Artificial Organs
32(12):962–998, Wiley Periodicals, Inc.
© 2008, Copyright the Authors
Journal compilation © 2008, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.
Thoughts and Progress
Score Model for the Evaluation of Dialysis should include platelets, leukocytes, and red cells, as
Membrane Hemocompatibility well as complement proteins, coagulation factors, and
a variety of other plasma proteins.
*Ansgar Erlenkötter, *Pia Endres, *Bernd Nederlof, Currently available membranes not only differ
†Carsten Hornig, and †Jörg Vienken with respect to their physicochemical properties,
*Fresenius Medical Care, St. Wendel; †Fresenius such as performance, thickness, pore size, porosity,
Medical Care, Bad Homburg, Germany and adsorptive capacities, they also show different
grades of hemocompatibility. Reasons for this are
Abstract: The interaction of blood with artificial surfaces is not only differences in polymer composition, but
of particular interest during hemodialysis treatments with also conditions of manufacturing and of sterilization
extracorporeal blood circuits. Components of the extracor- techniques (1).
poreal blood circuit are known to have only a moderate, The large surface area of a dialysis membrane
sometimes even an unfavorable hemocompatibility, and
and the chronic exposure of blood to a foreign
thus may provoke adverse biochemical or clinical sequelae.
This article describes a newly established hemocompatibil- surface area during hemodialysis treatment has been
ity assessment score. This score is based on on a standard- the motivation for extensive research on how to
ized series of in vitro tests and is applied to commercially improve the hemocompatibility pattern of dialysis
available hemodialysis membranes. It relates to a variety membranes. Standard investigations are still based on
of membrane polymers, such as regenerated cellulose,
in vitro test systems only.
diethylaminoethyl-modified cellulose, polyethersulfone/
polyarylate blends and polysulfone. In order to compare This article describes an in vitro assessment of the
different polymers used in the manufacturing of dialysis hemocompatibility pattern of five different polymers
membranes, a set of the following hemocompatibility used for commercially available dialysis membranes.
parameters was assessed and assembled to an overall Different polymer classes, such as regenerated
score: generation of complement factor 5a, thrombin-
cellulose, diethylaminoethyl cellulose (DEAE),
antithrombin III-complex, release of platelet factor 4, gen-
eration and release of elastase from polymorphonuclear polyethersulfone/polyarylate (PEPA) blends and
granulocytes, and platelet count. With respect to these polysulfone, as well as their different modes of ster-
parameters, the results reveal major differences between ilization, have been assessed and are compared. Test
the selected dialysis membranes. This new score model results of the individual polymers have led to a new
proves to be an efficient tool to derive objective results, and
score model that has been developed for this inves-
it may, thus, be used in the future to facilitate the selection
of membrane polymers with an appropriate hemo- tigation and allows for both to provide an objective
compatibility pattern for dialysis therapy. Key Words: assessment tool for the comparison of different poly-
Dialysis membrane—Dialyzer—Hemocompatibility— mers and to guide manufacturers for future mem-
Score model. brane development.
Biocompatibility has become the number one
factor for the characterization of dialyzers, and in MATERIALS AND METHODS
particular of the incorporated dialysis membrane.
Consequently, the target of membrane development Blood collection
in the past was to improve its bio(hemo)compatibility. 450 mL of whole blood was collected from
Markers to characterize membranes under investiga- healthy volunteers. A specially designed blood
tion must reflect both the cellular and the plasmatic bag system with a 17 G (1.5 mm) dialysis fistula
components of the blood–membrane interface. They needle (Fresenius Medical Care, Bad Homburg,
Germany) was applied for blood collection. Whole
blood was collected in these bags, which had been
doi:10.1111/j.1525-1594.2008.00657.x prefilled with 750 IU heparin (Ratiopharm, Ulm,
Received April 2007; revised October 2007. Germany) in a 50-mL saline solution, resulting in
Address correspondence and reprint requests to Dr. Ansgar a final heparin concentration of 1.5 IU/mL. This
Erlenkötter, Biocompatibility Laboratory, Fresenius Medical Care,
Frankfurter Straße 6-8, D—66606 St. Wendel, Germany. E-mail: heparin concentration was used in order to simulate
[email protected] clinical dialysis as close as possible and reflects an
962
THOUGHTS AND PROGRESS 963
Reservoir
Pven
FIG. 1. Experimental setup of the recircu-
Dialyzer lation model including a sample reservoir,
two pressure measure chambers, a pump
segment, and a polyvinylchloride tubing
(blood flow rate: QB = 200 mL/min, filling
Sampling
volume: VB = 150 mL, heparinization:
point 1.5 IU/mL). Two of these setups were used
Part in parallel in order to allow for membrane
comparisons under identical conditions.
Blood pump
appropriate anticoagulation for in vitro testing of lants (Sarstedt Monovetten, Sarstedt, Nümbrecht,
dialyzers. This heparin concentration further yielded Germany). Additional samples were taken directly
the most pronounced differences in hemocompatibil- from the blood bags immediately before the filling
ity pattern between different polymers. All hemo- procedure. These were used as prevalues (0 min).
compatibility experiments started within 30 min after
blood collection. Membrane polymers
For the assessment of dialysis membrane hemo-
Recirculation model compatibility, five different, commercially available
Figure 1 shows the in vitro recirculation model membranes made from polysulfone and cellulose
used in our study. The test system consists of a dia- were chosen: g-sterilized polysulfone (Toraysulfone,
lyzer and a tubing system specifically developed for BS-1.3 dialyzer, 1.3 m2, Toray, Tokyo, Japan),
our test system using standard bloodline components g-sterilized polyethersulfone/polyarylate (PEPA =
(Fresenius Medical Care). In each experiment, two polyester–polymer alloy, FLX-12GWS dialyzer,
test systems (test vs. reference dialyzer) were evalu- 1.2 m2, Nikkiso, Tokyo, Japan), steam-sterilized
ated in parallel in a standard incubator (Memmert, polysulfone (Helixone, FX60 dialyzer, 1.4 m2
Schwabach, Germany). The temperature was set at [Fresenius Medical Care]), regenerated cellulose
37°C in all experiments. (Cuprophan, GFS 12 dialyzer, 1.3 m2, Gambro AB,
The systems were prerinsed with saline solution Hechingen, Germany), and cellulose etherified with
(0.9% NaCl) for approximately 30 min. In order to DEAE groups (Hemophan, GFS Plus 12 dialyzer,
prevent any ultrafiltration, the dialyzate compart- 1.3 m2, Gambro AB). Membranes were assessed in
ment was filled with saline solution and clamped. their original housing. Thus, it was not possible to
After the prerinsing step, the test systems were care- achieve absolutely identical, but only similar mem-
fully filled with blood. Blood volume in the circuit brane surface areas. Differences are still significant as
was adjusted to 150 mL in order to keep the blood the dialyzer with the largest surface area yielded
volume for all test series constant. Each donor pro- better results. In each test, the dialysis membranes
vided blood for one double experiment. were tested in comparison with the same reference
Blood flow rate in the recirculation setup was estab- membrane, that is, polysulfone (Helixone [Fresenius
lished at 200 mL/min. Blood samples were taken at Medical Care]), because this material is one of the
several time points (15, 30, 60, 120, and 180 min) from most widely used membrane polymers in the world
the two circuits into syringes prefilled with anticoagu- today.
Artif Organs, Vol. 32, No. 12, 2008
964 THOUGHTS AND PROGRESS
FIG. 2. (a) and (c) course of C5a generation (a, n = 2) and platelet loss (c, n = 3) during a single experiment (dialyzer A: left graph/closed
circles, steam-sterilized polysulfone membrane (Helixone); dialyzer B: right graph/open circles, g-sterilized polysulfone membrane
(Toraysulfone). The gray area under the curve illustrates the integral generation of either C5a or platelet loss during the time course of
the experiment. The double-headed arrow represents the calculated mean C5a generation per hour or the mean platelet loss; (b)
correlation of C5a score versus normalized mean C5a generation per hour: three pairs of variables were predefined: point A represents
the mean C5a generation with no activation of the complement system (0¦0), point B represents the mean C5a generation of the reference
dialyzer (100¦20), and point C represents the maximum level of C5a generation by one of the test dialyzers (14 440¦100); (d) correlation
of PLT score versus mean PLT loss: three pairs of variates were predefined: point A represents no mean PLT loss (0%¦0), point B
represents the mean PLT loss of the reference dialyzer (IPLT,reference¦20), and point C represents the total loss of platelets (100%¦100). The
dotted line represents the correlation for a reference value of 5%; the dashed line represents correlation for a reference value of 75%.
Determination of hemocompatibility markers Data analysis
Hemocompatibility was assessed by measuring The assessment of hemocompatibility data should
complement factor 5a (C5a), thrombin-antithrombin be transparent and traceable. Therefore, we have
III-complex (TAT), platelet count (PLT), platelet developed a four-step procedure to analyze the raw
factor 4 (PF4), and release of elastase from polymor- data.
phonuclear granulocytes (PMN-elastase). These 1st step: Calculation of the mean C5a genera-
parameters have turned out to reflect a typical selec- tion per hour, the mean TAT generation per
tion of well-established hemocompatibility markers. hour, the mean platelet loss, the mean PF4 release
Platelet count and hematocrit (HCT) were measured per hour, and the mean PMN-elastase release
in triplicate by using an automatic hematology ana- per hour.
lyzer (K 4500, Sysmex, Norderstedt, Germany). C5a A typical course is shown in Fig. 2a for C5a-
(EIA-3327, DRG Instruments, Marburg, Germany), generation and in Fig. 2c for platelet loss during a
TAT (Enzygnost TAT micro, Dade Behring, Marburg, single experiment. The left graph shows the results of
Germany), PF4 (1875353 Asserachrom, Roche Diag- steam-sterilized polysulfone membrane (Helixone)
nostics, Mannheim, Germany), and PMN-elastase (dialyzer A), while the right graph represents data for
(K6840, Immundiagnostik AG, Bensheim, Germany) a g-sterilized polysulfone membrane (Toraysulfone)
plasma concentrations were determined in duplicate (dialyzer B). As a measure of the total C5a genera-
by applying commercially available enzyme-linked tion and platelet loss, the area under the curve
immunosorbent assay kits. (AUC) was calculated. A mean C5a generation rate
Artif Organs, Vol. 32, No. 12, 2008
THOUGHTS AND PROGRESS 965
per hour was achieved by multiplying with 120/TI2. Normalized values are similarly calculated for
This leads to the index IC5a (Eq. 1) TAT, PF4, and PMN-elastase. A normalized value
(IC5a = mean C5a generation per hour, cC5a,pre = C5a was not calculated for the PLT results.
concentration in blood bag, cC5a,HCT–cor. = HCT cor- 3rd step: Calculation of the hemocompatibility
rected C5a concentration, t = time, TI = total incuba- scores for C5a, TAT, PLT-loss, PF4, and PMN-
tion time): elastase.
Hemocompatibility scores were introduced in
120 ⎛ 2
t
⎞ order to compare the previously calculated values
2 ⎜ ∫ C 5 a , HCT − cor .
IC 5 a = c (t )dt − cC 5 a, pre ∗TI ⎟ (1) among the individual experiments (C5a score, TAT
TI ⎝ t1 ⎠ score, PLT score, PF4 score, and PMN-elastase score).
Figure 2b shows the correlation of a C5a score versus
This index represents the mean increase of the C5a normalized C5a generation per hour. This correlation
level during 1 h of recirculation. In Fig. 2a, it is rep- is calculated by predefining three pairs of variables.
resented by the double-headed arrow. The param- A normalized mean C5a increase per hour of zero
eters TAT, PF4, and PMN-elastase were calculated in results in a C5a-score of zero and reflects no activation
an analogous manner. of the complement system (see point A in Fig. 2b). A
Figure 2c represents the platelet loss in % versus score of 20 is purposely defined as a 100% normalized
time. Platelet loss is only quantified for the initial mean C5a increase per hour as it is predefined for the
phase of the in vitro recirculation (0–60 min). In a reference dialyzer (see point B).The dialyzer with the
later stage, platelet count increases following a highest degree of activation defines the last pair of
detachment of platelets from the surface. These pos- variables and is named “Reference 2.” In the example
sibly “dead” cells have lost their overall functionality of Fig. 2b, the normalized C5a increase per hour of
and thus cannot be used for the scoring analysis. Reference 2 was 14 440% (see point C). Equation 4
The mean platelet loss is calculated by dividing shows the mathematical solution of this correlation
the integral (gray area under the curve) as shown in (nIC5a,reference 2 = normalized mean C5a generation per
Eq. 2 by the initial incubation time (IPLT = mean hour of reference 2; IC5a = mean C5a generation per
platelet loss [%], t = time, TII = initial incubation hour of the test dialyzer).
time). In the figure, it is represented by the double-
headed arrow. ⎛ Ig 5 ⎞
⎜ 1+ + 2 Ig 5 − Ig nIC 5 a , reference 2 ⎟
C 5a score = 4⎝ Ig 4 ⎠
∗
t2 ⎛ Ig 5 ⎞
1+ 2
∫ Δc PLT (t )dt
(2) 5 ⎝⎜ Ig 4 + 2 Ig 5 − Ig nI C 5 a , reference 2 ⎠⎟
∗
(4)
I PLT = t1
⎛ Ig 5 ⎞
+ 2 Ig 5 − Ig nIC 5 a ,rreference 2 ⎟
TII ⎜
nIC 5 a ⎝ Ig 4 ⎠
2nd step: Calculation of the normalized mean C5a This kind of calculation is similarly used for TAT,
generation per hour, the normalized mean TAT gen- PF4, and PMN-elastase analyses.
eration per hour, the normalized mean PF4 release The calculation of the PLT score is slightly different
per hour, and the normalized mean PMN-elastase and illustrated in Fig. 2d. In this calculation, no plate-
release per hour, respectively. let loss (IPLT = 0%) results in a PLT score of zero (see
The blood donor characteristics have a strong point A). The total loss of platelets is defined as 100
influence on hemocompatibility results. Therefore, (see point C).The reference value is set as a PLT score
a normalization step is necessary and is always of 20 once again (see point B1 and B2). Depending on
advisable. The previously calculated values are nor- the mean platelet loss of the reference dialyzer, two
malized by defining a reference value of 100%. different correlations are shown in Fig. 2d.The dotted
For example, the normalized mean C5a generation line represents the correlation for a reference value
per hour is calculated as shown in Eq. 3 of 5%, and the dashed line for a reference value of
(nIC5a = normalized mean C5a generation per hour, 75%. The correlation of PLT-score and mean PLT
IC5a,reference dialyzer = mean C5a generation per hour of loss is described by Eq. 5 (IPLT = mean PLT loss,
the reference dialyzer, IC5a,test dialyzer = mean C5a gen- IPLT,REF = mean PLT loss of the reference dialyzer).
eration per hour of the test dialyzer):
⎛ Ig 5 ⎞ Ig 5
1−
⎝⎜ Ig 100 − Ig I PLT,reference ⎠
⎟ − Ig IPLT,reference
IC 5 a,test dialyzer PLT score = 100 ∗ IPLT Ig 100−
nIC 5 a = ∗ 100 (3)
IC 5 ,reference dialyzer (5)
Artif Organs, Vol. 32, No. 12, 2008
966 THOUGHTS AND PROGRESS
FIG. 3. Dialyzer membrane polymer versus (a) platelet score; (b) TAT score; (c) C5a score; and (d) PMN-elastase score. High scores
implicate an unfavorable hemocompatibility and low scores implicate an excellent hemocompatibility. The solid line indicated by the arrow
illustrates the predefined value for the standard reference. Data are means ⫾ SD (n = 3).
4th step: Calculation of the total hemocompatibil- RESULTS
ity score.
In all experiments, the steam-sterilized polysulfone
Five hemocompatibility scores could be calcu-
membrane (Helixone) in its original housing served
lated by the procedure described here. To create a
as a standard reference, because this material is cur-
total hemocompatibility score (THS) as requested
rently the most widely used membrane material.
for the selection of polymers for membrane manu-
Changes of hemocompatibility pattern are always
facturing, weighing factors are defined for each
expressed in relation to this reference. Figure 3 shows
hemocompatibility category. Following our experi-
the results of all hemocompatibility assessments
ence, the determination of the weighing factors is
comparing different membrane polymers.
arbitrarily taken for each hemocompatibility param-
eter as 0.25 for the plasmatic immune system (C5a),
cell activation (PMN-elastase), and coagulation PLT score (Fig. 3a)
activation (TAT). PF4 and platelet count are taken The highest degree of platelet loss in vitro was
equally, with their weight summing up to a final observed for the PEPA membrane (PLT score =
figure of 0.25. The THS is then calculated by 80 ⫾ 6, in Fig. 3a) leading to a score value of 80.
Eq. 6: Further, the g-sterilized polysulfone membrane
(Toraysulfone) showed a higher PLT loss (PLT
THS = fC 5 a ∗ C 5 a score + fTAT ∗TAT score + fPLT ∗ score = 46 ⫾ 18) in comparison with the reference
PLT score + fPF 4 ∗ PF 4 score + (6) material, while a lower PLT loss was recorded only
fPMN ∗ PMN score for the cellulosic membranes (Cuprophan, PLT score
Artif Organs, Vol. 32, No. 12, 2008
THOUGHTS AND PROGRESS 967
FIG. 4. Calculated THS for steam-steri-
lized polysulfone (Helixone), g-sterilized
polysulfone (Toraysulfone), PEPA, regen-
erated cellulose (Cuprophan), and DEAE
cellulose (Hemophan). High scores impli-
cate an unfavorable hemocompatibility and
low scores implicate an excellent hemo-
compatibility. The solid line indicated by the
arrow illustrates the predefined value for
the standard reference.
16 ⫾ 4 and Hemophan, PLT score = 12 ⫾ 3). The rial (PMN-elastase score: 23 ⫾ 11). A lower level
results for the PF4 score were comparable. could be measured for the g-sterilized polysulfone
membrane (Toraysulfone, PMN-elastase score:
TAT score (Fig. 3b) 14 ⫾ 7).
DEAE cellulose membrane (Hemophan) showed
the highest TAT score: 100 ⫾ 40. Steam-sterilized Total hemocompatibility score
polysulfone membrane (Helixone) showed a TAT From all results of the individual in vitro scoring
score of 17 ⫾ 5 as compared with a g-sterilized experiments, a THS could be derived as a summary
polysulfone membrane (Toraysulfone) with a TAT (see Fig. 4).
score of 28 ⫾ 16. The PEPA membrane exhibited a The THS of the steam-sterilized polysulfone mem-
TAT score of 32 ⫾ 25, and the regenerated cellulose brane (Helixone) was found to be 19.6. This closely
membrane (Cuprophan) a TAT score of 32 ⫾ 25. resembles the predefined value of 20 for the reference
membrane. The THS values further increased in the
C5a score (Fig. 3c) following order: g-sterilized polysulfone membrane
A lower complement activation score compared (Toraysulfone, THS: 32) < PEPA membrane (THS:
with the reference was only found for the PEPA 38) < DEAE cellulose membrane (Hemophan,
membrane (C5a score: 5.0 ⫾ 2.5). All other materials THS: 40) and < regenerated cellulose membrane
under investigation showed higher levels: g-sterilized (Cuprophan, THS: 63).
polysulfone (Toraysulfone, C5a score: 30.1 ⫾ 1.5),
DEAE-modified cellulose membrane (Hemophan) DISCUSSION
with a C5a score of 41.8 ⫾ 4.1, and the highest acti- In this study, the hemocompatibility pattern of five
vation was found for the regenerated cellulose mem- hemodialysis membranes composed of different
brane (Cuprophan, C5a score = 100 ⫾ 1). The C5a polymers (regenerated cellulose, DEAE cellulose,
score for the steam-sterilized polysulfone (Helixone) PEPA, and polysulfone) was assessed. A standard-
in a direct comparison was found to be 19.9 ⫾ 0.1. ized in vitro test system and a newly developed score
model was developed for this purpose. Based on the
PMN-elastase score (Fig. 3d) results of this in vitro evaluation and by means of
PMN-elastase release was found to be highest for the new score model, a total hemocompatibility
the regenerated cellulose membrane (Cuprophan, score could be derived for each membrane type and
PMN-elastase score: 100 ⫾ 106), followed by the allowed a comparison of its in vitro hemocompatibil-
DEAE cellulose membrane (Hemophan, PMN- ity performance.
elastase score: 11 ⫾ 5). The PEPA membrane was The selection of a certain polymer type prior to
found to be in the same range as the reference mate- the manufacturing of dialysis membrane influences
Artif Organs, Vol. 32, No. 12, 2008
968 THOUGHTS AND PROGRESS
the hemocompatibility pattern of a membrane. a substitution of hydroxylgroups by DEAE
However, further factors, such as potting material, groups, bears positive charges in a physiological
sterilization techniques, geometry of the membrane environment. This surface charge gives rise to the
capillaries, and the overall manufacturing conditions adsorption of negatively charged heparin to some
further influence the hemocompatibility profile. A extent. A lack of freely available heparin due to its
straightforward assessment of membranes in a com- adsorption at the membrane surface might explain
parable housing should facilitate the selection of the higher TAT values (Fig. 3b).
polymers for large-scale production, as revealed in It might be necessary to monitor the effects of
our study. As shown here, a membrane–polymer different materials used for the fabrication of dialysis
evaluation, even if limited to five parameters but membranes on the level of single parameters.
carried out in a recirculation model under standard- However, the complexity of biochemical reactions at
ized conditions with respect to blood collection, the blood–membrane interface implies that the
heparinization, blood flow rate and blood volume, overall hemocompatibility of a material may not be
tubing sets, time of investigation, and temperature appropriately represented by a single parameter
came across with results enabling the manufacturer alone (6). This is supported by the observations from
to make up his mind for actual or future polymer our study. Here, the individual materials showed a
selection. The use of mini-modules have proven to different behavior with respect to each parameter
be a practical tool in in vitro or ex vivo experiments, under investigation. Thus, it seems more convenient
(2,3) and thus have been used successfully in our to monitor the overall hemocompatibility of a mate-
membrane assessments. In addition, studies with rial, taking into account a set of hemocompatibility
full-size hemodialyzers in in vitro circuits are com- parameters in order to facilitate the decision in favor
parable with the experimental setup of our study or against a certain polymer. Therefore, the devel-
(4,5). In this study, we focused on parameters related oped score model, which incorporates and summa-
to the activation of the complement system, the acti- rizes the results of the individual experiments, offers
vation of blood coagulation, and the activation of an objective total hemocompatibility assessment for
cells, such as platelets and leukocytes. each tested polymer. Score models have also been
Striking differences for membranes could be employed by other groups before (7).
shown for some of the parameters.They can be attrib-
uted to both the molecular composition of the
CONCLUSIONS
membrane bulk as well as the permeability for low-
molecular-weight proteins. To avoid effects specific After calculation of the THS value for each
for low- or high-flux membranes, we clamped the polymer, it is possible to determine clear-cut differ-
ports for dialysis fluids in the respected dialyzer and ences between the polymers used. While regenerated
investigated biocompatibility parameters exclusively cellulose had the highest score value and thus the most
as a function of the membranes surface properties unfavorable overall hemocompatibility of all poly-
and not for their filtration effects. mers tested, other cellulose-based polymers behaved
In addition, some activated proteins that are sig- also less well than the three remaining synthetic poly-
nificant for blood compatibility might be adsorbed at mers. PEPA had a better overall hemocompatibility
the membrane surface. This is valid for the activated than regenerated cellulose or DEAE cellulose. Nev-
complement proteins C3a and C5a for some but not ertheless, in the group of synthetic polymers, this
all membrane polymers. The evaluated AUC, then, synthetic biomaterial showed the lowest hemo-
represents the difference between generation and compatibility. This study also revealed differences
loss through adsorption. There is a notion that mem- between the two polysulfone types investigated: The
brane biocompatibility is related to free and not g-sterilized polysulfone had a worse overall hemo-
to the amount of adsorbed activated complement compatibility than the steam-sterilized polysulfone.
proteins. As an example, polyacrylonitrile (PAN) The intention of this article was to establish an in
membranes, such as AN69 (Hospal, Lyon, France), vitro score model and not to offer generalized char-
are considered to be highly biocompatible due to acteristics of different polymers. This new score
their adsorptive capacity for complement proteins model reflects an overall hemocompatibility score for
and despite a considerable complement activation. each individual material. This score model delivers
Consequently, the assessment of complement activa- objective results and allows for the comparison of
tion in this article was done accordingly. materials that are not tested simultaneously but after
Further, Hemophan, the cellulosic membrane in certain time intervals. Further, with the help of such a
our investigations that has been modified through score model, the selection of membrane materials
Artif Organs, Vol. 32, No. 12, 2008
THOUGHTS AND PROGRESS 969
with an appropriate hemocompatibility pattern is
possible and can be applied even for other applica-
tions of membranes or biomaterials in medical device
technology.
Acknowledgments: The authors gratefully ack-
nowledge the excellent technical support of Ms. I.
Werth, M. Reis and H. Hoffmann.
REFERENCES
1. Masaki T, Gilson J, Leypoldt JK, Cheung AK. Effect of perme-
ability on indices of haemodialysis membrane biocompatibility.
Nephrol Dial Transplant 1999;14:1176–81.
2. Schmidt B, Mujais SK. Evaluation of dialysis membrane-blood
compatibility: experimental methods. In: Bonomini V, Berland
Y, eds. Dialysis membranes: structure and predictions. Contrib
Nephrol 1995;113:32–44.
3. Mulvihill J, Cazenave JP, Mazzucotelli JP, et al. Minimodule
dialyser for quantitative ex vivo evaluation of membrane hemo-
compatibility in humans: comparison of acrylonitrile copoly-
mer, cuprophan and polysulphone hollow fibres. Biomaterials
1992;13:527–36.
4. Yang MC, Lin CC. In vitro characterization of the occurrence of
hemolysis during extracorporeal blood circulation using a mini
haemodialyser. ASAIO J 2000;46:293–7.
5. Ward RA, McLeish KR. Haemodialysis with cellulose mem-
branes primes the neutrophil oxidative burst. Artif Organs
1995;19:801–7.
6. Boure T, Vanholder R. Which dialyser membrane to choose?
Nephrol Dial Transplant 2004;19:293–6.
7. Remuzzi A, Boccardo P, Benigni A. In vitro platelet adhesion to
dialysis membranes. Nephrol Dial Transplant 1991;6 (Suppl.
2):36–9.
Artif Organs, Vol. 32, No. 12, 2008
Didactical article
Reçu le :
3 avril 2012
Accepté le :
Dialyzers biocompatibility and efficiency
20 janvier 2013
determinants of sterilization method choice§
Mise au point sur la stérilisation des dialyseurs :
un des critères de choix à prendre en compte en 2012
B. Allard (Pharmacien gérant)a,*, R. Begri (Pharmacien adjoint)a, J. Potier
(Médecin praticien hospitalier)b, S. Coupel (Médecin adjoint centre
formation)c
Available online at a
ECHO, pharmacie à usage intérieur, 11, rue Félix-Éboué, BP55, 44401 Rezé, France
b
Service d’hémodialyse, centre hospitalier du Cotentin, 46, rue du Val-de-Sair,
ScienceDirect 50102 Cherbourg, France
c
www.sciencedirect.com ECHO, site du Confluent, 6, rue de la Gare, BP10095, 44402 Rezé cedex, France
Summary Résumé
Introduction. When selecting a dialyzer, the criteria of biocompa- Introduction. Les critères de biocompatibilité, entrant dans le
tibility entering the choice of the dialyzer are based on the nature of choix du dialyseur, reposent avant tout sur la nature du polymère
the polymer constituting the membrane, but rarely on the dialyzer de la membrane, et rarement sur le mode de stérilisation du
sterilization methods. dialyseur.
Methods. Thus, it seemed necessary to analyze the advantages and the Méthodes. Il est donc apparu nécessaire de définir les modes de
disadvantages of the modes of sterilization and to analyze their impact stérilisation et analyser leurs impacts sur la biocompatibilité et les
on the performances and the biocompatibility of dialyzers to integrate performances des dialyseurs, de façon à intégrer ce critère dans
this parameter to their investment choices for the kidney machine. l’arbre décisionnel de choix du dialyseur.
Results. Radiation sterilization induces structural modifications of Résultats. La stérilisation par irradiation s’accompagne de modifi-
the dialyzers components. Beta radiation causes fewer modifications cations structurelles des constituants des dialyseurs, l’irradiation
than gamma radiation; these structural modifications can release bêta-entraı̂nant moins de modifications que l’irradiation gamma, ces
cytotoxic components. Steam sterilization is a reference in terms of modifications structurelles pouvant amener la libération de composés
security for thermostable materials with an added benefit for steam cytotoxiques. Les modes de stérilisation par chaleur humide, auto-
process, which allows the removal of waste generated during the clave ou vapeur fluente, restent une référence en terme de sécurité
dialyzer’s manufacturing. pour les matériaux non thermosensibles, avec un avantage supplé-
Discussion. For the same membrane, performances of the dialyzers mentaire pour la vapeur fluente, permettant un effet de chasse des
change according to the methods of sterilization, inducing different résidus de fabrication.
performances and different uses. Discussion. Pour une même membrane, les performances des dia-
Conclusion. To conclude, beyond the choice of the polymer, the lyseurs sont modifiées en fonction du mode de stérilisation, entraı̂-
method of sterilization should consistently be taken into account in nant des performances et possibilités d’utilisation différentes.
therapeutic choices and purchase of dialyzers. Considering its very Conclusion. En conclusion, au-delà du choix du polymère, le mode
high biocompatibility and considering the acquired experience, de stérilisation devrait être systématiquement pris en compte dans
§
The editorial board of Le Pharmacien Hospitalier et Clinicien considers each published article for funding English translation and online publication of the most
relevant article per issue. A research grant was awarded to Allard et al. Pharm Hosp Clin 2013;48 [In press].
DOI of original article: http://dx.doi.org/10.1016/j.phclin.2013.01.011
* Corresponding author.
e-mail :
[email protected] (B. Allard).
2211-1042/$ - see front matter ß 2013 Elsevier Masson SAS. All rights reserved.
http://dx.doi.org/10.1016/j.phclin.2013.10.071 Le Pharmacien Hospitalier et Clinicien 2013;48:e15-e21
e15
B. Allard et al. Le Pharmacien Hospitalier et Clinicien 2013;48:e15-e21
moist heat sterilization is to be favored in the choice of dialyzers of les processus de choix thérapeutique et donc rentrer en compte dans
equivalent performances. Furthermore, non-heat sensitive products le processus d’achat.
have to be developed first and foremost with this mode of sterilization. ß 2013 Elsevier Masson SAS. Tous droits réservés.
ß 2013 Elsevier Masson SAS. All rights reserved.
Mots clés : Dialyseur, Stérilisation, Biocompatibilité, Performance
Keywords: Dialyzer, Sterilization, Biocompatibility, Dialyzer
performance
Introduction Dialyzers in their definitive packaging are exposed several
hours for the gamma radiation, or few minutes only under
Dialyzers (so called kidney machine) are sterile medical devi- the electron-beam for the beta brilliance. For short-term
ces that are made up of various polymers or cellulose acetate and surge requirements beta sterilization allows high
assembled between them in a synthetic polymer hull. productivity.
Good biocompatibility and performances for the clearance of Nevertheless, beta radiation allows a more precise radiation.
uraemic toxin are the matters of vital concern. Key performance During the gamma radiation sterilization, dialyzers are dry
results considered are permeability, surface and uraemic toxin (without an agent of filling) or filled with an aqueous solution
cleared according to the technique of dialysis used. Unlike the in the blood and dialysate compartments. The name ‘‘gamma
first three criteria, biocompatibility does not set the stage for wet’’ (wet) or ‘‘gamma dry’’ (dry) results from the state of the
manufacturers and prescribers performance reporting. dialyzer at the time of the sterilization up to its end use.
So, the criteria of biocompatibility entering the choice of the In the sterilization gamma, the agent by filing allows the
dialyzer are based on the nature of the polymer constituting distribution of the residues of manufacturing and products
the membrane, but rarely on the mode of sterilization of the generated during the sterilization in the aqueous solution.
dialyzer. The sterilization technology has evolved significantly Nonetheless, these dialyzers are much heavier and sensitive
over the past decades and has an influence on the biocom- to the temperature.
patibility of dialyzers as well as on their performances.
The table in supplementary data, Appendix 1 summarizes the
Impact of the radiation on the dialyzer constituting materials
various ranges of Dialyzers, classified according to the manu-
facturers of membrane, and specifies the various methods of Depending on the dose, radiation induces structural modifi-
sterilization as well as the places of manufacturing and the cations, splits and reticulations leading to a molecular rear-
type CE mark attached. rangement that can affect in a more or less important way
Thus, it seemed necessary to analyse the advantages and the polymers to be sterilized [1,2]. The absorbed energy is expres-
disadvantages of the modes of sterilization and to analyse sed in Gray unit.
their impact on the performances and the biocompatibility of Exposure to the beta radiation may induce fewer structural
Dialyzers to integrate this parameter to their investment modifications due to a shorter duration and a more precise
choices for the kidney machine. radiation dose. However, only scarce data are available [3]. The
structural changes observed after gamma radiation, can arise
even for microdoses. In the case of the polysulfone, these
Various methods of sterilization appear for a 1.66 KGy dose while the Pharmacopoeia sterilizing
dose required is 25 KGy [4]. According to the sterilization
There are numerous methods of sterilization as specified in
method, these modifications may impact biocompatibility
the plan below (fig. 1), but today, ethylene oxide sterilization
because of the new substances in the unknown toxicity gene-
gave way to radiation and heat sterilization (fig. 1).
rated and/or because of the alteration of performances of the
dialyzer compared to the non-sterilized initial product [5].
Radiation sterilization Furthermore, the structural modifications continue even after
the end of the radiation.
Principle and implementation These changes would be due to the presence of free radicals
Radiation sterilization bombard the medical device with very [1] formed within the polymer during the radiation process
high energy ionizing particles altering microorganisms DNA/ that causes accelerated oxidation after the conclusion of the
RNA and producing very reactive free radicals. sterilization. The presence of these free radicals can be mini-
The powerful ionizing radiation sources used for dialyzers are mized by radiation without oxygen.
mainly generated by the gamma radiation of cobalt 60 or by (tables Ia and Ib), according to the works of Uhlenbusch-
electron-beam for the beta radiation equipment. Körwer and Nyoman Ruspiasih, shows the main polymer
e16
Dialyzers determinants of sterilization method choice
Table Ia
Effects of the gamma radiation on polymers.
Effets de l’irradiation gamma sur les polymères.
Monomer/Polymer Gamma irradiation effect
Acrylic monomer (PAN membranes) Increased permeability of high molecular weight molecules
Cellulose Splitting of the molecular structure
Polymethylmethacrylate (PMMA) Splitting of the molecular structure, decreased permeability of high molecular
weight molecules (106 to 105 at 35 kGy)
Polysulfone (PSU) Simultaneous chemical bridging and splitting reactions above 25 kGY
Polyethylene (EVALW and hull) Chemical bridging: hydrocarbon compounds producing aldehydes and organic acid
Polycarbonate (hull) Browning
Polyvinylchloride (PVC Coque) Splitting of the molecular structure, browning, hydrochloric acid release, possible
infiltration of partially decomposed additives that is responsible for an unpleasant
smell and accelarated aging process
Polyurethane (PUR) Release of mutagenic compounds, especially from aromatic polyurethane
From Uhlenbusch-Körver et al. [2] and Nyoman Ruspiasih et al. [4].
modifications known for dialyzers after gamma radiation Moist heat sterilization processes
[2,4].
On empotage resins, mainly constituted by aromatic polyure- Principle and implementation
thane, gamma radiation induces toxic molecule formation This sterilization mode is based on the heat denaturation of
such as the carcinogenic and mutagenic MDA (4.4-methyl membranes and proteins of the infectious agents. For dialy-
alcohol dianiline) molecule [6–9]. Conversely, polyurethanes zers, two moist heat sterilization strategies are implemented:
of aliphatic origin that do not release MDA may be used. autoclave or steam flow.
However, other carcinogenic compounds are formed in the Autoclave: The vapor acts simply as coolant agent. The mem-
rat, although the outcome is less pronounced than the aro- brane of dialysis is not in direct contact with the sterilizing
matic polyurethane [7–12]. The cost of this raw material vapor. Sterilization consists of submitting the medical device
inhibits also its function. to the heat action for a definite time (20–90 min), in a high
Table Ib
Effects of the gamma radiation on polymers.
Effets de l’irradiation gamma sur les polymères.
Preferential alteration of polymer induced by gamma irradiation
Chemical bridging
[TD$INLE]
Polyamide
Polyethylene
Silicone elastomer
Splitting reaction
[TD$INLE]
Polymethylmethacrylate
Cellulose
Polyvinyl chlorure
From Uhlenbusch-Körver et al. [2].
e17
B. Allard et al. Le Pharmacien Hospitalier et Clinicien 2013;48:e15-e21
Table II
Possible sterilization methods according to polymer membrane type.
Modes de stérilisation possibles selon la nature de la membrane.
Types of polymer membrane Gamma Etylene oxide Moist heat
irradiation sterilization sterilization
Cellulosic polymer Cellulose, cellulose derivatives + + +
(cellulose acetate excepted)
Cellulose acetate + +
Polysulfone and polysulfone derivatives Polysulfone + + +
Polyarylethersulfone (ArvlaneW) + + +
PES (DiapesW) + + +
PolyamideSW/PolyamixW + + +
PEPA + + +
Other synthetic polymers PAN + +
PMMA +
EVALW + +
Polycarbonate + +
Polyamide + +
From Uhlenbusch-Körwer et al. [2].
temperature (superior to 121 8C) in a given time and in a host-cell response conditioned by the genetic heritage and
pressure superior to the atmospheric pressure (1 in 2 Bars); the environmental conditions of the cell.
there is a relation between the pressure and the temperature
in the middle of saturated vapor (Regnault’s table), as well as Comparison of cell toxicity according to sterilization
a conversely proportional correlation meanwhile and tempe- mode
rature [6,13,14].
In their biocompatibility and toxic risks analysis of the steri-
The term ‘‘vapor’’ used by certain manufacturers should be
lization processes for dialysis devices, Guedri et al. [15] compa-
replaced by ‘‘moist heat’’.
red the fragmentation of the DNA of lymphocytes taken at the
Steam flow: dialyzers are individually sterilized by the passage
beginning, the middle and at the end of the dialysis session
of a continuous flow of vapor in all the compartments of the
(by electrophoresis and spectrometry). The dosage in spec-
dialyzer (blood and dialysate) in a temperature of 121 8C during
trometry of malondialdehyde stemming from samples leav-
15 minutes. Compared with the sterilization by the heat, the
ing the fibres allowed estimating the production of free
vapor fluente present the advantage to eliminate by the
radicals. The results showed an increased fragmentation of
continuous flow of vapor inside and outside of the fibres of
the DNA and the production of free radicals after gamma
possible residues of manufacturing as well as the pyrogenic
sterilization. The method considered the most biocompatible
and the present endotoxins after the sterilization.
being the moist heat sterilization.
During a cytotoxicity study considering the proliferative capa-
Impact of moist heat on dialyzer constituents city of the V79/B2 cells of Chinese hamster, Bommer et al. [16]
This mode of sterilization by heat is not applicable for heat in contact with eluted samples of dialyzers cuprophane ste-
sensitive polymers, such as polymethylmetacrylate, poly- rilized by moist heat or by increasing doses of gamma rays.
acrylonytrile and cellulose triacetate. On the other hand, this The samples eluted from dialyzers sterilized by gamma radia-
effective and well-mastered process, is the gold standard in tion reduced the efficiency of the examination of cell culture
terms of safety, for non-heat sensitive materials [1,2,5] by 30% compared with the control. Similar effects on the
and should be recommended for dialyzers sterilization formation of colonies were observed. The cytotoxicity, mini-
(table II). mal after moist heat sterilization, increases with the gamma
radiation and depending the dose received. According to the
Influence of sterilization mode on authors, this phenomenon would be due to the presence of
cytotoxic compounds after gamma sterilization (table III).
dialyzers cytotoxicity
Effects of manufacturing and sterilization residues
A number of studies tried to estimate the in vitro impact of
on mitochondrial oxidative phosphorylation
the manufacturing/sterilization process on the cell toxicity.
These works present the advantage to have used a ‘‘standar- The presence of manufacturing and sterilization residues
dized’’ biological material, thus reducing the variability of the in dialyzer is not often analysed with regard to the
e18
[(Figure_1)TD$IG] Dialyzers determinants of sterilization method choice
Chemical Ethylene oxide
α Ray
(Noyau d’Hélium)
(Helium nucleus)
β Ray
Irradiation (Very high energy electron beam)
(ionizing radiation)
Dry (without filler)
γ Ray
(High-energy photons)
Wet (with filler)
Dry heat
Heat Autoclave
(Moist heat outside of the medical device)
Moist heat
Steam
(Steam inside the medical device)
Figure 1. Dialysers sterilization methods.
Méthodes de stérilisation des dialyseurs.
biocompatibility in hemodialysis. Brunet et al. [17,18] estimated The results showed differences in the capacity of mitochon-
the effect of manufacturing residues of the dialyzer extracted dria to synthesize ATP and respiratory control (capacity to
from the rinsing liquid in an in vitro model using isolated rat activate the phosphorylation and in the presence of ADP). The
liver mitochondria. most noxious effect was observed with the dialyzer with
The study of the oxidative phosphorylation was realized on cellulosic membrane. Among both membranes it polysulfone,
the third rinsing liter of a cellulosic dialyzer and two dialyzers the FX obtained a better score than the APS. According to the
polysulfone, one sterilized by fluente steam (FX) and the other author’s hypothesis, the flushing effect of vapor fluente
one by gamma radiation (APS) [17]. sterilization on the manufacturing residues may explain this
difference.
Table III
Comparative study of cytotoxicity according to the dialysers Environmental studies on the hemocompatibility
sterilization method.
Étude comparative de cytotoxicité selon le mode de stérilisation In a study of Erlenkotter et al. [19], two polysulfone mem-
des dialyseurs. branes, one sterilized by gamma radiation, and the other by
1h 4h 6h the moist heat, presented different scores of hemocompati-
Ethylene Oxide sterilization bility. This study is based on the analysis of a score of
Vapor hemocompatibility regarding the factor C5a of the comple-
27 kGy + + + ment, complex thrombin–antithrombin III, the release of the
56 kGy ++ ++ ++ platelet factor 4, the synthesis and the release of elastase
from polynuclears and platelet count. The polysulfone mem-
Bommer et al. [16].
Eluates effects after circulation of culture medium (500 mL during 1 h, 4 h and 6 h) in brane sterilized by gamma radiation presented a score of
dialysers sterilized by different methods [ethylene oxide (EtO) and steam gamma
irradiation (56 kGy and 27 kGy)] on the proliferative capacity of Chinese hamster cells hemocompatibility inferior to that of the moist heat sterilized
V79/B2. : absence of cytotoxicity; +: moderate toxicity; ++: greater toxicity. polysulfone membrane.
Effets des éluats obtenus après circulation d’un milieu de culture (500 mL pendant 1 h,
4 h et 6 h) dans des dialyseurs stérilisés selon différentes méthodes [oxyde d’éthylène Because of the more recent and less widespread use, there are
(EtO), vapeur et irradiation gamma (27 kGy et 56 kGy)] sur la capacité proliférative des few data concerning the hemocompatibility of dialyzers ste-
cellules V79/B2 de hamster chinois. : absence de cytotoxicité ; + : toxicité modérée ;
++ : toxicité plus importante. rilized by radiation beta.
e19
B. Allard et al. Le Pharmacien Hospitalier et Clinicien 2013;48:e15-e21
Table IV
Effect of the sterilization method on the performance of the Purema membrane.
Impact du mode de stérilisation sur les performances de la membrane Purema.
Name Manufacturer Polymer Membrane Sterilization Beta2-Microglobulin clearance Albumin losses (g/4 h)
mode Le Roy Potier Le Roy Potier
et al. [23] (%) et al. [24] (%) et al. [23] et al. [24]
XENIUM 210 Membrana Polyethersulfone PUREMA Gamma 80.0 80.8 0.59 <3g
(polyether irradiation
sulfone)
PHYLTER 22S Moist heat 82.4 82.4 6.14 5–10 g
sterilization
In 2011, Kiaii et al. reported [20] a thrombocytopenia risk sterilization mode, the Purema membrane exhibited different
associated with the use of dialyzers sterilized by beta radiation albumin permeability.
from a 22 months retrospective cohort study of 1832 patients. The sterilization mode evidenced an effect on the structure
The relative risk of thrombocytopenia – defined by a rate of and the permeability of the membrane to solute (and thus to
platelets lower than 100 103/mL and a reduction in more a toxin). There are also more marked differences depending
than 15% by the platelet number at the end of dialysis – was on the dialysis methods used (HD or HDF pre-dilution, post-
significantly more elevated with beta sterilized dialyzers (OR: dilution, and according to the convective volume), in particu-
3.57; CI 95%: 2.54–5.04, P < 0.001). However, this publication lar transmembrane pressures (PTM). The impact of the ste-
was very controversial [21,22], since mechanisms bringing to a rilization mode of the structure of the membrane makes some
thrombocytopenia were many, multifactorial and with difficult laboratories to propose dialyzers more adapted to the dialysis
to analyse through retrospective data. The possible role of the methods (table IV).
sterilization by beta radiation beta in thrombocytopenia is still The moist heat sterilization Phylther (Phylther SD) was elimi-
to be confirmed by more thorough studies. nated on a gamma sterilization mode (Phylther G) to defeat
the two important permeability of Phylther SD in HDF.
Influence of sterilization mode on Conversely, the heat sterilization version must be exclusively
reserved for a use in conventional HD.
performances of dialyzers
Differences of performances for same membrane, Performances stability over time for gamma
according to sterilization mode sterilized dialyzers
Studies comparing the efficiency of the membranes of dialysis It is probable that the dialysis membranes sterilized by gamma
for online hemodiafiltration (HDF) reported differences for the radiation are altered over time [1]. Indeed, the Cotentin Hospital
same polyethersulfone membrane according to its steriliza- in Cherbourg studied the losses of albumin in HDF of a poly-
tion mode by moist heat or gamma. The two criteria compa- sulfone derived membrane as a function of time since steriliza-
red were: the b2-microglobulin (PM: 11,800 Da) reduction tion of the dialyzer. This dialyzer exhibited first normal loss of
coefficient and the loss of albumin (PM: 66,000 Da). albumin (< 3 g by dialysis session). However, 4 months later,
The study of Le Roy et al., realized in HDF compared eleven exactly the same dialyzer taken from the same batch and
membranes of high permeability with a surface superior to secondary packaging (box) stored according to the manufac-
2 m2 with regard to the clearance of the small molecules, of turer recommendations, exhibited losses of albumin exceeding
the b2-microglobulin and the losses of albumin during 5 g by dialysis session. The tests were made on several dialyzers
4 hours dialysis sessions [23]. of the same batch with the same result.
Each of the dialyzers was tested on three patients during three As far as we conceive that the polymerization process of the
successive sessions by the patient. Among them, two dialyzers membrane is stable in the medium-term, this fortuitous
resulting from the same manufacturer, but from two different observation questions performance stability over the time
distributors, presented the same membrane PuremaW with this of the radiation sterilized products. We are conducting a
difference that one was moist heat sterilized (Phylther S) and prospective cohort study on this topic.
the other one of the gamma radiation (Xenium).
The losses of albumin hemodiafiltration with the heat steri-
lized membrane was 6.14 g/4 h while that of the gamma Conclusion
sterilized membrane was only 0.59 g/4 h.
These results confirmed the survey carried out in Cherbourg In the decision tree for dialyzers choice, criteria that should be
on the same membranes from ten trials [24]. Depending the taken in account are:
e20
Dialyzers determinants of sterilization method choice
performances regarding therapeutic objectives for clear- [7] Shintani H, Nakamura A. Formation of 4,40 -methylenedianiline
in polyurethane potting materials by either gamma-ray or
ance of uraemic toxin;
autoclave sterilization. J Biomed Mater Res 1991;25(10):1275–86.
practical implementation, established according to the
[8] Shintani H. Formation and elution of toxic compounds from
techniques used, to the equipment and to the patients sterilized medical products: methylenedianiline formation in
population; polyurethane. J Biomater Appl 1995;10(1):23–58.
biocompatibility, more difficult to circumscribe for a [9] Mujais SK, Ing T, Kjellstrand C. Acute complications of hemo-
heterogeneous population of polypathologic patients. dialysis and their prevention and treatment. In: Jacobs C,
Kjellstrand C, Koch KM, Winchester JF, editors. Replacement
Nonetheless, the sterilization mode may have a major impact of renal function by dialysis. 4th ed., Dordrecht: Kluwer Aca-
on materials used in the manufacturing of dialyzers. With demic Publishers; 1996. p. 688–725.
regard to the cytotoxicity of some sterilization modes and [10] Piento RJ, Shah MJ, Lebherz WB, Andrews AW. Correlation of
because of the exposure, three times a week in standard bacterial mutagenicity and hamster cell transformation with
tumorigenicity induced by 2,4-toluenediamine. Cancer Lett
dialysis to six times a week for the daily dialysis, this criterion
1977;3:45–52.
must be taken into account in the decision tree of choice of [11] Shanmugam K, Subrahmanyam S, Tarakad SV, et al. 2,4-
the dialyzer. Toluene diamines – their carcinogenicity, biodegradation, ana-
Considering its very high biocompatibility and considering the lytic techniques and an approach towards development of
acquired experience, moist heat sterilization is to be favoured biosensors. Anal Sci 2001;17(12):1369–74.
in the choice of dialyzers of equivalent performances. Fur- [12] Gogolewski S. Selected topics in biomedical polyurethanes.
Colloid Polym Sci 1989;267:757–85.
thermore, non-heat sensitive products have to be developed [13] Hoenich NA, Woffindin C, Ronco C. In: Jacobs C, Kjellstrand CM,
first and foremost with this mode of sterilization. Koch CM, Winchester JF, editors. Haemodialysers and asso-
ciated devices in replacement of renal function by dialysis. 4th
ed., Dordrecht: Kluwer Academis Publishers; 1996. p. 188–230.
Disclosure of interest [14] Takesawa S, Ohmi S, Konno Y, et al. Varying methods of
sterilisation, and their effects on the structure and perme-
The authors declare that they have no conflicts of interest ability of dialysis membranes. Nephrol Dial Transplant
concerning this article. 1987;1(4):254–7.
[15] Guedri Y, et al. Biocompatibilité et risques toxiques des pro-
cessus de stérilisation des dispositifs de dialyse. Nephrol Ther
Appendix A. Supplementary data 2008;4:523.
[16] Bommer J, Miltenburger HG, Hoerner V, Kilian S. Cytotoxicity of
Supplementary data associated with this article can be found, g-ray sterilised dialysis devices. J Am Soc Nephrol 1994;5:408.
in the online version, at doi:10.1016/j.phclin.2013.10.071. [17] Brunet P, Simon N, Roubicek C, Papa K, Berland Y, Bruguerolle B.
Les résidus de fabrication des hémodialyseurs diminuent la
phosphorylation oxydative des mitochondries isolées du rat.
References Nephrol Ther 2010;6:576–80.
[18] Tabouy LJ, Chauvet-Monges AMT, Brunet PJ, et al. In vitro
[1] Dawids S, Handlos VN. Practical aspects of sterility and medical mitochondrial test to assess hémodialyser biocompatibility.
device sterilization – polymers: their properties and blood Nephrol Dial Transplant 1997;12:1635–9.
compatibility. Kluwer Academic Publishers; 1989: 347–68. [19] Erlenkötter A, Endres P, Nederlof B, et al. Score model for the
[2] Uhlenbusch-Körwer I, Bonnie-Schorn E, Grassmann A, Vienken evaluation of dialysis membrane hemocompatibility. Artif
J. Dialyser sterilisation. In: Understanding membranes and Organs 2008;32(12):962–9.
dialysers. Ed PABST Science Publishers; 2004: 80–99. [20] Kiaii M, Djurdjev O, Farah M, Levin A, Jung B, MacRae J. Use of
[3] Landfield H. Sterilization of medical devices based on polymer electron-beam sterilized hemodialysis membranes and risk of
selection and stabilization techniques. In: Szycher M, editor. thrombocytopenia. JAMA 2011;306(15):1679–87.
Biocompatible polymers, materials and composites. PA, Tech- [21] Bieber S, Ahmad S. Electron-beam sterilized hemodialyzers and
nomic Publ Inc.; 1983. p. 975–99. thrombocytopenia. JAMA 2012;307(7):665–6.
[4] Nyoman Rupiasih N, Vidyasagar PB. Comparative study of effect [22] De Prada L, Lee J, Gillespie A, Benjamin J. Thrombocytopenia
of low and medium dose rate of g irradiation on microporous associated with one type of polysulfone hemodialysis mem-
polysulfone membrane using spectroscopic and imaging tech- brane: a report of 5 cases. Am J Kidney Dis 2013;61(1):131–3.
niques. Polymer Degradation Stability 2008;93(7):1300–7. [23] Le Roy F, Hanoy M, Claeyssens S, Bertrand D, Freguin C, Godin
[5] Guillot-Maachi I, Martin S, Pourieux A, Philip V, Saux MC. M. Beta2-microglobulin removal and albumin losses in post-
Biocompatibilité des membranes de dialyse. J Pharm Clin dilution hemodiafiltration: membrane effect. Clin Kidney J
1998;17(4):195–200. 2009;2(suppl 2):402.
[6] Shintani H, Nakamura A. Analysis of a carcinogen, 4,40 -methy- [24] Potier J, Besselièvre T, Queffeulou G. Quelle membrane en
lenedianiline, from thermosetting polyurethane during ster- hémodiafiltration on line (HDF OL) et pour quelle modalité
ilization. J Anal Toxicol 1989;13(6):354–7. postD ou preD ? Nephrol Ther 2008;4:517.
e21
Riigihangete vaidlustuskomisjon Teie:
[email protected]
Meie: 01.02.2024 nr 2-6/14-6
Vastus OÜ MP AlpinTech täiendavale seisukohale vaidlustuses riigihankele „Dialüüsi
tarvikute ostmine“, viitenumber 272185
Hankija: AS Lääne-Tallinna Keskhaigla
registrikood: 10822269
aadress: Paldiski mnt 68, 10617 Tallinn
Hankija esindaja: jurist Kaisa Suvi
telefon: 6507 399
e-post:
[email protected]
Vaidlustaja: OÜ MP AlpinTech
registrikood: 16209624
Vaidlustaja esindaja: advokaat Ringo Heidmets
Adokaadibüroo GLIMSTEDT Estonia OÜ
registrikood: 14233883
e-post:
[email protected]
telefon: 5663 2589
1. SENINE MENETLUS
OÜ MP AlpinTech (vaidlustaja) esitas 17.01.2024 vaidlustuse (vaidlustus) riigihankes
„Dialüüsi tarvikute ostmine“ viitenumbriga 272185 (riigihange). Vaidlustaja taotles
Riigihangete vaidlustuskomisjoni (VAKO) kohustama AS-i Lääne-Tallinna Keskhaigla
(hankija) viima riigihanke alusdokumendid osades 1, 6 ja 35 vaidlustaja soovidega kooskõlla.
Hankija muutis 23.01.2024 riigihanke alusdokumente riigihanke osade 6 ja 35 osas ning esitas
enda vastuse vaidlustusele. Vaidlustaja esitas 29.01.2024 oma täiendava seisukoha (täiendav
seisukoht), milles tunnistas, et hankija on riigihanke osade 6 ja 35 osas kõrvaldanud
õigusrikkumised vaidlustaja suhtes. Käesolevaga esitab hankija endapoolse vastuse vaidlustaja
täiendavale seisukohale.
AS Lääne-Tallinna Keskhaigla, Paldiski mnt 68, 10617 Tallinn, Eesti Vabariik
Telefon: 650 7393, Faks 650 7265, e-post
[email protected]
Registrikood: 10822269, Arvelduskonto: EE081010220022881019 SEB Pank
2. HANKIJA KAHTLUS VAIDLUSTAJA REAALSES VÕIMALUSES RIIGIHANKES
OSALEDA
Vaidlustaja on oma täiendavas seisukohas väitnud, et on planeerinud osaleda riigihankes
koostöös ühispakkujaga Rihanna OÜ (registrikood 11924606) (ühispakkuja). Esialgses
vaidlustuses pole ühispakkujast ja ühispakkumusest aga sõnagi.
Vaidlustaja sõnastusest lähtuvalt („...on planeerinud...“) on hankijal jätkuvalt kahtlus kas
vaidlustuse esitamise hetkel, s.o. 17.01.2024, eksisteeris vaidlustaja ning ühispakkuja vahel
tegelik tahtlus ühispakkumuse esitamiseks või esindas vaidlustaja vaidlustuses ainult
ühispakkuja huve ning potentsiaalne ühispakkumus on võte vältimaks vaidlustuse läbi vaatamata
jätmist RHS § 192 lg 3 p-i 6 alusel. Ühispakkuja on esitanud 26.01.2024 kinnituse, et
ühispakkujal on soov osaleda koos vaidlustajaga vaidlusaluses riigihankes ühispakkujatena.
Kinnituskiri on ühispakkuja poolt aga allkirjastatud alles peale hankija taotlust. Kui vaidlustaja
esindas vaidlustuses sisuliselt ühispakkuja huve ning vaidlustaja tahtlus esitada ühispakkumine
ühispakkujaga tekkis alles peale hankija poolt 23.01.2024 esitatud taotlust vaidlustuse läbi
vaatamata jätmiseks RHS § 192 lg 2 p 6 alusel, tuleb hankija hinnangul vaidlustus jätta läbi
vaatamata.
Vaidlustaja on oma täiendavas seisukohas väitnud, et vaidlustaja avastas vaidlustatud riigihanke
ning registreerus selle juurde 17.01.2024. Ei ole eluliselt usutav, et vaidlustaja avastas riigihanke
17.01.2024 kl 11:04, esitas vaidlustuse 17.01.2024 kl 21:36 ning vahepealse 10,5 tunni jooksul
jõudis sõlmida koostöökokkuleppe ühispakkujaga. Vaidlustaja menetluskulude nimekirjast
lähtuvalt on enne vaidlustuse esitamist jõutud veel 7 tunni ning 42 minuti ulatuses osutada ka
õigusabi vaidlustajale. Seega, kui vaidlustaja väidab, et ta tõepoolest alles 17.01.2024 riigihanke
avastas, on vaidlustaja vähem kui 3 tunniga jõudnud turult leida dialüüsitarvikute osas
koostööpartneri, sõlminud selles osas koostöökokkuleppe, süvenenud koos koostööpartneriga
(s.o ühispakkujaga) riigihanke alusdokumentidesse, leidnud alusdokumentidest neid ühiselt
häirivad puudujäägid ning kommunikeerinud omapooolsed mured vaidlustaja esindajale. Hankija
juhib tähelepanu, et vaidlustaja planeeritav koostööpartner, ühispakkuja, on hanke juurde
registreerunud juba 08.12.2023, s.o. 39 päeva enne vaidlustuse esitamist ning 47 päeva enne
25.01.2024 ehk pakkumuste esitamise tähtaega. Hankija tuletab meelde, et riigihanke juurde
registreerunud ettevõtjate nimekiri on nähtav ainult hankija meeskonna liikmetele ning sealt
vaidlustaja ega ühispakkuja potentsiaalset koostööpartnerit otsida ei saa. Hankijale ei tundu
selline situatsioon eluliselt usutav.
Hankijale tundub eluliselt usutav, et taoliste suuremahuliste koostööprojektide (esialgselt
vaidlustatud osade eeldatav maksumus 1 610 400 EUR, kogu riigihanke eeldatav maksumus 4
105 000,00 EUR), sõlmimiseks kulub kindlasti enam kui 2 tundi ja 50 minutit (aeg, mis jääb üle
lahutades 10 tunnist ja 32 minutist väidetav õigusabi osutamine 7 tunni ja 42 minuti ulatuses).
Juba ainuüksi koostöölepingu koostamine, rääkimata tõenäolisest pooltevahelistest
läbirääkimistest lepingutingimuste osas, võtab kauem aega kui 2 tundi ja 50 minutit.
AS Lääne-Tallinna Keskhaigla, Paldiski mnt 68, 10617 Tallinn, Eesti Vabariik
Telefon: 650 7393, Faks 650 7265, e-post
[email protected]
Registrikood: 10822269, Arvelduskonto: EE081010220022881019 SEB Pank
Hankija juhib tähelepanu, et vaidlustaja ainukene juhatuse liige Marko Pikani on vaidlustajat
esindava Adokaadibüroo GLIMSTEDT Estonia OÜ osanik ning vaidlustaja lepingulise esindaja,
vandeadvokaadi abi Ringo Heidmets’a, tööandja. Vaidlustaja on esitanud menetluskulude
nimekirja, mille kohaselt on lepingulise esindamise ajakulu kuni vaidlustuse esitamiseni 7 tundi
ja 42 minutit. Samas pole menetluskulude nimekirja järgi võimalik kontrollida kas lepingulise
esindamise eest tekkinud õigusabi kulud on tekkinud ühispakkuja või vaidlustaja poolsest
lepingulise esindaja poole pöördumisest, mis kuupäevadel on tekkinud õigusabikulud, kas
lepingulise esindamise eest on esitatud arved ning kellele.
Seetõttu on hankijal jätkuvalt kahtlus kas vaidlustajal oli vaidlustuse esitamise ajal reaalne
võimalus riigihankes osaleda.
Alternatiivselt esitab hankija oma seisukoha täiendavale seisukohale, kui vaidlustust ei peaks
läbi vaatamata jätma.
3. RIIGIHANKE OSA 1
Vaidlustaja on oma täiendavas seisukohas nõustunud hankijaga, et suures doosis kiirgus võib
dialüsaatorite materjalidele ja patsientidele negatiivselt mõjuda. Vaidlustaja on samuti oma
täiendavas seisukohas märkinud, et „kiirgusega steriliseerimise mõju on sügavam, steriliseeritud
tulemuse saavutamine oluliselt kindlam [...] erinevalt auruga steriliseerimisest.“ Samas pole
vaidlustaja, erinevalt hankijast, lisanud oma seisukoha tõendamiseks ühtegi tõendit. Hankija on
oma eelnevale 23.01.2024 vastusele lisanud teadusartikli hankija veendumuse kohta, et auruga
steriliseerimine on oluline biosobivuse seisukohast. Hankija lisab käesolevale vastusele lisaks
veel kaks teadusartiklit auruga steriliseerimise kohta, Lisa 11 ja Lisa 22.
Hankijal on õigus otsustada, millist kaupa ta hankida soovib ning tehnilise kirjelduse koostamisel
lai kaalutlusõigus. Vaidlustajal ei ole subjektiivset õigust nõuda, et hangitaks teistsugust kaupa,
mis oleks vaidlustaja hinnangul parem või otstarbekam. Vaidlustaja on hankija vabadust
hankeeseme määramise osas ka oma vaidlustuses ise möönanud. Hankija on oma eelnevale
23.01.2024 vastusele lisanud kahe juhtiva (ning konkureeriva) dialüsaatorite tootja brošüürid,
1
Erlenkötter, A.; Endres, P.; Nederlof, B.; Hornig, C.; Vienken, J. Score model for the evaluation of
dialysis membrane hemocompatibility. Artif. Organs 2008, 32, 962–969. Väljavõte lk 968: „: The γ-
sterilized polysulfone had a worse overall hemocompatibility than the steam-sterilized polysulfone.“
2
Allard B, Begri R., Potier J., Couple S. Dialyzers biocompatibility and efficiency. Determinants of
sterilization method choice. Pharm Hospet Clin. 2013;48(4):e15-e21. Väljavõte lk e15: „Radiation
sterilization induces structural modifications of the dialyzers components. Beta radiation causes fewer
modifications than gamma radiation; these structural modifications can release cytotoxic components.
Steam sterilization is a reference in terms of security for thermostable materials with an added benefit for
steam process, which allows the removal of waste generated during the dialyzer’s manufacturing.“
Väljavõte lk e20: „It is probable that the dialysis membranes sterilized by gamma radiation are altered
over time [1]. Indeed, the Cotentin Hospital in Cherbourg studied the losses of albumin in HDF of a
polysulfone derived membrane as a function of time since sterilization of the dialyzer. This dialyzer
exhibited first normal loss of albumin (< 3 g by dialysis session). However, 4 months later, exactly the
same dialyzer taken from the same batch and secondary packaging (box) stored according to the
manufacturer recommendations, exhibited losses of albumin exceeding 5 g by dialysis session. The tests
were made on several dialyzers of the same batch with the same result.“
AS Lääne-Tallinna Keskhaigla, Paldiski mnt 68, 10617 Tallinn, Eesti Vabariik
Telefon: 650 7393, Faks 650 7265, e-post
[email protected]
Registrikood: 10822269, Arvelduskonto: EE081010220022881019 SEB Pank
kes kasutavad auruga steriliseerimist. Seetõttu ei ole tõene vaidlustaja väide, et hankija on
riigihanke koostamisel aluseks võtnud vaid ühe konkreetse tootja tooted ning seeläbi
põhjendamatult piiranud konkurentsi. Hankijal on usaldus, et auruga steriliseerimise nõue vastab
hankija vajadustele, hankija on põhjendanud sellise nõude vajadust ning näidanud oma
23.01.2024 vastuses, et riigihanke koostamisel pole aluseks võetud vaid ühe konkreetse tootja
tooteid. Kui mitmel pakkujal on võimalik tehnilisele kirjeldusele vastavat toodet pakkuda, on
riigihange avatud konkurentsile.
Eeltoodust lähtuvalt on hankija seisukohal, et vaidlustaja vaidlustus riigihanke osas 1 tuleb jätta
rahuldamata.
4. MENETLUSKULUD
Hankija on seisukohal, et vaidlustaja menetluskulud tuleb jätta vaidlustaja kanda.
Riigihanke tingimused pidid ühispakkujale teada olema juba enne 17.01.2024, sest ühispakkuja
on riigihanke juurde registreerunud 08.12.2023, s.o. 39 päeva enne vaidlustuse esitamist ning 47
päeva enne 25.01.2024 ehk pakkumuste esitamise tähtaega. Vaidlustaja on oma täiendavas
seisukohas väitnud, et ühispakkumuses tugineb ta ühispakkuja valdkonnaspetsiifilisele
kogemusele ja võimekusele ning vaidlustaja osaks jääb lepingu täitmise administratiivne pool.
Tuginedes vaidlustaja väitele, on just ühispakkujal erialaspetsiifilised teadmised mõistmaks
riigihanke tehnilist kirjeldust ning vähemalt 39 päeva aega pöördumaks hankija poole läbi
riigihangete registri ettepanekuga riigihanke alusdokumentide muutmiseks. Ühispakkuja seda
pole aga teinud. Hankija hinnangul on 39 päeva enam kui piisav aeg hankijale ettepaneku
esitamiseks ning vastuse ära ootamiseks. Kui ühispakkuja oleks pöördunud hankija poole
omapoolse ettepanekuga, poleks riigihanke osade 6 ja 35 osas menetlustuskulusid üldse
tekkinud.
Hankija juhtis eelnevalt käesolevas vastuses tähelepanu, et vaidlustaja ainukene juhatuse liige
Marko Pikani on vaidlustajat esindava Adokaadibüroo GLIMSTEDT Estonia OÜ osanik ning
vaidlustaja lepingulise esindaja, vandeadvokaadi abi Ringo Heidmets’a, tööandja. VAKO poolt
välja mõistetavad õigusabikulud on otsene tuluallikas vaidlustaja juhatuse liikmele, kes omab
oma Adokaadibüroo GLIMSTEDT Estonia OÜ osaluse kaudu kontrolli ja motivatsiooni
VAKO-le esitatud õigusabikulude suuruse osas.
Eelnevast tulenevalt palub hankija vaidlustuskulud jätta vaidlustaja kandja. VAKO vastupidise
otsuse korral palub hankija VAKO-l hinnata õigusabikulude suuruse ja mahu põhjendatust
arvestades eelpool toodud asjaolusid.
Riigihanke osa 1 osas on hankija seisukohal, et vaidlustus tuleb jätta rahuldamata ning
menetluskulud jätta vaidlustaja kanda.
AS Lääne-Tallinna Keskhaigla, Paldiski mnt 68, 10617 Tallinn, Eesti Vabariik
Telefon: 650 7393, Faks 650 7265, e-post
[email protected]
Registrikood: 10822269, Arvelduskonto: EE081010220022881019 SEB Pank
Eeltoodud asjaolusid arvesse võttes palub hankija:
1. jätta vaidlustaja vaidlustus läbi vaatamata RHS § 192 lg 3 p 6 alusel;
2. jätta vaidlustuse esitamisega seotud kulud vaidlustaja kanda;
või alternatiivselt:
1. jätta vaidlustaja vaidlustus rahuldamata riigihanke osa 1 osas;
2. jätta vaidlustaja vaidlustus riigihanke osade 6 ja 35 osas läbi vaatamata RHS § 192
lg 3 p 5 alusel ning lõpetada vaidlustusmenetlus RHS § 197 lg 1 p 8 alusel;
3. jätta vaidlustuse esitamisega seotud kulud vaidlustaja kanda;
Lugupidamisega
(allkirjastatud digitaalselt)
Kaisa Suvi
Volitatud esindaja
Lisatud:
Lisa 1 Erlenkötter, A.; Endres, P.; Nederlof, B.; Hornig, C.; Vienken, J. Score model for the
evaluation of dialysis membrane hemocompatibility.
Lisa 2 Allard B, Begri R, Potier J, Couple S. Dialyzers biocompatibility and efficiency.
Determinants of sterilization method choice.