavad täiskomplekti kohta.“ „garanteeritud“ ja „garanteerimata“
sõnadega komplekte. Valdkonnas väljakujunenud
„Lisaks garanteeritud
„Garanteeritud tava on, et instrumendid kas antakse
instrumentide komplektidele
instrumentide tasuta kasutamiseks või hankija ostab
on hankijal õigus instrumente
komplektide arv instrumendid välja või kasutatakse
või komplekte ka täiendavalt
arvestatakse laenuinstrumente (instrumendid tuuakse
osta pakkumuses märgitud
/…/“ ja „Lisaks kohale, kui hankija selleks soovi
hindadega.“
garanteeritud avaldab). Pole aru saada, kes, kellele ja
instrumentide mida garanteerib.
komplektidele
Samuti on arusaamatu hankija nõue
/…/“.
saada endale 1 komplekt garanteeritud
instrumente 100 endoproteesi
täiskomplekti kohta. Kui pakkuja kinnitab
selle tingimusega nõustumust, peab ta
olema valmis tarnima hankeperioodi
jooksul kuni 20 komplekti instrumente
ühe osa kohta. Tavapäraselt saab hankija
hakkama 1—3 instrumentide komplektiga
ühe hanke osa kohta kogu hankeperioodi
jooksul (3—4 aastat). Kuna tegemist on
sedavõrd tavatu (ja ebaselge) nõudega,
11/13
ei ole aru saada, millele pakkuja peab
kinnituse andma.
Täiendavat segadust tekitab punkt,
millega hankija annab endale õiguse
lisaks garanteeritud instrumentide
komplektidele instrumente või komplekte
ka täiendavalt osta pakkumuses märgitud
hindadega. Mis mahus, mis tingimustel ja
mis alusel hankija lisatellimusi esitama
hakkab? Kuna juba mõiste „garanteeritud
instrumentide komplektid“ on ebaselge,
pole aru saada, millele pakkuja peab
kinnituse andma. Instrumentide vajadus
ja nende tarnimise tingimused peavad
olema selged.
C. TEHNILINE KIRJELDUS ON EBAPROPORTSIONAALNE
3.9. Viimaks vaidlustab vaidlustaja dokumendi „Juhised pakkujale“ punkte 6.18—6.18.2 ja 7.2:
Punkt 6.18: „Hankijal on õigus nõuda pakkuja poolt pakutud toodete näidiseid koos
paigaldamiseks vajalike instrumentidega nende toodete vastavuse tuvastamiseks (sh
hankijale eesmärgipäraseks kasutamiseks sobivuse määramiseks) läbi funktsionaalsete
omaduste praktilise testimise. Näidis peab olema originaalpakendis, varustatud toote
koodi/ref.nr-ga, toote nimetusega ja vastava pakkumuse osa numbriga.“
Punkt 6.18.1: „Näidis tuleb viia L. Puusepa 8, Tartu, ortopeediakliiniku ortopeedia osakonda
(ruum H 156) või saata postiga samale aadressile. Tellimuse täitmise tähtaeg on 7
(implantaadi näidis) või 14 (implantaadi näidis koos paigalduseks vajalike instrumentidega)
kalendripäeva alates tellimuse esitamisest. Tellimuse mittetähtaegsel täitmisel tunnistatakse
pakkumus mittevastavaks.“
Punkt 6.18.2. „Tootenäidiste maksumust ei hüvitata pakkujale. Näidise paigalduseks vajalikud
instrumendid tagastatakse pakkujale pärast hankemenetluse lõppemist, testimise tarbeks
kasutamise eest ei tasuta.“
Punkt 7.2: „Toodete vastavuse tuvastamiseks võib hankija vajadusel moodustada
ekspertkomisjoni, kes hindab näidiste vastavust praktilise testimise teel. Toodete praktiline
testimine protokollitakse. Kui toote praktilise testimise käigus ilmneb, et toode ei vasta
kirjeldatud omadustele või kasutustehnoloogiale või esineb tehnilisi tõrkeid toote kasutamisel
või toode ei täida ootuspäraselt eesmärki või omab varjatud puudusi, on hankijal õigus
tunnistada pakkumus mittevastavaks.“
3.10. Esiteks, tingimused on ebaproportsionaalsed (RHS § 3 p 1). Need nõuded annavad hankijale nt
õiguse nõuda kõikide komponentide kõikide suuruste kohaletoomist. Primaarproteeside puhul
tähendab selline nõue pakkujale orienteeruvalt 30—100 tuhande euro suurust kulutust.
Revisioon- või megaproteesi puhul võivad summad ulatuda sadade tuhandeteni. Sealjuures
peavad pakkujad suutma hankija nõude täita 7 või 14 kalendripäevaga. Olukorras, kus hankijal
on võimalik toodete vastavust kontrollida tehnilise dokumentatsiooni alusel – vajaduse korral
nõudes välja ka täiendavaid tõendeid –, on pakkumuste vastavuse kontrollimine sellisel viisil
pakkujad ilmselgelt ebamõistlikult koormav.
3.11. Teiseks, tingimused on läbipaistmatud (RHS § 3 p 1). Ebaselgeks jääb, kelle käest asub pakkuja
näidiste toomist nõudma. Vastavust tuleb kontrollida kõikide pakkumuste puhul (RHS § 114 lg 1).
Tingimuste sõnastusest tulenevalt on hankijal aga õigus nõuda näidiste toomist vaid ühe pakkuja
puhul, kusjuures nõude täitmata jätmine on eraldiseisev mittevastavuse alus ja sellega kaasnevat
kulu pakkujale ei hüvitata. On ilmselgelt õigusvastane olukord, milles hankija kontrollib
12/13
pakkumuste vastavust erineval moel. Lisaks, kui pakkuja ei pea näidiste toomisega kaasnevat
kulu äriliselt mõistlikuks (ja hankija teab seda), on hankijal sisuliselt võimalik tõrjuda
hankemenetlusest pakkuja, kes talle ei sobi. Pingereas teise pakkuja puhul pole aga hankija
kohustatud näidiste toomist nõudma, kuna tegemist on tema õigusega.
4. MENETLUSLIKUD KÜSIMUSED
4.1. Riigilõiv. Vaidlustaja on tasunud vaidlustuse eest riigilõivu 1280 eurot (RLS § 258 lg 1 p 2) (lisa
10).
4.2. Esindus. Vandeadvokaat Erki Fels ja vandeadvokaadi abi Gregor Saluveer kinnitavad, et on
vaidlustaja lepingulised esindajad (RHS § 190 lg 11).
4.3. Tähtaegsus. Pakkumuste esitamise tähtpäev on 19.06.2025. Vaidlustus on tähtaegne (RHS §
189 lg 2 p 2).
4.4. Läbivaatamine. Vaidlustaja on nõus kirjaliku menetlusega (RHS § 190 lg 1 p 6).
4.5. Esitamine. Vaidlustus on esitatud VAKO-le e-posti teel (
[email protected]).
Lugupidamisega
(digitaalselt allkirjastatud)
Erki Fels, Gregor Saluveer,
vandeadvokaat vandeadvokaadi abi
LISAD
Lisa 1 SymphoKnee väljavõte
Lisa 2 Gemini SL Surgical technique
Lisa 3 Link SymphoKnee Surgical technique
Lisa 4 OXINIUM-Knee-Brochure
Lisa 5 Legion TKS Surgical technique
Lisa 6 SL PLUS MIA surgical technique
Lisa 7 R3 Surgical technique
Lisa 8 SP-CL Product Rationale
Lisa 9 SP-CL Anatomically Adapted Cementless Hip System
Lisa 10 Riigilõivu tasumist tõendav maksekorraldus
13/13